K253461 · Meta Systems Co., Ltd. · EBZ · Oct 14, 2025 · Dental
Device Facts
Record ID
K253461
Device Name
Demi Pro
Applicant
Meta Systems Co., Ltd.
Product Code
EBZ · Dental
Decision Date
Oct 14, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6070
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Demi Pro is an LED visible light curing unit intended for polymerization of light-cured materials by dental professionals. It is also intended to provide illumination to aid in visualization during oral procedures.
Device Story
Demi Pro is a pen-style, cordless, rechargeable lithium-ion battery-powered LED curing light. It emits high-intensity blue light (380–515 nm) to activate photo-initiators in dental resins, initiating polymerization. It features multiple modes (Standard, High, Ortho) with varying intensities (up to 2500 mW/cm²) and duty cycles, plus a low-intensity diagnostic mode for oral visualization. Used by dental professionals in clinical settings, the device is operated via a two-button control system with visual status indicators. An optional intraoral transillumination tip is available. The device benefits patients by enabling efficient curing of restorative materials and adhesives, and improving clinical visibility during procedures. The housing is anodized aluminum with thermoplastic/polycarbonate optical components and a magnetized lens attachment system.
Clinical Evidence
No clinical trials were conducted. Substantial equivalence is supported by bench testing comparing wavelength, light intensity, and polymerization depth against the predicate VALO X. Biocompatibility was evaluated per ISO 10993-1:2018 and FDA guidance.
Technological Characteristics
LED-based photopolymerization; wavelength 380–515 nm; peak wavelengths 400 nm, 460 nm. Powered by 3.7V Li-ion battery. Housing: anodized aluminum; optical components: TPX and polycarbonate blends. Features magnetized lens attachment system. IPX4 waterproof rating (handpiece). Includes LED module temperature control. Operates via two-button control system.
Indications for Use
Indicated for dental professionals to polymerize light-cured dental materials and provide illumination for visualization during oral procedures.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 14, 2025
Meta Systems Co., Ltd.
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120A
Madison, New Jersey 07940
Re: K253461
Trade/Device Name: Demi Pro
Regulation Number: 21 CFR 872.6070
Regulation Name: Ultraviolet activator for polymerization
Regulatory Class: Class II
Product Code: EBZ
Dated: October 7, 2025
Received: October 7, 2025
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253461 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K253461 - Dave Yungvirt
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253461 | |
| Device Name Demi Pro | |
| Indications for Use (Describe) The Demi Pro is an LED visible light curing unit intended for polymerization of light-cured materials by dental professionals. It is also intended to provide illumination to aid in visualization during oral procedures. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Page 1
# 510(k) Summary K253461
10 October, 2025
## Contact Details
| Applicant Name | META SYSTEMS CO., LTD. |
| --- | --- |
| Applicant Address | #1214-18, Sicox tower 12F, 484, Dunchon-daero Jungwon-gu Seongnam-si Gyeonggi-do 13229 Korea, South |
| Applicant Contact Telephone | 82 31 731 7377 |
| Applicant Contact | Ray Jeon |
| Applicant Contact Email | bkjeon@meta-biomed.com |
## Device Name
| Device Trade Name | Demi Pro |
| --- | --- |
| Common Name | Ultraviolet activator for polymerization |
| Classification Name | Activator, Ultraviolet, For Polymerization |
| Regulation Number | 872.6070 |
| Product Code(s) | EBZ |
## Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K220471 | VALO X | EBZ |
| K170529 | Delight Ortho | EBZ |
## Device Description Summary
General information:
The Demi Pro is a pen-style cordless curing light used for the polymerization of light cured dental material and is powered by a rechargeable lithium-ion battery pack. An intraoral transillumination tip is available as an additional accessory. It is also intended to provide illumination to aid in visualization during oral procedures.
Principle of operation:
The LED Light Curing System operates by emitting high-intensity blue light from a powerful LED to activate the photo initiators contained in the resin, thereby initiating polymerization (curing).
{5}
Page 2
Wavelength range
- Usable wavelength range: 380 nm – 515 nm
- Peak wavelengths: 400 nm, 460 nm
Light intensity levels
- PLS STD: 1100 -1330 mW/cm²
- PLS HIGH: 2200 – 2500 mW/cm²
- ORTHO: 2500 mW/cm²
- DIAGNOSIS: 20 – 24 mW
Duty cycle
- PLS STD: 20 sec ON / 60 sec OFF
- PLS HIGH: 5 sec ON / 60 sec OFF
- ORTHO: 3 sec x 3 times ON / 60 sec OFF
- DIAGNOSIS: 60 sec ON / 60 sec OFF
Waterproof rating: IPX4 (Applies to Handpiece)
Intended Use/Indications for Use
The Demi Pro is an LED visible light curing unit intended for polymerization of light-cured materials by dental professionals. It is also intended to provide illumination to aid in visualization during oral procedures.
Indications for Use Comparison
Predicate Device #1 [VALO X, K220471]
VALO X curing light is a source of illumination for curing photo-activated dental restorative materials and adhesives. It is also intended to provide illumination to aid in visualization during oral procedures.
VALO X curing light accessory/diffusor lenses are not intended for complete cure of photo-activated materials and adhesives.
Predicate Device #2 [Delight Ortho, K170529]
For light curing polymerization of dental composites, luting materials, cements and other
{6}
light cured materials.
## Technological Comparison
The Demi Pro curing light demonstrates technological characteristics that are substantially equivalent to those of the predicate device, VALO X and Delight Ortho. All three devices are designed for the polymerization of light-cured dental materials using high-intensity LED light sources operating within a comparable wavelength range (380–515 nm). The curing modes provided by Demi Pro (Standard, High, and Ortho) are similar in function and output intensity to those of the predicate device, ensuring equivalent clinical performance.
All three devices feature an ergonomic design and portability and are intended for use by qualified dental professionals.
In particular, both the Demi Pro and VALO X are equipped with a two-button control system with visual indicators, enabling the user to select the curing mode and monitor the device status. In addition, each device includes a diagnostic light mode that utilizes similar light output and spectral characteristics to enhance clinical visualization.
In terms of construction, the materials used in both Demi Pro and VALO X are comparable. The main housing is made of anodized aluminum, and the optical components are constructed from similar thermoplastics and optical polymers, including TPX and polycarbonate blends. Magnetized lens attachment systems are present in both devices to facilitate easy interchange of accessories.
The Demi Pro device is powered by a rechargeable 3.7V lithium-ion battery with a capacity similar to those used in the VALO X and Delight Ortho devices. While VALO X offers additional AC power capability, the energy sources are functionally comparable for their intended use.
The fundamental operating principle of Demi Pro and VALO X is based on LED-driven photopolymerization, and both include a range of accessories that support clinical versatility, such as curing tips and transillumination aids.
In summary, the Demi Pro shares the key technological characteristics of the predicate devices in terms of design, materials, operating principle, power source, and performance specifications.
These similarities support the conclusion that the Demi Pro is substantially equivalent to the predicate devices (VALO X and Delight Ortho).
Page 3
{7}
Page 4
# Non-Clinical and/or Clinical Test Summary & Conclusions
Guidance for Industry and FDA Staff - Dental Curing Lights - Premarket Notification [510(k)] Submissions
The test results showed that Demi Pro has the same wavelength band, similar light intensity, and output characteristics as Valo X, and the polymerization depth measurement results also confirmed equivalent performance.
Both products use LED as a light source, adopt Li-ion batteries as a power source, and the LED module temperature control function to prevent risks due to temperature rise is also applied in the same way.
Therefore, Demi Pro is considered a medical device that is substantially equivalent in performance to the legally marketed predicate device [VALO X, K K220471].
No clinical trials have been conducted.
Study results and considering the provisions of ISO 10993-1:2018, as well as FDA Guidance "Use of International Standard ISO 10993, 'Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process", dated 20. September 2020, it is concluded that the device Demi Pro is evaluated as biocompatible if manufactured appropriately and applied in compliance with its intended purpose as outlined in the manufacturer's Instructions for Use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.