K253446 · Dk Medical Systems Co., Ltd. · KPR · Apr 10, 2026 · Radiology
Device Facts
Record ID
K253446
Device Name
AeroDR TX c02
Applicant
Dk Medical Systems Co., Ltd.
Product Code
KPR · Radiology
Decision Date
Apr 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
AeroDR TX c02 is a stationary X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment. AeroDR TX c02 is not intended for mammography, angiography, interventional, or fluoroscopy use.
Device Story
Stationary X-ray system; captures X-ray signals via detector; digitizes signals into DICOM images using Elui imaging software. Operated by licensed radiologic technologists in clinical environments. Supports patient postures including sitting, standing, or lying. Output used by clinicians for diagnostic screening of orthopedic, respiratory, or vertebral disc conditions. System components include X-ray tube, collimator, patient table, detector stand, and ceiling-suspended tube support. Benefits include diagnostic imaging of various anatomical parts.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, and IEC 60601-2-54 verified safety and EMC. Software, cybersecurity, and image quality performance were previously reviewed and cleared under K242119; no new clinical data required.
Indicated for obtaining radiographic images of anatomical parts (head, spine, chest, abdomen) in pediatric and adult patients for diagnostic screening of orthopedic, respiratory, or vertebral disc conditions. Contraindicated for mammography, angiography, interventional, or fluoroscopy.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 10, 2026
Dk Medical Systems Co., Ltd.
% Dong Ha Lee
RA Consultant
KMC, Inc.
#509, Daerung Post Tower, 43,
Digital-Ro 26-Gil, Guro-Gu,
SEOUL, 08389 SOUTH KOREA
Re: K253446
Trade/Device Name: AeroDR TX c02
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary x-ray system
Regulatory Class: Class II
Product Code: KPR
Dated: March 13, 2026
Received: March 13, 2026
Dear Dong Ha Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253446 - Dong Ha Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253446 - Dong Ha Lee
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by
GABRIELA M.
RODAL -S
for
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253446 | |
| Device Name AeroDR TX c02 | |
| Indications for Use (Describe) | |
| AeroDR TX c02 is a stationary X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment. AeroDR TX c02 is not intended for mammography, angiography, interventional, or fluoroscopy use. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
| 510(k) #: K253446 | 510(k) Summary | Prepared on: 2026-03-13 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | DK MEDICAL SYSTEMS Co., Ltd | |
| Applicant Address | 52, Chupalsandan 1-Gil, Paengseong-Eup Pyeongtaek-Si Gyeonggi-do 17998 Korea, South | |
| Applicant Contact Telephone | 82-2-535-7231 | |
| Applicant Contact | Mr. YoungWoo Kang | |
| Applicant Contact Email | ywkang@dk.co.kr | |
| Correspondent Name | KMC, Inc | |
| Correspondent Address | #509, Daerung Post tower, 43, Digital-ro 26-gil, Guro-gu Seoul 08389 Korea, South | |
| Correspondent Contact Telephone | 82-70-8965-5554 | |
| Correspondent Contact | Mr. DongHa Lee | |
| Correspondent Contact Email | dhlee@kmcerti.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | AeroDR TX c02 | |
| Common Name | System, X-Ray, Stationary | |
| Classification Name | Stationary x-ray system | |
| Regulation Number | 892.1680 | |
| Product Code(s) | KPR | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K242119 | INNOVISION-EXII | KPR |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| AeroDR TX c02 can receive X-ray signals from X-ray irradiation and digitize them into X-ray images by converting digital images to DICOM image format using Elui imaging software. AeroDR TX c02 is a general radiography X-ray system and not for mammography nor fluoroscopy. In addition, the system must be operated by a user who is trained and licensed to handle a general radiography X-ray system to meet the regulatory requirements of a Radiologic Technologist. Target areas for examinations include the head, spine, chest, and abdomen for diagnostic screening of orthopedic, respiratory, or vertebral discs. The system can capture a patient's postures, such as sitting, standing, or lying. This system can be used for patients of all ages. The x-ray detectors used with the system are unchanged from the predicate (K242119). The device software is unchanged from the predicate and it has been cleared as cybersecure under K242119. | | |
{5}
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| --- | --- |
| AeroDR TX c02 is a stationery X-ray system intended for obtaining radiographic images of various anatomical parts of the human body, both pediatrics and adults, in a clinical environment.AeroDR TX c02 is not intended for mammography, angiography, interventional, or fluoroscopy use. | |
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| Indications for use is the same as the predicate device (existing device) (K242119). | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| The predicate device (existing device) is an own manufactured device cleared by FDA (K242119).The subject device (Modified device) is the same as the predicate device (existing device) except for the below.1. Trade Name1) Trade name is changed. (AeroDR TX c02)2. X-ray System Components (Configuration)1) Alternate X-ray tube is changed. (E7884X)2) Alternate Collimator is added. (BL-80)3) Patient table is changed. (PT5-G1 K)4) X-ray detector stand is changed. (WS5-G1 K)5) Ceiling suspended x-ray tube support is changed. (CS5-G1 K)The differences (changes) do not affect to the x-ray image generated from the x-ray system.The differences (changes) between the subject device and predicate device were tested and verified about the safety and effectiveness accordance with FDA recognized standards. The test results show that it did not raise new safety and effectiveness concerns.The subject device is determined to be Substantially Equivalent (SE) to the predicate device. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| Non-Clinical Tests1. EMC and Safety test: EMC and Safety tests were conducted to verify that the subject device (modified device) met all design changes as it is Substantially Equivalent (SE) to the predicate device (existing device). We claim conformance to the following standards.1) IEC 60601-1 (Edition 3.2, 2020-08)2) IEC 60601-1-2 (Edition 4.1, 2020-09)3) IEC 60601-1-3 (Edition 2.2, 2021-01)4) IEC 60601-2-28 (Edition 3.0, 2017-06)5) IEC 60601-2-54 (Edition 2.0, 2022-09)2. Software and Cybersecurity test: The device software and cybersecurity characteristics are unchanged from the predicate and the test results have been reviewed under K242119.3. Performance test: X-ray image quality are unchanged form the predicate and the clinical image evaluation results have been reviewed under K242119.Conclusion: The differences (changes) do not affect to the x-ray image generated from the x-ray system.The differences (changes) between the subject device and predicate device were tested and verified about the safety and effectiveness accordance with FDA recognized standards. The test results show that it did not raise new safety and effectiveness concerns.The subject device is determined to be Substantially Equivalent (SE) to the predicate device. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.