Training & Validation Set: 3,623 subjects, 41,165 knees
>1 (qualified annotators)
Independent Test Set: 1,173 subjects, 11,383 knees
>1 (qualified annotators)
Indications for Use
MediAI-OA is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren & Lawrence Grading of standing, fixed-flexion radiographs of the knee. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, radiologists, orthopedists, physicians and medical technicians.
Device Story
MediAI-OA is an orthopedic image, computer-aided detection/diagnosis software; analyzes DICOM-compliant knee X-ray images (CR/DX); uses deep learning algorithms to perform knee detection, landmark detection, and joint space detection; outputs include KL grade (0-4), JSN status, osteophyte status, and sclerosis status; used in healthcare facilities by trained professionals (radiologists, orthopedists, physicians, technicians); provides summary analysis via screen, web report, and markup images; assists clinicians in diagnosing degenerative arthritis; does not replace full patient evaluation.
Clinical Evidence
Independent clinical performance test using 1,173 subjects (11,383 knees) from the Osteoarthritis Initiative (OAI) dataset, partitioned by site to prevent data leakage. Primary endpoints: sensitivity and specificity for KL grade, JSN, osteophytes, and sclerosis. Results: KL grade ≥2 sensitivity 0.876, specificity 0.865. JSN (≥OARSI grade 1) sensitivity 0.880, specificity 0.864. Osteophyte (≥OARSI grade 1) sensitivity 0.897, specificity 0.823. Sclerosis sensitivity 0.88, specificity 0.90. Performance metrics demonstrate substantial equivalence to manual annotations by trained readers.
Technological Characteristics
Automated radiological image processing software; deep learning algorithm; runs on server; inputs DICOM (CR/DX), JPEG, JPG, PNG; outputs static markup images and textual reports. Complies with IEC 62304 (software lifecycle), IEC 62366-1 (usability), ISO 14971 (risk management), and cybersecurity standards (SW96:2023, 81001-5-1, TIR57).
Indications for Use
Indicated for use by trained professionals (radiologists, orthopedists, physicians, technicians) to aid in the assessment of knee osteoarthritis (OA) features (sclerosis, joint space narrowing, osteophytes) and Kellgren & Lawrence (KL) grading in standing, fixed-flexion knee radiographs (CR or DX).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 15, 2026
CRESCOM Co., Ltd.
Rosa Han
Consultant
Room 801, 8, Seongnam-daero 331beon-gil, Bundang-gu
Seongnam-si, Gyeonggi-do 13558
Republic Of Korea
Re: K253172
Trade/Device Name: MediAI-OA
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: May 17, 2026
Received: May 18, 2026
Dear Rosa Han:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K253172 – Rosa Han
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K253172 – Rosa Han
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K253172
Device Name
MediAI-OA
Indications for Use (Describe)
MediAI-OA is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren & Lawrence Grading of standing, fixed-flexion radiographs of the knee.
It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.
The system is to be used by trained professionals including, but not limited to, radiologists, orthopedists, physicians and medical technicians.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
K253172
CRESCOM
## 1. 510(k) Summary
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
- • **Date** May 08, 2026
- • **Applicant / Submitter** CRESCOM Co., Ltd.
Room 801, 8, Seongnam-daero 331beon-gil, Bundang-gu, Seongnam-si, Gyeonggi-do, 13558, Republic of Korea
Phone +82-70-4820-3843
- • **Device Information**
Device(Trade) Name: MediAI-OA
Classification Name: Automated Radiological Image Processing Software
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: QIH
- • **Predicate Device**
Manufacturer: Alpha Intelligence Manifolds, Inc.
Device(Trade) Name: DeepXray
510(k) Number: K223621
Regulation Name: Medical image management and processing system
Regulatory Number: 21 CFR 892.2050
Regulatory Class: Class II
Product Code: QIH
- • **Identification of a Legally Marketed Predicate Device;**
MediAI-OA is substantially equivalent to the DeepXray marketed by Alpha Intelligence Manifolds, Inc., 510(k) Premarket Notification Number K223621, FDA Product Code QIH.
- • **General Description**
MediAI-OA is “Orthopedic image, computer aided detection/ diagnosis software” that automatically analyzes X-ray images of patients suspected of having osteoarthritis. It assists healthcare providers in making accurate and consistent diagnoses of knee osteoarthritis by providing information such as KL (Kellgren-Lawrence) grade, osteophyte grade, sclerosis, and JSN Grade.
- • **Indication for Use**
MediAI-OA is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren & Lawrence Grading of standing, fixed-flexion radiographs of the knee.
It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis.
The system is to be used by trained professionals including, but not limited to, radiologists, orthopedists, physicians and medical technicians.
1 / 6
{6}
# - • **Comparison of Technological Characteristics**
The MediAI-OA and the predicate device DeepXray (K223621) are orthopedic imaging detection and diagnostic support software that assists physicians in diagnosing degenerative arthritis by analyzing knee X-ray images. It displays the KL grade (Kellgren-Lawrence Grade) as well as the presence or absence of JSN, osteophytes, and sclerosis.
Both devices employ deep learning algorithms to analyze input images and the severity of arthritis.
They operate in healthcare facilities and hospitals, are intended for trained professionals including, but not limited to, radiologists, orthopedists, physicians and medical technicians, and are classified under 21 CFR 892.2050, Product Code QIH, Class II.
| | Subject Device | Predicate Device (K223621) |
| --- | --- | --- |
| **Device Name** | MediAI-OA | DeepXray |
| **Classification Name** | Automated Radiological Image Processing Software | Automated Radiological Image Processing Software |
| **Product Code** | QIH | QIH |
| **Intended Use /Indication for Use** | MediAI-OA is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren & Lawrence Grading of standing, fixed-flexion radiographs of the knee. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, radiologists, orthopedists, physicians and medical technicians. | DeepXray is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the measurement of minimum joint space width; the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren & Lawrence Grading of standing, fixed-flexion radiographs of the knee. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, radiologists, orthopedists, physicians and medical technicians. |
| **Anatomical Area** | Joint (knee) | Joint (knee) |
| **Image Input** | DICOM compliant images in either digitally computed (CR) or directly digital (DX) formats, jpeg, jpg, png. | DICOM compliant images in either digitally computed (CR) or directly digital (DX) formats |
| **Image Processing** | Knee detection; Landmark detection; Joint space detection | Knee detection; Landmark detection; Joint space detection |
| **Human Intervention for interpretation** | Required | Required |
| **Intended User** | Trained professionals | Trained professionals |
| **Output Format** | Summary analysis results through the screen, Web report, Markup images and textual report as static images | Web report with quality warning, markup images and editing interface |
| **Output Information** | - Knee OA status: KL grade 0/1, 2, 3 or 4 - JSN status: Absent/Present - Osteophyte status: Absent/Present - Sclerosis status: Absent/Present | - Knee OA status: KL grade ≥2 or ≤1 - JSN status: Absent/Present - Osteophyte status: Absent/Present - Sclerosis status: Absent/Present |
| **Runs on Server** | Yes | Yes |
2 / 6
{7}
# - • **Brief Summary of Non-Clinical Tests and Results**
The MediAI-OA complies with the following international and FDA-recognized consensus standards list in the below table.
| Recognition No. | Standard No. | Title of Standard | Remark |
| --- | --- | --- | --- |
| 13-79 | IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION | Medical device software - Software life cycle processes | |
| 5-129 | IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION | Medical devices - Part 1: Application of usability engineering to med | |
| 5-125 | ISO 14971 Third Edition 2019-12 | Medical devices - Application of risk management to medical device | |
| 13-131 | SW96:2023 | Standard for medical device security - Security risk management for device manufacturers | |
| 13-122 | 81001-5-1 Edition 1.0 2021-12 | Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle | |
| 13-83 | TIR57:2016 | Principles for medical device security - Risk management. | |
And MediAI-OA comply with the FDA guidance documents listed in the below table.
| Title of Guidance Document | Issue Date |
| --- | --- |
| Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] | July 28, 2014 |
| Content of Premarket Submissions for Device Software Functions, Guidance for Industry and Food and Drug Administration Staff | June 14, 2023 |
| Off-The-Shelf Software Use in Medical Devices | August 11, 2023 |
| Cybersecurity in Medical Devices Quality System Considerations and Content of Premarket Submissions | February 3, 2026 |
The risk analysis was completed and risk controls were implemented to mitigate identified hazards.
3 / 6
{8}
# • Brief Summary of Performance Tests and Results
CRESCOM conducted an independent clinical performance test of MediAI-OA. The test dataset was configured as follows. Site-based Data Partitioning To prevent data leakage and ensure the clinical validity of the algorithm, the study datasets were strictly partitioned by Clinical Site IDs rather than by individual patient randomization. The OAI protocol identifies multiple independent recruitment centers distributed across different geographic locations. Each site maintained its own participant pool and initial data files. The performance test was conducted using data from sites that were completely separate from those used for training and validation. The X-ray equipment manufacturers used for the images included Agfa, Fujifilm, GE and etc. (None of the cases used in this study were utilized for training or validation of the MediAI-OA model.)
| Characteristics | Category | | Training & Validation Set | | Independent Test Set | |
| --- | --- | --- | --- | --- | --- | --- |
| Gender | Female | | 2,156 (59.5%) | | 648 (55.2%) | |
| | Male | | 1,467 (40.5%) | | 525 (44.8%) | |
| Age (Years) | Age ≥ 60 | | 1,904 (52.6%) | | 681 (58.0%) | |
| | Age < 60 | | 1,719 (47.4%) | | 492 (42.0%) | |
| Race | Other Non-white | | 70 (2.0%) | | 12 (1.0%) | |
| | White | | 2,808 (77.5%) | | 982 (83.8%) | |
| | Black or African American | | 709 (19.6%) | | 165 (14.0%) | |
| | Asian | | 31 (0.8%) | | 14 (1.2%) | |
| | Unknown or not reported | | 5 (0.1%) | | 0 (0.0%) | |
| Total Subjects | | | 3,623 | | 1,173 | |
| KL | KL 0/1 | KL < 2 | 22,253 | | 6,598 | |
| | KL 2 | KL ≥ 2 | 10,212 | 18,912 | 2,799 | 4,785 |
| | KL 3 | | 5,813 | | 1,574 | |
| | KL 4 | | 2,887 | | 412 | |
| Total Knees | | | 41,165 | | 11,383 | |
| JSN | OARSI Grade 0 | | 53,441 | | 17,903 | |
| | OARSI Grade ≥1 | | 13,803 | | 4,861 | |
| Total JSN | | | 67,244 | | 22,764 | |
| Osteophyte | OARSI Grade 0 | | 20,138 | | 6,050 | |
| | OARSI Grade ≥1 | | 20,325 | | 5,056 | |
| Total Osteophyte | | | 40,493 | | 11,106 | |
| Sclerosis | Absent | | 28,357 | | 8,164 | |
| | Present | | 4,326 | | 2,822 | |
| Total Sclerosis | | | 32,683 | | 10,986 | |
4 / 6
{9}
| Characteristics | Category | Training & Validation Set | Independent Test Set |
| --- | --- | --- | --- |
| Modality | CR | 10,219 | 5,444 |
| | DX | 2,935 | 246 |
| | RG | 3,657 | 0 |
| | Unknown | 0 | 1 |
| X-ray Manufacturer | Agfa | 3,826 | 4,600 |
| | Fujifilm | 2,577 | 870 |
| | GE | 2,946 | 220 |
| | Konica-Minolta | 2,008 | 0 |
| | Pilips | 35 | 0 |
| | Siemens | 302 | 0 |
| | Swissray | 3,468 | 0 |
| | Others (Not reported) | 1,649 | 1 |
| Total DICOM | | 16,811 | 5,691 |
| Visiting Timepoint (Follow-up Period) | Baseline | 11,492 | 1,114 |
| | 12 months | 3,097 | 1,049 |
| | 24 months | 2,917 | 962 |
| | 36 months | 2,763 | 918 |
| | 48 months | 2,610 | 872 |
| | 72 months | 1,031 | 380 |
| | 96 months | 1,061 | 396 |
All annotations used as the reference standard were manually performed by qualified annotators who were rigorously trained in accordance with the OAI (Osteoarthritis Initiative) Standardized Reading Protocol. These annotations were used exclusively as ground truth during the model development and internal validation phases.
During the performance evaluation (inference) stage, the algorithm automatically generates all required diagnostic outputs, such as KL grade and sclerosis detection, without any manual intervention or prior knowledge of the reference labels. Furthermore, to ensure the objectivity of the performance evaluation, the datasets for training, validation, and testing were strictly separated, thereby eliminating any risk of data leakage.
The clinical performance test was conducted by comparing the software's osteoarthritis assessment results with the reference standard derived from the OAI dataset, using sensitivity and specificity as evaluation metrics. As a result, MediAI-OA was confirmed to meet all of the predefined performance criteria described below.
5 / 6
{10}
CRESCOM
- For KL grade classification, both sensitivity and specificity across all grades met performance targets comparable to those reported in previously conducted studies. In particular, for the detection of KL grade ≥ 2, the sensitivity was 0.876 (95% CI: 0.866–0.885) and the specificity was 0.865 (95% CI: 0.856–0.873). The lower bounds of the confidence intervals exceeded the predefined acceptance criteria, indicating strong discriminatory performance between normal and osteoarthritis cases.
Additionally, the sensitivity by KL grade was 0.865 (95% CI: 0.856–0.873) for grades 0 & 1, 0.717 (95% CI: 0.700–0.734) for grade 2, 0.750 (95% CI: 0.728–0.770) for grade 3, and 0.905 (95% CI: 0.873–0.930) for grade 4. The specificity was 0.876 (95% CI: 0.866–0.885) for grades 0 & 1, 0.862 (95% CI: 0.854–0.869) for grade 2, 0.972 (95% CI: 0.969–0.975) for grade 3, and 0.995 (95% CI: 0.993–0.996) for grade 4, demonstrating excellent discriminatory performance in distinguishing normal subjects from patients with osteoarthritis.
- For joint space narrowing (JSN) ≥ OARSI grade 1, the sensitivity and specificity were 0.880 (95% CI: 0.871–0.889) and 0.864 (95% CI: 0.860–0.869), respectively, demonstrating consistent and accurate detection of structural changes.
- For osteophyte detection ≥ OARSI grade 1, the sensitivity was 0.897 (95% CI: 0.888–0.905) and the specificity was 0.823 (95% CI: 0.813–0.833), indicating robust performance across varying severity levels.
- For sclerosis classification (presence/absence), the sensitivity and specificity were 0.88 (95% CI: 0.87–0.89) and 0.90 (95% CI: 0.89–0.90), respectively, demonstrating a high level of agreement with the reference standard.
Overall, MediAI-OA demonstrated accurate and reliable assessment of radiographic features of knee osteoarthritis, with clinical performance comparable to results obtained from annotations manually performed by qualified annotators who were rigorously trained in accordance with the OAI (Osteoarthritis Initiative) Standardized Reading Protocol.
# • Conclusion
Based on the non-clinical and clinical evidence, MediAI-OA is substantially equivalent to the legally marketed predicate device in terms of safety, effectiveness, and performance. The device introduces no new risks and demonstrates comparable functionality in aiding medical professionals in the assessment of radiographic features of knee osteoarthritis, supporting its suitability for regulatory clearance.
6 / 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.