K253396 · Apyx Medical Corporation · QPB · May 8, 2026 · General, Plastic Surgery
Device Facts
Record ID
K253396
Device Name
AYON Body Contouring System™ (AYON SYSTEM)
Applicant
Apyx Medical Corporation
Product Code
QPB · General, Plastic Surgery
Decision Date
May 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AYON™ Power Liposuction Sub-System is used for aesthetic body contouring.
Device Story
AYON™ Power Liposuction Sub-System is an electronic control system for aesthetic body contouring; adds power-assisted liposuction capability to the existing AYON Body Contouring System™. System inputs include user-defined speed settings via console touchscreen or handpiece buttons. Device transforms inputs into mechanical reciprocation of a cannula; reciprocates at 4000 ± 10% strokes/minute with 2.8 mm stroke distance. Used in operating environments by trained physicians. Output is mechanical movement of the cannula to facilitate fat aspiration. System integrates with infiltration and aspiration pumps; supports two handpieces simultaneously. Healthcare providers use the touchscreen GUI to monitor status and adjust parameters. Benefits include controlled, power-assisted tissue removal during lipoplasty.
Clinical Evidence
No clinical data. Substantial equivalence supported by mechanical and physical bench testing, including reciprocation performance, pull force, fatigue testing, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2, 60601-4-2), biocompatibility (ISO 10993-1), and sterilization/reprocessing validation (ISO 17665, ISO 11135).
Technological Characteristics
AC-powered console (120 VAC, 50-60 Hz). Features power-assisted handpieces with two-button control. Reciprocation mechanism: 4000 ± 10% strokes/min, 2.8 mm stroke distance. Connectivity: Touchscreen GUI, wireless footswitch. Materials: Patient-contacting materials evaluated per ISO 10993-1. Sterilization: Steam sterilization (autoclave) for reusable items. Software: ISO 62304 compliant.
Indications for Use
Indicated for aesthetic body contouring in patients undergoing lipoplasty procedures. Intended for use by trained physicians in an operating environment.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 8, 2026
Apyx Medical Corporation
Nikita Gajbhiye
Pre-Market Regulatory Affairs Manager
5115 Ulmerton Road
Clearwater, Florida 33760
Re: K253396
Trade/Device Name: AYON™ Power Liposuction Sub-System (AYON™ SYSTEM)
Regulation Number: 21 CFR 878.5040
Regulation Name: Suction Lipoplasty System
Regulatory Class: Class II
Product Code: QPB
Dated: December 19, 2025
Received: December 22, 2025
Dear Nikita Gajbhiye:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253396 - Angela Huber
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K253396 - Angela Huber
Page 3
Sincerely,
**Alicia Hemphill -S**
Digitally signed by Alicia Hemphill -S
Date: 2026.05.08 12:22:51 -05'00'
Alicia L. Hemphill (Johnson), MS
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253396 | ? |
| Please provide the device trade name(s). | | ? |
| AYON™ Power Liposuction Sub-System (AYON™ SYSTEM) | | |
| Please provide your Indications for Use below. | | ? |
| The AYON™ Power Liposuction Sub-System is used for aesthetic body contouring. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
Page 1 of 5 K253396
# 510(k) SUMMARY
## 1. General Information
Submitted by: Apyx Medical Corporation
5115 Ulmerton Road
Clearwater, Florida 33760-4004
United States of America
Contact Person: Nikita Gajbhiye
Regulatory Affairs Manager
(Pre-Market)
Phone: 857-210-6492
Email: nikita.gajbhiye@apyxmedical.com
Date Prepared: December 18th, 2025
Trade Name: AYON™ Power Liposuction Sub-System
(AYON™ SYSTEM)
Common Name: Liposuction System
Classification: Class II per 21CFR 878.5040 - Suction
Lipoplasty System
Product Code: QPB
Predicate Device: PAL System (K212024) by MicroAire Surgical
Instruments, LLC
## 2. Proposed Indications for Use
The AYON™ Power Liposuction Sub-System is used for aesthetic body contouring.
## 3. Device Description
The AYON Body Contouring System™ includes an Ultrasound Sub-system, Power Liposuction Sub-system, an Infiltration Sub-system, and an Aspiration Sub-system. The subject device for this submission is the AYON Body Contouring System™ addition of the Power Liposuction Sub-System. The Ultrasound, Aspiration, and Infiltration Sub-Systems are identical to the Predicate Device, the AYON Body Contouring System™ without the Power Liposuction Sub-System, cleared under K244050; there have been no changes made to these three subsystems. The Power Liposuction Sub-system includes Power Liposuction Handpieces, Power Liposuction cables Infiltration and Aspiration Tubing Adaptors. The Power Liposuction Handpiece can be used with commercially available cannulas. There are two connecting ports available on the AYON CONSOLE for this feature, and both may be used at the same
{6}
Page 2 of 5 K253396
time. The Power Liposuction handpieces have two buttons located on the handpieces. These can control the activation of the handpieces as well as either the aspiration or infiltration pump.
The AYON Body Contouring System™ may also accommodate other equipment, including commercially available electrosurgical generators owned by Apyx® Medical, such as the APYX-ONE Console. The controls for the Power Liposuction are integrated into a touch screen utilizing a graphical user interface on the AYON™ Console.
## 4. Performance Data
The relevant FDA guidance documents were followed for all aspects of performance testing.
### a. Mechanical and Physical Bench Testing
The Power Liposuction Sub-system was tested separately and as a combined system to demonstrate system compatibility of the subsystems and devices/accessories. The subsystem testing included;
- Power Liposuction Handpiece Reciprocation Performance and Dimensional Analysis, Pull Force and Fatigue testing
- Tubing Adaptor Mechanical Performance and conformity assessment to ISO 10079-1 and 10079-4.
### b. Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety testing was performed in accordance with IEC 60601-1-1. Electromagnetic Compatibility testing was performed in accordance with IEC 60601-1-2 and 60601-4-2. Wireless emissions testing was conducted for the wireless footswitch.
### c. Software and Cybersecurity
Software and Cybersecurity Verification and Validation testing were performed in accordance with ISO 62304.
### d. Biocompatibility Testing
Biological evaluation per ISO 10993-1 was conducted on the patient contacting materials.
### e. Reprocessing
The reprocessing of sterile products was tested in accordance with ISO 17665.
### f. Sterility
The processing of sterile products was tested in accordance with ISO 11135.
### g. Packaging
The product packaging was tested in accordance with ISO 11607.
{7}
Page 3 of 5 K253396
## 5. Clinical Performance
No clinical studies were required to support the substantial equivalence of the device. The non-clinical testing detailed in this submission support the substantial equivalence of the device.
## 6. Substantial Equivalence
The subject device for this submission is the AYON Body Contouring System™ addition of the Power Liposuction Sub-System. The identified Primary Predicate Device is the PAL System, manufactured by MicroAire Surgical Instruments, LLC under K212024.
Table 1. Substantial Equivalence Table
| | AYON™ Power Liposuction Sub-System Subject Device K253396 | MicroAire PAL System Primary Predicate Device K212024 | Device Comparison |
| --- | --- | --- | --- |
| Intended Use | Power Assisted Liposuction: For aesthetic body contouring. | Power Assisted Liposuction: For aesthetic body contouring. | Identical |
| Product Code/ Regulation Number | QPB, 21 CFR 878.5040 | QPB, 21 CFR 878.5040 | Identical |
| User | Used by trained Physicians in an operating environment. | Only trained and experienced healthcare providers/professionals should use this medical equipment. | Same |
| Power | AC Powered Voltage: 120 VAC Frequency: 50-60 Hz | AC Powered Voltage: 100-240 VAC Frequency: 50/60 Hz | Similar, accommodates United States voltage. |
| Mechanism of Action | Power Liposuction: AYON Power Liposuction Sub-System™ is an electronic control system designed to reciprocate a cannula in lipoplasty applications. | Power Assisted Liposuction: MicroAire power assisted liposuction is an electronic control system designed to reciprocate a cannula in lipoplasty applications. | Identical |
| User Interface/Controls | Single touchscreen utilizing a graphical user interface, on/off switches | The color touchscreen display provides an intuitive graphical user interface that allows users to view system status and | Identical |
{8}
Page 4 of 5
K253396
| | AYON™ Power Liposuction Sub-System Subject Device K253396 | MicroAire PAL System Primary Predicate Device K212024 | Device Comparison |
| --- | --- | --- | --- |
| | | set parameters with a touch of the screen. | |
| Operating Mode | Use the sliding bar on the Console display to set the speed of reciprocation of the Power Liposuction Handpiece (up to 100%) | Use the thumb-throttle on the PAL Handpiece to adjust the speed of reciprocation during the procedure; use the sliding bar on the Console display to set the maximum speed (up to 100%) | Similar, both control mechanisms provide the ability to adjust the handpiece speed. |
| Power Liposuction Handpiece | | | |
| Speed Control | Touch Screen and Handpiece | Touch Screen and Handpiece | Identical |
| Number of Attachable Devices | Two (2) Handpieces | Two (2) Handpieces | Identical |
| Electrical Safety Testing Passed | Meets applicable IEC requirements: IEC/EN 60601-1-2 IEC 60601-1 | Meets applicable IEC requirements: IEC/EN 60601-1-2 IEC 60601-1 | Identical |
| Performance | 4000 ± 10% strokes/ minute | 4000 – 5000 strokes/min. | Similar, the device operates within the range when set to high-speed. The subject device has a slightly tighter specification (3600 - 4400 strokes per minute). |
| Stroke Distance | 2.8 mm (±0.4 mm) | 2.8 mm (±0.4 mm) | Identical |
| Power Source | Console, detachable cable | Console, detachable cable | Identical |
| Duty Cycle | Duty cycle of 2 hours continuous use/2 hours off. | Duty cycle of 2 hours continuous use/2 hours off. | Identical |
| Sterilization | Re-usable items are autoclavable by end user, steam sterilization | Re-usable items are autoclavable by end user, steam sterilization | Identical |
| Tubing Adaptor accessory | | | |
4
{9}
Page 5 of 5 K253396
| | AYON™ Power Liposuction Sub-System Subject Device K253396 | MicroAire PAL System Primary Predicate Device K212024 | Device Comparison |
| --- | --- | --- | --- |
| Tubing Compatibility | Tubing Adaptor provides compatibility with standard, commercially available infiltration and aspiration tubing | Compatible with Aspiration Tubing and Infiltration Tubing | Same |
7. Substantial Equivalence Determination
The subject and primary predicate device have the same intended use and comply with the relevant standards and FDA guidance documents with regard to electrical safety, electromagnetic compatibility, software and cybersecurity, biocompatibility, transportation, packaging, sterilization.
The data provided shows the Power Liposuction Sub-system and associated handpiece reciprocates with the same performance (speed, distance) compared to the primary predicate device. The AYON Body Contouring System™ continues to operate as intended, following the addition of the fourth sub-system, the Power Liposuction Sub-system.
8. Conclusion
The performance data presented in this submission supports substantially equivalent performance between the subject device and the primary predicate device, MicroAire PAL System cleared under K212024.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.