K253315 · Institut Straumann AG · NHA · Jan 22, 2026 · Dental
Device Facts
Record ID
K253315
Device Name
Straumann Variobase Abutments XC for Bridge/Bar
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Jan 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations.
Device Story
Straumann Variobase Abutments XC are prosthetic components connecting Straumann dental implants to customized bridge/bar restorations. Devices are non-engaging, manufactured from Ti-6Al-7Nb (TAN) alloy, and feature a laser-texturized chimney surface to enhance cement retention. Available in multiple platform diameters (RB/WB, WB) and gingival heights, with straight or angled screw channel options. Used in dental clinics by clinicians; the abutment is screwed into the implant, and the prosthetic restoration is cemented onto the abutment chimney. Temporary restorations are used during the healing phase to shape soft tissue. The device provides a stable foundation for multi-unit restorations, facilitating functional and aesthetic rehabilitation for patients with missing teeth.
Clinical Evidence
Bench testing only. Performance evaluated via dynamic fatigue testing, pull-off force testing to validate cement retention on the laser-texturized surface, and SEM/EDX surface analysis. Biocompatibility assessed per ISO 10993-1. No clinical data provided.
Technological Characteristics
Material: Ti-6Al-7Nb (TAN) alloy. Surface: Partially anodized (violet) with laser-texturized chimney. Connection: TorcFit (non-engaging). Sterilization: Non-sterile, validated for end-user steam sterilization (autoclave). Dimensions: Platform diameters 3.8/4.5/5.5 mm; gingival heights 0.75/1.5/2.5/3.5 mm; chimney height 7 mm (customizable to 3.5 mm).
Indications for Use
Indicated for patients requiring screw-retained or cement-retained bridge restorations supported by Straumann dental implants. Temporary restorations may be used to maintain, stabilize, and shape soft tissue during healing, provided they are placed out of occlusion. Final abutments and restorations may be placed in occlusion once the implant is fully osseointegrated.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Straumann® BLX Variobase Abutments for Bar and Bridges (K190040)
Reference Devices
MRI Compatibility for Existing Straumann Dental Implant Systems (K190662)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 22, 2026
Institut Straumann AG
% Jennifer Jackson
Sr. Director, Regulatory Affairs and Quality
Straumann USA, LLC
60 Minuteman Road
Andover, Massachusetts 01810
Re: K253315
Trade/Device Name: Straumann Variobase Abutments XC for Bridge/Bar
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: September 29, 2025
Received: December 23, 2025
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253315 - Jennifer Jackson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K253315 - Jennifer Jackson
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253315 | |
| Device Name | |
| Straumann Variobase Abutments XC for Bridge/Bar | |
| Indications for Use (Describe) | |
| The Straumann Variobase Abutments XC for Bridge/Bar are prosthetic components placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann Variobase Abutments XC for Bridge/Bar are indicated for screw-retained or cement-retained bridge restorations. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize, and shape the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Final abutments and restorations may be placed in occlusion when the implant is fully osseointegrated. | |
| Type of Use (Select one or both, as applicable) | |
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
# 510(k) Summary
## Submitter’s Contact Information
Submitter: Straumann USA, LLC
60 Minuteman Road
Andover, MA 01810
Registration No.: 1222315 Owner/Operator No.: 9005052
On the behalf of :
Institut Straumann AG
Peter-Merian-Weg 12
4052 Basel, Switzerland
Contact Person: Jennifer M. Jackson, MS
Director of Regulatory Affairs and Quality
Phone Number: +1-978-747-2509
Fax Number: +1-978-747-0023
Prepared By & Alternate Contact: Olivier Russo
Associate Director Regulatory Affairs
Institut Straumann AG
Phone number: +41 61 965 1260
Date of Submission: January 21, 2026
## Name of the Device
Trade Names: Straumann Variobase Abutments XC for Bridge/Bar
Common Name: Straumann Variobase Abutments XC for Bridge/Bar
Classification Name: Endosseous dental implant abutment
Regulation Number: 21 CFR 872.3630
Device Classification: II
Product Code(s): NHA
Secondary Product Code: n/a
Review Panel: Dental
Proprietary Name: Straumann Variobase Abutments XC for Bridge/Bar
Institut Straumann AG
January 21, 2026
{5}
Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
# Predicate Device(s)
## Primary Predicate:
- K190040 - Straumann® BLX Variobase Abutments for Bar and Bridges
## Reference Devices:
- K190662 - MRI Compatibility for Existing Straumann Dental Implant Systems
- K230108 - Straumann® BLC and TLC Implants
# Device Description
Straumann Variobase Abutments XC for Bars and Bridges are intended to be placed into Straumann implants to provide support for multi-unit restorations.
The prosthetic restoration (bar/bridge) must be cemented onto the Variobase abutments, which is then screwed onto the implants. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they must be placed out of occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated.
The subject Variobase abutments are manufactured from Ti-6Al-7Nb (TAN) and are anodized in violet.
The Variobase Abutments for bar/bridges are non-engaging devices intended to support multiple-unit restorations.
The Variobase abutments are provided non-sterile with instructions for end user steam sterilization. The Variobase abutments are fixed in the implant by means of a basal screw which is also manufactured from Ti-6Al-7Nb (TAN). The subject Straumann Variobase Abutments XC for Bridge/Bar is provided in two different platform models, both available with straight and angled screw channel solutions:
Institut Straumann AG
January 21, 2026
{6}
Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
Table 10.4.a - Straumann Variobase Abutments XC for Bridge/Bar characteristics
| Platform | Ø (mm) | Gingival heights (mm) | Chimney height (mm) |
| --- | --- | --- | --- |
| RB/WB | 3.8 / 4.5 | 1.5 / 2.5 / 3.5 | 7 |
| WB | 5.5 | 0.75 / 1.5 | 7 |
## Intended Use
Prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations.
## Indications for Use
The Straumann Variobase Abutments XC for Bridge/Bar are prosthetic components placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann Variobase Abutments XC for Bridge/Bar are indicated for screw-retained or cement-retained bridge restorations. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize, and shape the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Final abutments and restorations may be placed in occlusion when the implant is fully osseointegrated.
## Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the following table:
Institut Straumann AG
January 21, 2026
{7}
Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
Table 10.7.a - Technological Characteristics
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | EQUIVALENCE DISCUSSION |
| --- | --- | --- | --- |
| K Number | K253315 - Straumann Variobase Abutments XC for Bridge/Bar | K190040 - Straumann® BLX Variobase Abutments for Bar and Bridges | K190040- BLX Variobase Abutments AS for Crown |
| Indications for Use | The Straumann Variobase Abutments XC for Bridge/Bar are prosthetic components placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann Variobase Abutments XC for Bridge/Bar are indicated for screw-retained or cement-retained bridge restorations. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize, and shape the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. Final abutments and restorations may be placed in occlusion when the implant is fully osseointegrated. | Straumann® Variobase™ prosthetic components directly or indirectly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. The prosthetic restoration (bridge or overdenture) can be cemented on the Straumann® Variobase™ prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. They may not be placed into occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated. | The Straumann® Variobase® for Crown AS is a titanium base placed onto Straumann dental implants to provide support for customized prosthetic restorations. Straumann Variobase for Crown AS are indicated for screw retained single tooth or cement retained single tooth and bridge restorations. A temporary restoration can be used prior to insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. Temporary restorations are indicated to be placed out of occlusion. All digitally designed copings and/or crowns for use with the Straumann® Variobase® for Crown AS are intended to be sent to Straumann for manufacture at a validated milling center. |
| Implant to Abutment Connection | TorcFit connection
New RB/WB Interface (Flat top with internal guidance within the conical interface)
New WB Interface (implant shoulder with internal guidance within the conical interface)
Non-engaging design | TorcFit connection
RB/WB Interface (sitting on implant shoulder)
WB Interface (sitting on implant shoulder)
Non-engaging design | TorcFit connection
RB/WB Interface (sitting on implant shoulder)
WB Interface (sitting on implant shoulder)
Engaging design |
| Abutment Material | Titanium alloy (Ti-6Al-7Nb, TAN) | Titanium alloy (Ti-6Al-7Nb, TAN) | Titanium alloy (Ti-6Al-7Nb, TAN) |
| Restoration Types Supported | Bridges and bars | Bridges and bars | Crowns |
| Compatible Implants | Straumann Bone Level implants having RB/WB and WB implant-to-abutment interface geometries
BLX and BLC implant systems:
• BLX implant diameters : 3.5 / 3.75 / 4 / 4.5 / 5 / 5.5 / 6.5
• BLC implant diameters : 3.3 / 3.7 / 4 / 4.5 / 5 / 5.5 / 6.5 | Straumann Bone Level BLX implants having RB/WB implant-to-abutment interface geometries | Straumann Bone Level BLX implants having RB/WB implant-to-abutment interface geometries |
| Maximum Angulation | 30° controlled in design software | 30° controlled in design software | 30° controlled in design software |
| Gingiva height | RB/WB Interface: 1.5 / 2.5 / 3.5 mm
WB Interface: 0.75 / 1.5 mm | RB/WB Interface: 1.5 mm
WB Interface: 1.5 mm | RB/WB Interface: 1.5 mm
WB Interface: 1.5 mm |
Institut Straumann AG
January 21, 2026
{8}
Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | | EQUIVALENCE DISCUSSION |
| --- | --- | --- | --- | --- |
| K Number | K253315 - Straumann Variobase Abutments XC for Bridge/Bar | K190040 - Straumann® BLX Variobase Abutments for Bar and Bridges | K190040- BLX Variobase Abutments AS for Crown | |
| Variobase Abutment Chimney Height | 7 mm (can be reduced to minimum 3.5 mm)
Note: the abutment post height is customizable. Adaptable levels are visually identifiable on the chimney section via grooves and spaced 1 mm apart, starting from the maximum height of the Variobase abutment | 5.5 mm (can be reduced to 3.5 mm) | 5.5mm | Equivalent to the primary predicate device.
The new Variobase XC portfolio offers more flexibility at the chimney height. |
| Variobase platform diameter range | RB, WB Interface: 3.8 / 4.5 mm
WB Interface: 5.5 mm | 4.5 mm | Ø 4.5mm and 5.5mm | Identical to the primary predicate |
| Screw Channel | Straight and Angled screw channel | Straight screw channel | Angled screw channel solutions | Identical to the primary predicate |
| Surface | Partially anodized
Roughened and structured chimney surface (laser-texturized) | Partially anodized | Partially anodized | Identical to the primary predicate for the anodized surface.
Roughened surface enhancing cement retention compared to currently available Variobase abutments was evaluated through biocompatibility testing and pull-off testing.
These assessments support the demonstration of substantial equivalence to the predicate device.
SEM/EDX analysis was performed to evaluate the surface characteristics. |
| Sterility | Provided non-sterile – terminally sterilized via autoclave prior to implantation. | Provided non-sterile – terminally sterilized via autoclave prior to implantation. | Provided non-sterile – terminally sterilized via autoclave prior to implantation. | Identical to the primary predicate |
Institut Straumann AG
January 21, 2026
{9}
Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
# Performance Testing
## Sterilization Validation and Shelf Life
The Straumann Variobase Abutments XC for Bridge/Bar subject devices are provided non-sterile and need to be sterilized by moist heat (steam) after milling into the patient-specific shape by the end-user. The recommended sterilization method has been validated according to ISO 17665-1 and applicable recommendations in the FDA guidance document “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015”. The sterilization methods and parameters are equivalent to the primary predicate and reference devices.
## Biocompatibility Testing
Biological assessment has been performed according to ISO 10993-1 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process” and to the FDA Guidance document “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process’, Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016” for each of the subject devices.
The subject devices are equivalent in material and manufacturing processes to the primary predicate and reference devices, therefore, no new issues regarding biocompatibility were raised.
## Electromagnetic Compatibility
There are no significant changes to the materials and dimensions from the currently marketed predicate devices. Therefore, no new issues of electromagnetic compatibility are raised for the subject devices, and they can be considered MR Conditional. K190662 - MRI Compatibility for Existing Straumann Dental Implant Systems was referenced for MRI testing and labelling.
Institut Straumann AG
January 21, 2026
{10}
Institut Straumann AG
January 21, 2026
5
# Traditional 510(k) Submission
## Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
## Performance Testing – Bench
### Dynamic Fatigue
The device design and performance testing submitted or referenced in support of this premarket notification were conducted according to the FDA guidance document “Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments” and are representative of the design and performance of the subject devices.
### Pull-off Testing
The Straumann Variobase Abutments XC for Bridge/Bar feature a roughened and structured chimney surface, which improves cement retention compared to the machined surface of current Variobase abutments. The effectiveness of this surface was evaluated by measuring the pull-off force required to separate the coping from the abutment after cementation.
A worst-case approach was applied to define the testing setup. The selection criteria focused on the abutment design, particularly the surface and geometry of the abutment post height, as cement is applied to this region. The minimum post height of 3.5 mm was chosen as the worst-case scenario, since it allows the least amount of cement and therefore represents the most challenging condition for adhesion.
Results demonstrate that the lasered-surface abutments generate higher pull-off force than machined surface abutments, confirming the enhanced cement retention of the new textured surface.
### Surface Analysis
The new roughened and structured chimney surface of the Straumann Variobase Abutments XC for Bridge/Bar was analyzed using SEM-EDX. Images demonstrate that the laser-texturized chimney exhibit altered surface topography compared to the machined surface, while retaining an equivalent elemental composition consistent with Ti-6Al-7Nb material.
{11}
Institut Straumann AG
January 21, 2026
6
# Traditional 510(k) Submission
Straumann Variobase Abutments XC for Bridge/Bar
510(k) Summary – K253315
## Conclusion
The conclusion drawn from the nonclinical tests submitted in this premarket notification demonstrates that the Straumann Variobase Abutments XC for Bridge/Bar are substantially equivalent to the primary predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.