The DeVilbiss 5 Liter (555) Oxygen Concentrator is intended to provide supplemental oxygen to adult patients requiring oxygen therapy. The device may be used in the home or an institutional setting. The device is not intended to be life sustaining or life supporting.
Device Story
Portable oxygen concentrator; produces oxygen-enriched gas via pressure swing adsorption (PSA) using molecular sieves to extract nitrogen from room air. Operates continuously; delivers oxygen at 0.5–5 LPM. Used in home or institutional settings; operated by patients or caregivers. Features internal compressor, intake/final HEPA filters, and control panel with flow settings and alarm displays. Connects to patient via nasal cannula. Provides visual/audible alarms for power failure, low oxygen concentration, flow obstruction, and system malfunctions. Healthcare providers use output to manage oxygen therapy for patients with respiratory needs; device benefits patients by providing consistent supplemental oxygen.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 18562 series, ISO 10993-17), electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), and performance testing per ISO 80601-2-69 and IEC 60601-1-8/11/6 standards.
Technological Characteristics
Pressure swing adsorption with molecular sieve. 120V, 60Hz power. 0.5-5 LPM flow. 8.5 PSIG outlet pressure. Class II, Type BF electrical classification. Biocompatible gas-pathway materials. Intake/final HEPA filtration. Dimensions: 18.4" x 14.6" x 9.9". Weight: 30 lbs. Software-controlled alarms for flow, concentration, and system integrity.
Indications for Use
Indicated for adult patients requiring supplemental oxygen therapy. Not for life-sustaining or life-supporting use.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 26, 2026
Devilbiss Healthcare, LLC
Susan Scott
Senior Director, Regulatory Affairs
100 Devilbiss Dr.
Somerset, Pennsylvania 15501
Re: K253549
Trade/Device Name: DeVilbiss 5 Liter Oxygen Concentrator (555)
Regulation Number: 21 CFR 868.5440
Regulation Name: Portable Oxygen Generator
Regulatory Class: Class II
Product Code: CAW
Dated: February 18, 2026
Received: February 18, 2026
Dear Susan Scott:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253549 - Susan Scott
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K253549 - Susan Scott
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K253549
Device Name
DeVilbiss 5 Liter Oxygen Concentrator (555)
Indications for Use (Describe)
The DeVilbiss 5 Liter (555) Oxygen Concentrator is intended to provide supplemental oxygen to adult patients requiring oxygen therapy. The device may be used in the home or an institutional setting. The device is not intended to be life sustaining or life supporting.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
Page 1 of 7
# DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator 510(k) Summary
## 1. SUBMITTER INFORMATION
Applicant: DeVilbiss Healthcare LLC
Contact: Susan Scott
Phone: (801) 931-1050
Email: sscott@drivemedical.com
Address: 100 DeVilbiss Dr
Somerset, PA 15501, USA
## 2. CORRESPONDENT INFORMATION
Contact: Susan Scott
Title: Senior Director, Regulatory Affairs
Firm: DeVilbiss Healthcare LLC
## 3. DATE PREPARED: NOVEMBER 13, 2025
## 4. DEVICE INFORMATION
Device Name: DeVilbiss 5 Liter (555) Oxygen Concentrator
Common Name: Portable oxygen generator
Regulation Number: 868.5440
Regulation Name: Generator, Oxygen, Portable
Product Code: CAW
Regulatory Class: Class II
## 5. PREDICATE DEVICE INFORMATION
Device Name: JUMAO Oxygen Concentrator
Common Name: Portable oxygen generator
510(k) Number: K230969
Manufacturer: Jiangsu Jumao X-Care Medical Equipment Co., Ltd.
The predicate device has not been subject to a design related recall.
## 6. DEVICE DESCRIPTION
The DeVilbiss 5-Liter (555) Oxygen Concentrator (hereafter referred to as “555”) is a portable device that produces an oxygen enriched gas mixture by drawing in room air and extracting nitrogen through a pressure swing adsorption process, allowing oxygen to be delivered at a range of prescribed flows to patients in need of supplemental oxygen.
{5}
Page 2 of 7
# DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator 510(k) Summary
## 7. INDICATIONS FOR USE
The DeVilbiss 5 Liter (555) Oxygen Concentrator is intended to provide supplemental oxygen to adult patients requiring oxygen therapy. The device may be used in the home or an institutional setting. The device is not intended to be life sustaining or life supporting.
{6}
Page 3 of 7
# DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator 510(k) Summary
## 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Table 1: Device Comparison Table
| Characteristic | Predicate Device: JUMAO Portable Oxygen Concentrator (K230969) | Subject Device: DeVilbiss 5 Liter (555) Oxygen Concentrator | Comparison |
| --- | --- | --- | --- |
| Product Code | CAW (generator, oxygen, portable) | CAW (generator, oxygen, portable) | Same |
| Indications for Use Statement | The JUMAO Oxygen Concentrator is intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve. It is not intended to sustain or support life. | The DeVilbiss 555 oxygen concentrator is intended to provide supplemental oxygen to adult patients requiring oxygen therapy. The device may be used in the home or an institutional setting. The device is not intended to be life sustaining or life supporting. | Equivalent. The differences do not result in a new intended use. |
| Intended patient population | The device is intended for use in adults. | The device is intended for use in adults. | Same |
| Proprietary name | Oxygen Concentrator | 5 Liter (555) Oxygen Concentrator | Same |
| Structure and main components | JMC5A Ni oxygen concentrator is made up of mainframe and flowmeter. The front panel of the device contains the controls and indicators. These include a standard barb fitting for attaching the oxygen tubing, the adjustable flow meter, a power light indicator, an elapsed time meter, and a standard on/off rocker type power switch. | The DeVilbiss 555 oxygen concentrator consists of a unit housing that includes a compressor and sieve beds. The device includes a control panel that provides Flow settings and alarm displays. A standard barb fitting provides for connection to a nasal cannula. The front panel includes a rocker style on/off power switch. | Equivalent. |
| Filters | Cabinet, out HEPA, compressor inlet | Intake HEPA, Internal Compressor, Final HEPA | Same |
| Material of main components | Materials demonstrated to be biocompatible for gas-pathway contact | Materials demonstrated to be biocompatible for gas-pathway contact | Equivalent The technological differences do not raise a different question of safety or effectiveness. |
| Technology | Pressure Swing Adsorption with a Molecular Sieve | Pressure Swing Adsorption with a Molecular Sieve | Same |
{7}
Page 4 of 7
# DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator
510(k) Summary
| Characteristic | Predicate Device: JUMAO Portable Oxygen Concentrator (K230969) | Subject Device: DeVilbiss 5 Liter (555) Oxygen Concentrator | Comparison |
| --- | --- | --- | --- |
| Power Requirements | AC120V, 60Hz;
Current: 3.5A;
Power: 450VA | 120 V~, 60 Hz,
2.6 A Max | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Oxygen concentration | 93%±3% at 0.5 to 5L/min (after turning on 5 minutes) | 87% - 96% at 0.5-5 LPM | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Oxygen flow | Continuous oxygen flow, adjustable
0.5~5L/min | Continuous oxygen flow, adjustable
0.5 to 5 LPM | Same |
| Outlet pressure | 38kPa± 5kPa | 8.5 PSIG regulated oxygen outlet pressure | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Noise Level | Sound level: ≤ 50.5dB (A);
Acoustic power level: 58.5dB (A) | A-weighted sound pressure level: <46 dBA
Sound power level @ 5 LPM: <56 dBA | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Weight | 16.1kg | 30 lbs (13.6kg) | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Dimensions | 330×260×540(mm) | 18.4” H x 14.6” W x 9.9” D
(46.6 x 37.0 x 25.0 cm) | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
{8}
# DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator
510(k) Summary
| Characteristic | Predicate Device: JUMAO Portable Oxygen Concentrator (K230969) | Subject Device: DeVilbiss 5 Liter (555) Oxygen Concentrator | Comparison |
| --- | --- | --- | --- |
| Electric classification | Class II, Type BF | Class II, Type BF | Same |
| Alarm | • Start-up fail alarm
• Low oxygen concentration alarm
• Power supply failure alarm
• Pressure failure alarm:
○ Outlet block, oxygen flowrate below 0.6 Lpm;
○ Compressor stop | • Power supply failure
• Low O2 concentration (start up period over)
• Malfunction – O2 gas temperature
• Malfunction – Obstruction of gas pathways, low flow
• Malfunction – Corrupted settings
• Malfunction – Non-recoverable valve error
• Malfunction – O2 oxygen sensor communication failure
• High flow
• Malfunction – Fan not running
• Startup period | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Mode of operation | Continuous duty | Continuous duty | Same |
| Normal operating ambient | Temperature range: 5°C~40°C (41°F~104°F)
Relative humidity: ≤80%
Atmospheric pressure: 86kPa~106kPa (12.47psi~15.37psi)
1828 meter (5997 feet) height above sea level | Temperature range: 41°F (5°C) to 95°F (35°C)
Humidity Range: 15% to 90%, non-condensing
Atmospheric pressure: 1013 hPa to 715 hPa
0-3000 m (sea level-9840 ft)
Across the voltage range: No degradation of performance | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Storage and transport ambient | Temperature Range: 0°C~+55°C (32°F~+131°F)
Relative Humidity Range: 10%~90%
Atmospheric pressure: 70kPa~106kPa (10.2psi~15.37psi) | Temperature Range: -13°F (-25°C) to 158°F (70°C)
Humidity Range: 15% to 90%, non-condensing | Equivalent
The technological differences do not raise a different question of safety or effectiveness. |
| Software control | Yes | Yes | Same |
| Patient interface style | Visual, direct contact type patient interface | Visual, direct contact type patient interface | Same |
Page 5 of 7
{9}
DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator 510(k) Summary
9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
Biocompatibility Testing
Biocompatibility testing to the following standards was conducted:
- ISO 18562-1:2017, Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2017, Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds
- ISO 10993-17:2023, Biological evaluation of medical devices - Part 17: Toxicological risk assessment of medical device constituents
- ISO 16000-3:2022, Indoor air — Part 3: Determination of formaldehyde and other carbonyl compounds in indoor and test chamber air — Active sampling method
- ISO 16000-6:2021, Indoor air — Part 6: Determination of organic compounds (VVOC, VOC, SVOC) in indoor and test chamber air by active sampling on sorbent tubes, thermal desorption and gas chromatography using MS or MS FID
- ISO 80601-2-69: 2020, Medical electrical equipment - Part 2-69: Particular requirements for the basic safety and essential performance of oxygen concentrator equipment
Electrical Safety
Electrical safety testing to the following standard was conducted:
- ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021], Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
Electromagnetic Compatibility (EMC)
EMC testing to the following standard was conducted:
- IEC 60601-1-2:2020 (consolidated version), Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Software
Software verification and validation was performed, and it was demonstrated that the software performs as intended in accordance with the FDA Guidance for the Content of Premarket Submissions for Device Software Functions
Performance Testing
Performance testing to the following standards was conducted:
Page 6 of 7
{10}
Page 7 of 7
# DeVilbiss Healthcare LLC: DeVilbiss 5 Liter (555) Oxygen Concentrator 510(k) Summary
- ISO 80601-2-69:2020 Particular requirements for the basic safety and essential performance of oxygen concentrator equipment.
- IEC 60601-1-8:2020 (consolidated version), Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral Standard: General requirements tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
- IEC 60601-1-11:2020 (consolidated version), Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISTA 3A:2018, Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
- IEC 60601-1-6:2020 (consolidated edition), Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- ANSI/AAMI/IEC 62366-1:2020 (consolidated edition), Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1
- IEC 62304:2015 (consolidated edition), Medical device software - Software life cycle processes
## Conclusion
The results of the performance testing described above demonstrate that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.