PhAST instrument and PhAST Blood Culture Gram-negative Panel
K253316 · Phast Corp. · SAN · Jun 26, 2026 · Microbiology
Device Facts
Record ID
K253316
Device Name
PhAST instrument and PhAST Blood Culture Gram-negative Panel
Applicant
Phast Corp.
Product Code
SAN · Microbiology
Decision Date
Jun 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1650
Device Class
Class 2
Attributes
PCCP
Indications for Use
The PhAST Blood Culture Gram-negative Panel (PhAST BC GN Panel) is an in vitro diagnostic assay for qualitative (breakpoint) determination of antimicrobial susceptibility testing (AST) of pathogenic gram-negative bacteria from positive blood cultures (BC) and is intended to be used with the PhAST instrument. The PhAST BC GN Panel is a phenotypic test that utilizes video microscopy and analysis of single-cell phenotypes to determine susceptibility based on categorical interpretation. The PhAST BC GN Panel does not provide organism identification. The PhAST BC GN Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and reported as monomicrobial gram-negative specimens by gram staining. Organism identity must be entered into the PhAST instrument or PhAST web application before categorical interpretation (Susceptible, Susceptible Dose-Dependent, Intermediate, Resistant, S/SDD/I/R) of antimicrobial susceptibility is reported. The PhAST BC GN Panel contains 15 antimicrobials. The PhAST BC GN Panel is intended to aid in the diagnosis and treatment of individuals suspected of bloodstream infection by a healthcare provider. Results should not be used as the sole basis for patient management decisions. This test is performed by laboratory health professionals in a clinical diagnostic setting. Results may be used as an aid to clinicians in determining appropriate antimicrobial therapy. Test results from the PhAST BC GN Panel should be interpreted in conjunction with other clinical and laboratory findings. Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing. Sub-culturing is necessary to support further testing for: bacteria and antimicrobials not on the PhAST BC GN Panel, inconclusive results, epidemiologic testing, recovery of organisms present in positive blood cultures samples, and susceptibility testing of bacteria in polymicrobial samples.
Device Story
Device performs qualitative antimicrobial susceptibility testing (AST) on positive blood culture samples; utilizes single-cell video microscopy to analyze bacterial phenotypes (growth, morphology, motility) in response to 15 antimicrobials. Workflow: operator lyses positive blood culture sample; loads into single-use cartridge; PhAST instrument automatically meters, dilutes, and routes sample to 56 imaging chambers; images captured at three time points; data sent to cloud for analysis. Diagnostic algorithm compares phenotypes in antimicrobial-free control chambers vs. antimicrobial-treated chambers to compute resistance scores; outputs categorical interpretation (S, SDD, I, R). Requires manual entry of organism identity (from external ID method) to generate report. Used in clinical labs; aids clinicians in selecting antimicrobial therapy; results must be interpreted with other clinical findings. Benefits include rapid AST results (approx. 91-154 minutes) directly from positive blood culture, facilitating faster targeted treatment.
Clinical Evidence
Clinical study evaluated 890 samples (prospective and contrived) across four sites. Performance compared to broth microdilution (BMD) reference method. Categorical agreement (CA) generally >90% for indicated combinations. Study identified specific antimicrobial/organism combinations with unacceptable performance, resulting in labeling limitations. No clinical sensitivity/specificity metrics reported as device is for AST, not ID.
Technological Characteristics
Single-cell, time-lapse video microscopy system. Cartridge contains 15 dried antimicrobials in 45 imaging chambers plus 8 control chambers. Automated sample preparation (lysis, dilution, mixing). Cloud-based data analysis. Software compliant with IEC 62304; cybersecurity per ANSI ISA 62443-4-1. Connectivity via PhAST web application.
Indications for Use
Indicated for qualitative antimicrobial susceptibility testing of pathogenic gram-negative bacteria (Acinetobacter spp., Enterobacterales, Pseudomonas aeruginosa) from positive, monomicrobial blood culture samples in patients suspected of bloodstream infection. Requires prior organism identification via FDA-cleared method.
Regulatory Classification
Identification
A cellular analysis system for multiplexed antimicrobial susceptibility testing is a multiplex qualitative and/or quantitative in vitro diagnostic device intended for the identification and determination of the antimicrobial susceptibility results of organisms detected in samples from patients with suspected microbial infections. This device is intended to aid in the determination of antimicrobial susceptibility or resistance when used in conjunction with other laboratory findings.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Detailed device description documentation, including the device components, ancillary reagents required but not provided, a detailed explanation of the methodology, including primer/probe sequence, design, rationale for sequence selection, and details of the antimicrobial agents, as applicable.
(ii) Detailed documentation from the following analytical and clinical performance studies: limit of detection, inclusivity, precision, reproducibility, interference, cross-reactivity, carryover, and cross-contamination, quality control and additional studies, as applicable to specimen type and assay intended use.
(iii) Detailed documentation from an appropriate clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) Limitations and protocols regarding the need for correlation of results by standard laboratory procedures, as applicable.
(ii) A detailed explanation of the interpretation of results and acceptance criteria.
(iii) A detailed explanation of the principles of operation and procedures for assay performance and troubleshooting.
Predicate Devices
Accelerate Pheno System, Accelerate PhenoTest BC Kit (DEN160032)
Submission Summary (Full Text)
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
# **510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT**
**I Background Information:**
**A 510(k) Number**
K253316
**B Applicant**
PhAST Corp.
**C Proprietary and Established Names**
PhAST instrument and PhAST Blood Culture Gram-negative Panel
**D Regulatory Information**
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| SAN | II | 21 CFR 866.1650 - A Cellular Analysis System For Multiplexed Antimicrobial Susceptibility | MI - Microbiology |
| LON | II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology |
**II Submission/Device Overview:**
**A Purpose for Submission:**
- To obtain a substantial equivalence determination for use of the PhAST instrument and PhAST Blood Culture Gram-negative Panel (PhAST BC GN Panel) for testing positive blood culture samples for qualitative (“breakpoint” device) determination of antimicrobial susceptibility testing (AST) of specific antimicrobials with specific gram-negative organisms
- To establish a Predetermined Change Control Plan (PCCP) for removal of existing limitations imposed due to insufficient resistant isolate testing.
**B Measurand:**
| Antimicrobial | | Categorical Interpretations* |
| --- | --- | --- |
| Amikacin | AMK | S, I, R |
| Ampicillin/sulbactam | SAM | S, I, R |
| Aztreonam | ATM | S, I, R |
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
| Cefepime | FEP | S, SDD, R |
| --- | --- | --- |
| Ceftazidime | TAZ | S, I, R |
| Ceftazidime/avibactam | CZA | S, R |
| Ceftriaxone | CRO | S, I, R |
| Ciprofloxacin | CIP | S, I, R |
| Ertapenem | ETP | S, I, R |
| Gentamicin | GEN | S, I, R |
| Levofloxacin | LVX | S, I, R |
| Meropenem | MEM | S, I, R |
| Piperacillin/tazobactam | TZP | S, I, R |
| Tobramycin | TOB | S, I, R |
| Trimethoprim/sulfamethoxazole | SXT | S, R |
*S – Susceptible; I – Intermediate; SDD – Susceptible-Dose Dependent; R – Resistant
### C Type of Test:
Qualitative antimicrobial susceptibility test (AST) system that utilizes single-cell video microscopy and analysis of single-cell phenotypes of organisms in positive blood culture samples to determine susceptibility based on categorical interpretation.
### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use.
#### B Indication(s) for Use:
The PhAST Blood Culture Gram-negative Panel (PhAST BC GN Panel) is an in vitro diagnostic assay for qualitative (breakpoint) determination of antimicrobial susceptibility testing (AST) of pathogenic gram-negative bacteria from positive blood cultures (BC) and is intended to be used with the PhAST instrument. The PhAST BC GN Panel is a phenotypic test that utilizes video microscopy and analysis of single-cell phenotypes to determine susceptibility based on categorical interpretation. The PhAST BC GN Panel does not provide organism identification. The PhAST BC GN Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and reported as monomicrobial gram-negative specimens by Gram staining. Organism identity must be entered into the PhAST instrument or PhAST web application before categorical interpretation (Susceptible, Susceptible-Dose Dependent, Intermediate, Resistant, S/SDD/I/R) of antimicrobial susceptibility is reported. The PhAST BC GN Panel contains 15 antimicrobials.
The PhAST BC GN Panel is intended to aid in the diagnosis and treatment of individuals suspected of bloodstream infection by a healthcare provider. Results should not be used as the sole basis for patient management decisions.
This test is performed by laboratory health professionals in a clinical diagnostic setting. Results may be used as an aid to clinicians in determining appropriate antimicrobial therapy. Test results from the PhAST BC GN Panel should be interpreted in conjunction with other clinical and laboratory findings. Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing. Sub-culturing is necessary to support further testing for: bacteria and antimicrobials not on the PhAST BC GN Panel,
K253316 - Page 2 of 34
{3}
inconclusive results, epidemiologic testing, recovery of organisms present in positive blood cultures samples, and susceptibility testing of bacteria in polymicrobial samples.
Testing is indicated for *Acinetobacter* spp., Enterobacterales, and *Pseudomonas aeruginosa*, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC). The PhAST BC GN Panel with the PhAST instrument has demonstrated acceptable performance with the following organisms:
Amikacin: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Serratia marcescens*)
Ampicillin/sulbactam: *Acinetobacter* spp. (*Acinetobacter baumannii* complex)
Aztreonam: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus vulgaris* group)
Cefepime: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*, *Serratia marcescens*) and *Pseudomonas aeruginosa*
Ceftazidime: *Acinetobacter* spp. (*Acinetobacter baumannii* complex), Enterobacterales (*Enterobacter cloacae* complex, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*, *Serratia marcescens*) and *Pseudomonas aeruginosa*
Ceftazidime/avibactam: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*, *Proteus vulgaris* group, *Serratia marcescens*) and *Pseudomonas aeruginosa*
Ceftriaxone: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*)
Ciprofloxacin: Enterobacterales (*Citrobacter freundii* complex, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*, *Serratia marcescens*) and *Pseudomonas aeruginosa*
Ertapenem: Enterobacterales (*Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*)
Gentamicin: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*, *Proteus vulgaris* group, *Serratia marcescens*)
Levofloxacin: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*,
K253316 - Page 3 of 34
{4}
*Klebsiella pneumoniae, Proteus mirabilis, Proteus vulgaris* group, *Serratia marcescens*) and *Pseudomonas aeruginosa*
Meropenem: *Acinetobacter* spp. (*Acinetobacter baumannii* complex), Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*, *Proteus mirabilis*) and *Pseudomonas aeruginosa*
Piperacillin/tazobactam: *Acinetobacter* spp. (*Acinetobacter baumannii* complex), Enterobacterales (*Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae*) and *Pseudomonas aeruginosa*
Tobramycin: Enterobacterales (*Citrobacter freundii* complex, *Citrobacter koseri*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Proteus vulgaris* group) and *Pseudomonas aeruginosa*
Trimethoprim/sulfamethoxazole: Enterobacterales (*Citrobacter koseri*, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella pneumoniae*, *Serratia marcescens*)
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
- • The PhAST BC GN Panel should not be used for samples with Gram stain results that indicate the positive blood culture contains yeast or gram-positive bacteria or includes more than one organism (i.e., polymicrobial samples).
- • The PhAST BC GN Panel does not replace the need to subculture the positive blood culture. A small percentage of polymicrobial samples may escape detection during Gram stain and/or the rapid ID method. The positive blood culture sample should be subcultured on a solid agar plate to confirm that the sample is monomicrobial. If the subcultured agar plate indicates that the sample is polymicrobial, the PhAST BC GN Panel results should be discarded. Susceptibility testing by alternate methods should be performed on each isolate type from the subcultured plate.
- • The performance of this test has only been evaluated using the following blood culture bottles:
- ○ BD BACTEC Standard/10 Aerobic/F
- ○ BD BACTEC Standard/10 Anaerobic/F
- ○ BD BACTEC PEDS PLUS/F
- ○ BD BACTEC Plus Aerobic/F
- ○ BD BACTEC Plus Anaerobic/F
- ○ BD BACTEC Lytic Anaerobic/F
- ○ bioMerieux BacT/ALERT SA Standard Aerobic
- ○ bioMerieux BacT/ALERT SN Standard Anaerobic
- ○ bioMerieux BacT/ALERT FA Plus Aerobic
- ○ bioMerieux BacT/ALERT FN Plus Anaerobic
- ○ bioMerieux BacT/ALERT PF Plus
- • Positive blood cultures should be tested immediately after a positive flag, where possible. In the case of unavoidable delays or if the need for re-testing arises, testing on the PhAST
K253316 - Page 4 of 34
{5}
BC GN Panel may be performed on positive blood culture bottles removed from the continuous monitoring system within 5 hours of bottle ring and subsequently kept at room temperature for up to 12 hours after removal.
- Positive blood culture samples must be loaded on the PhAST BC GN Panel cartridge and the run must be initiated within 10 minutes of lysing the positive blood culture sample.
The following limitations were added to the device labeling based on performance demonstrated in the current submission:
Perform an alternative method of testing prior to reporting results for the following antimicrobial/organism combination(s):
- Amikacin: P. mirabilis, P. vulgaris group
- Ampicillin/sulbactam: Enterobacterales
- Aztreonam: P. mirabilis, S. marcescens, P. aeruginosa
- Cefepime: E. cloacae complex, P. vulgaris group
- Ceftazidime: C. freundii complex, C. koseri, E. coli, P. vulgaris group
- Ceftriaxone: P. vulgaris group, S. marcescens
- Ciprofloxacin: C. koseri, P. vulgaris group
- Ertapenem: C. freundii complex, C. koseri, E. cloacae complex, P. vulgaris group, S. marcescens,
- Meropenem: E. cloacae complex, P. vulgaris group, S. marcescens
- Piperacillin/tazobactam: C. freundii complex, C. koseri, P. mirabilis, P. vulgaris, S. marcescens
- Tobramycin: K. pneumoniae, P. mirabilis, S. marcescens
- Trimethoprim/sulfamethoxazole: C. freundii complex, E. cloacae complex, K. oxytoca, P. mirabilis, P. vulgaris group
The ability of the PhAST BC GN Panel to detect resistance in the following antimicrobial/organism combinations is unknown because an insufficient number of resistant isolates were available during the clinical study:
- Amikacin: C. koseri, K. aerogenes, K. oxytoca
- Aztreonam: C. koseri, P. vulgaris group
- Cefepime: C. koseri
- Ceftazidime: P. mirabilis
- Ceftazidime/avibactam: C. freundii complex, C. koseri, K. aerogenes, P. mirabilis, P. vulgaris group,
- Ertapenem: P. mirabilis
- Gentamicin: C. koseri, P. vulgaris group, S. marcescens
- Levofloxacin: C. koseri, P. vulgaris group
- Meropenem: C. koseri, P. mirabilis
- Tobramycin: C. koseri, P. vulgaris group
- Trimethoprim/sulfamethoxazole: C. koseri
# D Special Instrument Requirements:
K253316 - Page 5 of 34
{6}
The PhAST BC GN Panel must be used with the PhAST instrument (PhAST instrument with the Phase 1 releases: cloud release dcs1.0.11 and instrument release v1.0.7.).
# IV Device/System Characteristics:
# A Device Description:
The PhAST BC GN Panel is an in vitro diagnostic assay for qualitative (breakpoint) determination of antimicrobial susceptibility testing (AST) of pathogenic gram-negative bacteria from positive blood cultures (BC) and is intended to be used with the PhAST instrument. The PhAST BC GN Panel is a phenotypic test that utilizes video microscopy and analysis of single-cell phenotypes to determine susceptibility based on categorical interpretation. The PhAST BC GN Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and reported as monomicrobial gram-negative specimens by Gram staining. The PhAST BC GN Panel does not provide organism identification. Organism identity, using an FDA-cleared method, must be entered into the PhAST instrument or PhAST web application before categorical interpretation (Susceptible, Susceptible-Dose Dependent, Intermediate, Resistant, S/SDD/I/R) of antimicrobial susceptibility is reported. The PhAST BC GN Panel contains 15 antimicrobials. The workflow from a direct positive blood culture specimen requires less than one minute of hands-on time. PhAST BC GN Panel cartridges are run on-demand utilizing one of four independent, random-access bays on the PhAST instrument.
# B Principle of Operation:
The PhAST instrument runs the PhAST BC GN Panel. The PhAST BC GN Panel is run by loading a positive blood culture sample onto an assay-specific, single-use cartridge, which is then inserted into the instrument. Each cartridge is loaded with one patient sample. The PhAST BC GN Panel cartridge contains 15 antimicrobials, each at 2 to 5 concentrations, dried down across 45 imaging chambers. An additional eight (8) chambers are used for positive controls and contain no antimicrobials. The PhAST BC GN Panel uses video-microscopy to automatically image and quantify a range of phenotypes of bacteria in the imaging chambers on the cartridge and how these phenotypes change in response to the antimicrobials relative to the (antimicrobial-free) growth control chambers. Video-microscopy data is gathered at three (3) time points and sent to PhAST's cloud for analysis. A susceptibility report is generated after the user enters the organism identity into the PhAST instrument or PhAST web application. The organism identity is determined via an external source, such as an FDA-cleared rapid molecular identification method or MALDI-TOF. The PhAST instrument can run four (4) PhAST BC GN Panel cartridges simultaneously and with random access.
To determine the susceptibility of organisms to a given antimicrobial, the Diagnostic Algorithm within the Data Analysis Pipeline compares phenotypes of the organisms in the no-antimicrobial, positive-growth control to the same phenotypes of the organisms under antimicrobial treatment at the different antimicrobial concentrations on the cartridge. From this comparison, the Diagnostic Algorithm computes a resistance score, which is used to determine the categorical interpretation (S, SDD, I, or R). The resistance score is not reported to the user. For samples where there remains significant ambiguity in the categorical interpretation (i.e., the S/SDD/I/R probabilities are below a given threshold), a "No Call: Inconclusive" result is reported to minimize the possibility of the user obtaining incorrect results. Cases of no-growth or low-growth, for which a reliable diagnostic output is precluded, are reported as "Inadequate Growth"
K253316 - Page 6 of 34
{7}
(for the entire sample). Cases in which a technical error prevents the result from being provided for a given antimicrobial, for example due to a bubble in the image or a focus problem, are reported as “Technical Error” (for that antimicrobial). In this case, the user is instructed to run the assay again one time using a new cartridge and, if the problem occurs again, to test with an alternate method.
The specific phenotypes utilized by the Diagnostic Algorithm vary by antimicrobial and by organism and are drawn from a broad range of phenotypes related to growth, morphology, and motility that are quantified from the video microscopy data at the single-cell level.
# C Instrument Description Information:
1. Instrument Name:
PhAST instrument
2. Specimen (Positive Blood Culture) Identification:
A specimen identification label is applied to the PhAST BC GN Panel cartridge in the designated accession number barcode area, and to initiate a run, the user enters the specimen identification into the PhAST instrument software via the provided external barcode scanner or by manually typing it using the touchscreen's onscreen keyboard.
3. Specimen Sampling and Handling:
The operator follows a streamlined workflow guided by the PhAST instrument touchscreen interface for sample preparation from positive blood cultures. The sample is prepared from a gram-negative and monomicrobial positive blood culture using a sterile collection method capable of obtaining approximately 1 mL of blood culture. The operator sterilizes the rubber septum with an alcohol pad and collects 1 mL of positive blood culture using either a 3 mL syringe with 18-gauge needle or subculture needle into a sterile 1.5 mL microcentrifuge tube.
A 60 μL aliquot of the positive blood culture is transferred to a lysis buffer tube containing red blood cell lysis buffer using a P200 micropipette. The tube is inverted 10 times to ensure proper mixing without vortexing, and visual inspection confirms adequate mixing. Using the same pipette with a new sterile tip, 60 μL of the culture-lysis buffer mixture is collected and dispensed into the PhAST BC GN Panel cartridge through the designated inlet. The cartridge inlet is sealed by folding over the attached blue cap. The operator scans the cartridge barcode and accession number barcode, removes the protective film from the cartridge underside, and loads the cartridge into an available bay on the PhAST instrument. The system automatically performs sample metering, dilution with distilled water, mixing with dried Mueller Hinton broth using a magnetic mixing ball, and routes the prepared sample through microchannels to 56 imaging chambers containing dried antimicrobials at various concentrations for automated antimicrobial susceptibility testing.
K253316 - Page 7 of 34
{8}
# 4. Calibration:
The PhAST instrument automatically performs comprehensive self-checks at start-up and every 24 hours, including camera exposure calibration, linear actuator homing, pressure sensor calibration, and continuous monitoring of pneumatic, optical, and electronic subsystems. At the start of each assay, the system conducts Assay Start quality control checks including cartridge positioning calibration and pressure sensor verification, and initiation of AST testing can proceed only if all system functions and quality control checks pass successfully.
# 5. Quality Control:
Quality control (QC) is performed by the operator using manufacturer-specified organisms (Escherichia coli ATCC 25922, Klebsiella pneumoniae ATCC 700603, Klebsiella pneumoniae CDC AR-0097, Klebsiella pneumoniae IHMA 23-084741-01, Enterobacter cloacae CDC AR-0432, Escherichia coli CDC AR-0434, and Escherichia coli ATCC NCTC13353) appropriate for each antimicrobial agent. The QC AST reports providing “Pass” or “Fail” results based on expected categorical interpretations for each strain and antimicrobial combination.
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
Accelerate Pheno System, Accelerate PhenoTest BC
# B Predicate 510(k) Number(s):
DEN160032
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K253316 (Device) | DEN160032 (Predicate) |
| --- | --- | --- |
| Device Trade Name | PhAST instrument and PhAST Blood Culture Gram-negative Panel | Accelerate Pheno System, Accelerate PhenoTest BC Kit |
| General Device Characteristic Similarities | | |
| Intended Use | The PhAST Blood Culture Gram-negative Panel (PhAST BC GN Panel) is an in vitro diagnostic assay for qualitative (breakpoint) determination of antimicrobial susceptibility testing (AST) of pathogenic gram-negative bacteria from positive blood cultures (BC) and is intended to be used with the PhAST instrument. The PhAST BC GN Panel is a phenotypic test that utilizes video microscopy and analysis of single-cell phenotypes to determine susceptibility based on categorical interpretation. The PhAST BC GN Panel does not provide organism | The Accelerate PhenoTest BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno system. The Accelerate PhenoTest BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the |
K253316 - Page 8 of 34
{9}
| | identification. The PhAST BC GN Panel is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system and reported as monomicrobial gram-negative specimens by Gram staining. Organism identity must be entered into the PhAST instrument or PhAST web application before categorical interpretation (Susceptible, Susceptible-Dose Dependent, Intermediate, Resistant, S/SDD/I/R) of antimicrobial susceptibility is reported. The PhAST BC GN Panel contains 15 antimicrobials. The PhAST BC GN Panel is intended to aid in the diagnosis and treatment of individuals suspected of bloodstream infection by a healthcare provider. Results should not be used as the sole basis for patient management decisions. This test is performed by laboratory health professionals in a clinical diagnostic setting. Results may be used as an aid to clinicians in determining appropriate antimicrobial therapy. Test results from the PhAST BC GN Panel should be interpreted in conjunction with other clinical and laboratory findings. Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing. Sub-culturing is necessary to support further testing for: bacteria and antimicrobials not on the PhAST BC GN Panel, inconclusive results, epidemiologic testing, recovery of organisms present in positive blood cultures samples, and susceptibility testing of bacteria in polymicrobial samples. | appropriate detected bacterial organisms. The Accelerate PhenoTest BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results. |
| --- | --- | --- |
| Sample Type | Blood cultures signaled as positive by a continuous monitoring blood culture system. | Same |
| Inoculation Method | Automated | Same |
| Read Method | Automated | Same |
| Technology | Single-cell, time-lapse microscopy | Similar |
K253316 - Page 9 of 34
{10}
| General Device Characteristic Differences | | |
| --- | --- | --- |
| IVD Functions | AST | ID and AST |
| Indicated Antimicrobials | Amikacin Ampicillin/sulbactam Aztreonam Cefepime Ceftazidime Ceftazidime/avibactam Ceftriaxone Ciprofloxacin Ertapenem Gentamicin Levofloxacin Meropenem Piperacillin/tazobactam Tobramycin Trimethoprim/ sulfamethoxazole | Amikacin Ampicillin/sulbactam Aztreonam Cefepime Ceftazidime Ceftriaxone Ciprofloxacin Ertapenem Gentamicin Meropenem Piperacillin/tazobactam Tobramycin Additional Gram-positive antimicrobials are also included on the Accelerate PhenoTest BC kit: Ampicillin Ceftaroline Daptomycin Linezolid Vancomycin |
| Tested Organism | Enterobacterales, *Pseudomonas aeruginosa*, *Acinetobacter* spp. | Enterobacterales, *Pseudomonas aeruginosa*, *Acinetobacter* spp., *Enterococcus* spp., *Staphylococcus* spp. |
| Sample Preparation | Short (<1 min) pre-processing step for red blood cell lysis prior to sample loading followed by automated reading | Automated |
| Time to AST Result | Range 91–154 minutes depending on drug and organism tested | Approximately 7 hours |
| AST Results | Report results as categorical interpretation | Report results as minimum inhibitory concentration (MIC) with categorical interpretation |
In addition to the similarities and differences between the candidate and predicate device listed in the table above, the candidate device has an authorized PCCP to support the removal of specific labeling limitations solely related to insufficient resistant isolate representation when additional resistant isolates become available. The plan outlines the specific procedures and acceptance criteria that PhAST Corp. intends to use to evaluate whether labeling limitations for specific antimicrobial/organism combinations on the PhAST BC GN Panel can be removed based on supplemental resistant isolate data. The PCCP requires a minimum of 5 resistant isolates by reference method with no observed very major errors among the supporting resistant isolates, and continued compliance with the AST Special Controls Guidance criteria including categorical agreement >89.9% and error rates (i.e., very major error rate ≤2%, major error rate ≤3%) for each combination seeking limitation removal. PhAST Corp. will update the PhAST BC GN Panel labeling to remove the specific limitation after validation confirms that performance criteria continue to be met.
K253316 - Page 10 of 34
{11}
## **VI Standards/Guidance Documents Referenced:**
- • FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009).
- • CLSI M100-Ed 36, Performance Standards for Antimicrobial Susceptibility Testing; 2025.
- • CLSI EP07 12$^{th}$ Edition, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically. 2024.
- • CLSI EP37 1$^{st}$ Edition, Supplemental Tables for Interference Testing in Clinical Chemistry.
- • IEC 61326-2-6 Edition 3.0 2020-10 19-43, Electrical equipment for measurement control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment.
- • IEC 61326-1 Edition 3.0 2020-10 19-42, Electrical equipment for measurement control and laboratory use - EMC requirements - Part 1: General requirements.
- • IEC 61010-1 Edition 3.1 2017-01 CONSOLIDATED VERSION 19-34, Safety requirements for electrical equipment for measurement control and laboratory use - Part 1: General requirements [Including: Corrigendum 1 (2019)].
- • ANSI AAMI IEC 62304:2006/A1:2016 13-79, Medical device software – Software life cycle processes [Including Amendment 1 (2016)].
- • ANSI AAMI IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, 5-129, Medical devices – Part 1: Application of usability engineering to medical devices.
- • ANSI AAMI ISO 14971 Third Edition 2019-12 5-125, Medical devices – Application of risk management to medical devices.
- • ANSI AAMI IEC TIR800002-1:2009, 13-34, Medical device software - Part 1: Guidance on the application of ISO 14971 to medical device software.
- • ANSI ISA 62443-4-1-2018, 13-119, Security for industrial automation and control systems Part 4-1: Product security development life-cycle requirements.
- • IEC 81001-5-1 Edition 1.0 2021-12, 13-122, Health software and health IT systems safety effectiveness and security - Part 5-1: Security - Activities in the product life cycle.
- • UL ANSI 2900-1 1$^{st}$ Edition 2017, 13-96, Standard for Safety Standard for Software Cybersecurity Network-Connectable Products Part 1: General Requirements.
- • AAMI TIR57:2016, 13-83, Principles for medical device security - Risk management.
- • AAMI TIR97:2019, 13-112, Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements.
- • ISO 20417 1$^{st}$ edition 2021-04 Corrected version 2021-12, 5-135, Medical Devices – Information to be supplied by the manufacturer.
## **VII Performance Characteristics (if/when applicable):**
K253316 - Page 11 of 34
{12}
# A Analytical Performance:
## 1. Precision/Reproducibility:
A reproducibility study of the PhAST BC GN Panel with the PhAST instrument was conducted using contrived positive blood culture bottles prepared with sixteen (16) bacterial strains from *Acinetobacter* spp., Enterobacterales, and *P. aeruginosa* with at least one susceptible and one resistant strain within each organism reporting group. Triplicate samples from each contrived blood culture were tested at three sites over three consecutive days to assess reproducibility across instruments, locations, operators, and testing days ($\geq 10$ isolates $\times 3$ sites $\times 3$ replicates $\times 3$ days = $\geq 270$ results per antimicrobial). Reproducibility was determined for each organism/antimicrobial combination as the Categorical Agreement (CA) between each result and the categorical mode across all tests for that organism/antimicrobial combination. Both inter-site and intra-site reproducibility were evaluated for each antimicrobial on the PhAST BC GN Panel.
The reproducibility results demonstrated acceptable CA performance of greater than 95% for each antimicrobial. A summary of the reproducibility results is provided in **Table 1**.
**Table 1. Summary of Reproducibility Results of the PhAST Blood Culture Gram-Negative Panel with the PhAST Instrument**
| Antimicrobial | No. isolates Tested | Total No. Tests | Categorical Agreement (CA) |
| --- | --- | --- | --- |
| Amikacin | 11 | 309 | 97.1% (300/309) |
| Ampicillin/sulbactam | 11 | 307 | 99.0% (304/307) |
| Aztreonam | 11 | 339 | 100% (339/339) |
| Cefepime | 13 | 393 | 99.7% (422/423) |
| Ceftazidime | 16 | 475 | 99.6% (473/475) |
| Ceftazidime/avibactam | 13 | 382 | 100% (382/382) |
| Ceftriaxone | 11 | 333 | 100% (333/333) |
| Ciprofloxacin | 13 | 394 | 100% (394/394) |
| Ertapenem | 9 | 284 | 97.5% (277/284) |
| Gentamicin | 11 | 335 | 100% (335/335) |
| Levofloxacin | 13 | 396 | 99.7% (395/396) |
| Meropenem | 16 | 467 | 99.8% (466/467) |
| Piperacillin/tazobactam | 16 | 473 | 97.3% (460/473) |
| Tobramycin | 13 | 382 | 98.7% (377/382) |
| Trimethoprim/sulfamethoxazole | 11 | 336 | 100% (336/336) |
## 2. Inoculum Density Study
An inoculum density study was performed to determine the effect of organism concentration in a positive blood culture on the performance of the PhAST BC GN Panel with the PhAST instrument and the ability of the PhAST instrument to prepare an appropriate inoculum regardless of the starting concentration. A panel of seventeen (17) bacterial strains including at least one susceptible and one resistant strain from each organism reporting group (i.e., Enterobacterales, *P. aeruginosa*, and *Acinetobacter* spp.) was tested at four target inoculum concentrations: 1, 10, 100, and 1000 CFU/mL. Each strain was spiked into whole human blood at each concentration and incubated in BacT/ALERT FA Plus Aerobic blood culture
K253316 - Page 12 of 34
{13}
bottles until flagged positive. Four biological replicates were prepared for each strain and inoculum concentration to meet the minimum target of 45 data points per antimicrobial and target concentration.
The organism concentration in each contrived sample was confirmed by colony count. PhAST BC GN Panel results were compared to modal reference broth microdilution (BMD) categorical interpretation results (S, SDD, I, or R). The PhAST BC GN Panel with the PhAST instrument demonstrated acceptable performance (Tables 2), with a CA of >90% for each antimicrobial and target inoculum concentrations.
Table 2. Categorical Agreement Results for each Target Inoculum Concentration
| Antimicrobial | Categorical Agreement (%) | | | |
| --- | --- | --- | --- | --- |
| | 1 CFU/mL | 10 CFU/mL | 100 CFU/mL | 1,000 CFU/mL |
| Amikacin | 45/45 (100%) | 46/46 (100%) | 45/46 (97.8%) | 46/47 (97.9%) |
| Ampicillin/sulbactam | 44/48 (91.7%) | 49/50 (98.0%) | 51/51 (100%) | 48/49 (98.0%) |
| Aztreonam | 48/48 (100%) | 48/48 (100%) | 48/48 (100%) | 47/47 (100%) |
| Cefepime | 52/56 (92.9%) | 49/53 (92.5%) | 52/56 (92.9%) | 51/55 (92.7%) |
| Ceftazidime | 67/67 (100%) | 63/64 (98.4%) | 67/67 (100%) | 67/67 (100%) |
| Ceftazidime/avibactam | 56/56 (100%) | 54/54 (100%) | 55/55 (100%) | 55/55 (100%) |
| Ceftriaxone | 47/47 (100%) | 48/48 (100%) | 48/48 (100%) | 46/46 (100%) |
| Ciprofloxacin | 55/55 (100%) | 52/52 (100%) | 54/54 (100%) | 54/54 (100%) |
| Ertapenem | 46/46 (100%) | 46/46 (100%) | 46/46 (100%) | 45/45 (100%) |
| Gentamicin | 48/48 (100%) | 47/47 (100%) | 48/48 (100%) | 46/46 (100%) |
| Levofloxacin | 56/56 (100%) | 53/54 (98.1%) | 54/54 (100%) | 54/54 (100%) |
| Meropenem | 68/68 (100%) | 63/63 (100%) | 67/67 (100%) | 63/64 (98.4%) |
| Piperacillin/tazobactam | 65/66 (98.5%) | 66/66 (100%) | 65/68 (95.6%) | 66/66 (100%) |
| Tobramycin | 45/46 (97.8%) | 46/47 (97.9%) | 51/51 (100%) | 48/50 (96.0%) |
| Trimethoprim/sulfamethoxazole | 48/48 (100%) | 48/48 (100%) | 47/47 (100%) | 47/47 (100%) |
### 3. Linearity:
Not applicable
### 4. Analytical Specificity/Interference:
#### Endogenous/Exogenous Interference Studies:
An interfering substances study was performed to evaluate if substances naturally present or artificially introduced into blood culture bottles affect the performance of the PhAST BC GN Panel with the PhAST instrument. A panel of seventeen (17) bacterial strains including at least one susceptible and one resistant strain from each organism reporting group (i.e., Enterobacterales, P. aeruginosa, and Acinetobacter spp.) was evaluated. The nine interfering substances were spiked into BacT/ALERT FA Plus Aerobic blood culture bottles at or above clinically relevant concentrations along with representative organisms and tested in triplicate. Control bottles were seeded with organism and no potential interferent. Bottles were incubated in a continuous monitoring blood culture system until positivity. As this was a method-to-method comparison determined by comparing categorical interpretation results (S,
K253316 - Page 13 of 34
{14}
SDD, I, or R) from interferent-containing samples with modal results from control samples, CA of $\geq 95\%$ was deemed acceptable.
Overall, the presence of high concentrations of the tested interferents in blood cultures did not significantly interfere with PhAST BC GN Panel results, with most combinations achieving acceptable performance (CA $\geq 95\%$) (**Tables 3 and 4**). Specific antimicrobial/organism combinations that demonstrated CA $< 95\%$ are discussed below.
Ertapenem, ampicillin/sulbactam, piperacillin/tazobactam, and tobramycin in the presence of triglycerides at 5 mg/mL had a CA $< 95\%$ for Enterobacterales species. Repeat testing of triglycerides at 1.67 mg/mL demonstrated acceptable performance.
Tobramycin in the presence of conjugated bilirubin at 400 mg/L had a CA $< 95\%$ for *P. aeruginosa*. Repeat testing of conjugated bilirubin at 133 mg/L demonstrated acceptable performance.
Footnotes in the performance tables (**Tables 3 and 4**) were included in device labeling to address the performance concerns of these interfering substances.
**Table 3. Overall Performance with Potential Interfering Substances (Part 1)**
| Antimicrobial | Categorical Agreement (%) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Conjugated Bilirubin ($\geq 400$ mg/L) | Unconjugated Bilirubin ($\geq 400$ mg/L) | Gamma Globulin ($\geq 50$ g/L) | Triglycerides ($\geq 5$ mg/mL) | Red Blood Cells ($\geq 20$ g/dL) |
| Amikacin | 41/41 (100%) | 38/38 (100%) | 38/38 (100%) | 33/34 (97.1%) | 37/37 (100%) |
| Ampicillin/sulbactam | 36/36 (100%) | 38/38 (100%) | 38/38 (100%) | 35/38 (92.1%)^{a} | 37/37 (100%) |
| Aztreonam | 36/36 (100%) | 38/38 (100%) | 36/36 (100%) | 36/36 (100%) | 38/38 (100%) |
| Cefepime | 42/42 (100%) | 38/38 (100%) | 39/39 (100%) | 42/42 (100%) | 44/44 (100%) |
| Ceftazidime | 50/50 (100%) | 47/47 (100%) | 48/48 (100%) | 49/50 (98.0%) | 50/50 (100%) |
| Ceftazidime/avibactam | 42/42 (100%) | 38/38 (100%) | 39/39 (100%) | 42/42 (100%) | 44/44 (100%) |
| Ceftriaxone | 36/36 (100%) | 38/38 (100%) | 36/36 (100%) | 36/36 (100%) | 38/38 (100%) |
| Ciprofloxacin | 42/42 (100%) | 38/38 (100%) | 39/39 (100%) | 40/40 (100%) | 41/41 (100%) |
| Ertapenem | 37/37 (100%) | 37/37 (100%) | 38/38 (100%) | 34/36 (94.4%)^{a} | 38/38 (100%) |
| Gentamicin | 41/41 (100%) | 36/37 (97.3%) | 35/36 (97.2%) | 37/37 (100%) | 38/38 (100%) |
| Levofloxacin | 40/41 (97.6%) | 36/37 (97.3%) | 37/37 (100%) | 42/42 (100%) | 41/41 (100%) |
| Meropenem | 50/50 (100%) | 44/44 (100%) | 48/48 (100%) | 49/49 (100%) | 49/50 (98.0%) |
| Piperacillin/tazobactam | 48/50 (96.0%) | 46/46 (100%) | 47/47 (100%) | 42/46 (91.3%)^{a} | 48/49 (98.0%) |
K253316 - Page 14 of 34
{15}
| Tobramycin | 36/39 (92.3%)^{b} | 40/41 (97.6%) | 42/42 (100%) | 33/35 (94.3%)^{a} | 40/40 (100%) |
| --- | --- | --- | --- | --- | --- |
| Trimethoprim/ sulfamethoxazole | 41/41 (100%) | 38/38 (100%) | 36/36 (100%) | 36/36 (100%) | 38/38 (100%) |
$^{a}$ Ertapenem, ampicillin/sulbactam, tobramycin and piperacillin/tazobactam demonstrated <95% Categorical Agreement (CA) at the tested concentration due to Enterobacterales isolates. When retested with 1.67 mg/mL triglycerides, performance was acceptable.
$^{b}$ Tobramycin demonstrated <95% CA at the tested concentration due to one P. aeruginosa isolate. When retested with 133 mg/L conjugated bilirubin, performance was acceptable.
Table 4. Overall Performance with Potential Interfering Substances (Part 2)
| Antimicrobial | Categorical Agreement (%) | | | |
| --- | --- | --- | --- | --- |
| | White Blood Cells (≥12,000 WBC/μL) | Platelets (≥400,000/μL) | Heparin (≥3000 Units/L) | SPS (≥0.1% w/v) |
| Amikacin | 37/37 (100%) | 36/36 (100%) | 41/41 (100%) | 36/36 (100%) |
| Ampicillin/sulbactam | 48/49 (98.0%) | 48/49 (98.0%) | 51/51 (100%) | 45/45 (100%) |
| Aztreonam | 36/36 (100%) | 36/36 (100%) | 39/39 (100%) | 38/39 (97.4%) |
| Cefepime | 42/42 (100%) | 42/42 (100%) | 42/42 (100%) | 39/39 (100%) |
| Ceftazidime | 42/42 (100%) | 35/36 (97.2%) | 48/48 (100%) | 47/47 (100%) |
| Ceftazidime/avibactam | 42/42 (100%) | 42/42 (100%) | 42/42 (100%) | 39/39 (100%) |
| Ceftriaxone | 50/50 (100%) | 51/51 (100%) | 60/60 (100%) | 48/48 (100%) |
| Ciprofloxacin | 42/42 (100%) | 42/42 (100%) | 42/42 (100%) | 39/39 (100%) |
| Ertapenem | 38/38 (100%) | 36/36 (100%) | 41/41 (100%) | 36/36 (100%) |
| Gentamicin | 36/36 (100%) | 36/36 (100%) | 36/36 (100%) | 40/40 (100%) |
| Levofloxacin | 38/38 (100%) | 36/36 (100%) | 41/41 (100%) | 36/36 (100%) |
| Meropenem | 41/42 (97.6%) | 42/42 (100%) | 41/42 (97.6%) | 39/39 (100%) |
| Piperacillin/tazobactam | 38/38 (100%) | 36/36 (100%) | 41/41 (100%) | 36/36 (100%) |
| Tobramycin | 39/39 (100%) | 40/40 (100%) | 38/40 (95.0%) | 36/36 (100%) |
| Trimethoprim/ sulfamethoxazole | 51/51 (100%) | 48/49 (98.0%) | 51/51 (100%) | 48/48 (100%) |
### Antimicrobial Interference Studies:
A study was performed to evaluate if antimicrobials present in blood culture bottles affect the performance of PhAST BC GN Panel with the PhAST instrument using seeded positive blood culture samples with and without interfering antimicrobials. Four representative antimicrobials from different drug classes were tested including ciprofloxacin (fluoroquinolones), ceftriaxone (cephalosporins), ampicillin/sulbactam (β-lactam/β-lactamase
K253316 - Page 15 of 34
{16}
inhibitor), and amikacin (aminoglycosides). A panel of ten (10) bacterial strains including at least one resistant strain from each organism reporting group (i.e., Enterobacterales, P. aeruginosa, and Acinetobacter spp.) was tested for each interfering antimicrobial. Antimicrobial interferents were spiked into resin-free BacT/ALERT SA Standard Aerobic blood culture bottles at the highest drug concentrations as recommended by CLSI EP37. Control bottles were seeded with organisms and no antimicrobial interferent. As this was a method-to-method comparison determined by comparing categorical interpretation results (S, SDD, I, or R) from interferent-containing samples with modal results from control samples, CA of ≥95% was deemed acceptable.
Overall, the presence of high concentrations of the tested interferents in blood cultures does not significantly interfere with the results of the PhAST BC GN Panel, with most combinations achieving acceptable performance (CA ≥95%) (Table 5). Specific antimicrobial/organism combinations that demonstrated CA <95% are discussed below.
Ceftazidime/avibactam in the presence of amikacin had a CA <95% due to a single very major error with one E. cloacae isolate, which was determined to be a technical error (i.e., focus problem) rather than antimicrobial interference.
Piperacillin/tazobactam in the presence of ceftriaxone had a CA <95% due to a single minor error with one replicate of a single K. pneumoniae isolate.
Tobramycin in the presence of ciprofloxacin had a CA <95% due to minor errors with one P. aeruginosa isolate and one K. pneumoniae isolate. Supplemental testing of two additional P. aeruginosa isolates as well as one K. pneumoniae isolate in the presence of ciprofloxacin at three-fold lower concentration (0.13 mg/dL) demonstrated acceptable performance.
Tobramycin in the presence of amikacin had a CA <95% due to three (3) minor errors with one P. aeruginosa isolate. Supplemental testing of two additional P. aeruginosa isolates demonstrated acceptable performance.
Amikacin in the presence of ciprofloxacin (0.4 mg/dL) and ampicillin/sulbactam (2.5/1.25 mg/dL) had a CA <95% due to one K. oxytoca isolate. Repeat testing with three-fold lower concentrations of ciprofloxacin (0.13 mg/dL) or ampicillin/sulbactam (0.83/0.42 mg/dL), demonstrated <95% CA due to minor errors of the same K. oxytoca isolate.
Footnotes in the performance table (Table 5) were included in device labeling to address the performance concerns of these specific interferences.
Table 5. Overall Performance with Potential Interfering Antimicrobials.
| Antimicrobial | Categorical Agreement (%) | | | |
| --- | --- | --- | --- | --- |
| | Ciprofloxacin (0.4 mg/dL) | Ceftriaxone (28 mg/dL) | Ampicillin/ sulbactam (2.5/1.25 mg/dL) | Amikacin (4.8 mg/dL) |
| Amikacin | 16/18 (88.9%)^{a} | 15/15 (100%) | 15/18 (83.3%)^{a} | 15/15 (100%) |
| Ampicillin/sulbactam | 17/17 (100%) | 16/16 (100%) | 20/20 (100%) | 15/15 (100%) |
K253316 - Page 16 of 34
{17}
| Aztreonam | 20/20 (100%) | 19/19 (100%) | 19/19 (100%) | 15/15 (100%) |
| --- | --- | --- | --- | --- |
| Cefepime | 25/25 (100%) | 19/19 (100%) | 19/19 (100%) | 20/20 (100%) |
| Ceftazidime | 25/25 (100%) | 19/19 (100%) | 25/25 (100%) | 30/30 (100%) |
| Ceftazidime/avibactam | 24/24 (100%) | 18/18 (100%) | 18/18 (100%) | 14/15 (93.3%)^{b} |
| Ceftriaxone | 20/20 (100%) | 18/18 (100%) | 20/20 (100%) | 17/17 (100%) |
| Ciprofloxacin | 25/25 (100%) | 19/19 (100%) | 18/18 (100%) | 20/20 (100%) |
| Ertapenem | 20/20 (100%) | 19/19 (100%) | 19/19 (100%) | 17/17 (100%) |
| Gentamicin | 20/20 (100%) | 19/19 (100%) | 20/20 (100%) | 15/15 (100%) |
| Levofloxacin | 24/25 (96.0%) | 19/19 (100%) | 20/20 (100%) | 20/20 (100%) |
| Meropenem | 24/24 (100%) | 18/18 (100%) | 25/25 (100%) | 27/28 (96.4%) |
| Piperacillin/tazobactam | 24/24 (100%) | 18/19 (94.7%)^{c} | 25/25 (100%) | 28/28 (100%) |
| Tobramycin | 16/22 (72.7%)^{d} | 14/14 (100%) | 18/18 (100%) | 17/20 (85.0%)^{e} |
| Trimethoprim/sulfamethoxazole | 20/20 (100%) | 19/19 (100%) | 20/20 (100%) | 15/15 (100%) |
$^{a}$ Amikacin in the presence of ciprofloxacin (0.4 mg/dL) or ampicillin/sulbactam (2.5/1.25 mg/dL) had a CA <95% due to one K. oxytoca strain tested. Repeat testing with three-fold lower concentrations of ciprofloxacin (0.13 mg/dL) or ampicillin/sulbactam (0.83/0.42 mg/dL), demonstrated CA <95% due to minor errors of the same K. oxytoca strain.
$^{b}$ Ceftazidime/avibactam in the presence of amikacin had a CA <95% due to a single very major error, which was determined to be a technical error.
$^{c}$ Piperacillin/tazobactam in the presence of ceftriaxone had a CA <95% due to a single minor error with one replicate of a single K. pneumonia isolate.
$^{d}$ Tobramycin in the presence of ciprofloxacin (0.4 mg/dL) had a CA <95% due to minor errors with one P. aeruginosa isolate and one K. pneumoniae isolate. Supplemental testing of two additional P. aeruginosa isolates as well as one K. pneumoniae isolate at three-fold lower ciprofloxacin concentration (0.13 mg/dL) demonstrated acceptable performance.
$^{e}$ Tobramycin in the presence of amikacin had a CA <95% due to three minor errors with one P. aeruginosa isolate. Supplemental testing of two additional P. aeruginosa isolates demonstrated acceptable performance.
# 5. Media Equivalency Study:
The PhAST BC GN Panel with the PhAST instrument was tested for compatibility with eleven different blood culture bottles from two manufacturers (BD BACTEC and bioMérieux BacT/ALERT). A panel of seventeen (17) bacterial strains including at least one susceptible and one resistant strain from each organism reporting group (i.e., Enterobacterales, P. aeruginosa, and Acinetobacter spp.) were individually mixed with human whole blood, seeded directly into blood culture bottles, and incubated for growth in continuous monitoring blood culture systems positivity. At least five (5) replicates of each resulting positive blood culture were tested using the PhAST BC GN Panel and PhAST instrument. Strict aerobic species (P. aeruginosa and Acinetobacter spp.) were not tested in anaerobic bottles.
K253316 - Page 17 of 34
{18}
PhAST BC GN Panel results of each bottle type with each antimicrobial/organism combination were compared to the modal reference broth microdilution (BMD) categorical interpretation results (S, SDD, I, or R). PhAST BC GN Panel results were also combined and analyzed based on blood culture bottle function and composition (i.e., aerobic with resin, aerobic without resin, pediatric aerobic with resin, anaerobic with resin, anaerobic without resin, and lytic anaerobic). The number of tests for each aggregated bottle type for each antimicrobial ranged from 51 to 82. Results for all antimicrobial/organism combinations demonstrated acceptable performance (≥90% CA) for all bottle types.
# 6. Detection Limit and Assay Reportable Range:
Not applicable
# 7. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Quality control (QC) testing was performed each day that testing was conducted. The QC panel consists of seven (7) CLSI recommended or independently validated QC strains, which included strains to provide one susceptible (S) and one resistant (R) QC result for each of the 15 antimicrobials on the panel. QC strains for each antimicrobial were tested a sufficient number of times (i.e., at least 20 times/site) at each testing site using the PhAST BC GN Panel with the PhAST instrument, including the reference site using the broth microdilution reference method. QC organisms were tested on a rotating basis.
QC expected ranges and results for the PhAST BC GN Panel are summarized in Table 6. For all antimicrobials, greater than 95% of results obtained during the clinical study were within the expected categorical interpretation (i.e., susceptible or resistant), which is acceptable.
Table 6. QC Results for the PhAST Blood Culture Gram-Negative Panel
| Antimicrobial | QC Strains | Expected PhAST BC GN Panel Categorical Interpretation | PhAST BC GN Panel Results in Agreement (%) | Expected CLSI/Validated Reference MIC Range (μg/mL) | Reference MIC Results In Range (%) |
| --- | --- | --- | --- | --- | --- |
| Amikacin | E. coli ATCC 25922 | S | 93/94 (98.9%) | 0.5 - 4 | 142/142 (100%) |
| | K. pneumoniae CDC AR-0097^{a} | R | 92/94 (97.9%) | 128 - 512 | 142/142 (100%) |
| Ampicillin/sulbactam | E. coli ATCC 25922 | S | 93/95 (97.9%) | 2/1 - 8/4 | 142/142 (100%) |
| | K. pneumoniae IHMA 23-084731-01^{a} | R | 93/93 (100%) | 64 - 256 | 150/150 (100%) |
| Aztreonam | E. coli ATCC 25922 | S | 93/93 (100%) | 0.06 - 0.5 | 142/142 (100%) |
| | E. coli NCTC 13353^{a} | R | 93/94 (98.9%) | 128 - 512 | 146/146 (100%) |
| Cefepime | K. pneumoniae ATCC 700603 | S | 93/93 (100%) | 0.5 - 2 | 144/144 (100%) |
| | E. coli NCTC 13353^{a} | R | 93/94 (98.9%) | ≥64 | 146/146 (100%) |
| Ceftazidime | E. coli ATCC 25922 | S | 93/93 (100%) | <0.12 - 0.5 | 142/142 (100%) |
| | E. coli NCTC 13353^{a} | R | 93/94 (98.9%) | 32 - 256 | 146/146 (100%) |
| Ceftazidime/avibactam | E. coli NCTC 13353^{a} | S | 93/93 (100%) | 0.12 - 0.5 | 146/146 (100%) |
| | E. coli CDC AR-0434^{a} | R | 91/93 (97.8%) | 32 - 128 | 152/152 (100%) |
| Ceftriaxone | E. coli ATCC 25922 | S | 93/93 (100%) | 0.03 - 0.12 | 142/142 (100%) |
| | K. pneumoniae ATCC 700603^{a} | R | 93/93 (100%) | 2 - 16 | 144/144 (100%) |
| Ciprofloxacin | E. cloacae CDC AR-0432^{a} | S | 93/93 (100%) | 0.03 - 0.12 | 144/144 (100%) |
| | E. coli NCTC 13353^{a} | R | 93/93 (100%) | 8 - 32 | 146/146 (100%) |
| Ertapenem | E. coli NCTC 13353^{a} | S | 93/93 (100%) | 0.06 - 0.5 | 146/146 (100%) |
K253316 - Page 18 of 34
{19}
| | K. pneumoniae IHMA 23-084731-01^{a} | R | 93/94 (98.9%) | 2 - 8 | 150/150 (100%) |
| --- | --- | --- | --- | --- | --- |
| Gentamicin | E. cloacae CDC AR-0432^{a} | S | 93/95 (97.9%) | 0.25 - 1 | 144/144 (100%) |
| | E. coli NCTC 13353^{a} | R | 93/94 (98.9%) | 128 - 512 | 146/146 (100%) |
| Levofloxacin | E. cloacae CDC AR-0432^{a} | S | 93/93 (100.0%) | 0.03 - 0.125 | 144/144 (100%) |
| | K. pneumoniae CDC AR-0097^{a} | R | 93/94 (98.9%) | 32 - 128 | 142/142 (100%) |
| Meropenem | K. pneumoniae IHMA 23-084731-01^{a} | S | 92/92 (96.8%) | 0.06 - 0.5 | 144/150 (96.0%) |
| | K. pneumoniae CDC AR-0097^{a} | R | 93/94 (98.9%) | 32 - 128 | 142/142 (100%) |
| Piperacillin/tazobactam | E. coli ATCC 25922 | S | 93/93 (100%) | 1 - 8 | 142/142 (100%) |
| | K. pneumoniae IHMA 23-084731-01^{a} | R | 93/93 (100%) | 128 - 512 | 150/150 (100%) |
| Tobramycin | E. cloacae CDC AR-0432^{a} | S | 93/93 (100%) | 0.25 - 1 | 144/144 (100%) |
| | K. pneumoniae CDC AR-0097^{a} | R | 93/93 (100%) | 32 - 128 | 142/142 (100%) |
| Trimethoprim/sulfamethoxazole | E. cloacae CDC AR-0432^{a} | S | 93/93 (100%) | 0.06 - 0.25 | 144/144 (100%) |
| | E. coli CDC AR-0434^{a} | R | 92/92 (100%) | 16 - 64 | 152/152 (100%) |
$^{a}$ Validation performed for non-CLSI recommended QC strain
### Validation of New QC Strains
To validate the use of seven (7) QC strains (i.e., E. coli ATCC 25922, K. pneumoniae CDC AR-0097, K. pneumoniae IHMA 23-084731-01, E. coli NCTC 13353, K. pneumoniae ATCC 700603, E. coli CDC AR-0434, and E. cloacae CDC AR-0432) with the PhAST BC GN Panel and PhAST instrument, the BMD reference method was performed to establish the expected BMD MIC range. Testing was performed with at least 80 replicates and two reagent lots over two days. For all antimicrobials, >95% of results were within the newly established MIC reference ranges, which is acceptable. The QC strains were then tested with the PhAST BC GN Panel using at least 80 replicates. For all antimicrobials, >95% of results were in agreement with the expected categorical interpretation, which is acceptable. Taken together, the acceptable performance supports the use of the seven QC strains with the PhAST BC GN Panel and PhAST instrument.
### Purity Check
Purity plates were prepared from the positive blood cultures of every sample tested. AST results were only reported for pure isolates; data generated from plates that generated multiple colony morphologies was excluded from analyses.
### Device Failure
Several types of device or test failures were observed and accounted for in the study:
- Assay Start QC Failures: Twenty tests did not initiate due to an automated quality check failure. The system detected these issues, and all 20 samples were successfully re-tested with a new cartridge, yielding valid results.
- User-Cancelled Tests: Two tests were cancelled by the user at external clinical sites.
- The combined “Technical Error” and “No Call: Inconclusive” result rates for each antimicrobial/organism group combination were assessed during the clinical study and was ≤10% unless otherwise noted.
K253316 - Page 19 of 34
{20}
## Growth Failure Rate
There were 17 “Inadequate Growth” results from the total samples tested (1.9%). These full sample results were not included in the performance analysis.
### 8. Sample Stability Study
The purpose of this study was to demonstrate that positive blood culture samples that remain in the continuous monitoring system for up to 5 hours after bottle ring and/or held at room temperature for up to 12 hours after removal from the continuous monitoring system could provide accurate categorical interpretation results using the PhAST GN BC Panel with the PhAST instrument.
Contrived positive blood culture specimens (prepared in BacT/ALERT FA Plus Aerobic blood culture bottles) containing the recommended blood volume were tested across three storage conditions: immediate testing (T0), extended incubator storage (14 hours at 35°C), and mixed storage simulating clinical workflow delays (5 hours at 35°C followed by 14 hours at room temperature). A panel of sixteen (16) bacterial strains including at least one susceptible and one resistant strain from each organism reporting group (i.e., Enterobacterales, P. aeruginosa, and Acinetobacter spp.) was evaluated under three different storage conditions for each of the 15 antimicrobial agents on the panel. Each strain was tested in triplicate across all conditions. Categorical interpretation results (S, SDD, I, or R) at different storage timepoints were compared to the modal PhAST BC GN Panel results determined at T0. As this was a method-to-method comparison, CA ≥95% for each antimicrobial agent was considered acceptable.
The study demonstrated acceptable performance of ≥95% CA for both extended storage conditions (i.e., 14 hours at 35°C, and 5 hours at 35°C followed by 14 hours at room temperature) for each antimicrobial when compared to results obtained at T0 (Table 7).
As noted in the instructions for use, positive blood cultures should be tested immediately after a positive flag, where possible. In the case of unavoidable delays or if the need for re-testing arises, testing on the PhAST BC GN Panel may be performed on positive blood culture bottles removed from the continuous monitoring system within 5 hours of bottle ring and subsequently kept at room temperature for up to 12 hours after removal.
Table 7. Summary of Sample Stability Results of the PhAST Blood Culture Gram-Negative Panel with the PhAST Instrument
| Antimicrobial | Categorical Agreement (%) | |
| --- | --- | --- |
| | 14h 35°C vs. T0 | 5h 35°C + 14h RT vs. T0 |
| Amikacin | 31/31 (100%) | 32/32 (100%) |
| Ampicillin/sulbactam | 28/29 (96.6%) | 30/30 (100%) |
| Aztreonam | 31/31 (100%) | 32/32 (100%) |
| Cefepime | 37/37 (100%) | 37/37 (100%) |
| Ceftazidime | 46/46 (100%) | 47/47 (100%) |
| Ceftazidime/avibactam | 37/37 (100%) | 38/38 (100%) |
| Ceftriaxone | 31/31 (100%) | 31/31 (100%) |
| Ciprofloxacin | 36/36 (100%) | 38/38 (100%) |
| Ertapenem | 33/34 (97.1%) | 34/35 (97.1%) |
| Gentamicin | 31/31 (100%) | 32/32 (100%) |
| Levofloxacin | 37/37 (100%) | 38/38 (100%) |
K253316 - Page 20 of 34
{21}
| Meropenem | 46/46 (100%) | 46/46 (100%) |
| --- | --- | --- |
| Piperacillin/tazobactam | 45/45 (100%) | 46/46 (100%) |
| Tobramycin | 32/32 (100%) | 35/35 (100%) |
| Trimethoprim/sulfamethoxazole | 31/31 (100%) | 31/31 (100%) |
# 9. Assay Cut-Off:
Not applicable
# 10. Cross Contamination/Carry-Over:
The Carryover/Cross-Contamination Study was designed to evaluate the possibility of sample carryover on the PhAST instrument with the PhAST BC GN Panel across samples run at different times within the same bay or cross-contamination between the instrument's four bays. Contrived blood culture samples seeded with susceptible or resistant isolates were tested with the PhAST BC GN Panel in alternating patterns to detect carryover or cross-contamination. Testing was performed across 5 runs and 4 bays in a predetermined alternating pattern designed to detect both cross-contamination between bays within a single run and carryover on a single bay across multiple runs. All tested antimicrobials achieved 100% categorical agreement compared to reference BMD results, which were acceptable.
# B Comparison Studies:
# 1. Method Comparison with Predicate Device:
Clinical performance testing using the PhAST BC GN Panel with PhAST instrument was conducted at four sites (three geographically diverse US clinical sites and one internal site). For instances in which testing was required to supplement existing data from the original study and support specific claims, testing was performed using the PhAST BC GN Panel with the PhAST instrument at one internal site. Performance was evaluated using prospectively collected, fresh leftover positive blood culture samples from patients with suspected bacteremia. Samples were confirmed by Gram stain to contain only gram-negative bacteria prior to testing on the PhAST BC GN Panel. Additionally, blood culture bottles containing fresh human donor blood were contrived with clinical stock isolates and challenge isolates to supplement the fresh prospective blood cultures for lower prevalence species and resistance profiles. Subcultures of all positive blood cultures onto appropriate media (Tryptic Soy Agar with 5% sheep blood) were used to check for purity prior to preparing a frozen stock for shipment to the central reference laboratory. Polymicrobial samples were excluded from analysis. Organism identification was obtained from an FDA-cleared MALDI-TOF method for input into the PhAST web application to generate antimicrobial susceptibility test (AST) results.
A total of 1028 samples were enrolled in the study, of which 890 samples were included in the performance analysis (256 fresh prospective positive blood culture samples, 121 samples contrived with clinical stock isolates, and 513 samples contrived with challenge isolates). In total, 138 samples were excluded from final performance analyses due to not meeting inclusion criteria, invalid QC results, or other protocol deviations.
Performance was determined by comparing the categorical interpretation results (S, SDD, I, or R) of the PhAST BC GN Panel to the modal categorical interpretation results based on the minimum inhibitory concentration (MIC) values obtained from the reference broth
K253316 - Page 21 of 34
{22}
microdilution (BMD) method as described in the CLSI document Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically, M07, 12th edition. Performance was generally based on criteria outlined in the Class II Special Controls Document: Antimicrobial Susceptibility Test (AST) Systems including categorical agreement (CA) and categorical errors (minor, major, and very major errors) for each antimicrobial agent. CA was calculated as the percentage of PhAST BC GN Panel categorical interpretive results (S, SDD, I, or R) that were identical to the interpretive categories of the reference BMD result.
A summary of the performance of the PhAST BC GN Panel is described below for each antimicrobial agent with indicated species. Complete details including CA and error rate analysis per organism group are summarized in Table 8.
The time to result (TTR) of the PhAST BC GN Panel with the PhAST instrument was assessed across samples tested during the clinical study. The TTR was measured from the time the cartridge was loaded into the instrument to the generation of the susceptibility report following organism identity entry. The observed mean TTR was as follows:
- Enterobacterales: 99 ± 7 minutes (mean ± standard deviation), ranged from 91 minutes to 130 minutes.
- Acinetobacter spp.: 99 ± 7 minutes (mean ± standard deviation), ranged from 91 minutes to 116 minutes.
- Pseudomonas aeruginosa: 141 ± 6 minutes (mean ± standard deviation), ranged from 135 minutes to 154 minutes.
Amikacin/Enterobacterales. A total of 508 Enterobacterales (27 C. freundii complex, 37 C. koseri, 39 E. cloacae complex, 164 E. coli, 40 K. aerogenes, 37 K. oxytoca, 81 K. pneumoniae, 36 P. mirabilis, 9 P. vulgaris group, and 38 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 95.2%, which was acceptable. There were 20 minor errors, 6 major errors (6/457 = 1.3%), and 1 very major error (1/34 = 2.9%).
When evaluating results by individual species, P. mirabilis demonstrated a CA of 86.1%, which is not acceptable. P. mirabilis had one major error (1/3 = 33.3%), which was considered a random error due to the limited number of resistant isolates tested. P. vulgaris group demonstrated a CA of 88.9%, which is not acceptable. Due to the unacceptable CA, these antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
o Amikacin: P. mirabilis, P. vulgaris group
A limitation is included in the device labeling to address the lack of testing with resistant Citrobacter koseri, K. aerogenes and K. oxytoca isolates.
Ampicillin/sulbactam/Enterobacterales. A total of 332 Enterobacterales (19 C. koseri, 162 E. coli, 25 K. oxytoca, 76 K. pneumoniae, 36 P. mirabilis, and 14 P. vulgaris group) were
K253316 - Page 22 of 34
{23}
evaluated. The combined results from clinical and challenge testing demonstrated a CA of 58.7%, which is not acceptable. There were 69 minor errors, 68 major errors (68/251 = 27.1%), and no very major errors. Due to the unacceptable CA and major error rate, this antimicrobial/organism group combination is not indicated for use with the PhAST BC GN Panel, and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combination(s):
- Ampicillin/sulbactam: Enterobacterales
**Ampicillin/sulbactam/Acinetobacter spp.** A total of 62 *A. baumannii* complex were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 91.9%. There were 4 minor errors, 1 major error (1/35 = 2.9%), and no very major errors. Overall performance is acceptable.
**Aztreonam/Enterobacterales.** A total of 494 Enterobacterales isolates (25 *C. freundii* complex, 18 *C. koseri*, 43 *E. cloacae* complex, 170 *E. coli*, 31 *K. aerogenes*, 26 *K. oxytoca*, 82 *K. pneumoniae*, 14 *P. vulgaris* group, 44 *P. mirabilis*, and 41 *S. marcescens*) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 96.6%. There were 14 minor errors, 3 major errors (3/357 = 0.8%), and no very major errors. Overall performance is acceptable.
When evaluating results by individual species, *S. marcescens* had 2 major errors (2/35 = 5.7%), which is not acceptable. A limitation is included in the device labeling to address the unacceptable major error rate. In addition, *P. mirabilis* met all acceptance criteria in the clinical study; however, based on other results from supplemental analytical testing, this antimicrobial/organism combination was removed from the Indications for Use. These antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel, and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
- Aztreonam: *P. mirabilis*, *S. marcescens*
A limitation is included in the device labeling to address the lack of testing with resistant *C. koseri* and *P. vulgaris* group isolates.
**Aztreonam/Pseudomonas aeruginosa.** A total of 34 *P. aeruginosa* were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 70.6%, which is not acceptable. There were 10 minor errors, no major errors, and no very major errors. Due to the unacceptable CA, this antimicrobial/organism group combination is not indicated for use with the PhAST BC GN Panel, and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
- Aztreonam: *P. aeruginosa*
K253316 - Page 23 of 34
{24}
**Cefepime/Enterobacterales.** A total of 525 Enterobacterales (46 *C. freundii* complex, 36 *C. koseri*, 173 *E. coli*, 32 *K. aerogenes*, 27 *K. oxytoca*, 80 *K. pneumoniae* group, 41 *P. mirabilis*, 42 *S. marcescens*, 38 *E. cloacae* complex, and 10 *P. vulgaris* group) were evaluated. The susceptible-dose dependent (SDD) category is treated as an Intermediate (I) category for performance analysis. The combined results from clinical and challenge testing demonstrated a CA of 94.1%. There were 27 minor errors, 4 major errors (4/414 = 1.0%), and no very major errors. Overall performance is acceptable.
When evaluating results by individual species, *E. cloacae* complex demonstrated a CA of 81.6%, which is not acceptable. There were 6 minor errors, 1 major error (1/22 = 4.5%) and no very major errors. *P. vulgaris* group demonstrated a CA of 80.0%, which is not acceptable. Due to the unacceptable CA and major error rate of *E. cloacae* and unacceptable CA of *P. vulgaris* group, these antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
o Cefepime: *E. cloacae* complex, *P. vulgaris* group
A limitation is included in the device labeling to address the lack of testing with resistant *C. koseri* isolates.
**Cefepime/*Pseudomonas aeruginosa*.** A total of 54 *P. aeruginosa* isolates were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 96.3%. There were 2 minor errors, no major errors, and no very major errors. Overall performance is acceptable.
**Ceftazidime/Enterobacterales.** A total of 553 Enterobacterales isolates (59 *E. cloacae* complex, 70 *K. aerogenes*, 28 *K. oxytoca*, 80 *K. pneumoniae*, 40 *P. mirabilis*, 42 *S. marcescens*, 27 *C. freundii* complex, 19 *C. koseri*, 176 *E. coli*, and 12 *P. vulgaris* group) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 92.6%, which is acceptable. There were 27 minor errors, 13 major errors (13/378 = 3.4%), and 1 very major error (1/164 = 0.6%).
When evaluating results by individual species, *C. freundii* complex had 1 major error (1/11 = 9.1%), which is not acceptable. *C. koseri* demonstrated a CA of 68.4%, which is not acceptable. *E. coli* demonstrated a CA of 89.8%, which is not acceptable. There were 9 minor errors, 9 major errors (9/136 = 6.6%) and no very major errors. *P. vulgaris* group demonstrated a CA of 75.0%, which is not acceptable. There were 2 minor errors, 1 major error (1/12 = 8.3%) and no very major errors. Due to the unacceptable CA and the major error rate of *E. coli* and *P. vulgaris* group, the unacceptable CA of *C. koseri*, and unacceptable major error rate of *C. freundii* complex, these antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combination(s):
o Ceftazidime: *C. freundii* complex, *C. koseri*, *E. coli*, *P. vulgaris* group
K253316 - Page 24 of 34
{25}
A limitation is included in the device labeling to address the lack of testing with resistant P. mirabilis isolates.
**Ceftazidime/Pseudomonas aeruginosa.** A total of 55 P. aeruginosa isolates were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 98.2%. There was 1 minor error, no major errors, and no very major errors. Overall performance is acceptable.
**Ceftazidime/Acinetobacter spp.** A total of 44 A. baumannii complex samples were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 100.0%. There were no minor errors, no major errors, and no very major errors. Overall performance is acceptable.
**Ceftazidime/avibactam/Enterobacterales.** A total of 516 Enterobacterales isolates (24 C. freundii complex, 19 C. koseri, 43 E. cloacae complex, 180 E. coli, 47 K. aerogenes, 28 K. oxytoca, 80 K. pneumoniae, 42 P. mirabilis, 11 P. vulgaris group and 42 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 99.8%. There was 1 major error (1/484 = 0.2%) and no very major errors. Overall performance is acceptable.
A limitation is included in the device labeling to address the lack of testing with resistant C. koseri, K. aerogenes, P. mirabilis, C. freundii complex and P. vulgaris group isolates.
**Ceftazidime/avibactam/Pseudomonas aeruginosa.** A total of 55 P. aeruginosa isolates were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 100%. There were no major errors and no very major errors. Overall performance is acceptable.
**Ceftriaxone/Enterobacterales.** A total of 502 Enterobacterales isolates (26 C. freundii complex, 28 C. koseri, 43 E. cloacae complex, 176 E. coli, 32 K. aerogenes, 26 K. oxytoca, 81 K. pneumoniae, 40 P. mirabilis, 9 P. vulgaris group, and 41 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 99.4%. There were no minor errors, 3 major errors (3/338 = 0.9%), and no very major errors. Overall performance is acceptable.
When evaluating results by individual species, performance P. vulgaris group demonstrated a CA of 88.9%, which is not acceptable. There were no minor errors, 1 major error (1/9 = 11.1%), and no very major errors. This major error is considered a random error due to the limited number of susceptible isolates tested. S. marcescens had 2 major errors (2/35 = 5.7%), which is not acceptable. Due to the unacceptable CA of P. vulgaris group and the unacceptable major error rate of S. marcescens, these antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
o Ceftriaxone: P. vulgaris group, S. marcescens
K253316 - Page 25 of 34
{26}
**Ciprofloxacin/Enterobacterales.** A total of 547 Enterobacterales isolates (26 *C. freundii* complex, 40 *E. cloacae* complex, 217 *E. coli*, 46 *K. aerogenes*, 27 *K. oxytoca*, 78 *K. pneumoniae*, 44 *P. mirabilis*, 39 *S. marcescens*, 19 *C. koseri*, and 11 *P. vulgaris* group) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 95.2%. There were 20 minor errors, 6 major errors (6/357 = 1.7%), and no very major errors. Overall performance is acceptable.
When evaluating results by individual species, *C. koseri* demonstrated a CA of 57.9%, which is not acceptable. There were 4 minor errors, 4 major errors (4/19 = 21.1%) and no very major errors. *P. vulgaris* group demonstrated a CA of 54.5%, which is not acceptable. There were 4 minor errors, 1 major error (1/11 = 9.1%) and no very major errors. Due to the unacceptable CA and major error rate of *C. koseri* and *P. vulgaris* group, these antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
- Ciprofloxacin: *C. koseri*, *P. vulgaris* group
**Ciprofloxacin/Pseudomonas aeruginosa.** A total of 51 *P. aeruginosa* isolates were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 100%. There were no major errors and no very major errors. Overall performance is acceptable.
**Ertapenem/Enterobacterales.** A total of 488 Enterobacterales isolates (192 *E. coli*, 41 *K. aerogenes*, 27 *K. oxytoca*, 81 *K. pneumoniae*, 42 *P. mirabilis*, 21 *C. freundii* complex, 16 *C. koseri*, 40 *E. cloacae* complex, 6 *S. marcescens*, and 22 *P. vulgaris* group) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 92.6%. There were 13 minor errors, 23 major errors (23/426 = 5.4%), and no very major errors.
When evaluating results by individual species, *C. freundii* complex demonstrated a CA of 71.4%, which is not acceptable. There were 3 minor errors, 3 major errors (3/13 = 23.1%) and no very major errors. *C. koseri* demonstrated a CA of 75.0%, which is not acceptable. There were 2 minor errors, 2 major errors (2/15 = 13.3%) and no very major errors. *E. cloacae* complex demonstrated a CA of 75.0%, which is not acceptable. There were 4 minor errors, 6 major errors (6/23 = 26.1%) and no very major errors. *S. marcescens* demonstrated a CA of 86.4%, which is not acceptable. There were no minor errors, 3 major errors (3/21 = 14.3%) and no very major errors. A limitation is included in the device labeling to address the unacceptable CA and major error rate of *C. freundii* complex, *C. koseri*, *E. cloacae* complex, and *S. marcescens*. In addition, *P. vulgaris* group met all acceptance criteria in the clinical study. However, this antimicrobial/organism combination was removed from the Indications for Use due to the concern for potential limited clinical utility as a high rate of Technical Error (TE) and No Call: Inconclusive (NC) results (53.3%) was observed during the clinical study. These antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel and the following limitation is included in the device labeling:
K253316 - Page 26 of 34
{27}
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
○ Ertapenem: C. freundii complex, C. koseri, E. cloacae complex, S. marcescens, P. vulgaris group
A limitation is included in the device labeling to address the lack of testing with resistant P. mirabilis isolates.
Gentamicin/Enterobacterales. A total of 570 Enterobacterales isolates (27 C. freundii complex, 18 C. koseri, 42 E. cloacae complex, 249 E. coli, 33 K. aerogenes, 28 K. oxytoca, 81 K. pneumoniae, 43 P. mirabilis, 11 P. vulgaris group, and 38 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 97.4%. There were 13 minor errors, no major errors, and 2 very major errors (2/150 = 1.3%). Overall performance is acceptable.
A limitation is included in the device labeling to address the lack of testing with resistant C. koseri, P. vulgaris group and S. marcescens isolates.
Levofloxacin/Enterobacterales. A total of 493 Enterobacterales isolates (27 C. freundii complex, 19 C. koseri, 43 E. cloacae complex, 168 E. coli, 31 K. aerogenes, 27 K. oxytoca, 78 K. pneumoniae, 44 P. mirabilis, 14 P. vulgaris group, and 42 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 97.0%. There were 15 minor errors, no major errors, and no very major errors. Overall performance is acceptable.
A limitation is included in the device labeling to address the lack of testing with resistant C. koseri and P. vulgaris group isolates.
Levofloxacin/Pseudomonas aeruginosa. A total of 51 P. aeruginosa isolates were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 98.0%. There was 1 minor error, no major errors, and no very major errors. Overall performance is acceptable.
Meropenem/Enterobacterales. A total of 534 Enterobacterales isolates (39 C. freundii complex, 36 C. koseri, 173 E. coli, 31 K. aerogenes, 43 K. oxytoca, 82 K. pneumoniae, 43 P. mirabilis, 36 E. cloacae complex, 12 P. vulgaris group, and 39 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 95.5%. There were 4 minor errors, 14 major errors (14/481 = 2.9%), and 5 very major errors (5/52 = 9.6%).
When evaluating results by individual species, E. cloacae complex demonstrated a CA of 88.9%, which is not acceptable. There was 1 minor error, 2 major errors (2/24 = 8.3%) and one very major error (1/11 = 9.1%). P. vulgaris group demonstrated a CA of 66.7%, which is not acceptable. There were 2 minor errors, 2 major errors (2/12 = 16.7%) and no very major errors. S. marcescens demonstrated a CA of 82.1%, which is not acceptable. There were no minor errors, 3 major errors (3/35 = 8.6%) and 4 very major errors (4/4 = 100%). Due to the unacceptable CA, major error rate and very major error rate of E. cloacae complex, the unacceptable CA and major error rate of P. vulgaris group, and the unacceptable CA and
K253316 - Page 27 of 34
{28}
very major error rate of S. marcescens, these antimicrobial/organism combinations are not indicated for use with the PhAST BC GN Panel, and the following limitation is included in the device labeling:
- Perform an alternative method of testing prior to reporting results for the following antibiotic/organism combinations:
o Meropenem: E. cloacae complex, P. vulgaris group, S. marcescens
A limitation is included in the device labeling to address the lack of testing with resistant C. koseri and P. mirabilis isolates.
Meropenem/Pseudomonas aeruginosa. A total of 52 P. aeruginosa isolates were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 100%. There were no minor errors, no major errors, and no very major errors. Overall performance is acceptable.
Meropenem/Acinetobacter spp. A total of 42 A. baumannii complex samples were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 97.6%. There was 1 minor error, no major errors, and no very major errors. Overall performance is acceptable.
Piperacillin/tazobactam/Enterobacterales. A total of 497 Enterobacterales isolates (51 E. cloacae complex, 175 E. coli, 32 K. aerogenes, 28 K. oxytoca, 92 K. pneumoniae, 23 C. freundii complex, 15 C. koseri, 35 P. mirabilis, 9 P. vulgaris group, and 37 S. marcescens) were evaluated. The combined results from clinical and challenge testing demonstrated a CA of 90.7%. There were 44 minor errors, 2 major errors (2/370 = 0.5%), and no very major errors. Overall performance is acceptable.
When evaluating results by individual species, C. koseri demonstrated a CA of 53.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.