Midmark 800 Digital Vital Signs Device (1-300-0100, 1-300-0200, 1-300-0300, 1-300-0400)
Applicant
Midmark Corporation
Product Code
MWI · Cardiovascular
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Midmark 800 Digital Vital Signs Device is intended for use by clinicians and other medically qualified personnel to measure specific physiologic parameters in adult and pediatric patients. The device provides the following non-invasive measurements for patients 2 years and older, unless otherwise specified: • Non-Invasive Blood Pressure (NIBP): for patients aged 3 years and older • Pulse Rate • Body Temperature (temporal artery) • Functional Oxygen Saturation (SpO₂ - optional) • Respiration Rate from Photoplethysmogram (RRp - optional) The Midmark 800 Digital Vital Signs Device is intended for screening, spot-checking, or confirmation of vital signs, with data transfer to an electronic medical record (EMR) system. It is indicated for use in ambulatory care clinics, hospitals, and other points of care, under the direction of trained medical personnel.
Device Story
Non-invasive multi-parameter monitor for spot-checking vital signs; inputs include NIBP (oscillometric), pulse rate, temporal artery temperature (infrared), SpO₂ (pulse oximetry), and RRp (photoplethysmogram). Device integrates OEM modules (Exergen, Nellcor, or Masimo) without modification. Operated by clinicians in ambulatory clinics/hospitals; data transferred via USB-C to EMR. Provides real-time physiological measurements to assist clinical decision-making; replaces continuous monitoring predicate with spot-check functionality; improves UI readability/navigation.
Clinical Evidence
Clinical validation of NIBP module performed per ISO 81060-2:2018/Amd 2:2024 in non-randomized, dual-observer, single-center study (adult/pediatric). Device met all accuracy criteria for systolic/diastolic BP using Linear and Step algorithms. Pulse rate accuracy validated against single-lead ECG reference; results demonstrated equivalent performance.
Technological Characteristics
Multi-parameter monitor; NIBP (oscillometric), SpO₂ (Nellcor/Masimo modules), temporal artery temperature (infrared), RRp (photoplethysmogram). Connectivity: USB-C wired. Power: 100-240VAC or 7.4V Li-ion battery. Standards: IEC 60601-1, IEC 60601-1-2, ISO 81060-2, ISO 80601-2-56, ISO 80601-2-61, IEC 62304.
Indications for Use
Indicated for adult and pediatric patients (NIBP 3+ years; thermometer, SpO₂, RRp 2+ years) for screening, spot-checking, or confirmation of vital signs in clinical settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Masimo Radical-7 Pulse CO-Oximeter and Accessories (K193242)
Submission Summary (Full Text)
{0}
June 12, 2026
Midmark Corporation
Jessica Hembrey
Regulatory Affairs Specialist
60 Vista Dr.
Versailles, Ohio 45380
Re: K253196
Trade/Device Name: Midmark 800 Digital Vital Signs Device (1-300-0100, 1-300-0200, 1-300-0300, 1-300-0400)
Regulation Number: 21 CFR 870.2300
Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm)
Regulatory Class: Class II
Product Code: MWI, DQA, DXN, FLL, BZQ
Dated: May 13, 2026
Received: May 14, 2026
Dear Jessica Hembrey:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253196 - Jessica Hembrey
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253196 - Jessica Hembrey
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**STEPHEN C. BROWNING -S**
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253196 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Midmark 800 Digital Vital Signs Device (1-300-0100, 1-300-0200, 1-300-0300, 1-300-0400) | | |
| Please provide your Indications for Use below. | | ? |
| The Midmark 800 Digital Vital Signs Device is intended for use by clinicians and other medically qualified personnel to measure specific physiologic parameters in adult and pediatric patients. | | |
| The device provides the following non-invasive measurements for patients 2 years and older, unless otherwise specified: • Non-Invasive Blood Pressure (NIBP): for patients aged 3 years and older • Pulse Rate • Body Temperature (temporal artery) • Functional Oxygen Saturation (SpO₂ - optional) • Respiration Rate from Photoplethysmogram (RRp - optional) | | |
| The Midmark 800 Digital Vital Signs Device is intended for screening, spot-checking, or confirmation of vital signs, with data transfer to an electronic medical record (EMR) system. | | |
| It is indicated for use in ambulatory care clinics, hospitals, and other points of care, under the direction of trained medical personnel. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Page 1 of 1
{4}
# Midmark 800 Digital Vital Signs Device 510(k) Summary

Designing better care.™
Midmark Corporation
60 Vista Drive
Versailles, Ohio 45380
1-800-MIDMARK
midmark.com
In accordance with 21 CFR §807.92, a 510(k) summary is provided below with the required information.
# 1. Administrative Information
| Table 1. Administrative Information | |
| --- | --- |
| **Date Summary Prepared** | 09/26/2025 |
| **510(k) Sponsor Address** | Midmark Corporation 60 Vista Drive Versailles, Ohio 45380 |
| **Contact Person** | Jessica Hembrey Regulatory Affairs Specialist Email: jhembrey@midmark.com Phone: 847.415.9742 Fax: 937.526.8482 |
| **Trade Name** | Midmark 800 Digital Vital Signs Device (1-300-0100, 1-300-0200, 1-300-0300, 1-300-0400) |
| **Common Name** | Vital Signs Device |
| **Classification Name** | Monitor, Physiological, Patient (Without Arrhythmia Detection or Alarms) (21 CFR 870.2300, Product Code MWI) |
| **Associated Product Codes** | DQA, DXN, FLL, BZQ |
| **Device Classification** | Class II |
| **Review Panel** | Cardiovascular |
# 2. Equivalent Predicate Comparators
| Table 2. Predicate Devices | | | | |
| --- | --- | --- | --- | --- |
| Manufacturer Name | Predicate | Trade Name | 510(k) No. | Decision Date |
| Midmark Corporation | Primary | IQvitals Zone | K161909 | November 17, 2016 |
| Masimo Corporation | Secondary | Masimo Radical-7 Pulse CO-Oximeter and Accessories | K193242 | February 27, 2020 |
Midmark Corporation
Page 1 of 10
K253196
{5}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
### 3. Indications for Use
The Midmark 800 Digital Vital Signs Device is intended for use by clinicians and other medically qualified personnel to measure specific physiologic parameters in adult and pediatric patients.
The device provides the following non-invasive measurements for patients 2 years and older, unless otherwise specified:
- Non-invasive Blood Pressure (NIBP): for patients aged 3 years and older
- Pulse Rate
- Body Temperature (temporal artery)
- Functional Oxygen Saturation (SpO₂ - optional)
- Respiration Rate from Photoplethysmogram (RRp - optional)
The Midmark 800 Digital Vital Signs Device is intended for screening, spot-checking, or confirmation of vital signs, with data transfer to an electronic medical record (EMR) system.
It is indicated for use in ambulatory care clinics, hospitals, and other points of care, under the direction of trained medical personnel.
### 4. Device Models
Table 3. Midmark 800 Digital Vital Signs Device Configurations
| Finished Device Kit Number | Digital Vital Signs Device Model Number | Features | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | DVSD | Exergen Temporal Thermometer | Nellcor SpO₂ | Masimo SpO₂ | Masimo SpO₂ w/RRp |
| 4-000-0801 | 1-300-0100 | X | X | - | - | - |
| 4-000-0802 | 1-300-0200 | X | X | X | - | - |
| 4-000-0803 | 1-300-0300 | X | X | - | X | - |
| 4-000-0804 | 1-300-0400 | X | X | - | - | X |
### 5. Device Description
The Midmark 800 Digital Vital Signs Device is a non-invasive, multi-parameter monitoring system intended for use by clinicians and medically qualified personnel to obtain and display physiological measurements in adult and pediatric patients. The device is designed for spot-checking, screening, or confirmation of vital signs, with optional data transfer to an electronic medical record (EMR) system.
The device measures non-invasive blood pressure (NIBP), pulse rate, body temperature using an Exergen temporal artery thermometer cleared under K011291, and functional oxygen saturation (SpO₂) using either a Nellcor SpO₂ module cleared under K123581 or a Masimo SpO₂ module cleared under K193242. The OEM modules and sensors are used
Midmark Corporation
Page 2 of 10
K253196
{6}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
without modification to their technological characteristics and intended use, and have been validated for compatibility and performance with the Midmark 800 Digital Vital Signs Device. When configured with compatible components, the device also provides optional measurement of respiration rate derived from the photoplethysmogram (RRp).
The Midmark 800 is available in multiple model variants offering different combinations of measurement parameters and accessories, as shown in **Table 3**.
Device specifications and technological Characteristics are provided in **Table 4**.
## 6. Device Specifications and Technological Characteristics
**Table 4** provides a comparison of the intended use and technological characteristics of the Midmark 800 Digital Vital Signs Device and its predicate devices. As shown, the Midmark 800 has the same intended use as the primary predicate device and incorporates similar technological characteristics as multi-parameter vital signs monitors, each measuring non-invasive blood pressure (NIBP), pulse rate, body temperature, and optional SpO$_{2}$.
The Midmark 800 Digital Vital Signs Device differs from the primary predicate in that it is designed for spot-check monitoring rather than continuous monitoring, eliminates wireless communication in favor of USB-C wired connectivity, and incorporates user interface updates to improve readability and navigation. In addition, the Midmark 800 supports an optional Respiration Rate from Photoplethysmogram (RRp) feature through integration of a compatible Masimo SpO$_{2}$ module. The Masimo Radical-7 serves as the secondary predicate for this sole purpose. Validation testing, risk analysis, and comparative evaluation confirm that these technological differences do not raise new questions of safety or effectiveness.
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate |
| --- | --- | --- | --- |
| **Device Name** | **Midmark 800 DVSD** | **Midmark IQvitals Zone** | **Masimo Radical-7** |
| Manufacturer | Midmark Corporation | Midmark Corporation | Masimo |
| 510(k) Number | TBD | K161909 | K193242 |
| Device Classification | Class II | Class II | Class II |
| Product Code(s) | **MWI, DXN, DQA, FLL, BZQ** | **MWI, DXN, DQA, FLL** | **MWI, DXN, DQA, FLL, BZQ, CCK, DPZ** |
| Intended Use / Indications for Use | The Midmark 800 Digital Vital Signs Device is intended for use by clinicians and other medically qualified personnel to measure specific physiologic parameters in adult and pediatric patients. The device provides the following non-invasive measurements for patients 2 years and older, unless otherwise specified: | The IQvitals Zone is intended to be used by clinicians and medically qualified personnel for measuring and monitoring: • Noninvasive blood pressure for adult and pediatric patients (3 years and above) • Pulse rate for adult and pediatric patients • Noninvasive functional oxygen saturation of arterial | The Radical-7 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO_{2}) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well |
Midmark Corporation
Page 3 of 10
K253196
{7}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
Table 4. Substantial Equivalence Table – Comparison of Technological Characteristics
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate |
| --- | --- | --- | --- |
| | • Non-invasive Blood Pressure (NIBP): for patients aged 3 years and older • Pulse Rate • Body Temperature (temporal artery) • Functional Oxygen Saturation (SpO₂ - optional) • Respiration Rate from Photoplethysmogram (RRp - optional) The Midmark 800 Digital Vital Signs Device is intended for screening, spot-checking, or confirmation of vital signs, with data transfer to an electronic medical record (EMR) system. It is indicated for use in ambulatory care clinics, hospitals, and other points of care, under the direction of trained medical personnel. | hemoglobin (SpO₂) for adult and pediatric patients • Body temperature measured at Temporal Artery for adult and pediatric patients The most likely location for IQvitals Zone device to be used is for monitoring patients in general medical locations, hospitals and alternative care environments. | or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. The Radical-7 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Masimo Radical-7 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms. The Radical-7 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions in hospitals and hospital-type facilities. The Radical-7 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no |
Midmark Corporation
Page 4 of 10
K253196
{8}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
Table 4. Substantial Equivalence Table – Comparison of Technological Characteristics
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate |
| --- | --- | --- | --- |
| | | | motion conditions in hospitals and hospital-type facilities. The Radical-7 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities. The Radical-7 and Accessories are indicated for the continuous non-invasive monitoring of Respiratory Rate from photoplethysmogram (RRp) for adult and pediatric patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities. |
| Patient Population | Adult & Pediatric (3+ years for NIBP; 2+ years for thermometer, SpO₂ and RRp) | Adult and pediatric (ages 3+) | Adult and pediatric |
| Accessories | SpO₂ sensors, temporal scanner, BP cuffs | SpO₂ sensors, temporal scanner, BP cuffs | SpO₂ sensors |
| Available Module Configurations | Non-invasive Blood Pressure (NIBP) Pulse Oximeter Pulse Oximeter with optional RRp Temperature | Noninvasive Blood Pressure (NIBP) Pulse Oximeter Temperature | Pulse Oximeter |
| Connectivity | USB, wired data transfer | USB, Bluetooth, wireless capability | USB, Ethernet, optional wireless |
| Biocompatibility | Monitor housing is non-patient-contacting. OEM accessories (SpO₂ sensors, temperature probes, BP cuffs) have FDA-cleared biocompatibility. | Monitor housing is non-patient-contacting. Accessories have FDA-cleared biocompatibility. | SpO₂ accessory has FDA-cleared biocompatibility. |
Midmark Corporation
Page 5 of 10
K253196
{9}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
Table 4. Substantial Equivalence Table – Comparison of Technological Characteristics
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate |
| --- | --- | --- | --- |
| **Non-Invasive Blood Pressure (NIBP) Module Specifications** | | | |
| Operating Principle / Measurement Method | Oscillometric | Oscillometric | N/A |
| NIBP Range | Systolic: 50-260 mmHg Mean: 30-230 mmHg Diastolic: 20-210 mmHg | Systolic: 50-260 mmHg Mean: 30-230 mmHg Diastolic: 20-210 mmHg | N/A |
| NIBP Measurement Accuracy | ±5 mmHg (standard deviation ≤8 mmHg) | ±5 mmHg (standard deviation < 8 mmHg) | N/A |
| NIBP Measurement Resolution | 1 mmHg | 1 mmHg | N/A |
| Pulse Rate Range | 40-200 bpm | 40-200 bpm | N/A |
| Pulse Rate Accuracy | ±5 bpm or ±5%, whichever is greater | ±3 bpm or ±5%, whichever is greater | N/A |
| Pulse Rate Resolution | 1 bpm | 1 bpm | N/A |
| Initial Cuff Pressure | Automatic or User-selectable | Automatic or User-selectable | N/A |
| Maximum Cuff Pressure | 280 ± 5 mmHg | 280 ± 5 mmHg | N/A |
| Overpressure Cutoff | 300 ± 30 mmHg | 300 ± 30 mmHg | N/A |
| Cuff Types | Reusable | Reusable and single use cuffs | N/A |
| Cuff Sizes | Child Small Adult Adult Adult Long Large Adult Large Adult Long Thigh | Child Small Adult Adult Adult Long Large Adult Large Adult Long Thigh | N/A |
| **Masimo Pulse Oximetry Specifications** | | | |
| Pulse Oximeter | Rainbow SET Technology Board | MX Board (Rainbow SET) | Rainbow SET Technology Board |
| SpO₂ Modes | Spot Monitoring | Continuous Monitoring | Continuous and Spot Monitoring |
| SpO₂ Sensor Types | Reusable | Reusable and Disposable | Reusable |
| SpO₂ Sensor Sizes | Adult and Pediatric | Adult and Pediatric | Adult and Pediatric |
| SpO₂ Measurement Range | 0 - 100% | 0 - 100% | 0 - 100% |
| SpO₂ Measurement Accuracy | No Motion: 70-100% (±2%) With Motion: 70-100% (±3%) | No Motion: 70 - 100% (± 2%) With Motion: 70 - 100% (± 3%) | No Motion: 70-100% (±2%) |
Midmark Corporation
Page 6 of 10
K253196
{10}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
Table 4. Substantial Equivalence Table – Comparison of Technological Characteristics
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate |
| --- | --- | --- | --- |
| | | | With Motion: 70-100% (±3%) |
| SpO₂ / Pulse Rate Resolution | SpO₂: 1% Pulse Rate: 1 bpm | SpO₂: 1% Pulse Rate: 1 bpm | SpO₂: 1% Pulse Rate: 1 bpm |
| Pulse Rate Measurement Range | 25-240 bpm | 25-240 bpm | 25-240 bpm |
| Pulse Rate Measurement Accuracy | No motion: ±3 bpm With Motion: ±5 bpm | No motion: ±3 bpm With Motion: ±5 bpm | No motion: ±3 bpm With Motion: ±5 bpm |
| Respiration Rate Measurement Method | Respiration rate from photoplethysmogram (RRp) | N/A | Respiration rate from photoplethysmogram (RRp) |
| RRp Measurement | 4 - 70 rpm | N/A | 4 – 70 rpm |
| RRp Measurement Accuracy | ±3 rpm ARMS, ±1 rpm mean error | N/A | ±3 rpm ARMS, ±1 rpm mean error |
| **Nellcor Pulse Oximetry Specifications** | | | |
| Pulse Oximeter | Nellcor NELL-1 | Nellcor NELL-1 | N/A |
| SpO₂ Modes | Spot Monitoring | Continuous Monitoring | N/A |
| SpO₂ Sensor Types | Reusable | Reusable | N/A |
| SpO₂ Sensor Sizes | Adult and Pediatric | Adult and Pediatric | N/A |
| SpO₂ Measurement Range | 1 - 100% | 0 - 100% | N/A |
| SpO₂ Measurement Accuracy | No Motion: 70 - 100% (±2%) With Motion: 70 - 100% (±3%) | No Motion: 70 - 100% (± 2%) With Motion: 70 - 100% (±3%) | N/A |
| SpO₂ / Pulse Rate Resolution | SpO₂: 1% Pulse Rate: 1 bpm | SpO₂: 1% Pulse Rate: 1 bpm | N/A |
| Pulse Rate Measurement Range | 20 - 250 bpm | 20 - 240 bpm | N/A |
| Pulse Rate Measurement Accuracy | No motion: ±3 bpm With Motion: ±5 bpm | No motion: ±3 bpm With Motion: ±5 bpm | N/A |
| **Temperature Measurement Method** | | | |
| Temperature Sensor | Exergen Temporal Scanner | Exergen Temporal Scanner | N/A |
| Measurement Method | Infrared temporal artery measurement | Infrared temporal artery measurement | N/A |
| Temperature Measurement Range | 15.5°C - 43°C (60°F - 110°F) | 15.5°C - 43°C (60°F - 110°F) | N/A |
| Clinical Accuracy | ±0.1°C (±0.2°F) | ±0.1°C (±0.2°F) | N/A |
| Response Time | ~0.04 seconds | ~0.04 seconds | N/A |
Midmark Corporation
Page 7 of 10
K253196
{11}
## Midmark 800 Digital Vital Signs Device 510(k) Summary
**Table 4. Substantial Equivalence Table – Comparison of Technological Characteristics**
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate |
| --- | --- | --- | --- |
| Display Time | 30 seconds | 30 seconds | N/A |
| Resolution | 0.1°C / 0.1°F | 0.1°C / 0.1°F | N/A |
| **Additional Specifications for the Midmark 800** | | | |
| Alarms / Conditions | No patient alarms. Informational alerts (e.g., measurement error, peripheral connection) provided via user interface. | High priority: NIBP systolic, MAP, diastolic high or low; SpO_{2} high or low, and some high priority technical alarms Medium priority: SpO_{2} high or low; pulse rate high or low, and some medium priority technical alarms Low priority: Temperature high or low, and some low priority technical alarms. | N/A |
| Product Weight | 3.9 lb | 3.9 lb | N/A |
| Dimensions | 9.3' x 6.9' x 2.9' (LxWxH) | 10.5' x 4' x 7' (LxWxH) | N/A |
| Power Supply | 100-240VAC, 50/60Hz OR Rechargeable lithium-ion battery (7.4V) | 100-240VAC, 50-60 Hz OR Lithium-ion battery (7.2V) | N/A |
| Environmental Operating Conditions | Temperature: 10°C to 40°C Humidity: 15%-90% RH (non-condensing) Altitude: Up to 10,000 feet | Temperature: 10°C to 40°C Humidity: 15%-90% RH (non-condensing) Altitude: Up to 10,000 feet | N/A |
| **Comparison of Compliance with FDA-Recognized Standards** | | | |
| Compliance with FDA-Recognized Consensus Standards | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 80601-2-30 IEC 80601-2-49 ISO 81060-2 ISO 80601-2-56 ISO 80601-2-61 IEC 62133-2 IEC 62304 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-8 IEC 80601-2-30 IEC 80601-2-49 ISO 81060-2 ISO 80601-2-56 ISO 80601-2-61 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 60601-1-8 ISO 80601-2-61 IEC 62304 |
### 7. Non-Clinical Performance Data
Non-clinical testing and evaluations were performed for the Midmark 800 Digital Vital Signs Device. A summary of evaluations is shown in **Table 5**. Based on the results of studies performed, the Midmark 800 Digital Vital Signs Device does not raise new issues of safety or effectiveness.
Midmark Corporation
Page 8 of 10
K253196
{12}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
**Table 5. Non-Clinical Performance Data**
| Standard | Standard Name |
| --- | --- |
| ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. Amd2:2021] | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)] |
| IEC 60601-1-2 Edition 4.1:2020-09 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
| IEC 60601-1-6 Edition 3.2 2020-07 | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance – Collateral Standard: Usability |
| IEC 62366-1 Edition 1.1 2020-06 | Medical devices - Part 1: Application of usability engineering to medical devices |
| IEC 80601-2-30: Edition 2.0 2018-03 | Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers |
| IEC 80601-2-49 Edition 1.0 2018 | Medical electrical equipment – Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitors |
| ISO 80601-2-56 Edition 2.1 2021-02 | Medical electrical equipment – Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement |
| ISO 80601-2-61 Edition 2.0 2017-12 | Medical electrical equipment - Part 2-61: Particular requirements for the basic safety and essential performance of pulse oximeter equipment |
| ISO 10993-1 Edition 5.0 2018-08 | Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process |
| IEC 62133-2 Edition 1.1 2021-07 | Safety requirements for portable sealed secondary lithium batteries – Part 2: Lithium systems |
## 8. Clinical Performance Data
Clinical validation of the Midmark Non-Invasive Blood Pressure (NIBP) Module was conducted under ISO 81060-2:2018/Amd 2:2024 in a non-randomized, dual-observer, single-center study. Adult and pediatric subjects participated across multiple studies evaluating the Linear and Step algorithms. The device met all accuracy criteria for systolic and diastolic blood pressure in both adult and pediatric populations. Following the validation procedure outlined in ISO 81060-2:2018/AMD 1:2020/AMD 2:2024 Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type, the Midmark NIBP Module with the Step and Linear algorithms passed the criteria systolic and diastolic requirements for all study subsets.
A second single-center clinical study was conducted to evaluate pulse rate derived from the Midmark 800 NIBP module. Pulse rate accuracy was assessed using a single-lead ECG reference device with simultaneous acquisition in adult and pediatric participants. The results demonstrated equivalent pulse rate performance of the subject device relative to
Midmark Corporation
Page 9 of 10
K253196
{13}
# Midmark 800 Digital Vital Signs Device 510(k) Summary
the ECG reference device for the evaluated cuff sizes and measurement methods. Both the Step and Linear measurement methods were evaluated for pulse rate. For the Linear method, data from a single cuff size were not included in the analysis.
A summary of evaluations is shown in **Table 6**.
**Table 6. Clinical Performance Data**
| Standard | Standard Name |
| --- | --- |
| IEC 80601-2-30: Edition 2.0 2018-03 | Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers |
| ISO 81060-2:2018/Amd2:2024 | Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type |
| ISO TS 81060-5 First Edition 2020-02 | Non-invasive sphygmomanometers – Part 5: Requirements for the repeatability and reproducibility of NIBP simulators for testing of automated non-invasive sphygmomanometers |
| ISO 14155:2020-07 | Clinical investigation of medical devices for human subjects – Good clinical practice |
## 9. Conclusion
The results from non-clinical and clinical testing demonstrate that the Midmark 800 Digital Vital Signs Device is substantially equivalent, is as safe and effective, and performs as well as the legally marketed predicate devices, the Midmark IQvitals Zone (K161909) and the Masimo Radical-7 (K193242).
Midmark Corporation
Page 10 of 10
K253196
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.