K253301 · Time IS Brain, S.L. · GWF · Jun 27, 2026 · Neurology
Device Facts
Record ID
K253301
Device Name
BraiN20® (BraiN20)
Applicant
Time IS Brain, S.L.
Product Code
GWF · Neurology
Decision Date
Jun 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1870
Device Class
Class 2
Indications for Use
The BraiN20® is intended for use in monitoring neurological status by recording the N20 response of somatosensory evoked potentials (SSEP). The device provides objective electrophysiological information and is not intended to provide a diagnosis independently of clinical assessment.
Device Story
BraiN20 is a portable, non-invasive, battery-powered medical device for monitoring neurological status via SSEP N20 response. System comprises main unit with integrated touchscreen, stimulation module, and acquisition module. Device delivers electrical pulses to patient skin via disposable electrodes and simultaneously records electrophysiological signals. Dual-core processor architecture: one core manages acquisition, interface, storage, and external communication; second core dedicated to N20 SEP detection and analysis algorithm. Device performs real-time EEG/EMG analysis, identifying N20 presence, amplitude, and latency. Results displayed on integrated screen. Used in clinical settings for monitoring; provides objective data to assist clinicians in neurological assessment. No wireless connectivity; data stored on internal SD card.
Clinical Evidence
No clinical studies were required or relied upon. Substantial equivalence supported by bench performance testing, including system functionality, signal acquisition, stimulation performance, battery operation, reliability, and hardware integration. Compliance with IEC 60601-2-40, IEC 60601-1-2, IEC 62304, ISO 10993, and IEC 62366-1/60601-1-6 demonstrated.
Technological Characteristics
Active, portable, non-invasive device. Materials: disposable electrodes (biocompatible per ISO 10993). Energy: rechargeable battery. Connectivity: standalone (no wireless), SD card storage. Interface: integrated touchscreen. Connectors: 1.5mm DIN (EEG/EMG), MINI-MED (stimulation). Software: proprietary embedded OS. Sterilization: non-sterile, single-use electrodes. Stimulation: current-controlled, 1-10 Hz, 45 mA max, 180 V max, 100-500 us pulse width.
Indications for Use
Indicated for monitoring neurological status in patients via recording of N20 somatosensory evoked potential (SSEP) responses.
Regulatory Classification
Identification
An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 27, 2026
Time is Brain, S.L.
Carlos Navarro
RAQA Manager
Carretera de Canyet s/n
Badalona, Barcelona 08916
Spain
Re: K253301
Trade/Device Name: BraiN20® (BraiN20)
Regulation Number: 21 CFR 882.1870
Regulation Name: Evoked Response Electrical Stimulator
Regulatory Class: Class II
Product Code: GFW, GXY, IKN
Dated: January 22, 2026
Received: May 26, 2026
Dear Carlos Navarro:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253301 - Carlos Navarro
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
{3}
K253301 - Carlos Navarro
Page 3
# Patrick Antkowiak -S
for
Jay Gupta
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253301
Device Name
BraiN20® (BraiN20)
Indications for Use (Describe)
The BraiN20® is intended for use in monitoring neurological status by recording the N20 response of somatosensory evoked potentials (SEP).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
timeisbrain
K253301
## SUBMITTER
**Manufacturer:** Time is Brain S.L.
**Address:** Carretera del Canyet s/n, 08916 Badalona, Barcelona, Spain
**Contact Person:** Carlos Navarro Sobrino
**Telephone Number:** (+34) 609 199 384
**Email:** cnavarro@tibtimeisbrain.com
**Trade Name:** BraiN20®
**Date Prepared:** 26/June/2026
## DEVICE
**Device Name:** BraiN20®
**Common Name:** Evoked Response Electrical Stimulator / Somatosensory Evoked Potential Monitoring System
**Classification Name:** Evoked Response Electrical Stimulator
**Regulation Number:** 21 CFR 882.1870
**Product Code:** GWF
**Device Class:** Class II
## PREDICATE DEVICES
**Primary Predicate Device:**
- EPAD Device (K132616)
**Secondary Predicate Device:**
- NeuroEPG System (K123843)
| ITEM | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | ADDITIONAL PREDICATE DEVICE | EQUIVALENCE |
| --- | --- | --- | --- | --- |
| | BraiN20® | EPAD (Evoked Potential Assessing Device) (K132616) | NeuroEPG System (K123843) | |
| Regulation | 21 CFR 882.1870 | 21 CFR 882.1870 | 21 CFR 882.1870 | Equivalent |
| Product Code | GWF, GXY, IKN | GWF, GXY, IKN | GWF, GWJ, GXY | Equivalent |
Page 1 of 9
{6}
timeisbrain
K253301
| ITEM | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | ADDITIONAL PREDICATE DEVICE | EQUIVALENCE |
| --- | --- | --- | --- | --- |
| | BraiN20® | EPAD (Evoked Potential Assessing Device) (K132616) | NeuroEPG System (K123843) | |
| Indications for Use | The BraiN20® is intended for use in monitoring neurological status by recording the N20 response of somatosensory evoked potentials (SSEP). | The EPAD is intended for use in monitoring neurological status by recording somatosensory evoked potentials (SSEP) or assessing the neuromuscular junction (NMJ). | The NeuroEPG System is an evoked response testing and diagnostic device that is capable of eliciting, acquiring and measuring auditory and somatosensory evoked potentials. The intended use of the NeuroEPG System is to objectively record evoked responses from patients 18 years of age and older, upon the presentation of sensory stimuli. The product is indicated for use as a diagnostic agent adjunctive tool and sensory related disorders (i.e., auditory, somatosensory) and in surgical procedures for intraoperative monitoring. The NeuroEPG system was intended to be used by trained personnel in a hospital, clinic, EEG/EP technologists, surgeons, or physician's office, operating room, or other appropriate setting. The anatomical sites of contact for auditory evoked potential testing are the patients ear canal (with the contact object being a sound delivery earbud and the patient's head) and possibly other body sites (with the contact object being electrodes that are capable of measuring bio potentials). The anatomical sites of contact for somatosensory evoked potential testing are the patients upper slash lower limbs (with the contact object being skin surface electrodes) and the patient's head and possibly other body sites (with the contact object being electrodes that are capable of measuring bio potentials). | Substantially Equivalent |
| Device Description | BraiN20® is an active, portable, and non-invasive medical device with a rechargeable battery, | The EPAD Headbox contains a complete data acquisition system that has built-in amplifiers, | The NeuroEPG system is an evoked response testing and diagnostic device that is capable of eliciting, | BraiN20® operates without wireless communication. |
Page 2 of 9
{7}
timeisbrain
K253301
| ITEM | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | ADDITIONAL PREDICATE DEVICE | EQUIVALENCE |
| --- | --- | --- | --- | --- |
| | BraiN20® | EPAD (Evoked Potential Assessing Device) (K132616) | NeuroEPG System (K123843) | |
| | integrated touchscreen, embedded software, where patient skin contact occurs exclusively through disposable accessory electrodes (stimulation and recording) applied to the skin surface for transient use (<24 h) for SEP somatosensory evoked potential) monitoring. The objective of the BraiN20® is to generate and identify the presence or absence of the N20 SEP component generated (and in case of detection, its amplitude and latency), so that through the processing of the registration data information can be offered that will be presented on the screen to the MD user. The BraiN20® medical device is a portable MD capable of stimulating and recording at the same time and in a synchronized manner to process the data and robustly analyze the N20 SEP component. BraiN20® consists of the main device, CMAP stimulation and recording, cables for CMAP stimulation and recording electrodes and acquisition (or recording) electrodes with its textile support holding. The BraiN20® medical device allows you to generate electrical pulses, carry out the acquisition, configure and control the analysis of the recorded signal and display the results on the screen. It is mainly composed of the stimulation module and the acquisition module, which are physically | analog to digital converters, and digital signal processors. User interface is via tablet touchscreen computer provided with the EPAD System and running the Android operating system. The EPAD software application is preloaded onto the tablet. Data can be transferred to an external computer for archiving purposes. Communication between the EPAD Headbox and tablet is via Bluetooth wireless or USB connection. Electrode cables are provided for left and right stimulation and left and right acquisition, color coded for correct connection to the EPAD Headbox: Custom cutaneous electrodes for stimulus and acquisition are provided by SafeOp for use with the EPAD. A total of 11 electrodes are applied for full patient monitoring (upper and lower limbs). The SafeOp electrodes are wet gel, single patient use, disposable, and biocompatible for short term (<24 hours) use on intact skin. | acquiring, and measuring auditory and somatosensory evoked potentials. The NeuroEPG System is intended to objectively record evoked response from patients 18 years of age and older, upon presentation of sensory stimuli. The product is indicated for use as a diagnostic aid and adjunctive tool in sensory related disorders (i.e., auditory, somatosensory) and in surgical procedures for intraoperative monitoring. | Substantially Equivalent |
Page 3 of 9
{8}
timeisbrain
K253301
| ITEM | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | ADDITIONAL PREDICATE DEVICE | EQUIVALENCE |
| --- | --- | --- | --- | --- |
| | BraiN20® | EPAD (Evoked Potential Assessing Device) (K132616) | NeuroEPG System (K123843) | |
| | isolated and also have independent battery power systems. The BraiN20® medical device has a dual-core processor in the acquisition module that allows one of them to be dedicated to acquisition management, interface, storage, communication with the acquisition module and external systems, and the other core exclusively for the processing and detection algorithm of the N20 SEP. In the stimulation module there is a second processor dedicated to stimulation control. BraiN20® medical device analyses EEG and EMG data in real time on the device itself. In addition, it has an SD card for data storage. For the acquisition of EEG and EMG, the BraiN20® has standardized DIN 1.5mm type connectors and for stimulation it has the MINI-MED connector suitable for clinical applications (to which the specific cable for the type of electrode must be connected). Finally, the BraiN20® medical device has an embedded touch screen that allows you to configure and launch both stimulation and acquisition and subsequently display the N20 SEP amplitude and latency information, and a graph of the N20 SEP measured. | | | |
| Modalities | SSEP | SSEP NMJ | SSEP AEP | Substantially Equivalent BraiN20® mode SSEP is a subset |
| Operating | The device operates with a | Google Android | Microsoft Windows (2000 or | The subject |
Page 4 of 9
{9}
timeisbrain
K253301
| ITEM | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | ADDITIONAL PREDICATE DEVICE | EQUIVALENCE |
| --- | --- | --- | --- | --- |
| | BraiN20® | EPAD (Evoked Potential Assessing Device) (K132616) | NeuroEPG System (K123843) | |
| System | proprietary embedded operating system | | XP) | device operates with a proprietary embedded operating system, whereas the predicate devices use commercial operating systems such as Google Android or Microsoft Windows Substantially Equivalent |
| Interface Connection | Not connected | Bluetooth or USB | USB | BraiN20® is not connected |
| User Interface | Portable Touchscreen | Tablet Touchscreen | Not stated | BraiN20® integrates display. Substantially Equivalent to Primary Predicate |
| Operating Environment | Temperature: 20 – 28 °C. Relative humidity: 30 – 60 % RH. Pressure: 860 – 1060 hPa. | Temperature: 10° - 40°C Relative Humidity: 30-75% | Not stated | Substantially Equivalent to Primary Predicate |
| **Waveform Acquisition** | | | | |
| Number of Amplifier Channels | 4 | 8 | Not stated | Substantially Equivalent to Primary Predicate |
| Waveform Display | Yes | Yes | Yes | Equivalent |
| **Stimulation** | | | | |
| Number of Stimulation Channels | 2 | 6 | Not stated | Substantially Equivalent to Primary Predicate |
| Stimulator Type | Current controlled | Current controlled | Current controlled | Equivalent |
| Maximum Stimulation Intensity | 45 mA | 100 mA | 25 mA | Substantially Equivalent to Primary Predicate |
| Stimulation Frequency | 1-10 Hz | 0.1-50 Hz | 0.1-100 Hz | Substantially Equivalent |
| Maximum Voltage | 180 V with 10 kΩ load impedance | 380 VDC with a 3.8 kΩ load impedance | Not stated | Substantially Equivalent to Primary |
Page 5 of 9
{10}
timeisbrain
K253301
| ITEM | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | ADDITIONAL PREDICATE DEVICE | EQUIVALENCE |
| --- | --- | --- | --- | --- |
| | BraiN20® | EPAD (Evoked Potential Assessing Device) (K132616) | NeuroEPG System (K123843) | |
| | | | | Predicate |
| Pulse Width | 100-500 us | 100-300 us | 10-1000 usec | Substantially Equivalent to Primary Predicate |
| Waveform | Monophasic | Monophasic | Monophasic or Biphasic | Equivalent to Primary Predicate |
| Low Frequency Filter | 30 Hz | 0.1 – 500 Hz | Not stated | Substantially Equivalent to Primary Predicate |
| High frequency filter | 230 Hz | 30 – 3000 Hz | Not stated | Substantially Equivalent to Primary Predicate |
| **Electrodes** | | | | |
| Anatomical Sites | SSEP: Upper limbs and head | SSEP: Upper/lower limbs and head/neck | SSEP: Upper/lower limbs and head/neck AEP: Head | Equivalent |
| Type | Custom cutaneous electrodes for use with BraiN20® only | Custom cutaneous electrodes for use with EPAD only | Custom cutaneous electrodes for use with NeuroEPG only | Substantially Equivalent |
| Conductive Gel | Wet gel | Wet gel | Not stated | Equivalent to Primary Predicate |
| EEG Connectors | 1.5mm DIN connector | Nicomatic three pin | Not stated | Standardized connector Substantially Equivalent to Primary Predicate |
| Current Density | Not stated | <0.75 mArms/cm² | Not stated | Recording electrodes are standard and approved by FDA. Stimulation electrodes are a well established technology. |
| Sterility | Non-sterile, single patient use, disposable | Non-sterile, single patient use, disposable | Non-sterile, single patient use, disposable | Equivalent |
### Conclusions Regarding Substantial Equivalence
Based on a comprehensive comparison with the primary and secondary predicate devices (EPAD Device, K132616, and NeuroEPG System, K123843), the BraiN20® device is substantially equivalent in terms of intended use, fundamental technology, and overall safety and effectiveness.
Page 6 of 9
{11}
timeisbrain
K253301
While BraiN20® includes certain design differences—such as a limited testing mode (SSEP only), absence of wireless communication, a proprietary embedded operating system, integrated display, and multiple safety and usability enhancements. These differences do not raise new questions of safety or effectiveness.
Performance testing demonstrated compliance with the standards identified in this submission.
## **DEVICE DESCRIPTION**
BraiN20® is an active, portable, non-invasive medical device intended for monitoring neurological status through the acquisition and analysis of somatosensory evoked potentials (SSEP), specifically the N20 cortical response.
The system consists of a central unit incorporating embedded software, rechargeable batteries, stimulation and signal acquisition electronics, and an integrated touchscreen display. Patient contact is achieved through disposable stimulation and recording electrodes applied to intact skin for transient use (<24 hours).
BraiN20® delivers controlled electrical stimulation and acquires electrophysiological signals in a synchronized manner. Embedded software analyzes the recorded signals in real time to identify the presence or absence of the N20 response and, when detected, determine its latency and amplitude. Results are displayed directly on the device.
## **INTENDED USE**
The BraiN20® is intended for use in monitoring neurological status by recording the N20 response of somatosensory evoked potentials (SSEP).
The device provides objective electrophysiological information and is not intended to provide a diagnosis independently of clinical assessment.
## **PERFORMANCE TESTING**
Performance testing was conducted to demonstrate that BraiN20® performs as intended and is substantially equivalent to the predicate devices.
### **Electrical Safety**
Electrical safety testing was performed in accordance with applicable requirements of:
- IEC 60601-2-40
Page 7 of 9
{12}
timeisbrain
K253301
### Electromagnetic Compatibility
Electromagnetic compatibility (EMC) testing was conducted in accordance with:
- IEC 60601-1-2
### Software Verification and validation
Software verification and validation activities were performed in accordance with:
- IEC 62304
### Biocompatibility
Patient-contacting materials were evaluated for biocompatibility in accordance with:
- ISO 10993-1
- ISO10993-5
- ISO 10993-10
### Bench Performance Testing
Bench testing was performed to evaluate system functionality, signal acquisition performance, stimulation performance, battery operation, reliability, hardware integration, and overall system operation. (IEC 62133-2 for battery operation).
### Shelf Life Testing
Shelf-life testing, including accelerated aging evaluations, was conducted to support a 24-month shelf-life for applicable device components.
Additional testing demonstrated that the device maintains performance throughout its intended service life when maintained according to manufacturer recommendations.
### Usability Evaluation
Usability testing was conducted in accordance with applicable human factors and usability engineering principles, including IEC 62366-1 and IEC 60601-1-6
### Clinical Testing
No clinical studies were required or relied upon to support a determination of substantial equivalence for this submission.
### CONCLUSION
The information provided in this submission demonstrates that BraiN20® has the same intended use and similar technological characteristics as the identified predicate devices.
Performance testing confirmed that the device meets applicable safety, electromagnetic compatibility, software, biocompatibility, usability, shelf-life, and functional performance requirements.
Page 8 of 9
{13}
timeisbrain
K253301
Any differences between BraiN20® and the predicate devices do not raise new questions of safety or effectiveness. Therefore, BraiN20® is substantially equivalent to the legally marketed predicate devices identified in this submission.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.