K253254 · E.M.S Electro Medical Systems S.A · ELC · Jan 28, 2026 · Dental
Device Facts
Record ID
K253254
Device Name
GBT Machine Airflow Prophylaxis Master
Applicant
E.M.S Electro Medical Systems S.A
Product Code
ELC · Dental
Decision Date
Jan 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GBT Machine AIRFLOW Prophylaxis Master combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis. The GBT Machine AIRFLOW Prophylaxis Master is intended for use in the following dental and periodontal applications: - Removing supra and subgingival calculus deposits and stains from teeth - Periodontal pocket lavage with simultaneous ultrasonic tip movement - Scaling and root planing The GBT Machine AIRFLOW Prophylaxis Master is intended for use in the cleaning and polishing of teeth by the Projection of water, air, and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth. The GBT Machine AIRFLOW Prophylaxis Master can be used for the following cleaning procedures: - plaque removal for placement of sealants - surface preparation prior to bonding/cementation of inlays, onlays, crowns and veneers - surface preparation prior to placing composite restorations - effective plaque and stain removal for orthodontic patients - cleaning prior to bonding ortho brackets - cleaning implant fixture prior to loading - stain removal for shade determination - plaque removal prior to fluoride treatment - plaque and stain removal prior to whitening procedure The GBT Machine AIRFLOW Prophylaxis Master is also intended for use as an air-polisher in patients suffering from periodontal disease. The GBT Machine AIRFLOW Prophylaxis Master is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
Device Story
Integrated dental system combining ultrasonic scaler and air-polishing unit; central control unit with interchangeable handpieces (PIEZON, AIRFLOW, PERIOFLOW), powder chambers, irrigation system, wireless Bluetooth foot pedal. Clinician-operated in dental clinic; adjusts power/irrigation via control panel. Ultrasonic function (24–32kHz) uses piezoelectric vibration for calculus/stain removal; air-polishing function sprays air/water/powder mixture for cleaning. Connectivity (Wi-Fi/LTE) enables usage monitoring, remote diagnostics, and software updates. RFID-based handpiece recognition preconfigures settings. Output allows clinicians to perform periodontal maintenance, surface preparation, and prophylaxis, improving patient oral hygiene and treatment outcomes.
Clinical Evidence
No clinical data. Bench testing only. Verification included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), wireless coexistence (IEEE/ANSI C63.27), cybersecurity (static analysis/penetration testing), and performance testing per ISO 18397, ISO 7494-1, and ISO 7494-2.
Technological Characteristics
Integrated dental system; piezoelectric ultrasonic generator (24–32kHz); air-polishing delivery system. Features: electronic water flow control, RFID handpiece recognition, Wi-Fi/LTE connectivity. Standards: IEC 60601-1, IEC 80601-2-60, IEC 60601-1-2, ISO 18397, ISO 7494-1/2. Wireless: Bluetooth foot pedal, Wi-Fi/LTE for data/updates.
Indications for Use
Indicated for dental/periodontal patients requiring supra/subgingival calculus/stain removal, scaling, root planing, and cleaning/polishing of teeth. Used for plaque removal (sealants, ortho, restorations, whitening, fluoride), surface prep (bonding/cementation), and non-surgical subgingival plaque removal in periodontal pockets up to 5 mm post-initial treatment.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 28, 2026
E.M.S. Electro Medical Systems S.A.
% Sheila Hemeon-Heyer
President
Heyer Regulatory Solutions
125 Cherry Lane
Amherst, Massachusetts 01002
Re: K253254
Trade/Device Name: GBT Machine Airflow Prophylaxis Master
Regulation Number: 21 CFR 872.4850
Regulation Name: Ultrasonic Scaler
Regulatory Class: Class II
Product Code: ELC, KOJ
Dated: September 29, 2025
Received: September 29, 2025
Dear Sheila Hemeon-Heyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253254 - Sheila Hemeon-Heyer
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253254 - Sheila Hemeon-Heyer
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253254 | ? |
| Please provide the device trade name(s). | | ? |
| GBT Machine Airflow Prophylaxis Master | | |
| Please provide your Indications for Use below. | | ? |
| The GBT Machine AIRFLOW Prophylaxis Master combines the functions of an ultrasonic scaler and air-
polishing unit within a single chassis. | | |
| The GBT Machine AIRFLOW Prophylaxis Master is intended for use in the following dental and periodontal
applications: | | |
| - Removing supra and subgingival calculus deposits and stains from teeth | | |
| - Periodontal pocket lavage with simultaneous ultrasonic tip movement | | |
| - Scaling and root planing | | |
| The GBT Machine AIRFLOW Prophylaxis Master is intended for use in the cleaning and polishing of teeth
by the Projection of water, air, and dental powders onto the tooth surface. The device removes dental
plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of
teeth. | | |
| The GBT Machine AIRFLOW Prophylaxis Master can be used for the following cleaning procedures: | | |
| - plaque removal for placement of sealants | | |
| - surface preparation prior to bonding/cementation of inlays, onlays, crowns and veneers | | |
| - surface preparation prior to placing composite restorations | | |
| - effective plaque and stain removal for orthodontic patients | | |
| - cleaning prior to bonding ortho brackets | | |
| - cleaning implant fixture prior to loading | | |
| - stain removal for shade determination | | |
| - plaque removal prior to fluoride treatment | | |
| - plaque and stain removal prior to whitening procedure | | |
| The GBT Machine AIRFLOW Prophylaxis Master is also intended for use as an air-polisher in patients
suffering from periodontal disease. The GBT Machine AIRFLOW Prophylaxis Master is indicated for the
non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) #: K253254
510(k) Summary
Prepared on: 2026-01-14
| Contact Details | 21 CFR 807.92(a)(1) |
| --- | --- |
| Applicant Name | E.M.S. Electro Medical Systems S.A. |
| Applicant Address | Ch. de la Vuarpillière 31 Nyon 1260 Switzerland |
| Applicant Contact Telephone | +41224046476 |
| Applicant Contact | Mr. Florian Bertrand |
| Applicant Contact Email | fbertrand@ems-ch.com |
| Correspondent Name | Heyer Regulatory Solutions |
| Correspondent Address | 125 Cherry Lane Amherst MA 01002 United States |
| Correspondent Contact Telephone | 413-330-8578 |
| Correspondent Contact | Ms. Sheila Hemeon-Heyer |
| Correspondent Contact Email | Sheila@Heyer-Regulatory.com |
| Device Name | 21 CFR 807.92(a)(2) |
| --- | --- |
| Device Trade Name | GBT Machine Airflow Prophylaxis Master |
| Common Name | Ultrasonic scaler |
| Classification Name | Scaler, Ultrasonic |
| Regulation Number | 872.4850 |
| Product Code(s) | ELC, KOJ |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| --- | --- |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K190124 | AIRFLOW Prophylaxis Master | ELC, KOJ |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| --- | --- |
The GBT Machine AIRFLOW® Prophylaxis Master is a dental device that combines ultrasonic scaling and air-polishing functions into a single integrated system. It is a modified version of the AIRFLOW Prophylaxis Master previously cleared under 510(k) K190124. The device is compatible with EMS ultrasonic scaling instruments cleared under 510(k) K190124 and with prophylaxis powders including AIR-FLOW CLASSIC (K190124), AIR-FLOW PERIO (K190124), and AIR-FLOW PLUS (K171189).
The device is composed of a central control unit, interchangeable handpieces (PIEZON, AIRFLOW, and PERIOFLOW), powder chambers, an irrigation system, and a wireless Bluetooth foot pedal. The control unit contains the power supply, ultrasonic generator, powder and fluid delivery circuits, and a microprocessor control system.
Treatment is activated via the wireless foot pedal, and the control panel allows the clinician to adjust power and irrigation levels. The ultrasonic scaling function operates in the 24–32kHz range using piezoelectric vibration to remove calculus and stains, while the air-
{5}
polishing function sprays a mixture of air, water, and powder for supragingival and subgingival cleaning.
## Intended Use/Indications for Use
**21 CFR 807.92(a)(5)**
The GBT Machine AIRFLOW Prophylaxis Master combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis.
The GBT Machine AIRFLOW Prophylaxis Master is intended for use in the following dental and periodontal applications:
- Removing supra and subgingival calculus deposits and stains from teeth
- Periodontal pocket lavage with simultaneous ultrasonic tip movement
- Scaling and root planing
The GBT Machine AIRFLOW Prophylaxis Master is intended for use in the cleaning and polishing of teeth by the Projection of water, air, and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth.
The GBT Machine AIRFLOW Prophylaxis Master can be used for the following cleaning procedures:
- plaque removal for placement of sealants
- surface preparation prior to bonding/cementation of inlays, onlays, crowns and veneers
- surface preparation prior to placing composite restorations
- effective plaque and stain removal for orthodontic patients
- cleaning prior to bonding ortho brackets
- cleaning implant fixture prior to loading
- stain removal for shade determination
- plaque removal prior to fluoride treatment
- plaque and stain removal prior to whitening procedure
The GBT Machine AIRFLOW Prophylaxis Master is also intended for use as an air-polisher in patients suffering from periodontal disease. The GBT Machine AIRFLOW Prophylaxis Master is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.
## Indications for Use Comparison
**21 CFR 807.92(a)(5)**
The subject device has a subset of the indications for use for the periodontal applications and the same indications for use for the cleaning applications as compared to those cleared for the predicate device.
## Technological Comparison
**21 CFR 807.92(a)(6)**
The GBT Machine AIRFLOW® Prophylaxis Master shares the same fundamental technological characteristics as the predicate device, the AIRFLOW® Prophylaxis Master (K190124). Both devices integrate ultrasonic scaling and air-polishing functions within a single control unit. The scaling function operates on the basis of piezoelectric technology, while the air-polishing function projects a controlled mixture of air, water, and powder to remove plaque and stains.
Several technological improvements have been introduced in the GBT Machine to enhance usability, connectivity, and ergonomics. These include:
- An updated ultrasonic generator module with improved performance but same frequency range and power
- Electronic water flow control replacing the mechanical flow regulator
- Improved water bottle irrigation system
- New cart design
- Simplified user settings providing the same power/pressure to promote consistent use
- Default GBT setting providing optimum power and water flow settings for the majority of treatments
- RFID-based handpiece recognition to preconfigure settings automatically.
- Wi-Fi and LTE connectivity, enabling secure data transmission for usage monitoring, remote diagnostics, and software updates.
These connectivity features (Wi-Fi, LTE, RFID) do not impact the treatment functionality or the indications for use of the device. No new safety or effectiveness concerns are introduced by these additions. The device intended use, indications for use, principles of operation, anatomical sites, patient-contacting materials, and key performance characteristics remain unchanged.
{6}
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Design verification and validation activities were performed on the GBT Machine AIRFLOW Prophylaxis Master to ensure conformity with applicable requirements and to demonstrate substantial equivalence to the predicate device. These activities included:
- Electrical safety testing to IEC 60601-1:2005 + A1:2012 + A2:2020 (Edition 3.2) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 80601-2-60:2019 Medical electrical equipment Part 2: Particular requirements for basic safety and essential performance of dental equipment
- Electromagnetic compatibility (EMC) testing to IEC 60601-1-2:2014+AMD1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
- Wireless communication coexistence and testing to IEEE/ANSI C63.27-2021 American National Standard for Evaluation of Wireless Coexistence
- Cybersecurity testing (Static Analyses and Penetration Testing)
- Performance testing according to ISO 18397:2016 Dentistry Powered scalers, ISO 7494-1:2018 Dentistry Part 1: General requirements and test methods, and ISO 7494-2:2022 Dentistry Part 2: Water and air supply
- Bench testing to internal company protocols to verify the implementation of the changes introduced since the predicate device resulting in the GBT Machine Airflow Prophylaxis Master
- Software testing to verify correct operation of the five discrete software systems in the device
No clinical tests were submitted, referenced, or relied on in this 510(k) for a determination of substantial equivalence.
The design modifications introduced in the GBT Machine AIRFLOW Prophylaxis Master were specifically evaluated to ensure they do not negatively impact device performance, reliability, or safety. These modifications do not alter the intended use or raise new questions of safety or effectiveness compared to the predicate device. User instructions and labeling have been updated accordingly to reflect these changes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.