K253472 · Mighty Oak Medical · PQC · Feb 25, 2026 · Orthopedic
Device Facts
Record ID
K253472
Device Name
FIREFLY® Pedicle Screw Navigation Guide
Applicant
Mighty Oak Medical
Product Code
PQC · Orthopedic
Decision Date
Feb 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Indications for Use
The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below: • Pedicle screw’s shank is straight along its longitudinal axis (i.e. not curved) • Pedicle screw’s major and minor thread diameters are centered about the longitudinal axis • Pedicle screw’s longitudinal axis matches the direction of insertion • Pedicle screw is intended to be inserted into a pilot hole • Pedicle screw’s diameter is larger than the pilot hole created with FIREFLY® • Patient’s pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan • Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide o Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity o Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be guided to tap pilot holes o All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system’s instructions for use This device is intended for single use only.
Device Story
Patient-specific surgical navigation guide; assists in accurate pedicle screw placement during posterior spinal fixation. Input: preoperative CT or MRI-derived synthetic CT imaging data. Operation: surgeon uses imaging to define trajectories; system generates patient-specific guides via additive manufacturing; guides fit onto patient anatomy to direct surgical instruments (sounding probes, taps) for pilot hole creation. Output: physical guide for intraoperative use. Context: used in OR by surgeons; guides removed before final screw insertion. Benefit: improved accuracy of pedicle screw placement via patient-specific anatomical alignment.
Clinical Evidence
Bench testing only. Quantitative and qualitative validations performed comparing MRI-derived synthetic CT inputs to standard CT scan inputs. Metrics included segmentation output quantitative comparison and presurgical plan qualitative comparison. Results confirmed acceptance criteria were met, demonstrating equivalent performance.
Technological Characteristics
Patient-specific guides manufactured via additive manufacturing (epoxy resin or polymer powder). Non-patient-specific components manufactured per ASTM F899 or F136. Single-use, non-sterile. Operates as a mechanical guide based on preoperative imaging-derived anatomical landmarks.
Indications for Use
Indicated for patients undergoing non-cervical open, posterior spinal fixation (T1-S2/S2AI and ilium) requiring fusion, where pedicle screws are used. Requires patient pedicles to be dimensionally adequate for screw placement as determined by preoperative imaging.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
February 25, 2026
Mighty Oak Medical
Mark Wylie
VP Quality and Regulatory
750 W. Hampden Avenue, Suite 120
Englewood, Colorado 80110
Re: K253472
Trade/Device Name: FIREFLY® Pedicle Screw Navigation Guide
Regulation Number: 21 CFR 888.3070
Regulation Name: Thoracolumbosacral Pedicle Screw System
Regulatory Class: Class II
Product Code: PQC
Dated: October 6, 2025
Received: October 10, 2025
Dear Mark Wylie:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253472 - Mark Wylie
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253472 - Mark Wylie
Page 3
Sincerely,
EILEEN
CADEL -S
for
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253472 | ? |
| Please provide the device trade name(s). | | ? |
| FIREFLY® Pedicle Screw Navigation Guide | | |
| Please provide your Indications for Use below. | | ? |
| The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below:
• Pedicle screw’s shank is straight along its longitudinal axis (i.e. not curved)
• Pedicle screw’s major and minor thread diameters are centered about the longitudinal axis
• Pedicle screw’s longitudinal axis matches the direction of insertion
• Pedicle screw is intended to be inserted into a pilot hole
• Pedicle screw’s diameter is larger than the pilot hole created with FIREFLY®
• Patient’s pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan
• Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide
o Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity
o Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be guided to tap pilot holes
o All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system’s instructions for use
This device is intended for single use only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K253472 - Page 1 of 5
# 510(K) SUMMARY

# MIGHTY OAK
# MEDICAL
FIREFLY® Pedicle Screw Navigation Guide
## Submitter:
Mighty Oak Medical
750 W. Hampden Ave., Suite 120
Englewood, CO 80110
(720) 398-9703
Contact: Mark A. Wylie, VP of Quality and Regulatory
Date Prepared: 06OCT2025
## Device
Trade Name: FIREFLY® Pedicle Screw Navigation Guide
Common Name: Pedicle Screw Placement Guide
Device Classification: Class II
Regulation, Name: 21 CFR 888.3070, Thoracolumbosacral pedicle screw system
Device Product Code: PQC
## Type of 510(k)
Original Submission: Traditional.
## Predicate Device(s):
| 510(k) | Product Code | Trade Name | Manufacturer |
| --- | --- | --- | --- |
| Primary Predicate Device | | | |
| K181399 | PQC | FIREFLY® Pedicle Screw Navigation Guide | Mighty Oak Medical |
## Introduction
This 510(k) filing seeks to expand the FIREFLY® indications for use with MRI-derived synthetic CT scans as an input. The FIREFLY® design is the same as was originally cleared in K143222, K162419, and K181399.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging or MRI-Derived Synthetic CT data.
{5}
K253472 - Page 2 of 5
# Description
The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT or MRI-Derived Synthetic CT imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models may also be provided.
# Indications for Use
The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below:
- Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved)
- Pedicle screw's major and minor thread diameters are centered about the longitudinal axis
- Pedicle screw's longitudinal axis matches the direction of insertion
- Pedicle screw is intended to be inserted into a pilot hole
- Pedicle screw's diameter is larger than the pilot hole created with FIREFLY®
- Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan
- Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide
- Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity
- Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be guided to tap pilot holes
- All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use
This device is intended for single use only.
# Materials
The material used to manufacture patient-specific components is an epoxy resin for use in stereolithography systems or polymer powder for use in multi-jet fusion systems. Patient contacting materials used for non-patient-specific components are manufactured in accordance with ASTM F899 or F136.
# Performance Data
Quantitative and Qualitative validations were completed to demonstrate that MRI-Derived synthetic CT inputs generate equivalent FIREFLY Pedicle Screw Navigation Guide system components when compared to CT scan inputs, including: Segmentation Output Quantitative Comparison, Presurgical Plan Qualitative Comparison, and MRI-derived sCT Qualification. The results demonstrated that the acceptance criteria were met and that the FIREFLY® Pedicle Screw Navigation Guide's performance is adequate to perform as intended.
{6}
K253472 - Page 3 of 5
# Technological Characteristics
The subject FIREFLY® Pedicle Screw Navigation Guide possesses the same technological characteristics as the predicate device. These include:
- Performance
- Manufacturing process
- Sterilization
- Biocompatible materials
- Basic design
Technological characteristics which are different have been supported with descriptive information and/or performance data. Therefore the fundamental scientific technology of FIREFLY® is the same as the previously cleared device.
# Substantial Equivalence Discussion
The table below provides a descriptive comparison of the similarities and differences for the subject and predicate device.
Table 1 - Comparison of Devices
| Device--Features↓ | FIREFLY Pedicle Screw Navigation Guide (This Submission) | FIREFLY Pedicle Screw Navigation Guide | Equivalency |
| --- | --- | --- | --- |
| Premarket notification | K253472 (Current Submission) | K181399 | N/A |
| Manufacturer | Mighty Oak Medical | Mighty Oak Medical | No change |
| Device Description | The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's preoperative imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models may also be provided. | The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models may also be provided. | Addition of MRI-Derived Synthetic CT imaging data as an input |
{7}
K253472 - Page 4 of 5
| Device→ Features↓ | FIREFLY Pedicle Screw Navigation Guide (This Submission) | FIREFLY Pedicle Screw Navigation Guide | Equivalency |
| --- | --- | --- | --- |
| Indications for Use Statement | The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below:
• Pedicle screw’s shank is straight along its longitudinal axis (i.e. not curved)
• Pedicle screw’s major and minor thread diameters are centered about the longitudinal axis
• Pedicle screw’s longitudinal axis matches the direction of insertion
• Pedicle screw is intended to be inserted into a pilot hole
• Pedicle screw’s diameter is larger than the pilot hole created with FIREFLY®
• Patient’s pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan
• Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide
o Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity
o Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be guided to tap pilot holes
o All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system’s instructions for use
This device is intended for single use only. | The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below:
• Pedicle screw’s shank is straight along its longitudinal axis (i.e. not curved)
• Pedicle screw’s major and minor thread diameters are centered about the longitudinal axis
• Pedicle screw’s longitudinal axis matches the direction of insertion
• Pedicle screw is intended to be inserted into a pilot hole
• Pedicle screw’s diameter is larger than the pilot hole created with FIREFLY®
• Patient’s pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan
• Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide
o Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity
o Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be guided to tap pilot holes
o All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system’s instructions for use
This device is intended for single use only. | No change |
| General intended use | A pedicle screw placement guide is intended to be used as a surgical instrument to assist in the intra-operative placement of pedicle screws that are used to anchor pedicle screw spinal systems onto the posterior elements of the spine. The device utilizes anatomic landmarks that are identifiable on preoperative imaging scans. | A pedicle screw placement guide is intended to be used as a surgical instrument to assist in the intra-operative placement of pedicle screws that are used to anchor pedicle screw spinal systems onto the posterior elements of the spine. The device utilizes anatomic landmarks that are identifiable on CT imaging scans. | Update imaging scan language to be applicable to CT or MRI-Derived Synthetic CT imaging data |
| Product Classification | Pedicle Screw Placement Guide | Pedicle Screw Placement Guide | No change |
| Regulatory Class | Class II | Class II | No change |
| Product Code | PQC | PQC | No change |
{8}
K253472 - Page 5 of 5
| Device→ Features | FIREFLY Pedicle Screw Navigation Guide (This Submission) | FIREFLY Pedicle Screw Navigation Guide | Equivalency |
| --- | --- | --- | --- |
| Anatomic location | Posterior Spine (T1-S2/S2AI and ilium) | Posterior Spine (T1-S2/S2AI and ilium) | No change |
| Technological characteristics | Assists in placement of any pedicle screws that meet conditions specified in the indications for use.
All Guides T1-S2/S2AI and ilium are included and remain unchanged from K181399 as it relates to form, fit and function. | Assists in placement of any pedicle screws that meet conditions specified in the indications for use. | No change |
| Manufacturing process | Additive manufacturing | Additive manufacturing | No change |
| Basic design | Patient-specific surgical guide | Patient-specific surgical guide | No change |
| Contact duration | Limited (< 24 hours) | Limited (< 24 hours) | No change |
| Sterilization | Non-sterile | Non-sterile | No change |
| Method of use | Pedicle preparation for pedicle screw placement | Pedicle preparation for pedicle screw placement | No change |
| Image Acquisition Modality | CT Scan and MRI-Derived synthetic CT Scan | CT Scan | Addition of MRI-Derived Synthetic CT imaging data as an input |
## Conclusion
The FIREFLY® Pedicle Screw Navigation Guide possesses the same intended use and technological characteristics as the predicate device. Therefore the FIREFLY® Pedicle Screw Navigation Guide is substantially equivalent for its intended use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.