Elde Medical Electrodes (T5/ 5 cm x 5cm electrode T10/ 5 cm x 10cm electrode)
K253237 · Elde Medikal Sanayi ve Ticaret Anonim Sirketi · GXY · May 8, 2026 · Neurology
Device Facts
Record ID
K253237
Device Name
Elde Medical Electrodes (T5/ 5 cm x 5cm electrode T10/ 5 cm x 10cm electrode)
Applicant
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Product Code
GXY · Neurology
Decision Date
May 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Elde Medical electrode is intended to transmit electrical current to the patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limited to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-the-Counter) or Prescription use.
Device Story
Elde Medical Electrodes are cutaneous electrodes used as accessories for TENS or EMS devices. The device consists of a top cover, lead wire, conductive carbon film, and conductive hydrogel. It functions by transmitting electrical current from a stimulation unit through a lead wire to the conductive hydrogel, which is applied to the patient's skin. The device is non-sterile and intended for single-patient, multiple-application use. It is available in square (5cm x 5cm) and rectangular (5cm x 10cm) sizes. The device is used in clinical or home settings by patients or clinicians. The output is the delivery of electrical stimulation to the skin, which facilitates nerve or muscle stimulation for therapeutic purposes.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-5, -10, -23), accelerated aging (ASTM F1980-21), shipping performance (ASTM D4169-23), peel strength, and electrical impedance/safety testing (IEC 60601-2-2, IEC 60601-1).
Technological Characteristics
Materials: PET carrier liner, conductive carbon film, conductive hydrogel, top cover. Sensing/Actuation: Passive electrical conduction. Energy: Passive interface for external TENS/EMS stimulators. Dimensions: 50x50mm and 50x100mm. Connectivity: Lead wire with female socket. Sterilization: Non-sterile. Standards: ISO 10993-5/10/23, ASTM F1980-21, ASTM D4169-23, IEC 60601-1, IEC 60601-2-2.
Indications for Use
Indicated for adult patients requiring transcutaneous electrical nerve stimulation (TENS) or electrical muscular stimulation (EMS). Intended for single-patient, multiple-application use. No specific contraindications listed.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 8, 2026
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Deniz Ozgulbas
Regulatory Consultant
Necip Fazil Bulvari Keyap Carsi Sitesi No: 44
D-1 Blok Dukkan No: 53/2
Umraniye/Istanbul, 34775
Turkey
Re: K253237
Trade/Device Name: Elde Medical Electrodes (T5/ 5 cm x 5cm electrode T10/ 5 cm x 10cm electrode)
Regulation Number: 21 CFR 882.1320
Regulation Name: Cutaneous electrode
Regulatory Class: Class II
Product Code: GXY
Dated: September 28, 2025
Received: September 29, 2025
Dear Deniz Ozgulbas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253237 - Deniz Ozgulbas
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253237 - Deniz Ozgulbas
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253237 | ? |
| Please provide the device trade name(s). | | ? |
| Elde Medical Electrodes (T5/ 5 cm x 5cm electrode; T10/ 5 cm x 10cm electrode) | | |
| Please provide your Indications for Use below. | | ? |
| The Elde Medical electrode is intended to transmit electrical current to the patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limited to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-the-Counter) or Prescription use. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
ELDEMEDICAL
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Necip Fazil Bulvari Keyap Carsi Sitesi No: 44 D-1 Blok Dukkan No: 53/2
Umraniye Istanbul 34775 TURKEY
# K253237 510(k) Summary
## Applicant – 510(k) Owner
Name: Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Address: Necip Fazil Bulvari Keyap Carsi Sitesi No: 44 D-1 Blok Dukkan No: 53/2
Umraniye Istanbul 34775 TURKEY
Phone: +90 216 660 0074
Contact Person: Nurettin Eren
Prepared date: March 15, 2026
## Device Information
Trade name: Elde Medical Electrodes
Common name: Cutaneous Electrodes
Classification Name: Electrode, Cutaneous (21 CFR 882.1320)
Product Code: GXY
Regulatory Class: Class II
## Predicate Device
Trade name: Electrode Pad
Applicant: ShenZhen Deliduo Medical Technology Co., Ltd.
510(k) Number: K241512
Product Code: GXY
Regulatory Class: Class II
## Device Description
The Elde Medical electrode is used as an accessory to the TENS or EMS device unit, which transmits electrical current to patient skin. The electrical current of Elde Medical electrode is first transmitted via the lead wire then transmitted to the conductive gel which is adhered to patient skin.
1 of 5
{5}
ELDEREDICH
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Necip Fazil Bulvari Keyap Carsi Sitesi No: 44 D-1 Blok Dukkan No: 53/2
Umraniye Istanbul 34775 TURKEY
The Elde Medical electrode is composed of a top cover, lead wire, conductive carbon film, conductive hydrogel media, and a carrier liner. The carrier liner is made of PET (polyethylene terephthalate).
The Elde Medical electrode is non-sterile and intended for single adult patient multiple application use. The Elde Medical electrode has various shapes and sizes.
## Indications for Use
The Elde Medical electrode is intended to transmit electrical current to the patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limited to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over- The -Counter) or Prescription use.
## Comparison of Technological Characteristics with the Predicate Device
The following table compares technological and other characteristics of the subject and predicate devices.
| | Subject Device Elde Medical Electrode | Predicate Device (K241512) | Comparison |
| --- | --- | --- | --- |
| Trade Name | Elde Medical Electrode | Electrode Pad | N/A |
| Common Name | Cutaneous electrode | Cutaneous electrode | Same |
| Classification Name | Cutaneous electrode | Cutaneous electrode | Same |
| Product Code | GXY | GXY | Same |
| Regulation Number | 21 CFR 882.1320 | 21 CFR 882.1320 | Same |
| Medical Specialty | Neurology | Neurology | Same |
| Device Class | II | II | Same |
| Indications for Use | The Elde Medical electrode is intended to transmit electrical current to the patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limited to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS | The Electrode Pad is intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limited to | Same |
{6}
ELDEREDEN
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Necip Fazil Bulvari Keyap Carsi Sitesi No: 44 D-1 Blok Dukkan No: 53/2
Umraniye Istanbul 34775 TURKEY
| | (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use. | not limited to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use | |
| --- | --- | --- | --- |
| Target Population | Single patient use and multiple application | Single patient use and multiple application | Same |
| Prescription Use | OTC and Prescription | OTC and Prescription | Same |
| Design (Shape & Connection) | Rectangle and square, lead wire with female socket | Electrode Pad: rectangle, square, according to customized specification. Lead wire with female socket, or snap button with male snap connector | Similar |
| Materials | • Top cover material • Electrically conductive carbon film • Biocompatible conductive hydrogel • Protective film (Electrode carrier liner) | • Top cover material • Electrically conductive carbon film • Biocompatible conductive hydrogel Protective film (Electrode carrier liner) | Same |
| Electrode Pad Size | Square (50*50mm) Rectangular (50*100mm) | Square (50*50mm) Rectangular (50*100mm) | Same |
| Patient Contact are of Electrode Pad | Square: 2500 mm² Rectangle: 5000mm² | Square: 2500 mm² Rectangle: 5000mm² | Same |
| Hydrogel thickness | 1.0mm ±0.1mm | 0.75mm ±0.15mm | Similar |
| Electrode Impedance Testing | <300Ω | 426-635Ω | Different*1 |
| Conductive carbon film | PU/PE conductive film | PU/PE conductive film | Same |
| Shelf Life | 2 years | 2 years | Same |
| Stainless Steel Adhesion (180° peel) | 136 grams minimum (≅1.3N) | 136 grams minimum (≅1.3N) | |
3 of 5
{7}
ELDEREDICH
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Necip Fazil Bulvari Keyap Carsi Sitesi No: 44 D-1 Blok Dukkan No: 53/2
Umraniye Istanbul 34775 TURKEY
| Biocompatibility | ISO 10993-5:2009 ISO 10993-10:2010 ISO 10993-23:2021 | ISO 10993-5:2009 ISO 10993-10:2010 ISO 10993-23:2021 | Same |
| --- | --- | --- | --- |
| Sterilization | Non-sterile | Non-sterile | Same |
| Electrical safety | Lead wire meets clause 8.5.2.3 of IEC 60601-1:2005+AMD1:2012+AMD2:2020 | Lead wire meets Clause 8.5.2.3 of AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012 | Similar |
## Difference*1 analysis:
The subject device demonstrated impedance values of less than 300 ohms, while the predicate device reported a range of 426–635 ohms. Both devices passed electrical safety and impedance testing, confirming that the difference does not present new or increased risks. The lower impedance of the subject device supports efficient current transfer and does not alter the effectiveness of the electrical stimulation. Differences in reported values may also be attributable to variations in measurement methodology, as the predicate device’s test parameters are not specified. In conclusion, the differences in impedance values does not impact the subject device being as safe and effective as the predicate device.
## Performance Data (Non-Clinical Tests)
Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device or reference device. The test results demonstrated that the subject device complies with the following standards:
- ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- ISO 10993-10:2021 Biological Evaluation of Medical Devices - Part 10: Tests For Skin Sensitization
- ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation
- ASTM F1980-21, Standard guide for accelerated aging of sterile barrier systems for medical devices.
- ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems
- Peels strength testing(180° stainless steel peel)
- Impedance test per IEC 60601-2-2
- Lead wires test per IEC 60601-1
4 of 5
{8}
ELDEREDEN
Elde Medikal Sanayi ve Ticaret Anonim Sirketi
Necip Fazil Bulvari Keyap Carsi Sitesi No: 44 D-1 Blok Dukkan No: 53/2
Umraniye Istanbul 34775 TURKEY
## Clinical Tests
None.
## Conclusions
Based on successful biocompatibility testing of subject device, the electrical performance of the insulated lead wire components and Impedance testing, Elde Medical electrodes are safe and effective when used as an interface between a user's skin and an approved nerve and muscle stimulation device. The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed devices identified in the submission. Thus the subject device is substantially equivalent to the predicate devices under K241512.
5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.