K253337 · Inter-Med, Inc. · EMA · Apr 15, 2026 · Dental
Device Facts
Record ID
K253337
Device Name
REGEN Bioactive Cement
Applicant
Inter-Med, Inc.
Product Code
EMA · Dental
Decision Date
Apr 15, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Indications for Use
Designed for luting of indirect restorations (inlays, onlays, crowns, bridges, and endodontic posts) made of metal alloys, composite resins, porcelain, lithium disilicate or zirconia.
Device Story
REGEN Bioactive Cement is a dual-cure (anaerobic self-polymerization and photopolymerization) resin-based luting cement; used by dental professionals to bond indirect restorations (inlays, onlays, crowns, bridges, endodontic posts) to dentin or enamel. Input: two-component paste (base and catalyst) delivered via dual-barrel syringe and mixing tip. Transformation: monomers polymerize via radical-release mechanism to create permanent bond; releases calcium and phosphate ions while bonded. Output: cured cement layer providing structural adhesion and radiopacity. Used in clinical dental settings; applied by dentists. Benefits: provides durable restoration retention and bioactive ion release.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including: ISO 4049 compliance (setting time, flexural strength, water sorption, solubility, film thickness), shear bond strength, polymerization shrinkage, radiopacity, cytotoxicity (ISO 10993-5), transit testing, and verification of calcium/phosphate ion release.
Technological Characteristics
Dual-component, non-aqueous, resin-based paste. Dual-cure (anaerobic self-polymerization and photopolymerization). External communicating device for permanent contact (>30 days) per ISO 7405:2018. Packaged in dual-barrel syringes with mixing tips and foil-sealed pouch. Non-sterile. Complies with ISO 4049 for dental polymer-based restorative materials.
Indications for Use
Indicated for luting indirect dental restorations including inlays, onlays, crowns, bridges, and endodontic posts made of metal alloys, composite resins, porcelain, lithium disilicate, or zirconia in patients requiring permanent restoration cementation.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 15, 2026
Inter-Med, Inc.
Alex Johnson
VP of Engineering, Regulatory & Quality
2200 S. St.
Racine, Wisconsin 53404
Re: K253337
Trade/Device Name: REGEN Bioactive Cement
Regulation Number: 21 CFR 872.3275
Regulation Name: Dental Cement
Regulatory Class: Class II
Product Code: EMA
Dated: September 26, 2025
Received: September 30, 2025
Dear Alex Johnson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253337 - Alex Johnson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253337 - Alex Johnson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253337 | ? |
| Please provide the device trade name(s). | | ? |
| REGEN Bioactive Cement | | |
| Please provide your Indications for Use below. | | ? |
| Designed for luting of indirect restorations (inlays, onlays, crowns, bridges, and endodontic posts) made of metal alloys, composite resins, porcelain, lithium disilicate or zirconia. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K253337 | 510(k) Summary | Prepared on: 2026-04-14 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Inter-Med, Inc. | |
| Applicant Address | 2200 South St Racine WI 53404 United States | |
| Applicant Contact Telephone | 262-631-5306 | |
| Applicant Contact | Mr. Alex Johnson & Mr. Dominic Egizi | |
| Applicant Contact Email | ajohnson@vista-dental.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | REGEN Bioactive Cement | |
| Common Name | Dental cement | |
| Classification Name | Cement, Dental | |
| Regulation Number | 872.3275 | |
| Product Code(s) | EMA | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K210259 | Parkell Self-Adhesive Cement | EMA |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| REGEN Bioactive Cement is a bioactive luting cement with a dual-cure system, anerobic self-polymerization and photopolymerization, designed to create a durable and permanent bond between dentin or enamel and indirect restorations such as dentures and crowns. While bonded to the tooth, REGEN Bioactive Cement releases calcium and phosphate.
Five (5) grams of each component (base paste and catalyst paste) are packaged into a foil-sealed dual-barrel syringe totaling 10 grams of product mass. This device is for prescription use (Rx) only.
This is the only 510(k) for this medical device, no prior 510(k)s have been submitted. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| Designed for luting of indirect restorations (inlays, onlays, crowns, bridges, and endodontic posts) made of metal alloys, composite resins, porcelain, lithium disilicate or zirconia. | | |
| Indications for Use Comparison | | 21 CFR 807.92(a)(5) |
| The indications for use are the same. | | |
| Technological Comparison | | 21 CFR 807.92(a)(6) |
{5}
Similarities between the subject device (REGEN Bioactive Cement) and the predicate device (Parkell Self-Adhesive Cement).
- REGEN Bioactive Cement has identical indications for use as its predicate device (Parkell Self-Adhesive Cement).
- REGEN Bioactive Cement is classified under product code EMA, with respect to 21 CFR 872.3820 and the common name "self-adhesive resin cement", which is identical to that of the predicate device (Parkell Self-Adhesive Cement).
- REGEN Bioactive Cement has identical intended population use as the predicate device (Parkell Self-Adhesive Cement) and is used in the same anatomical location.
- Identical to the predicate device (Parkell Self-Adhesive Cement), REGEN Bioactive Cement is a two-component self-adhesive resin device capable of self-polymerization and photopolymerization.
- REGEN Bioactive Cement uses an identical mechanism of action for its indications compared to the predicate device (Parkell Self-Adhesive Cement), namely, both materials utilize monomers polymerized through a radical-release mechanism to provide structural and adhesive function, and radiopacity to indirect restorations.
- Identical to the predicate device (Parkell Self-Adhesive Cement), REGEN Bioactive Cement is classified as an external communicating device for permanent (>30 days) contact per ISO 7405:2018.
- REGEN Bioactive Cement composition consists of similar ingredients to the predicate device (Parkell Self-Adhesive Cement).
- Any differences in composition do not raise any concerns as biocompatibility testing confirms the product is biocompatible for its intended use.
- Identical to the predicate device (Parkell Self-Adhesive Cement), REGEN Bioactive Cement is a dual-component, non-aqueous paste-like material that is commercialized as non-sterile. Both the subject device and predicate device are stored in dual barrel syringes and are applied through identical mixing tips.
- Identical to the predicate device (Parkell Self-Adhesive Cement), REGEN Bioactive Cement is for prescription use (Rx) only.
This is the only 510(k) for this medical device, no prior 510(k)s have been submitted.
Discussion of Differences between the subject device (REGEN Bioactive Cement) and the predicate device (Parkell Self-Adhesive Cement).
- The formulation of REGEN Bioactive Cement is slightly different than the predicate device (Parkell Self-Adhesive Cement).
- The differences in composition do not raise any concerns as biocompatibility testing (ISO 10993-5:2009 cytotoxicity testing) and benchtop testing (510(k) Section Benchtop testing) confirms the product is substantially equivalent to the predicate device.
- Therefore, the subject device remains substantially equivalent to its predicate device despite differences in composition.
- REGEN Bioactive Cement is stored differently than Parkell Self-Adhesive Cement.
- A heat-sealed foil pouch is used with REGEN Bioactive Cement to promote long term stability of the product. Parkell Self-Adhesive Cement syringes are not stored with a secondary seal, instead placed directly into a box. The addition of a foil pouch as another packaging layer is not considered a significant change and substantial equivalence to the predicate device is maintained.
- REGEN Bioactive Cement's predicted shelf life of 15 months is shorter than the predicate device's (Parkell Self-Adhesive Cement) stated shelf life of 24 months.
- This difference is not considered significant as the expiration of dental products varies widely in the field. As all products maintain an expiration date on their labeling, this information is sufficiently communicated to users.
- Therefore, the subject device remains substantially equivalent to its predicate device despite differences in composition.
## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The following non-clinical tests were conducted to evaluate the functionality, performance, safety, and substantial equivalence of REGEN Bioactive Cement to the predicate device (Parkell Self-Adhesive Cement):
- Real-Time Shelf-Life Summary
- Given the available data for REGEN Bioactive Cement, a room-temperature shelf-life of 15 months is assigned.
- Setting Time and Flowability
- This test confirms that REGEN Bioactive Cement conforms to the ISO 4049 set time requirements for Class 3 devices of no longer than 10 minutes and matches Parkell Self-Adhesive Cement's claim of sub-5-minute self-cure. Therefore, the subject device is substantially equivalent to the predicate device in terms of set time.
- Flexural Strength
- The formulation of REGEN Bioactive Cement "passed" the test criteria of ISO 4049 and provided results that support a substantial equivalence designation of the subject device to the predicate device Parkell Self-Adhesive Cement.
- Water Sorption and Solubility
- The formulation of REGEN Bioactive Cement "passed" the test criteria of ISO 4049 and provided results that support a substantial equivalence designation of the subject device to the predicate device Parkell Self-Adhesive Cement.
- Shear Bond Testing
- REGEN Bioactive Cement "passed" the test criteria and provided results that were substantially equivalent to the predicate device, Parkell Self-Adhesive Cement.
- Film Thickness
{6}
o REGEN Bioactive Cement "passed" the test criteria and provided results that were substantially equivalent to the predicate device, Parkell Self-Adhesive Cement.
- Polymerized Shrinkage
o REGEN Bioactive Cement yielded shrinkage results in line with the predicate device, Parkell Self-Adhesive Cement. Therefore, the subject device is substantially equivalent to the predicate device in terms of shrinkage.
- Radiopacity
o REGEN Bioactive Cement is concluded to be equivalent to Parkell Self-Adhesive Cement in terms of radiopacity. Therefore, substantial equivalence is supported.
- Cytotoxicity
o REGEN Bioactive Cement yielded substantially equivalent cytotoxicity results compared to the predicate device (Parkell Self-Adhesive Cement).
- Transit Testing
o This testing confirmed REGEN Bioactive Cement's packaging is robust and passed simulated transit testing parameters.
- Phosphate and Calcium Ion Release
o REGEN Bioactive Cement's claim of Calcium and Phosphate ion release is verified. The test results support a substantial equivalence designation of the subject device to the predicate device (Parkell Self-Adhesive Cement).
Clinical testing was not deemed necessary for this 510(k).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.