← Product Code [NBC](/productcode/NBC) · K253539

# Alere NT-proBNP for Alinity i (K253539)

_Axis-Shield Diagnostics, Ltd. · NBC · Feb 18, 2026 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K253539

## Device Facts

- **Applicant:** Axis-Shield Diagnostics, Ltd.
- **Product Code:** [NBC](/productcode/NBC.md)
- **Decision Date:** Feb 18, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1117
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K253539 · Feb 18, 2026 | Alere NT-proBNP for Alinity i | Axis-Shield Diagnostics, Ltd. | Adjudicated clinical cohort from K241176 | The sponsor used an existing adjudicated clinical cohort to evaluate the device's prognostic value for 12-month risk stratification of all-cause mortality, cardiac-related mortality, and heart-related hospitalization in patients with new onset or worsening heart failure. | Prognostic performance; Heart failure; Risk stratification; Retrospective analysis |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of a previously adjudicated prospective cohort; Follow-up/Duration: 12 months | 861 subjects presenting to the ED with new onset or worsening heart failure; Sample Size: 861 | Not applicable for this study | All-cause mortality, cardiac-related mortality, and heart-related hospitalization |

## Indications for Use

The Alere NT-proBNP for Alinity i assay is a chemiluminescent microparticle immunoassay (CMIA) used for the in vitro quantitative determination of N-terminal pro B-type natriuretic peptide (NT-proBNP) in human serum and plasma on the Alinity i system. The Alere NT-proBNP for Alinity i assay can also be used as an aid in 12-month risk stratification for prognosis of all-cause mortality, cardiac-related mortality, and heart-related hospitalization in patients presenting to the emergency department, diagnosed with new onset or worsening HF. The assay may further be used in the risk stratification of patients with acute coronary syndrome (ACS).

## Device Story

Automated two-step chemiluminescent microparticle immunoassay (CMIA) for quantitative NT-proBNP measurement in human serum/plasma; used on Alinity i system in clinical laboratory settings. Sample incubated with anti-NT-proBNP coated paramagnetic microparticles; washed; incubated with acridinium-labeled anti-NT-proBNP conjugate; washed; trigger solutions added. Chemiluminescent reaction measured as relative light units (RLU) proportional to analyte concentration. Results interpreted by clinicians alongside medical history and physical exam to aid HF diagnosis and risk-stratify patients for 12-month adverse outcomes (mortality/hospitalization) or ACS. Benefits include objective prognostic data for ED triage and management of HF/ACS patients.

## Clinical Evidence

Prospective 12-month follow-up study of 861 subjects with adjudicated HF diagnosis presenting to the ED. Primary endpoint: composite of all-cause mortality, cardiac-related mortality, or heart-related hospitalization. Kaplan-Meier analysis and Cox proportional hazards regression showed statistically significant association between higher NT-proBNP quartiles and increased risk of composite events. Adjusted hazard ratios for Q4 vs Q1 were 1.56 (95% CI: 1.14, 2.12) for composite outcome and 8.24 (95% CI: 2.66, 25.50) for cardiac-related mortality. Nonclinical performance (precision, linearity, specificity) previously established in K241176.

## Technological Characteristics

Chemiluminescent microparticle immunoassay (CMIA); quantitative measurement; sandwich immunoassay principle. Measuring interval: 15.8 to 35,000.0 pg/mL. No hook effect up to 372,620 pg/mL. Designed for use on the Alinity i system. Traceability per CLSI EP32-R.

## Regulatory Identification

The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.

## Special Controls

*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”

## Predicate Devices

- Roche Elecsys proBNP II ([K092649](/device/K092649.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K253539

B Applicant

Axis-Shield Diagnostics Limited

C Proprietary and Established Names

Alere NT-proBNP for Alinity i

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NBC | Class II | 21 CFR 862.1117 - B-Type Natriuretic Peptide Test System | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

Expansion of Indications for Use

B Measurand:

NT-proBNP

C Type of Test:

Quantitative immunoassay

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K253539 - Page 2 of 8

# III Intended Use/Indications for Use:

## A Intended Use(s):

See Indications for Use below.

## B Indication(s) for Use:

The Alere NT-proBNP for Alinity i assay is a chemiluminescent microparticle immunoassay (CMIA) used for the in vitro quantitative determination of N-terminal pro B-type natriuretic peptide (NT-proBNP) in human serum and plasma on the Alinity i system.

The Alere NT-proBNP for Alinity i assay can also be used as an aid in 12-month risk stratification for prognosis of all-cause mortality, cardiac-related mortality, and heart-related hospitalization in patients presenting to the emergency department, diagnosed with new onset or worsening HF.

The assay may further be used in the risk stratification of patients with acute coronary syndrome (ACS).

## C Special Conditions for Use Statement(s):

Rx – For Prescription Use Only

## D Special Instrument Requirements:

Alinity i system

# IV Device/System Characteristics:

## A Device Description:

Unchanged since K241176

## B Principle of Operation:

Unchanged since K241176

## V Substantial Equivalence Information:

## A Predicate Device Name(s):

Elecsys ProBNP II Stat Immunoassay

## B Predicate 510(k) Number(s):

K092649

{2}

K253539 - Page 3 of 8

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K251440 | K092649  |
| --- | --- | --- |
|  Device Trade Name | Alere NT-proBNP for Alinity i | Elecsys proBNP II STAT Immunoassay  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | For the in vitro quantitative determination of N-terminal pro B-type natriuretic peptide (NT-proBNP) in human serum and plasma. This assay may be used in the risk stratification of patients with acute coronary syndrome (ACS). | Same  |
|  Specimen Type | Human serum and plasma | Same  |
|  Test Principle | Sandwich immunoassay | Same  |
|  Measurement Type | Quantitative | Same  |
|  Detection Technology | Chemiluminescence | Same  |
|  General Device Characteristic Differences |  |   |
|  Principle of Operation | Chemiluminescent microparticle immunoassay (CMIA) | Electro chemiluminescent microparticle immunoassay (ECLIA)  |
|  Measuring Interval | 15.8 to 35,000.0 pg/mL | 5 to 35,000 pg/mL  |
|  Hook Effect | No hook effect up to 372,620 pg/mL | No hook effect up to 300,000 pg/mL  |

VI Standards/Guidance Documents Referenced:

CLSI EP32-R (Formerly X05-R): Metrological Traceability and Its Implementation; A Report

CLSI EP09c 3rd Edition: Measurement Procedure Comparison and Bias Estimation Using Patient Samples

{3}

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:
Unchanged since K241176.

2. Linearity:
Unchanged since K241176.

3. Analytical Specificity/Interference:
Unchanged since K241176.

4. Detection Limit and Assay Reportable Range:
Unchanged since K241176.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Unchanged since K241176.

6. Assay Cut-Off:
Unchanged since K241176.

B Comparison Studies:

1. Method Comparison with Predicate Device:
Not applicable.

2. Matrix Comparison:
Unchanged since K241176.

C Clinical Studies:

1. Clinical Sensitivity:
Not applicable.

2. Clinical Specificity:
Not applicable.

K253539 - Page 4 of 8

{4}

K253539 - Page 5 of 8

3. Clinical Cut-Off:

See “Other Clinical Supportive Data (When 1. and 2. Are Not Applicable)”.

4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Risk of Adverse Events in Patients Diagnosed with New Onset or Worsening HF Presenting to the ED

A clinical study was performed to evaluate the candidate device as an aid in 12-month risk stratification for prognosis of all-cause mortality, cardiac-related mortality, and heart-related hospitalization in patients presenting to the emergency department, diagnosed with new onset or worsening HF using the adjudicated cohort described in K241176. Of the 880 subjects with an adjudicated diagnosis, 861 were prospectively followed for 6±1 and 12±2 months after the initial ED admission. (The remaining 19 subjects did not have a follow-up evaluation.) The study included 495 (57.5%) male subjects and 366 (42.5%) female subjects, with an overall mean age of 62.0 years. Specimens were collected at admission to the ED and tested with the Alere NT-proBNP for Alinity i assay. Study population quartiles for NT-proBNP were calculated in pg/mL from the included population (n = 861):

|  Study Population Quartile | First Quartile (Q1) | Second Quartile (Q2) | Third Quartile (Q3) | Fourth Quartile (Q4)  |
| --- | --- | --- | --- | --- |
|  NT-proBNP Concentration (pg/mL) | ≤ 1214.7 | > 1214.7 to 3261.3 | > 3261.3 to 7031.3 | > 7031.3  |

The primary outcome was a composite outcome, defined as the first occurrence of all-cause mortality, cardiac-related mortality, or heart-related hospitalization. The secondary outcomes included the individual components of the composite outcome. The prognostic value of NT-proBNP quartiles was assessed via Kaplan-Meier survival analysis, Cox proportional hazards regression, and event rate analysis (absolute risk) for the primary and secondary outcomes.

Kaplan-Meier Survival Analysis

The Kaplan-Meier curves demonstrate that the risk of the composite outcome within 12 months is higher for higher NT-proBNP population quartiles.

{5}

![img-0.jpeg](img-0.jpeg)
NT-proBNP Q1 -Q2 Q3 Q4

# Absolute Risk

Absolute risk of each outcome at 12 months was estimated using univariable Cox proportional hazards models. All evaluated outcomes (all-cause mortality, cardiac-related mortality, heart-related hospitalization, and the composite) demonstrate an increasing trend across NT-proBNP quartiles. The results are summarized in the following table.

|  Outcome | Absolute Risk (%) (95% CI)  |   |   |   |
| --- | --- | --- | --- | --- |
|   |  Q1 | Q2 | Q3 | Q4  |
|  Composite Outcome | 44.03 (36.86, 50.39) | 49.40 (42.25, 55.66) | 56.49 (49.37, 62.60) | 63.08 (55.99, 69.02)  |
|  Mortality | 10.21 (5.97, 14.25) | 11.45 (7.03, 15.66) | 18.94 (13.41, 24.12) | 25.50 (19.33, 31.19)  |
|  Cardiac Mortality | 2.03 (0.04, 3.97) | 6.84 (3.31, 10.24) | 7.24 (3.51, 10.83) | 15.78 (10.50, 20.75)  |
|  Heart-Related Hospitalization | 39.69 (32.44, 46.16) | 46.74 (39.43, 53.17) | 47.89 (40.34, 54.49) | 53.41 (45.55, 60.13)  |

K253539 - Page 6 of 8

{6}

Cox Proportional Hazards Regression

Univariable results and multivariable results are summarized for the composite outcome and each individual outcome in the table below.

|  Outcome | NT-proBNP Quartile^{a} | Unadjusted Hazard Ratio (95% CI) | Adjusted Hazard Ratio (95% CI)^{b}  |
| --- | --- | --- | --- |
|  Composite | Q2 | 1.17 (0.89, 1.56) | 1.16 (0.86, 1.56)  |
|   |  Q3 | 1.43 (1.09, 1.89) | 1.44 (1.08, 1.93)  |
|   |  Q4 | 1.72 (1.31, 2.25) | 1.56 (1.14, 2.12)  |
|  All-Cause Mortality | Q2 | 1.13 (0.63, 2.03) | 1.15 (0.63, 2.10)  |
|   |  Q3 | 1.95 (1.15, 3.32) | 1.89 (1.08, 3.30)  |
|   |  Q4 | 2.73 (1.65, 4.53) | 2.13 (1.19, 3.80)  |
|  Cardiac-Related Mortality | Q2 | 3.46 (1.14, 10.52) | 3.69 (1.19, 11.42)  |
|   |  Q3 | 3.67 (1.21, 11.15) | 3.84 (1.22, 12.04)  |
|   |  Q4 | 8.39 (2.96, 23.77) | 8.24 (2.66, 25.50)  |
|  Heart-Related Hospitalization | Q2 | 1.25 (0.92, 1.68) | 1.23 (0.89, 1.69)  |
|   |  Q3 | 1.29 (0.95, 1.75) | 1.30 (0.94, 1.80)  |
|   |  Q4 | 1.51 (1.12, 2.04) | 1.44 (1.02, 2.03)  |

a NT-proBNP Q1 is the reference category for the hazard ratios presented for Q2, Q3, and Q4.
b The multivariable model was adjusted for age, sex, BMI, smoking, diabetes mellitus, hypertension, eGFR, and NYHA class. Only subjects with complete data for all covariates (n=816) were included

## Risk Stratification of Patients with Acute Coronary Syndrome (ACS)

The sponsor provided information to support the risk stratification of patients with ACS claim based on the following clinical guideline and peer-reviewed literature references:

- Rao SV, O'Donoghue ML, Ruel M, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of patients with acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation 2025;151:e771-e862.

K253539 - Page 7 of 8

{7}

- Jernberg T, Stridsberg M, Venge P, et al. N-terminal pro brain natriuretic peptide on admission for early risk stratification of patients with chest pain and no ST-segment elevation. J Am Coll Cardiol 2002;40(3):437-445.
- James SK, Lindahl B, Siegbahn A, et al. N-terminal pro-brain natriuretic peptide and other risk markers for the separate prediction of mortality and subsequent myocardial infarction in patients with unstable coronary artery disease; a Global Utilization of Strategies to Open occluded arteries (GUSTO)-IV substudy. Circulation 2003;108(3):275-281.

## D Expected Values/Reference Range:

Unchanged since K241176.

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K253539 - Page 8 of 8

---

**Source:** [https://fda.innolitics.com/device/K253539](https://fda.innolitics.com/device/K253539)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
