Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A)

K253537 · Bright Uro, Inc. · EXQ · Apr 16, 2026 · Gastroenterology, Urology

Device Facts

Record IDK253537
Device NameGlean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A)
ApplicantBright Uro, Inc.
Product CodeEXQ · Gastroenterology, Urology
Decision DateApr 16, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1620
Device ClassClass 2

Indications for Use

The Glean Urodynamics System is a urodynamics analyzer system that is intended to quantify the pressure and flow characteristics of the lower urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment.

Device Story

Multi-channel urodynamics system; measures bladder and abdominal pressure to calculate detrusor pressure; performs Uroflow, CMG, UPP, and Micturition Studies. System components: Bladder Sensor, Abdominal Sensor, Insertion Tool, Uroflowmeter. Sensors are single-use, silicone-housed, inserted into bladder and anorectal canal; store data locally; wirelessly transmit data to clinician tablet/app after removal. Clinician uses Glean Mobile App to prepare sensors, log symptoms, and download data; Glean Web App used for data viewing/analysis. Patient uses Glean Mobile App as digital voiding diary. Enables ambulatory urodynamic monitoring in clinical settings. Output assists clinicians in diagnosing incontinence, retention, and micturition disorders.

Clinical Evidence

Bench testing only. Evaluations included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), software verification and validation, cybersecurity risk management, human factors/usability engineering, packaging/shelf-life testing, and bioburden testing (USP <62>). Battery service life and performance verification tests were also conducted. All pre-determined acceptance criteria were met.

Technological Characteristics

Multi-channel urodynamics system. Components: silicone-housed bladder/abdominal sensors, insertion tool, uroflowmeter. Sensing principle: electronic pressure sensors (0-320 cm H2O range, 20 Hz sampling). Connectivity: wireless data transmission to mobile app. Software: Glean Mobile App (Clinician/Patient) and Glean Web App. Sterilization: not specified (packaging testing per ISO 11607).

Indications for Use

Indicated for adult patients requiring urodynamic testing, including Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP), and Micturition Studies, for the diagnosis of urinary incontinence, abnormal urinary retention, or neurological micturition disorders.

Regulatory Classification

Identification

A urodynamics measurement system is a device used to measure volume and pressure in the urinary bladder when it is filled through a catheter with carbon dioxide or water. The device controls the supply of carbon dioxide or water and may also record the electrical activity of the muscles associated with urination. The device system may include transducers, electronic signal conditioning and display equipment, a catheter withdrawal device to enable a urethral pressure profile to be obtained, and special catheters for urethral profilometry and electrodes for electromyography. This generic type of device includes the cystometric gas (carbon dioxide) device, the cystometric hydrualic device, and the electrical recording cystometer, but excludes any device that uses air to fill the bladder.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION April 16, 2026 Bright Uro, Inc. Suranjan Roychowdhury, Ph.D., RAC Chief Product Development Officer 3 Goddard Irvine, CA 92618 Re: K253537 Trade/Device Name: Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A) Regulation Number: 21 CFR§ 876.1620 Regulation Name: Urodynamics Measurement System Regulatory Class: II Product Code: EXQ, EXY Dated: November 18, 2025 Received: November 18, 2025 Dear Suranjan Roychowdhury: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253537 - Suranjan Roychowdhury, Ph.D., RAC Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K253537 - Suranjan Roychowdhury, Ph.D., RAC Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Negeen Haghighi -S for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253537 | ? | | Please provide the device trade name(s). | | ? | | Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Glean Urodynamics System Abdominal Sensor (GUS-1000-A) | | | | Please provide your Indications for Use below. | | ? | | The Glean Urodynamics System is a urodynamics analyzer system that is intended to quantify the pressure and flow characteristics of the lower urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K253537 Page 1 of 5 # 510(k) Summary ## 1. Submitter Information 510 (k) submitter: Bright Uro, Inc. Address: 3 Goddard Irvine, CA 92618, United States Applicant Contact Person: Dr. Suranjan Roychowdhury Chief Product Development Officer Phone: 949-216-0873 Email: suranjan@brighturo.com Preparation date: April 15, 2026 ## 2. Device Name Trade Name of the Device: Glean Urodynamics System Male Delivery System (GUS-1000-M); Glean Urodynamics System Female Delivery System (GUS-1000-F); Glean Urodynamics System Uroflowmeter (UFM-1000 (Uroflowmeter)); Abdominal Sensor (GUS-1000-A) Common Name: Urodynamics measurement system Classification Name: Cystometer, Electrical Recording Classification Regulation: 21 CFR 876.1620 Device Class: II Panel: Gastroenterology/Urology Product Code: EXQ, EXY ## 3. Predicate Devices | | 510(k)# | Trade Name | | --- | --- | --- | | Primary Predicate | K243052 | Glean Urodynamics System | | Secondary Predicate | 510(k) Exempt | Laborie Goby System | The predicate devices have not been subject to a design related recall. No reference devices were used in this submission. ## 4. Device Description The subject device is a modified Glean Urodynamics System. The specific modification to the subject device consists of adding an Abdominal Sensor component (GUS-1000-A) to the system. The Abdominal Sensor {5} K253537 Page 2 of 5 measures abdominal pressure and enables detrusor pressure to be calculated using the Glean Urodynamics System (GUS) to conduct multi-channel urodynamics studies. Addition of the Abdominal Sensor component enables the subject device to perform multi-channel urodynamics studies. The modified Glean Urodynamics System (GUS) is a multi-channel urodynamic system indicated for standard Urodynamic tests such as Uroflow (UF), Cystometrogram (CMG), Urethral Pressure Profile (UPP), and Micturition Studies (MS). The Glean Urodynamics System obtained market clearance through the original 510(k) Premarket Notification K243052. No changes were made to the intended use, indications for use, or technological characteristics of the predicate device's three physical components, the Bladder Sensor, Insertion Tool, and Uroflowmeter. The Glean Mobile App (Clinician) and Glean Web App software were modified to include use with the Abdominal Sensor. The Glean Mobile App (Patient) was not modified. The multi-channel GUS consists of four physical component elements: Bladder Sensor, Abdominal Sensor, Insertion Tool, and Uroflowmeter, as well as three software applications: Glean Mobile App (Clinician), Glean Mobile App (Patient), and Glean Web App. The patient may use the Glean Mobile App as a digital voiding diary, logging fluid input, leakage, urgency, and other urological symptoms. The clinician may use the Glean Mobile App to prepare the Sensor for insertion, log symptoms, and download data. The Glean Web App may be used by clinicians to view and analyze data. The Bladder Sensor is inserted through the urethra into the bladder using the Insertion Tool. Once inserted, the Sensor has a Removal String that hangs out of the urethra to enable removal of the Sensor. The Sensor may stay in the bladder for the entire duration of monitoring while collecting data. The Sensor stores data that may be wirelessly transmitted to the Glean Mobile App (Clinician) once it is removed from the body. The Abdominal Sensor is manually inserted into the anorectal canal. Once inserted, the Abdominal Sensor has a removal string that extends from the anus to enable removal of the Sensor. The Sensor may stay in the rectum for the entire duration of monitoring while collecting data. The Sensor stores data that may be wirelessly transmitted to the Glean Mobile App used by the clinician once it is removed from the body. The Uroflowmeter is used to measure voided volume and flow. The Glean Mobile App (Clinician) wirelessly receives data from the Uroflowmeter after the patient has completed a voiding cycle. ## 5. Indications For Use The Glean Urodynamics System is a urodynamics analyzer system that is intended to quantify the pressure and flow characteristics of the lower-urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment. {6} K253537 Page 3 of 5 # 6. Comparison of Technological Characteristics with Predicate Device | Technological/ Performance Characteristics | Subject Device | Primary Predicate Device | Secondary Predicate Device | | --- | --- | --- | --- | | | Glean Urodynamic System with Abdominal Sensor | Glean Urodynamic System | Laborie Goby System (Exempt) | | Intended Use / Indications for Use | The Glean Urodynamics System is a urodynamic analyzer system that is intended to quantify the pressure and flow characteristics of the lower-urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment. | The Glean Urodynamics System is a urodynamic analyzer system that is intended to quantify the pressure and flow characteristics of the lower-urinary tract. The system can be used in adult patients only to perform standard urodynamic tests such as Uroflow, Cystometrogram (CMG), Urethral Pressure Profile (UPP) and Micturition Studies. The major application of urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of micturition disorder. The system is intended to be used as medical diagnostic equipment. | The Goby™ system is a Urodynamic Analyzer system that is intended to quantify the pressure, flow, and EMG characteristics of the lower urinary tract. Using the available transducers, the system can be used for performing standard Urodynamic tests such as Uroflow, CMG, UPP, and Micturition Studies. The major application of Urodynamics is the diagnosis of uncontrolled loss of urine (incontinence), abnormal urinary retention, or neurological cases of Micturition disorder. The device is intended to be used as medical diagnostic equipment. | | Principles of Operation | An electronic pressure sensor is used to measure pressure in the bladder and in the abdomen. This sensor is connected to an electrical recording device that stores measurements from the sensor. Once monitoring is complete, the sensor is removed from the patient's body, measurements are transferred from the sensor to a tablet, and the sensor discarded. | An electronic pressure sensor is used to measure pressure in the bladder. This sensor is connected to an electrical recording device that stores measurements from the sensor. Once monitoring is complete, the sensor is removed from the patient's body, measurements are transferred from the sensor to a tablet, and the sensor discarded. | A catheter is used to measure pressure in the bladder and in the abdomen. The catheter is connected to a sensor outside the patient. The sensor is connected to an electrical recording device that stores measurements from the sensor. Once monitoring is complete, the catheter is removed from the patient's body and discarded. | | Anatomical location of use | The silicone housing with sensor and electronics are placed in the patient's bladder and anorectal cavity. | The silicone housing with sensor and electronics are placed in the patient's bladder. | The catheter tip is positioned in the bladder and anorectal cavity. The proximal portion of the catheter is taped to the patient's body. | | Mode of Operation | Continuous | Continuous | Continuous | | Pressure (Range) | 0 to 320 cm H2O | 0 to 320 cm H2O | 0 to 320 cm H2O (PLM001) -50 to 200 cm H2O (COM762) | | Pressure Sampling Rate | 20 Hz (minimum rate) | 20 Hz | 20 Hz | | Sensor/Catheter diameter | Less than 5.35 mm or 16 Fr (bladder) 17.5 mm maximum (abdominal) | Less than 5.35 mm or 16 Fr (bladder) | 7 Fr (bladder) 7 Fr (abdominal) | | Sensor/ Catheter Material | Silicone | Silicone | TPU-PVD | | Environment of Use | Clinic | Clinic | Clinic | | Enables ambulatory urodynamic | Yes | Yes | Yes | {7} K253537 Page 4 of 5 | Technological/ Performance Characteristics | Subject Device | Primary Predicate Device | Secondary Predicate Device | | --- | --- | --- | --- | | monitoring | Glean Urodynamic System with Abdominal Sensor | Glean Urodynamic System | Laborie Goby System (Exempt) | | Single Use Sensor/ Catheter | Yes | Yes | Yes | | Method of Removal | String | String | Catheter | | Integrated EMG | Yes | Yes | Yes | | Integrated Uroflowmetry | Yes | Yes | Yes | As evidenced by the above table, both the subject and the predicate devices have the same intended use, but the subject and predicate devices have different technological characteristics. Additionally, several types of performance testing were conducted on the subject device to evaluate its safety and effectiveness. Overall, it was established that the differences in technological characteristics between the subject and the predicate devices do not raise different questions of safety or effectiveness. ## 7. Non-Clinical Testing Below is a list of the evaluations and tests that were performed and successfully completed for the subject device per the specified guidance and standards: - Biocompatibility evaluation according to ISO 10993-1:2018 - Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process and FDA Guidance “Use of International Standard ISO 10993-1” (2016). - Electrical Safety testing according to IEC 60601-1: 2020 - Medical electrical equipment – Basic safety and essential performance - Electromagnetic Compatibility testing according to IEC 60601-1-2: 2020 - General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests - Software Verification and Validation Testing according to FDA’s Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices” - Cybersecurity risk management activities according to FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued September 27, 2023 - Human Factors Engineering/Usability Engineering (HFE/UE) processes were used as part of the design and development process for the subject device according to FDA Guidance, “Applying Human Factors and Usability Engineering to Medical Devices” - Packaging testing according to ISO 11607-1:2019/Amd. 1:2023- Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including Amendment 1 (2023)] and Shelf Life testing - Bioburden (for abdominal sensor only): USP &lt;62&gt; Microbiological Examination of Nonsterile Products: Tests for Specified Microorganisms Additionally, battery service life and performance verification test data were submitted to establish performance and durability of the subject device. {8} K253537 Page 5 of 5 All pre-determined acceptance criteria were met. ## 8. Conclusions Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...