Profoject Enteral/Oral Feeding Syringe; Profoject Reusable Enteral/Oral Feeding Syringe (Model A, Model B); Profoject ENFit Adaptor; Profoject Enteral Feeding Syringes Cap
Applicant
CMT Health PTE., Ltd.
Product Code
PNR · Gastroenterology, Urology
Decision Date
Apr 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Profoject™ Enteral/Oral Feeding Syringe: The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical and non-clinical settings by users ranging from clinicians to laypersons in all age groups. Profoject™ Reusable Enteral/Oral Feeding Syringe: The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in clinical and non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only. Profoject™ ENFit Adaptor: The device is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel. Profoject™ Enteral Feeding Syringes Cap: The Profoject™ Enteral Feeding Syringes Cap is used to prevent fluid loss and contamination of syringe contents until ready for use.
Device Story
Profoject enteral/oral feeding syringes (disposable/reusable) and accessories (ENFit adaptors, caps) facilitate fluid delivery into body orally or enterally. Syringes consist of barrel, plunger, and stopper/gasket; available in various volumes (0.5-60 mL). ENFit adaptors connect enteral giving sets to funnel-type catheters; caps prevent fluid loss/contamination. Used in clinical/non-clinical settings by clinicians or laypersons. Devices comply with ISO 80369-3 for ENFit connectivity to mitigate misconnection risks. Healthcare providers use syringes to measure/dispense fluids; adaptors enable compatibility between legacy and ENFit systems. Benefits include standardized, secure enteral connections and reduced contamination risk.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including performance testing (ISO 80369-3, ISO 7886-1, ISO 20695), biocompatibility (ISO 10993-1, -5, -10, -23, -4, -11, USP <151>, ASTM F756), and sterilization validation (ISO 11135, ISO 11737-1).
Indicated for all age groups in clinical and non-clinical settings for oral or enteral fluid delivery, measuring, and dispensing. Includes reusable options for single-patient use, ENFit adaptors for connection to enteral catheters, and syringe caps for contamination prevention.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Oral/Enteral Syringes with ENFit connector (12 mL to 60 mL), Low Dose Tip Oral/Enteral Syringes with ENFit connector (1 mL to 6 mL) (K183540)
Cedic Enteral Distal End ENFit Transition Connector, Cedic Enteral ENFit Transition Connector for Medication Port (K140581)
NeoMed NeoConnect Enteral Syringes with ENFit Connector and compatible NeoSecure Tip Caps (K152857)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 8, 2026
CMT Health Pte. Ltd
Monica Ma
Official Correspondent
150 Beach Road, #28-05, Gateway West
Singapore, 189720
Singapore
Re: K253268
Trade/Device Name: Profoject Enteral/Oral Feeding Syringe; Profoject Reusable Enteral/Oral Feeding Syringe (Model A, Model B); Profoject ENFit Adaptor; Profoject Enteral Feeding Syringes Cap
Regulation Number: 21 CFR 876.5980
Regulation Name: Gastrointestinal Tube And Accessories
Regulatory Class: Class II
Product Code: PNR, PIO
Dated: September 29, 2025
Received: September 29, 2025
Dear Monica Ma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253268 - Monica Ma
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K253268 - Monica Ma
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANTHONY LEE -S
Anthony Lee, Ph.D., M.B.A.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253268 | ? |
| Please provide the device trade name(s). | | ? |
| Profoject™ Enteral/Oral Feeding Syringe;
Profoject™ Reusable Enteral/Oral Feeding Syringe (Model A, Model B);
Profoject™ ENFit Adaptor;
Profoject™ Enteral Feeding Syringes Cap | | |
| Please provide your Indications for Use below. | | ? |
| Profoject™ Enteral/Oral Feeding Syringe:
The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is
intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical and
non-clinical settings by users ranging from clinicians to laypersons in all age groups.
Profoject™ Reusable Enteral/Oral Feeding Syringe:
The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is
intended to be used to deliver fluids into the body orally or enterally. It is intended to be used multiple times
in clinical and non-clinical settings by users ranging from clinicians to laypersons in all age groups. The
device is indicated for single patient use only.
Profoject™ ENFit Adaptor:
The device is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter
with funnel.
Profoject™ Enteral Feeding Syringes Cap:
The Profoject™ Enteral Feeding Syringes Cap is used to prevent fluid loss and contamination of syringe
contents until ready for use. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
K253268 - page 1/12
510(k) Summary
[As required by 21 CFR 807.92]
1. Submission Information
510(k) Number: K253268
Date: September 29, 2025
Type of 510(k) Submission: Traditional
Submitter: CMT HEALTH PTE. LTD.
150 BEACH ROAD, #28-05, GATEWAY WEST, SINGAPORE, 189720
Contact Person: Monica Ma
E-mail: ra@cmthealth.com
Tel: +65 6846 1379
2. Device Description
Proprietary Name: Profoject™ Enteral/ Oral Feeding Syringe, Profoject™ Reusable Enteral/ Oral Feeding Syringe, Profoject™ ENFit Adaptor, Profoject™ Enteral Feeding Syringes Cap
Common Name: Enteral Syringes With Enteral Specific Connectors, Enteral Specific Transition Connectors
Regulation Name: Gastrointestinal tube and accessories
Product Code: PNR, PIO
Device Class: II
Regulation Number: 21CFR 876.5980
Review Panel: Gastroenterology/Urology
Indications for use: Profoject™ Enteral/ Oral Feeding Syringe:
The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or non-clinical settings by users ranging from clinicians to laypersons in all age groups.
Profoject™ Reusable Enteral/ Oral Feeding Syringe:
The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in clinical and non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.
Profoject™ ENFit Adaptor:
The Profoject™ ENFit Adaptor is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel.
Profoject™ Enteral Feeding Syringes Cap:
The Profoject™ Enteral Feeding Syringes Cap is used to prevent fluid loss and contamination of syringe contents until ready for use.
{6}
K253268 - page 2/12
Device Description:
Projoject™ Enteral/ Oral Feeding Syringe:
Projoject™ Enteral/Oral Feeding Syringe is a disposable device designed for enteral and oral feeding, available in transparent and amber options and offered in multiple sizes ranging from 0.5 mL to 60 mL. It consists of plunger, plunger stopper, barrel, and it is used to deliver fluids into the body orally or connected to an enteral access device with male ENFit connector. Available with and without adaptor or syringes cap.
Projoject™ Reusable Enteral/ Oral Feeding Syringe:
The Projoject™ Reusable Enteral/Oral Feeding Syringe is a reusable device designed for enteral and oral feeding, available in multiple sizes ranging from 1 mL to 60 mL. It is offered in two models—Model A and Model B—with each model available in both transparent and amber products. The syringe comprises a plunger, O-ring gasket, and barrel, and is intended for the administration of fluids either orally or via connection to an enteral access device using a male ENFit connector.
Projoject™ ENFit Adaptor:
Projoject™ ENFit Adaptor is a disposable adaptor used to connect an enteral giving set with an ENFit connector to an enteral catheter with funnel. To facilitate the interconnection between newer devices having end ENFit connectors and the funnel feeding ports with existing ('legacy') end connectors, the adaptor is designed with two ports. One port is a male connector that complies with the ISO 80369-3 standard and can be directly connected to the corresponding female ENFit connector; the other port is a stepped connector that can adapt to previously used end connectors. The adaptor can supply with the feeding syringes or sold separately (sterile or non-sterile).
Projoject™ Enteral Feeding Syringes Cap:
Projoject™ Enteral Feeding Syringes Cap is a disposable tip cap used to prevent fluid loss and contamination of syringe contents until ready for use. The device has four model types: Model I, Model II, Model III and Model IV. The cap can supply with the feeding syringes or sold separately (sterile or non-sterile).
The proposed devices are available in sterile and non-sterile types. Sterile devices are sterilized by ethylene oxide gas to achieve a SAL of 10-6. The validity periods of both sterile and non-sterile products have been verified to be 5 years.
{7}
K253268 - page 3/12
## 3. Predicate Device Identification
K183540 - Oral/Enteral Syringes with ENFit connector (12 mL to 60 mL), Low Dose Tip Oral/Enteral Syringes with ENFit connector (1 mL to 6 mL)
K140581 - Cedic Enteral Distal End ENFit Transition Connector, Cedic Enteral ENFit Transition Connector for Medication Port
K152857 - NeoMed NeoConnect Enteral Syringes with ENFit Connector and compatible NeoSecure Tip Caps
## 4. Non-Clinical Test Conclusion
Non-clinical verification of the proposed devices has been conducted to evaluate their safety, performance, and functionality. The results of these tests have demonstrated the overall safety of the proposed device and its effectiveness in accordance with relevant test methods and ultimately support a substantial equivalence determination. Particularly, the following was conducted to adequately demonstrate the effectiveness of the proposed device in accordance with relevant test methods cited below:
## Performance Testing
1. ISO 20695 First edition 2020-03 Enteral feeding systems - Design and testing
2. ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
3. ISO 80369-1 Second edition 2018-11 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
4. ISO 80369-3 First Edition 2016-07-01 Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications [Including AMENDMENT 1 (2019)].
5. ISO 80369-20 First edition 2015-05-15 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
6. ISO 11737-1 Third edition 2018-01 [Including AMD1:2021] Sterilization of health care products - Microbiological methods - Part 1: Determination of a population of microorganisms on product [Including Amendment 1 (2021)]
7. USP - NF <62> Microbiological Examination of Nonsterile Products: Tests for Specified Microorganisms
8. ASTM F1886/F1886M-16 (2024) Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
9. ASTM F1140/F1140M-13 (2020) Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages
10. ASTM F88/F88M-23 Standard Test Method for Seal Strength of Flexible Barrier Materials
11. ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
12. USP - NF <71> Sterility Tests
13. ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
14. ISO 11135 Second edition 2014-07-15 Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]
{8}
K253268 - page 4/12
15. ISTA 3A: 2018 Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or Less
16. Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications - Guidance for Industry and Food and Drug Administration Staff
17. Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff
18. ASTM ST98:2022 Cleaning validation of health care products – requirements for development and validation of a cleaning process for medical devices
19. USP - NF <1111> Microbiological Examination of Nonsterile Products: Acceptance Criteria for Pharmaceutical Preparations and Substances for Pharmaceutical Use
20. USP - NF <61> Microbiological Examination of Nonsterile Products: Microbial Enumeration Tests
## Biocompatibility Testing
The biocompatibility evaluation for the proposed devices was conducted in accordance with ISO 10993-1:2018 Biological Evaluation of Medical Devices - Part 1. Evaluation and Testing within a Risk Management Process, as recognized by FDA. The proposed devices are classified as a surface device, mucosal membrane contact.
1. ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
2. ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
3. ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
4. USP <151> Rabbit Pyrogen Study
5. ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
6. ISO 10993-4 Third edition 2017-04 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood
7. ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
## 6. Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data was not included in this submission
{9}
K253268 - page 5/12
# 7. Substantially Equivalent Comparison Conclusion
Table 1 Device comparison table for Profoject™ Enteral/Oral Feeding Syringe and
Profoject™ Reusable Enteral/Oral Feeding Syringe
| Parameters | | Proposed Device | Predicate Device | Remark |
| --- | --- | --- | --- | --- |
| 1 | 510(k) Number | K253268 | K183540 | -- |
| 2 | 510(k) Holder | CMT HEALTH PTE. LTD. | NeoMed, Inc. | -- |
| 3 | Trade Name | Profoject™ Enteral/ Oral Feeding Syringe, Profoject™ Reusable Enteral/ Oral Feeding Syringe | Oral/Enteral Syringes with ENFit connector (12 mL to 60 mL), Low Dose Tip Oral/Enteral Syringes with ENFit connector (1 mL to 6 mL) | -- |
| 4 | Product Code | PNR | PNR | Same |
| 5 | Regulation Number | 21CFR 876.5980 | 21CFR 876.5980 | Same |
| 6 | Review Panel | Gastroenterology/Urology | Gastroenterology/Urology | Same |
| 7 | Device Class | II | II | Same |
| 8 | Indications for use | Profoject™ Enteral/Oral Feeding Syringe: The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical and non-clinical settings by users ranging from clinicians to laypersons in all age groups. Profoject™ Reusable/Oral Enteral Feeding Syringe: The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in clinical and non-clinical settings by users | Single Use Oral/Enteral Syringes with ENFit Connector (provided sterile and non-sterile): The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used to deliver fluids into the body orally or enterally. It is intended to be used in clinical and non-clinical settings by users ranging from clinicians to laypersons in all age groups. Reusable Oral/Enteral Syringes with ENFit Connector (provided non-sterile): The device is intended for use as a dispenser, a measuring device and an oral fluid transfer device. It is intended to be used | Similar Note 1 |
{10}
K253268 - page 6/12
| | | ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only. | multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only. | |
| --- | --- | --- | --- | --- |
| 9 | Target Population | All age groups | All age groups | Same |
| 10 | Single Use | Single use or single patient use | Single use or single patient use | Same |
| 11 | Configuration | Plunger stopper or gasket Plunger Barrel with ENFit connector Syringes cap (optional) ENFit adaptor (optional) | Gasket Plunger Barrel with ENFit connector | Similar Note 2 |
| 12 | Material | Polypropylene Polyisoprene rubber or Silicone Purple color additive Amber color additive | Polypropylene Silicone Polydimethylsiloxane White Colorant | Similar Note 3 |
| 13 | Product Size (nominal volumes) | Low dose tip ENFit syringe: 0.5ml, 1ml, 3ml, 5ml, 6ml | Low dose tip ENFit syringe: 1ml to 6ml | Similar Note 4 |
| | | Standard ENFit syringe: 5ml, 6ml, 10ml, 12ml, 20ml, 30ml, 35ml, 50ml, 60ml | Standard ENFit syringe: 12ml to 60ml | |
| 14 | Product Performance | Complied with: ISO 80369-3 ISO 7886-1 | Complied with: ISO 80369-3 ISO 7886-1 | Same |
| 15 | Biocompatibility | No Cytotoxicity
No Irritation
No Sensitization No
Acute Toxicity No
Pyrogenicity
No Subacute Toxicity | No Cytotoxicity
No Irritation
No Sensitization No
Acute Toxicity | Similar Note 5 |
| --- | --- | --- | --- | --- |
| 16 | Sterile | Sterile or non-sterile | Sterile or non-sterile | Same |
| 17 | Sterile Method | EO Sterilized (Only sterile device) | EO Sterilized (Only sterile device) | Same |
| 18 | Sterilization | EO (ethylene gas) to SAL=10-6(Only sterile device) | EO (ethylene gas) to SAL=10-6(Only sterile device) | Same |
{11}
K253268 - page 7/12
Note 1
For sterilized reusable products, performance tests are conducted in accordance with ISO 80369-3 and ISO 7886-1 standards, and residual levels of ethylene oxide (EO) and ethylene chlorohydrin (ECH) are tested per ISO 10993-7:2008/Amd.1:2019 standard. For non-sterile reusable products, performance tests are carried out in line with ISO 80369-3 and ISO 7886-1 standards, and microbial testing is performed per ISO 11737-1 standard.
Through cleaning validation and maximum re-use cycle verification, the proposed device can meet the needs of multiple uses without introducing additional cross-contamination risks, thus being suitable for use in both clinical and non-clinical settings.
Note 2
The configuration of proposed device is different from predicate device, but the proposed device of the structure has passed the verification of ISO 80369-3, ISO 80369-20 and ISO 7886-1. According to ISO 20695, enteral syringes shall consist of at least the following: a) a graduated container; b) unless the enteral syringe is designed for gravity use, there shall be a means to create pressure (e.g. a plunger or a bulb); c) an outlet port. Syringe cap and ENFit Adaptor are not required components, it will not affect the safety and effectiveness of the proposed device.
Note 3
Through the biocompatibility test, the results confirmed that the materials of the proposed device will not cause any safety issues and effectiveness issues.
Note 4
The product size for proposed device is different from predicate device, and this difference is just in infusion capacity and dose not effect indication for use, the physician can select by per patient's condition. And through performance test reports that proposed device will not raise safety and effectiveness issues.
Note 5
According to the guideline documents, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", proposed device added pyrogen test and subacute toxicity test. The test results comply with the requirements of the standard.
{12}
K253268 - page 8/12
Table 2 Device comparison table for Profoject™ ENFit Adaptor
| Parameters | | Proposed Device | Predicate Device | Remark |
| --- | --- | --- | --- | --- |
| 1 | 510(k) Number | K253268 | K140581 | -- |
| 2 | 510(k) Holder | CMT HEALTH PTE. LTD. | CEDIC S.R.L. | -- |
| 3 | Trade Name | Profoject™ ENFit Adaptor | Cedic Enteral Distal End ENFit Transition Connector, Cedic Enteral ENFit Transition Connector for Medication Port | -- |
| 4 | Product Code | PIO | PIO | Same |
| 5 | Regulation Number | 21CFR 876.5980 | 21CFR 876.5980 | Same |
| 6 | Review Panel | Gastroenterology/Urology | Gastroenterology/Urology | Same |
| 7 | Device Class | II | II | Same |
| 8 | Indications for use | The Profoject™ ENFit Adaptor is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel. | The Cedic Enteral Distal End ENFit Transition Connector is intended for connecting an enteral giving set with an ENFit connector to an enteral catheter with funnel.
Cedic Enteral ENFit Transition Connector For Medication Port: Intended for connecting a male oral tip syringe to an enteral giving set or enteral catheter with an ENFit medication port. Available with and without end cap.
Cedic Enteral Funnel ENFit Transition Connector: Intended for connecting an enteral giving set equipped with a stepped enteral distal end to an enteral catheter equipped with an ENFit connector. Available with and without end cap. | Same |
| 9 | Environment of Use | Hospital - Home | Hospital - Home | Same |
| 10 | User population | Healthcare professionals or layusers | Healthcare professionals or layusers | Same |
| 11 | Single Use | Single use | Single use | Same |
| 12 | Rx vs. OTC | Rx | Rx | Same |
{13}
K253268 - page 9/12
| 13 | Material | ABS; LDPE; Purple color additive | ABS HF 380 LDPE (Riblene MM20) Soft PVC (Nakan FMA919N) Remafin Violet PE43076356-ZT (2%) | Similar Note 1 |
| --- | --- | --- | --- | --- |
| 14 | Product Performance | Complied with: ISO 80369-3 ISO 80369-20 | Complied with: ISO 80369-3 ISO 80369-20 | Same |
| 15 | Expiration Date | 5 years | 3 years | Similar Note 2 |
| 16 | Sterile | Sterile or non-sterile | Non-sterile | Similar Note 3 |
Note 1
Through the biocompatibility test, the results confirmed that the materials of the proposed device will not cause any safety issues and effectiveness issues.
Note 2
The expiration date for proposed device is different from predicate device and performance testing proves that proposed devices still meet ISO 80369-3 and ISO 80369-20 requirements after 5 years of accelerated aging. Therefore, this difference does not raise new safety and effectiveness issues.
Note 3
For non-sterile proposed devices, performance testing has demonstrated that non-sterile proposed devices meet ISO 80369-3 and ISO 80369-20 requirements, and microbiological testing has demonstrated that microbial counts are within acceptable limits.
For sterile proposed devices, perform performance tests in accordance with ISO 80369-3 and ISO 7886-1 standards and test the residual levels of EO and ECH in accordance with ISO 10993-7:2008/Amd.1:2019 standards. The test results all meet the requirements of the corresponding standards; therefore, this difference does not raise new safety and effectiveness issues.
{14}
K253268 - page 10/12
Table 3 Profoject™ Enteral Feeding Syringes Cap
| Parameters | | Proposed Device | Predicate Device | Remark |
| --- | --- | --- | --- | --- |
| 1 | 510(k) Number | K253268 | K152857 | -- |
| 2 | 510(k) Holder | CMT HEALTH PTE. LTD. | NEOMED, INC. | -- |
| 3 | Trade Name | Profoject™ Enteral Feeding Syringes Cap | NeoMed NeoConnect Enteral Syringes with ENFit Connector and compatible NeoConnect™ NeoSecure™ Tip Cap | -- |
| 4 | Product Code | PNR | PNR | Same |
| 5 | Regulation Number | 21CFR 876.5980 | 21CFR 876.5980 | Same |
| 6 | Review Panel | Gastroenterology/Urology | Gastroenterology/Urology | Same |
| 7 | Device Class | II | II | Same |
| 8 | Indications for use | The Profoject™ Enteral Feeding Syringes Cap is used to prevent fluid loss and contamination of syringe contents until ready for use. | A NeoConnect™ Enteral Syringe with ENFit™ connector accessory used to prevent fluid loss and contamination of syringe contents until ready for use. | Same |
| 9 | Rx vs OTC | OTC | Rx | Note 1 |
| --- | --- | --- | --- | --- |
| 10 | Target Population | All age groups | All age groups | Same |
| 11 | Single Use | Single use | Single use | Same |
| 12 | Material | LDPE
TPE
Polypropylene
Purple color additive
Orange color additive | Polypropylene
Color additive | Similar
Note 2 |
| 13 | Connection | Male ENFit | Male ENFit | Same |
| 14 | Product Performance | Complied with:
ISO 80369-3
ISO 20695 | Unknown | Note 3 |
| 15 | Biocompatibility | No Cytotoxicity
No Intracutaneous
Reactivity No Skin | Unknown | Note 4 |
{15}
K253268 - page 11/12
| | | Sensitization | | |
| --- | --- | --- | --- | --- |
| 16 | Sterile | Sterile or non-sterile | Sterile or non-sterile | Same |
| 17 | Sterile Method | EO Sterilized (Only sterile device) | EO Sterilized (Only sterile device) | Same |
| 18 | SAL | 10^{-6} (Only sterile device) | 10^{-6} (Only sterile device) | Same |
Note 1
Based on the risk management, the device can be used safely and effectively by laypersons as OTC device.
Note 2
Through the biocompatibility and performance test according to ISO 80369-3:2016, the results confirmed that the materials of the proposed device will not cause any safety issues and effectiveness issues.
Note 3
The proposed device is designed in accordance with the ENFit specification and complies with ISO 80369-3, a recognized consensus standard for reducing the risk of misconnection. Validation testing conducted in accordance with ISO 20695:2020 and ISO 80369-3:2016 demonstrates that all results meet the requirements of these standards and therefore, the proposed device does not cause safety issues and effectiveness issues.
Note 4
According to ISO 10993-1:2018 and guidance of FDA’s “Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" | FDA”, the proposed device is adjudicated surface device-mucosal membrane indirect-contact time ≤ 24h, tests to be performed are cytotoxicity, skin sensitization and irritation or intracutaneous reactivity. The test results compliant with standard requirements and therefore, the proposed device does not cause safety issues and effectiveness issues.
{16}
K253268 - page 12/12
## The Conclusions:
The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices identified in the submission. Thus, the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.