mobiCARE Cardiac Monitoring System (mobiCARE-MC200M, mobiCARE-MC200ML, mobiCARE-MC200M7, mobiCARE-MC200ML7)
Applicant
Seers Technology Co., Ltd.
Product Code
MWJ · Cardiovascular
Decision Date
Jun 25, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Indications for Use
mobiCARE Cardiac Monitoring System is a wearable, ambulatory ECG recorder. It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. mobiCARE Cardiac Monitoring System is intended for use by patients 18 years or older. The device is not intended for use with critical care patients or for real-time monitoring.
Device Story
Single-channel, wearable, reusable ambulatory ECG recorder; acquires/stores ECG signals via disposable electrodes for 6-9 days. Home user initiates measurement via mobile app; clinician uploads recorded data via cradle and UPLoader software to 3rd-party viewing platform. Device not for real-time monitoring. Output used by clinicians to assess cardiac rhythm; aids diagnosis of transient symptoms. Benefits include extended ambulatory monitoring for symptom correlation.
Clinical Evidence
Clinical performance test of electrode placement conducted per ISO 14155. Additional evidence includes bench performance testing, software V&V (ANSI AAMI IEC 62304), cybersecurity testing, biocompatibility (ISO 10993-1), human factors validation (ANSI AAMI IEC 62366-1), and electrical safety/EMC testing (IEC 60601 series).
Technological Characteristics
Single-channel wearable ECG; Type CF applied part. Bluetooth Low Energy (2402-2480 MHz) and USB 2.0 connectivity. AES-CCM encryption. Complies with IEC 60601-2-47 (ECG dynamic range ±10% of ±5mV), IEC 60601-1 (electrical safety), and ISO 10993-1 (biocompatibility). Reusable device with disposable electrodes; disinfected between users.
Indications for Use
Indicated for patients 18+ years old, asymptomatic or experiencing transient symptoms (palpitations, dyspnea, dizziness, lightheadedness, pre-syncope, syncope, fatigue, chest pain, anxiety). Not for critical care or real-time monitoring.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
S-Patch Ex Wearable ECG Patch, Wellysis Corp (K231289)
Submission Summary (Full Text)
{0}
June 25, 2026
Seers Technology Co., Ltd.
% Diane Horwitz
Regulatory Consultant
Cardiovascular Biometrics Sourcing, LLC
6703 West 131st St.
Overland Park, Kansas 66209
Re: K253384
Trade/Device Name: mobiCARE Cardiac Monitoring System (mobiCARE-MC200M, mobiCARE-MC200ML, mobiCARE-MC200M7, mobiCARE-MC200ML7)
Regulation Number: 21 CFR 870.2800
Regulation Name: Electrocardiograph, Ambulatory (Without Analysis)
Regulatory Class: Class II
Product Code: MWJ
Dated: May 24, 2026
Received: May 26, 2026
Dear Diane Horwitz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253384 - Diane Horwitz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253384 - Diane Horwitz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**JENNIFER W. SHIH -S**
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253384
Device Name
mobiCARE Cardiac Monitoring System (mobiCARE-MC200M, mobiCARE-MC200ML, mobiCARE-MC200M7, mobiCARE-MC200ML7)
Indications for Use (Describe)
mobiCARE Cardiac Monitoring System is a wearable, ambulatory ECG recorder. It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. mobiCARE Cardiac Monitoring System is intended for use by patients 18 years or older. The device is not intended for use with critical care patients or for real-time monitoring.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K253384
SEERS
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
1. GENERAL INFORMATION
1.1 Submitter and 510(k) Owner
SEERS Co., Ltd.
(17707) 291-13, Dongbu-daero, Jinwi-myeon,
Pyeongtaek-si, Gyeonggi-do, Republic of Korea
1.2 Official Correspondent
Diane Horwitz, Ph.D.
Cardiovascular Biometrics Sourcing, LLC
6703 West 131st St., Overland Park, KS 66209
Phone (703) 307-2921
E-mail: dmh@mandellhorwitzconsulting.com
1.3 Date of Preparation
June 25, 2026
2. NAME OF THE DEVICE
2.1.1 Trade/Proprietary Name
mobiCARE Cardiac Monitoring System (mobiCARE-MC200M series)
Models: mobiCARE-MC200M, mobiCARE-MC200ML, mobiCARE-MC200M7, mobiCARE-MC200ML7
2.1.2 Classification Information
Classification Name: Electrocardiograph, Ambulatory (Without Analysis)
Classification Regulation: 21 CFR 870.2800
Class: II
Product Code: MWJ
Panel: Cardiovascular
3. PREDICATE DEVICE
The predicate device is the S-Patch Ex Wearable ECG Patch, Wellysis Corp, K231289
4. DESCRIPTION OF THE DEVICE
mobiCARE-MC200M/mobiCARE-MC200ML/mobiCARE-MC200M7/mobiCARE-MC200ML7 ("mobiCARE-MC200M series") is single-channel, wearable, mobile and reusable ambulatory ECG recorder. It can continuously acquire and store the ECG signals through disposable electrodes for up to 6 or 9 days (expected continuous measurement time is different by the model). The mobiCARE-MC200M series is not intended for real-time monitoring.
1
{5}
K253384
SEERS
ECG measurement is activated after generating the measurement code using the mobiCARE Cardio Hospital mobile phone application. Status for the patch connection, battery level and patch signal is confirmed using this mobile application.
The home user then wears the mobiCARE-MC200M device for the prescribed period of time, usually 6 days or 9 days (depending on the model). After the measurement is completed, recorded ECG signal can be uploaded to a compatible ECG viewing (and/or analyzing) platform by qualified clinicians using the mobiCARE Cradle and mobiCARE Cardio UPLoader.
The mobiCARE-MC200M series is a prescription device intended to be used with direction from Healthcare Professionals in accordance with the user manual.
## 5. INDICATIONS FOR USE
mobiCARE Cardiac Monitoring System is a wearable, ambulatory ECG recorder. It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. mobiCARE Cardiac Monitoring System is intended for use by patients 18 years or older. The device is not intended for use with critical care patients or for real-time monitoring.
## 6. INDICATIONS FOR USE COMPARED TO THE PREDICATE
The intended use and indications for use of the subject and predicate are the same with the exception of the product name.
## 7. TECHNOLOGY CHARACTERISTICS COMPARED TO THE PREDICATE
The mobiCARE-MC200M series devices are similar in most characteristics to the predicate device. The slight technological differences (i.e., uploading of data and battery size) do not raise new issues of safety or effectiveness, as the subject device has been demonstrated to meet product requirements and specifications.
| | mobiCARE Cardiac Monitoring System (mobiCARE-MC200M series) Subject Device | S-Patch Ex Wearable ECG Patch K231289 Predicate Device |
| --- | --- | --- |
| Product Code Regulation Review Panel | MWJ 21 CFR 870.2800 Cardiovascular | DSH 21 CFR 870.2800 Cardiovascular |
| Intended Use / Indications for Use Statement | mobiCARE Cardiac Monitoring System is a wearable, ambulatory ECG recorder. It is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. mobiCARE Cardiac Monitoring System is intended for use by patients 18 years or older. The device is not intended for use | S-Patch Ex wearable ECG patch is indicated for use on patients who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light headedness, pre-syncope, syncope, fatigue, chest pain and/or anxiety. S-Patch Ex wearable ECG patch is intended for use by patients 18 years or older. |
2
{6}
K253384
SEERS
| | mobiCARE Cardiac Monitoring System (mobiCARE-MC200M series) Subject Device | S-Patch Ex Wearable ECG Patch K231289 Predicate Device |
| --- | --- | --- |
| | with critical care patients or for real-time monitoring. | |
| Monitoring duration | Up to 9 days, depending on model | 100 hours (4.2 days) |
| Multiple patient use | Yes; device is disinfected between users | Yes; device is disinfected between users |
| ECG Dynamic Range | ±10 % (± 5mV); complies with IEC 60601-2-47 standard | ±10 mV |
| Applied Part Category | Type CF (cardiac floating) | Type CF (cardiac floating) |
| Data Storage and Transfer | mobiCARE Cardio UPLoader software on Healthcare Professional PC | Mobile gateway on home user's phone described in Instructions for Use. |
| Viewing Software Platform | 3^{rd} Party ECG Viewing Software | 3^{rd} Party ECG Viewing Software |
| Communication Protocol | - Main Body: Bluetooth Low Energy (2402 to 2480 MHz) - Cradle: USB 2.0 | Bluetooth Low Energy (2402 – 2480 MHz) |
| Data Encryption | Advanced Encryption Standard-CCM mode | Advanced Encryption Standard-CCM mode |
## 8. PERFORMANCE TESTING
Bench performance testing, a clinical performance test, human factors engineering testing, biocompatibility and disinfection and transportation validation support the conclusion that any differences between the mobiCARE-M200M series and the predicate device do not impact safety or effectiveness.
Performance tests included:
- Software verification and validation testing in compliance with ANSI AAMI IEC 62304:2006/A1:2016.
- Cybersecurity testing in compliance with FDA Guidance, Cybersecurity in Medical Devices, 2026.
- Biocompatibility testing of patient-contacting cable in compliance with ISO 10993-1 Sixth edition 2025-11.
- Transportation testing
- Expected service life evaluation
- Human factors validation testing in compliance with ANSI AAMI IEC 62366-1:2015+AMD1:2020.
- Electromagnetic compatibility and electrical safety testing in compliance with IEC 60601-1 Edition 3.2 2020-08, ANSI AAMI IEC 60601-1-2:2014, IEC 60601-1-6 Edition 3.2 2020-07, IEC 60601-1-11 Edition 2.1 2020-07, ANSI AAMI IEC 60601-2-47:2012/(R)2016.
- Compliance with defibrillation-proof requirements.
- Disinfection validation Reprocessing Medical Devices in Health Care Settings, 2015.
- Clinical performance test of electrode placement in compliance with ISO 14155 Third edition 2020-07.
3
{7}
K253384
SEERS
## 9. CONCLUSIONS
The bench and clinical performance testing including software verification and validation, cybersecurity testing, biocompatibility, transportation, disinfection validation testing, and human factors validation testing provide evidence to support the conclusion that the mobiCARE MC200M device is substantially equivalent to the predicate device.
---4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.