G-BLOCK is intended to be used for the fabrication of permanent and temporary restorations such as dental implant prostheses, crowns, bridges, inlays, onlays and veneers.
Device Story
G-BLOCK is a graphene-doped PMMA polymer block used for fabricating dental prostheses via CAD/CAM milling. The device is supplied as a block attached to an aluminum pin for fixation in milling equipment. It is intended for use by dental lab technicians or dentists in a clinical or laboratory setting. The material provides high modulus, elastic limit, and deformation resistance, allowing for smaller prosthesis sections and reduced risk of cracks or fractures. The device is available in monochrome and multichroma variants to mimic natural dentition. By milling the block into specific anatomical shapes, clinicians create durable, lightweight, and color-stable restorations. The device is non-sterile and for single-patient use.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing, including mechanical performance (flexural modulus, flexural strength, compressive strength) and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, carcinogenicity) per ISO 20795-1, ISO 10477, ISO 7405, and ISO 10993.
Technological Characteristics
Material: Graphene-doped PMMA resin. Form factor: Polymer block with aluminum pin. Processing: CAD/CAM milling. Standards: ISO 20795-1, ISO 10477, ISO 48-4, ASTM E384, ISO 5833, ISO 13320. Properties: Flexural modulus >2000 MPa, flexural strength >65 MPa, water absorption <32 μg/mm³, residual monomer <2.2%. Non-radiopaque. Non-sterile.
Indications for Use
Indicated for patients requiring permanent or temporary dental restorations, including implant prostheses, crowns, bridges, inlays, onlays, and veneers. Suitable for all patient populations.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 29, 2025
Graphenano Dental S.L.
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120A
Madison, New Jersey 07940
Re: K253236
Trade/Device Name: G-Block
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth shade resin material
Regulatory Class: Class II
Product Code: EBF, EBG
Dated: September 26, 2025
Received: September 29, 2025
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253236 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253236 - Dave Yungvirt
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253236 | |
| Device Name G-BLOCK | |
| Indications for Use (Describe) G-BLOCK is intended to be used for the fabrication of permanent and temporary restorations such as dental implant prostheses, crowns, bridges, inlays, onlays and veneers. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K253236
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
I. SUBMITTER (21 CFR 807.92(a)(1))
DATE PREPARED: 28 September 2025
SUBMITTER NAME: Graphenano Dental S.L.
SUBMITTER ADDRESS: Polígono Industrial Táctica. Calle 2A, n.° 1. 46980 Paterna, Valencia (SPAIN)
CONTACT PERSON: Eva María Navarro Sáez, QA/RA Manager
TELEPHONE: +34 965 108 102
e-mail: Evanavarro@graphenanodental.com
II. DEVICE (21 CFR 807.92(a)(2))
DEVICE TRADE NAME: G-BLOCK
COMMON NAME: Tooth Shade Resin Material
CLASSIFICATION NAME: Material, tooth shade, resin
REGULATION DESCRIPTION: Tooth Shade Resin Material
CLASS: Class II (Special Controls)
REGULATION NUMBER: 21 CFR 872.3690
PRODUCT CODE: EBF – Material, Tooth Shade, Resin
SUBSEQUENT PRODUCT CODES: EBG – Crown and Bridge Temporary Resin
III. PREDICATE DEVICE (21 CFR 807.92(a)(3))
Primary Predicate:
K220329 G-CAM
Reference Predicate:
K160425 CAD/CAMouflage Milling Block
IV. DEVICE DESCRIPTION (21 CFR 807.92(a)(4))
G-BLOCK is a compact block made of graphene-doped polymer suitable for the manufacture of dental prostheses using CAD/CAM technology. The polymer is based on polymethyl methacrylate (PMMA) doped with graphene (an allotrope of carbon).
G-BLOCK is intended to be used for the fabrication of permanent and temporary restorations such as dental implant prostheses, crowns, bridges, inlays, onlays and veneers.
The previous indications for G-BLOCK are supported by the following concepts:
{5}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
- G-BLOCK, compounded of polymethyl methacrylate (PMMA) doped with Graphene, is manufactured using the heat-curing method.
- G-BLOCK presents high modulus and elastic limit to ensure that the tensions generated during biting and chewing do not cause permanent deformations, and it is possible to manufacture prosthesis of smaller sections.
- G-BLOCK presents high deformation resistance and stress limit, thus avoiding the formation of cracks and fractures.
- G-BLOCK is low density making the prosthesis lightweight.
- G-BLOCK increases the material hardness comparing with acrylic resins used in dentistry.
- G-BLOCK has colour stability.
- G-BLOCK has wide chromatic range, even within the same piece, making it look extremely natural.
- G-BLOCK is chemically inert.
- G-BLOCK water absorption is 4 µg/mm3 and a solubility of 0.5 µg/mm2. The release of residual monomer is minimum, with a percentage of 0.004% of residual monomer. Thanks to these physical properties G-BLOCK offers a durable and safety treatment.
G-BLOCK is available in different formats, sizes and colors, having all variations the same physicochemical characteristics.
# DEVICE DESING DESCRIPTION
The device is presented in the form of a polymer block glued to an aluminum pin that allows the device to be fixed on the milling machine so that the block can be milled to obtain dental prostheses with the use of specific CAD/CAM equipment.
G-BLOCK is available in 2 different formats: MONOCHROMA and MULTICHROMA. Monochrome and Multichroma blocks may be both used for anatomical monolithic restorations.
When machined, Monochroma and Multichroma present a different visual effect:
- Monochroma, is made of a pure colour matching the dentin.
{6}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
- Multichroma, has a chromatic spectrum based on the natural colour imitating the optical effects of the natural dentition.
Both variants are presented in two sizes: 14 and 40.
G-BLOCK device is available in the following colours, A1, A2, A3, A3.5, B1, B2, BL1, BL2, C2 (according to G-CAM shade guide).
G-BLOCK are provided non-sterile and as a single use device.
G-BLOCK must be used only by professionals as dental lab technicians and / or dentist.
G-BLOCK should be stored at room temperature in its original packaging, in dry storage and avoid exposure to direct sunlight. PMMA polymer from which the G-BLOCK is made is stable and can be stored for an extended period of 5-year shelf life.
{7}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
## V. INTENDED USE / INDICATIONS FOR USE (21 CFR 807.92(a)(5))
As established in the Indications for Use Statement:
> "G-BLOCK is intended to be used for the fabrication of permanent and temporary restorations such as dental implant prostheses, crowns, bridges, inlays, onlays and veneers."
G-BLOCK is classified under two of the three product codes included in the primary predicate device, G-CAM. Those codes are the following: EBF – Material, Tooth Shade, Resin; EBG – Crown and Bridge Temporary Resin. The intended use of G-BLOCK is reduced compared to G-CAM due to the different shape and dimensions, but both are equivalent. Specifically, one indication of G-BLOCK is that it is intended for final crowns and bridges, inlays, onlays and veneers restorations which means that can be used as a direct monolithic solution with no need of using other materials to make aesthetics restorations.
The reference device, CAD/CAMouflage Milling Block has the same physical presentation as G-BLOCK as is presented as well by a block shape. The block has smaller dimensions than the disc (primary predicate device shape). Its usefulness is intended for small parts so that is only classified as EBF- Material, Tooth Shade, Resin and EBG – Crown and Bridge Temporary Resin as the device proposed.
The differences with G-BLOCK and those two predicate devices do not affect the safety and effectiveness of the device when used as labeled.
## VI. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE PROPOSED DEVICE AND THE PREDICATE DEVICES (21 CFR 807.92(a)(6))
Graphenano Dental has used the FDA's Guidance for Industry and FDA Staff – “Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions” for comparing its G-BLOCK with
{8}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
the Primary Device, G-CAM, and Reference device, CAD/CAMouflage Milling Block. Both are legally marketed devices under 21 CFR 807.92(a)(3) classified as Class II. The technological characteristics comparison table (Table 1) outline and provides the difference, similarities and the substantial equivalency of G-BLOCK and the cited predicate and reference devices.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE DISCUSSION:
The G--BLOCK technological characteristics are the same as the predicate device G-CAM. Both are thermoplastic materials, designed to be processed with a CAD/CAM system. The main difference is that they are available in different shapes and, therefore, the anchorage to the milling machine is different as well. The primary predicate has a disc shape, and the proposed G-BLOCK has a block shape but is composed of exactly the same raw materials. Both devices have been evaluated under the applicable standards for their intended use, and they achieve the required values established by those.
When compared to the primary and reference predicate devices, the intended use and technological characteristics of G-BLOCK doesn't raise new issues related to safety and effectiveness. Based on the information provided in this submission, G-BLOCK is substantially equivalent to the primary predicate and reference devices.
The differences in physical properties and chemical compositions between G-BLOCK and the predicate devices do not impact safety and effectiveness. The finished product presented is biocompatible regardless of the shape variation.
In general, we consider the proposed device to be substantially equivalent to the predicate device and consider that any differences between the proposed and predicate and reference devices will not pose new concern.
{9}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
Table.1 Comparison with predicate and reference Device
| Feature | Proposed Device: G-BLOCK | Primary Predicate: G-CAM | Reference Device: CAD/CAMouflage Milling Block | Similarities & Differences (when difference with the predicate an explanation will be given) |
| --- | --- | --- | --- | --- |
| Manufacturer | Graphenano Dental, S.L. | Graphenano Dental, S.L. | Prismatik Dentalcraft, Inc. | Same as Primary Predicate |
| 510(k) N° | - | K220329 | K160425 | - |
| Class | Class II | Class II | Class II | Same |
| Device description | Polymer dental block | Polymer dental disc | Composite Restorative Material | Same as Primary Predicate
Equivalent to Reference Device |
| Product code- Regulation description | EBF – Material, Tooth Shade Resin
EBG – Crown and Bridge Temporary Resin | EBF – Material, Tooth Shade Resin
EBG – Crown and Bridge Temporary Resin
EBI – Denture Relining, Repairing, Rebasing Resin | EBF – Material, Tooth Shade Resin
EBG – Crown and Bridge Temporary Resin | Same as Reference Device
Equivalent to Primary Predicate
EBI not included for G-BLOCK as larger structures could be milled on a disc (i.e. dentures) not on a block. |
| Classification | 21 CFR 872.3690
21 CFR 872.3770 | 21 CFR 872.3690
21 CFR 872.3770
21 CFR 872.3760 | 21 CFR 872.3690
21 CFR 872.3770 | Same as Reference Device
Equivalent to Primary Predicate |
{10}
G-BLOCK
510(k) Premarket Notification Submission
510(k) SUMMARY
| Indications for Use | | G-BLOCK is intended to be used for the fabrication of permanent and temporary restorations such as dental implant prostheses, crowns, bridges, inlays, onlays and veneers | G-CAM discs are intended to be used for the manufacture of full and partial removable dentures, implant overdentures as well as permanent and temporary restorations such as anterior or posterior crowns and bridges, inlays, onlays, veneers, copings and substructures. | CAD/CAMouflage Milling Block is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces. The CAD/CAMouflage Milling Block is made for fabricating temporary and permanent restorations such as inlays, onlays, veneers and full crown restorations. | Partially same
G-Blocks have shorter indications for use than the primary predicate device which means that all the indications for use from the proposed device are indications for the primary predicate. |
| --- | --- | --- | --- | --- | --- |
| Technical characteristics | Flexural Modulus
ISO20795
>2000MPa | 3200 MPa +/-7%
(ISO 20795) | 3200 MPa +/-7%
(ISO 20795) | - | Same as Primary predicate and equivalent to reference device |
| | Flexural strength
ISO20795 >65MPa
ISO10477
≥ 50MPa | 140 MPa +/-7%
(ISO 20795)
148 MPa
(ISO 10477) | 140 MPa +/-7%
(ISO 20795)
148 MPa
(ISO 10477) | >100 MPa
(ISO 10477) | |
| | Surface hardness | 88 ShoreD
(ISO 48-4) | 88 ShoreD
(ISO 48-4)
19,5 KHN
(ASTM E384) | - | |
| | | 19,5 KHN
(ASTM E384) | | | |
| | Water absorption
ISO20975 <32 μg/mm³
ISO10477 ≤40 μg/mm³ | 4 μg/mm³
(ISO 20795) | 4 μg/mm³
(ISO 20795) | Water absorption
< 40 μg/mm³
(ISO 10477) | |
| | Residual monomer
ISO20795
<2,2% | <0.004%
(ISO 20795) | <0.004%
(ISO 20795) | - | |
| | Water solubility | <0.5 μg/mm³ | <0.5 μg/mm³ | <7.5 μg/mm³ | |
{11}
G-BLOCK
510(k) Premarket Notification Submission
510(k) SUMMARY
| | ISO20975
<1,6 μg/mm³
ISO10477
≤7.5 μg/mm³ | (ISO 20795) | (ISO 20795) | (ISO 10477) | |
| --- | --- | --- | --- | --- | --- |
| | Compressive strength (MPa) | 156 MPa
(ISO 5833) | 156 MPa
(ISO 5833) | - | |
| | Filler particle size distribution (μ) | 21,5 μ
(ISO 13320) | 21,5 μ
(ISO 13320) | - | |
| | Radio-opacity (mm Al) | Non-radiopaque | Non-radiopaque | - | |
| Performance testing | | Conforms with ISO 20795-1
Conforms with ISO 10477 | Conforms with ISO 20795-1
Conforms with ISO 10477 | Conforms with ISO 4049
Conforms with ISO 10477 | Same |
| Wearing time | | Permanent and temporary | Permanent and temporary | Permanent and temporary | Same |
| Usage | | Single patient | Single patient | Single patient | Same |
| Patient population | | Suitable for everyone | Suitable for everyone | Suitable for everyone | Same |
| Prescription/OTC | | Prescription Use Only | Prescription Use Only | Prescription Use Only | Same |
| Intended as an intraoral device | | Yes | Yes | Yes | Same |
| Device components | | No components | No components | No components | Same |
| Fixed/removable | | Fixed / Removable | Fixed / Removable | Fixed / Removable | Same |
| Design | | Block | Disc | Block | Same as Reference Device
Different to Primary Predicate |
| Size | | Size14: 12 x 14 x 18mm
Size40: 15.5 x 19 x 39 mm | Diameter 98mm
Diameter 95mm | Size14: 12 x 14 x 18mm | Same as Reference Device
Different to Primary Predicate |
| Appliance design | | CAD/CAM milling | CAD/CAM milling | CAD/CAM milling | Same |
| Supplied Sterile / Nonsterile | | Non- sterile | Non- sterile | Non- sterile | Same |
| Material | | Polymethylmethacrylate PMMA (Polymethylmethacrylate) resin doped with graphene | Polymethylmethacrylate PMMA (Polymethylmethacrylate) resin doped with graphene | Ceramic-filled nanohybrid Polymer resin | Same as Primary predicate and equivalent to reference device |
{12}
G-BLOCK
510(k) Premarket Notification Submission
510(k) SUMMARY
| Raw material: | Monomer Resin Graphene GNF-LS | Monomer Resin Graphene GNF-LS | Polymer resin with fillers and Pigments | Same as Primary predicate and equivalent to reference device |
| --- | --- | --- | --- | --- |
| Manufacturing | Powder + Liquid methacrylate-based resins with graphene powder mixed and heat cured | Powder + Liquid methacrylate-based resins with graphene powder mixed and heat cured | Powder + Liquid methacrylate-based resins with fillers and pigments mixed and heat cured | Equivalent to Reference device as new steps are added to mill the block shape and to add the anchorage pin to the block.
Same to Primary predicate |
| Maintenance | Oral hygiene | Oral hygiene | Oral hygiene | Same |
| Cleaning | Not specified | Not specified | Not specified | Same |
| Biocompatibility | Biocompatible | Biocompatible | Biocompatible | Same |
| Shelf life | 5 years | 5 years | Not specified | Same
The G-BLOCK shelf life is based on G-CAM shelf life calculation as each block is made by the same raw materials and method as G-CAM and, as well, the storage conditions are the same and the packaging equivalent, only changing in design. |
9
{13}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
## IDENTIFICATION OF THE RISK ANALYSIS METHOD
G-BLOCK has conducted a preliminary hazard analysis which identify risks, including risk of mechanical failure, toxicity and adverse tissue reaction, improper use, and incompatibility with other dental devices. The Failure Modes Effect Analysis (FMEA) showed that all risks, when reduced as far as possible, were acceptable. There were no so severe risks found to cause severe damage or that lead to the death of a patient. The device mechanical properties are comparable to predicate devices. The biocompatibility testing showed that G-BLOCK is biocompatible. The labeling of the device is designed to reduce the risk of improper use.
## DISCUSSION OF THE DEVICE CHARACTERISTICS
To reduce the risk of mechanical failure, G-BLOCK was designed according to applicable standard to its intended use. In addition, each production is subjected to exhaustive quality control to check that the product is in accordance with its technical specifications. Also, to reduce the risk of mechanical failure cause of milling errors, specific information is given to the professional clients in documents like Instructions for Use, Work process, Design Parameters Sheet and Labelling. Moreover, G-BLOCK is a thermoplastic solid material, so additional curing steps are not needed which prevents improper fabrication due to varying curing or working times.
To reduce the risk of toxicity and adverse tissue reaction, the material that is milled was tested against biocompatibility standards as specified by ISO 7405:2008, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry. Moreover, G-BLOCK labeling is designed to reduce the risk of improper use.
## DESCRIPTION OF THE PERFORMANCE ASPECTS (21 CFR 807.92 (b)(1)(2))
The testing of the performance aspects was performed to recognized standards, such as ISO 20795-1 and ISO 10477. (see Table 1 for specific application knowledge). There are no deviations to the procedure of the standard for the tests performed.
{14}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
## VII. NON-CLINICAL PERFORMANCE DATA
## SUMMARY DISCUSSION OF NON-CLINICAL DATA:
The subject device has been subject to bench to determine conformance to performance specifications and requirements taking account of its intended use and following all recommendations set out in FDA Document “Dental Composite Resin Devices-Premarket Notification [510(k)] Submissions”.
Non-clinical testing performed in foreseeable operating conditions showed correct operation of the device as per its intended use and specifically included mechanical performance.
Clinical testing is not required to demonstrate substantial equivalence as the indications for use are like legally marketed devices and the design is like designs previously cleared under a premarket notification.
## RELIANCE OF STANDARDS
The G-BLOCK product has demonstrated conformance with the non-clinical performance requirements through evaluation and testing in accordance with the following standards:
- Chemical and physical testing having under considerations ISO 4049; ISO 20795-1 and ISO 10477 (Compression, Elastic modulus, Bending strength, Water absorption and Residual monomer, radiopaque...)
- Biocompatibility check for material composition per ISO 7405 where the use of ISO 10993 is specified. Results for Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Subchronic Toxicity by Subcutaneous Implantation, Material-Mediated Pyrogenicity, Genotoxicity and Carcinogenicity concluded that the device is biocompatible.
The results of this non-clinical testing show that the strength of the G-BLOCK product is sufficient for its intended use and is substantially equivalent to the legally marketed predicate device.
{15}
G-BLOCK
510(k) Premarket Notification Submission
# 510(k) SUMMARY
## SUMMARY DISCUSSION OF CLINICAL DATA
Non-clinical test data are submitted to support this premarket notification and to establish substantial equivalence. No clinical studies are submitted.
## VIII. CONCLUSIONS
The results of the above-described studies demonstrate that the G-BLOCK is substantially equivalent in safety and effectiveness to the cleared primary predicate device and to the reference device based on the results of the physical properties and biocompatibility testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.