Swoop® Portable MR Imaging® System

K253489 · Hyperfine, Inc. · LNH · Dec 12, 2025 · Radiology

Device Facts

Record IDK253489
Device NameSwoop® Portable MR Imaging® System
ApplicantHyperfine, Inc.
Product CodeLNH · Radiology
Decision DateDec 12, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesAI/ML, Pediatric

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
DWI image qualityDeep learning-based Advanced Gridding + FISTAMean CNR of Advanced Reconstruction > mean CNR of Linear Reconstruction (p < 0.05)Mean CNR of Advanced Reconstruction >= mean CNR of Linear Reconstruction——Contrast-to-Noise Ratio Validation: 45 DWI images (41 multi-direction, 4 single-direction) with visible pathology—
DWI image quality (reader improvement)Deep learning-based Advanced Gridding + FISTAMedian score >= 0 on Likert scale relative to Linear Reconstruction; >= 1 in at least one quality categoryMedian score of 2 in all categories——Advanced Reconstruction Image Validation: 34 sets of DWI images (30 multi-direction, 4 single-direction)4 (ABR-certified radiologists)

Indications for Use

The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Device Story

Swoop is a portable, ultra-low field (approx. 64 mT) permanent magnet MRI system for head imaging at the point of care (e.g., ER, ICU, rehab). Operated by healthcare professionals, it uses touch-screen interfaces to execute exams and view images. The system acquires MRI data (T1W, T2W, FLAIR, DWI) and employs a deep learning-based reconstruction algorithm to enhance image quality, reduce noise, and improve sharpness. Output images are used by physicians to assist in clinical diagnosis. The device supports cloud storage and DICOM output. Benefits include providing diagnostic-quality head imaging in environments where traditional high-field MRI is impractical.

Clinical Evidence

No clinical trials; performance validated via bench testing and expert reader studies. Advanced reconstruction models were validated using 34 sets of DWI images (102 total images) reviewed by four ABR-certified radiologists. Metrics included NMSE, SSIM, and CNR. Results showed statistically significant improvement in CNR (p<0.05) and median Likert scores of 2 (improved quality) across all categories (noise, sharpness, contrast, geometric fidelity, artifact) compared to linear reconstruction. Dataset included diverse pathologies (stroke, tumor, TBI, ICH) across adult age groups.

Technological Characteristics

Portable, ultra-low field (64.9 mT nominal) permanent magnet MRI. Features transmit/receive linear RF coils. Patient weight capacity 1.6kg-200kg. Connectivity includes DICOM output and cloud storage. Software utilizes deep learning-based reconstruction for T1W, T2W, FLAIR, and DWI sequences. Complies with NEMA MS standards for image performance and IEC 60601 series for electrical safety/EMC. Biocompatibility per ISO 10993.

Indications for Use

Indicated for adult and pediatric patients (≥ 0 years) requiring head imaging where full diagnostic examination is not clinically practical.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD &amp; DRUG ADMINISTRATION December 12, 2025 Hyperfine, Inc. Kristen Evenson Sr. Manager, Regulatory Affairs 351 New Whitfield St Guilford, Connecticut 06437 Re: K253489 Trade/Device Name: Swoop® Portable MR Imaging® System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, MOS Dated: October 23, 2025 Received: October 24, 2025 Dear Kristen Evenson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K253489 - Kristen Evenson Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K253489 - Kristen Evenson Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253489 | ? | | Please provide the device trade name(s). | | ? | | Swoop® Portable MR Imaging® System | | | | Please provide your Indications for Use below. | | ? | | The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} K253489 # HYPERFINE ## 510(k) Summary Swoop® Portable MR Imaging® System ## 510(k) SUBMITTER Company Name: Hyperfine, Inc. Company Address: 351 New Whitfield St Guilford, CT 06437 ## CONTACT Name: Kristen Evenson Telephone: (612) 251-3030 Email: kevenson@hyperfine.io Date Prepared: December 5, 2025 ## DEVICE IDENTIFICATION Trade Name: Swoop® Portable MR Imaging® System Common Name: Magnetic Resonance Imaging Regulation Number: 21 CFR 892.1000 Classification Name: System, Nuclear Magnetic Resonance Imaging Coil, Magnetic Resonance, Specialty Product Code: LNH; MOS Regulatory Class: Class II ## PREDICATE DEVICE INFORMATION The subject Swoop Portable MR Imaging System is substantially equivalent to the primary predicate Swoop System (K250236), and the secondary predicate Swoop System (K251276). ## DEVICE DESCRIPTION The Swoop System is portable, ultra-low field MRI device that enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. The main interface is a commercial off-the-shelf device that is used for operating the system, providing access to patient data, exam setup, exam execution, viewing MRI image data for quality control purposes, and cloud storage interactions. The system can generate MRI data sets with a broad range of contrasts. The Swoop system user interface includes touch screen menus, controls, indicators, and navigation icons that allow the {6} operator to control the system and to view imagery. The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, FLAIR, and DWI sequences. The subject Swoop System described in this submission includes software modifications related to the pulse sequences. # INDICATIONS FOR USE The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. # SUBSTANTIAL EQUIVALENCE DISCUSSION The table below compares the subject device to the predicate. | Specification | Subject Swoop Portable MR Imaging System | Primary Predicate Swoop Portable MR Imaging System Model 2 (K250236) | Secondary Predicate Swoop Portable MR Imaging System Model 1 (K251276) | | --- | --- | --- | --- | | Intended Use/Indications for Use: | The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. | Same | Same | | Patient Population: | Adult and pediatric patients (≥ 0 years) | Same | Same | | Anatomical Sites: | Head | Same | Same | | Environment of Use: | At the point of care in professional health care facilities such as emergency rooms, intensive/critical care units, hospitals, outpatient, or rehabilitation centers. | Same | Same | | Energy Used and/or delivered: | Magnetic Resonance | Same | Same | | Magnet: | | | | {7} | Specification | Subject Swoop Portable MR Imaging System | Primary Predicate Swoop Portable MR Imaging System Model 2 (K250236) | Secondary Predicate Swoop Portable MR Imaging System Model 1 (K251276) | | --- | --- | --- | --- | | Field Strength | Model 1 Swoop System 63.3 ± 2.0 mT Model 2 Swoop System 64.9 mT (nominal) | Same as Model 2 of subject device | Same as Model 1 of subject device | | Type | Permanent magnet | Same | Same | | Patient accessible bore size | Model 1 Swoop System 24.0 in. width, 12.4 in. height Model 2 Swoop System 36.0 in. width, 13.4 in. height | Same as Model 2 of subject device | Same as Model 1 of subject device | | Magnet weight | Model 1 Swoop System 705 lbs Model 2 Swoop System 712 lbs | Same as Model 2 of subject device | Same as Model 1 of subject device | | Gradient System: | | | | | Maximum gradient amplitude | Model 1 Swoop System X: 24 mT/m, Y: 23 mT/m, Z: 39 mT/m Model 2 Swoop System X: 33.9 mT/m, Y: 33.2 mT/m, Z: 66.2 mT/m | Same as Model 2 of subject device | Same as Model 1 of subject device | | Rise Time | Model 1 Swoop System X: 2.1 ms, Y: 2.0 ms, Z: 3.8 ms Model 2 Swoop System X: 1.8 ms, Y: 1.8 ms, Z: 5.1 ms | Same as Model 2 of subject device | Same as Model 1 of subject device | | Slew Rate | Model 1 Swoop System X: 24 T/m/s, Y: 22 T/m/s, Z: 21 T/m/s Model 2 Swoop System X: 18.8 T/m/s, Y: 18.4 T/m/s, Z: 13.0 T/m/s | Same as Model 2 of subject device | Same as Model 1 of subject device | | RF Coils: | | | | | Coil Type | Transmit/receive | Same | Same | | Coil Design | Linear | Same | Same | | Other: | | | | | Patient Weight Capacity | 1.6kg-200 kg | Same | Same | {8} | Specification | Subject Swoop Portable MR Imaging System | Primary Predicate Swoop Portable MR Imaging System Model 2 (K250236) | Secondary Predicate Swoop Portable MR Imaging System Model 1 (K251276) | | --- | --- | --- | --- | | Operation Temperature | 15-30 C | Same | Same | | Warm Up Time | <3 minutes | Same | Same | | Temperature Control | No | Same | Same | | Humidity Control | No | Same | Same | | Sequences: | | | | | T1W sequences | T1 (Standard), T1 (Gray/White) Advanced Gridding reconstruction | Same | Same | | T2W sequences | T2, T2 (Fast) Advanced Gridding reconstruction | Same | Same | | FLAIR sequences | Model 1 Swoop System FLAIR Advanced Gridding reconstruction Model 2 Swoop System FLAIR, FLAIR (Fast) Advanced Gridding reconstruction | Same as Model 2 of subject device | Same as Model 1 of subject device | | DWI sequences | Model 1 Swoop System Single Direction DWI/ADC Multi-direction DWI/ADC Advanced Gridding + FISTA Model 2 Swoop System Single Direction DWI/ADC, Multi-direction DWI/ADC Advanced Gridding + FISTA | Single Direction DWI/ADC, Single Direction DWI/ADC (Fast) Advanced Gridding + FISTA | Single Direction DWI/ADC Advanced Gridding + FISTA | | Image Post-Processing (All sequences) | Advanced Denoising Image orientation transform Geometric distortion correction Receive coil intensity correction Advanced Interpolation ADC/Trace output (DWI) DICOM output | Advanced Denoising Image orientation transform Geometric distortion correction Receive coil intensity correction Advanced Interpolation ADC output (DWI) DICOM output | Advanced Denoising Image orientation transform Geometric distortion correction Receive coil intensity correction Advanced Interpolation ADC output (DWI) DICOM output | {9} The subject device and the predicate device have the same intended use, operating principles, and similar technological characteristics. There are minor differences between the subject device and the predicate in pulse sequences. These differences do not raise new questions of safety and efficacy as compared to the predicate. # NON-CLINICAL PERFORMANCE As part of demonstrating substantial equivalence to the predicate, a risk-based assessment was completed to identify the risks associated with the modifications. Based on the risk assessment, the following testing was performed. The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence. | Test | Test Description | Applicable Standard(s) | | --- | --- | --- | | Software Verification | Software verification testing in accordance with the design requirements to ensure that the software requirements were met. | IEC 62304:2016 FDA Guidance, “Content of Premarket Submissions for Device Software Functions” | | Image Performance | Testing to verify the subject device meets all image quality criteria. | NEMA MS 1-2008 (R2020) NEMA MS 3-2008 (R2020) NEMA MS 9-2008 (R2020) NEMA MS 12-2016 American College of Radiology standards for named sequences | | Cybersecurity | Testing to verify cybersecurity controls and management. | FDA Guidance, “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” | | Software Validation | Validation to ensure the subject device meets user needs and performs as intended. | FDA Guidance, “Content of Premarket Submissions for Device Software Functions” | The following testing was leveraged from the predicate device. Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing. | Test | Test Description | Applicable Standard(s) | | --- | --- | --- | | Biocompatibility | Biocompatibility testing of patient-contacting materials. | ISO 10993-1:2018 ISO 10993-5:2009 ISO 10993-10:2010 | | Cleaning/Disinfection | Cleaning and disinfection validation of patient-contacting materials. | FDA Guidance, “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling” ISO 17664:2017 ASTM F3208-17 | | Safety | Electrical Safety, EMC, and Essential Performance testing. | ANSI/AAMI ES 60601-1:2005/(R)2012 IEC 60601-1-2:2014 IEC 60601-1-6:2013 | {10} | Performance | Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems. | NEMA MS 8-2016 | | --- | --- | --- | # ADVANCED RECONSTRUCTION PERFORMANCE ANALYSIS AND VALIDATION Performance analysis and validation of the subject device Advanced Reconstruction models was performed for the updated DWI sequence. No images from the test dataset were used for model training. Only the DWI test dataset was updated as no changes were made to T1, T2, or FLAIR sequences, and no changes were made to any of the advanced reconstruction models. Performance analysis and validation of the T1/T2/FLAIR models were leveraged from the predicate device. The test summaries below describe the updated DWI test datasets. The test datasets for the T1, T2, and FLAIR are described in the previous clearances. In all cases, models are trained and validated with MRI data and images as the only inputs and outputs; there are no confounding factors and clinical subgroups are not defined or considered. While gender and age are available for most subjects, age, gender, ethnic background, and pathology are not expected to influence model architecture. # Performance Analysis: Study Design: Advanced Reconstruction was assessed for robustness, stability, and generalizability over a variety of subjects, design parameters, artifacts, and scan conditions using reference-based metrics. A set of images including Swoop data, high field images, and synthetic contrast images, was used as ground truth target images. Test input data (synthetic k-space generated from the target images) was reconstructed using both Advanced and Linear Reconstruction, and the similarity to the original ground truth image was compared between the two reconstruction methods. Reconstruction outputs with motion and zipper artifacts were qualitatively assessed. # Reference Standard and Metrics: Normalized mean squared error (NMSE) and structural similarity index (SSIM) were used to compare the ability of Advanced Reconstruction to reproduce the ground truth image compared to Linear Reconstruction. # Dataset and Sample Size per Model: The Swoop DWI dataset was updated to include multi-direction DWI images. None of these test images were used in model training. The demographics of the DWI subset are shown below. {11} | Model /SequenceGroup | #Patients | #Images | Demographics | | --- | --- | --- | --- | | DWI | 8 | 31 | Gender: | | Female | Male | Unknown | | 13% | 87% | 0% | | Age: | | 0-2 | 2-18 | 18-35 | 35-60 | 60+ | Unknown | | 0% | 0% | 25% | 0% | 75% | 0% | | Ethnicity data not recorded. | | Number of sites: 6 | | Equipment: | | Model 1 Swoop System (V1.9) | Model 2 Swoop System | | 13% | 87% | | Included pathology: Post-resection Tumor, ICH, Stroke, TBI, Post-Craniotomy. | # Study Results: NMSE and SSIM with the included multi-direction DWI dataset were similar to previous results. For all models and all test datasets NMSE was reduced and SSIM was improved for Advanced Reconstruction test images compared to Linear Reconstruction test images. # Contrast-to-Noise Ratio Validation Study Design: Regions of interest (ROI) encompassing pathologies were annotated, and the annotations were reviewed for accuracy by an American Board of Radiology (ABR) certified radiologist. The contrast-to-noise of hyper- and hypo- intense pathologies were measured with respect to healthy white matter tissue from the same image. The inclusion criterion for images used for this study was at least one visible pathology. Reference Standard and Metrics: Linear Reconstruction was used as the reference standard for the comparison. Contrast-to-Noise Ratio (CNR) between pathology and healthy tissues was measured to quantify how accurately pathology features are preserved by Advanced Reconstruction. The mean CNR of Advanced Reconstruction was required to be greater than the mean CNR of the baseline Linear Reconstruction at statistical significance level of 0.05 for each sequence type. {12} # Dataset and Sample Size: 45 DWI images (41 multi-direction, 4 single-direction) were included for lesion annotation. Inclusion criteria were that the images had at least one visible pathology. The demographics of the DWI dataset are shown below. | Patients | 12 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Images | 45 | | | | | | | ROIs | 145 | | | | | | | Demographics and other Variability | Gender: | | | | | | | | Female | | Male | | Unknown | | | | 25% | | 42% | | 33% | | | | Age: | | | | | | | | 0-2 | 2-18 | 18-35 | 35-60 | 60+ | Unknown | | | 0% | 0% | 8% | 42% | 33% | 17% | | | Ethnicity data not recorded. Number of sites: 5 | | | | | | | | Equipment type: | | | | | | | | Model 1 Swoop System (V1.8) | | Model 1 Swoop System (V1.9) | | Model 2 Swoop System | | | | 22% | | 22% | | 56% | | | | Included pathology: Acute/Subacute Stroke Stroke, Tumor, Post-operative Tumor, Cerebellar metastatic disease, white matter disease, Acute/Subacute Infarct | | | | | | Study Results: In all cases, CNR of Advanced Reconstruction was greater than or equal to Linear Reconstruction for both hyper- and hypo-intense pathologies. The study result demonstrates that Advanced Reconstruction does not unexpectedly modify, remove, or reduce the contrast of pathology features. # Advanced Reconstruction Image Validation Study Design: Four external, ABR-certified radiologists representing clinical users were asked to review side-by-side clinical image sets taken with the subject Swoop System, reconstructed with both Advanced and Linear Reconstruction. The reviewers rated the images using a five-point scale for image quality and the consistency of diagnosis using both methods in the categories of noise, sharpness, contrast, geometric fidelity, artifact, and overall image quality. # Reference Standard and Metrics: Linear Reconstruction was used as the reference standard for the comparison. Advanced Reconstruction was required to perform at least as well as Linear Reconstruction in all categories (median score $\geq 0$ on Likert scale) and perform better ( $\geq 1$ on Likert scale) in at least one of the quality-based categories. {13} # Dataset and Sample size: 34 sets of DWI images (30 multi-direction, 4 single-direction) were rated. A set consisted of DWI $b = 0$ , DWI (trace-weighted or single direction), and ADC. | Patients | 34 | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Images | 102 | | | | | | | Demographics and other Variability | Gender: | | | | | | | | Female | | Male | | Unknown | | | | 41% | | 35% | | 24% | | | | Age: | | | | | | | | 0-2 | 2-18 | 18-35 | 35-60 | 60+ | unknown* | | | 0% | 0% | 15% | 26% | 32% | 26% | | | *anonymized | | | | | | | | Ethnicity data not recorded. | | | | | | | | Number of sites: 8 | | | | | | | | Equipment type: | | | | | | | | Model 1 Swoop System (V1.8) | | Model 1 Swoop System (V1.9) | | Model 2 Swoop System | | | | 20% | | 10% | | 70% | | | | Included pathology: Acute Stroke, Subacute Stroke, Multiple Sclerosis, White matter disease, Metastatic disease, Post-operative glioma, Tumor, Hydrocephalus, ICH, IVH, | | | | | | Test Results: Advanced Reconstruction achieved a median score of 2 (the most positive rating scale value) in all categories. This scoring indicates reviewers found Advanced Reconstruction improved image quality while maintaining diagnostic consistency relative to Linear Reconstruction. # CONCLUSION Based on the intended use, technological characteristics, performance results, and comparison to the predicate, the subject Swoop Portable MR Imaging System has been shown to be substantially equivalent to the predicate device identified in this submission and does not present any new issues of safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...