Fule Metallic Lockable Intramedullary Nail Systems
K253155 · Beijing Fule Science & Technology Development Co., Ltd. · HSB · Jun 17, 2026 · Orthopedic
Device Facts
Record ID
K253155
Device Name
Fule Metallic Lockable Intramedullary Nail Systems
Applicant
Beijing Fule Science & Technology Development Co., Ltd.
Product Code
HSB · Orthopedic
Decision Date
Jun 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intramedullary Nail System of The Fule Metallic Lockable Intramedullary Nail is intended to provide temporary stabilization of various types of fractures, malunions and nonunion of the tibia. The nails are inserted using an open or closed technique and can be statically, dynamically and compressed locked. It is indicated for long bone fracture fixation, specifically tibial fracture fixation, which may include the following: - Open and closed tibial fractures - Pseudoarthrosis and correction osteotomy - Pathologic fractures, impending pathologic fractures, and tumor resections - Nonunion and malunion. The Fule Proximal Femoral Nail Systems (PFNA System and SISF System) of the Fule Metallic Lockable Intramedullary Nail are intended for the treatment of stable and unstable fractures of the proximal femur, including pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations thereof. - Short nails (PFNA:170–240 mm / SISF: 180–240 mm): - Pertrochanteric fractures - Intertrochanteric fractures - Basal neck fractures - Combinations of these fracture types - Long nails (PFNA:300–420 mm): - Subtrochanteric fractures - Pertrochanteric fractures associated with shaft fractures - Pathologic fractures (including prophylactic use) in both trochanteric and diaphyseal regions - Long subtrochanteric fractures - Proximal or distal non-unions - Malunions - Revisions
Device Story
Metallic orthopedic internal fixation system; includes intramedullary nails, proximal fixation components, compression nails/blades, locking heads, and interlocking screws. Used for temporary stabilization of long bone fractures (tibia and proximal femur). Inserted via open or closed surgical technique by orthopedic surgeons. System provides mechanical support to bone segments during healing. Implants are titanium alloy; reusable instruments are stainless steel. Supplied non-sterile; requires cleaning and steam sterilization by healthcare facility prior to use. Output is a stabilized fracture construct; facilitates bone union and patient mobility.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing (mechanical performance) and biocompatibility evaluation. Bench testing included static/dynamic four-point bending (ASTM F1264-24), torsion (ASTM F1264-24), locking screw bending (ASTM F1264-24), axial pullout/driving torque (ASTM F543-23), and compression bending (ASTM F384-17).
Technological Characteristics
Titanium alloy implants; stainless steel reusable instruments. Intramedullary nail-based fixation. Dimensions vary by fracture indication. Non-sterile, requires steam sterilization. Mechanical performance validated per ASTM F1264-24, ASTM F543-23, and ASTM F384-17. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring temporary stabilization of tibial fractures (open/closed, pseudoarthrosis, osteotomy, pathologic, nonunion, malunion) or proximal femoral fractures (stable/unstable pertrochanteric, intertrochanteric, basal neck, subtrochanteric, and associated shaft fractures).
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
T2 Tibial Nailing System, T2 Femoral Nail System, T2 Supracondylar Nail System, T2 Recon Nail System, T2 Greater Trochanter Nail (GTN), T2 Ankle Arthrodesis Nail, T2 Arthrodesis Nail System (K200880)
DePuy Synthes Trauma Orthopedic Nail Implants / Proximal Femoral Nail System (K182783)
{0}
June 17, 2026
Beijing Fule Science & Technology Development Co., Ltd.
Xiaoliang Chen
International Sales Manager
No. 50, Mafang West Industry Zone, Pinggu District
Beijing, 101101
China
Re: K253155
Trade/Device Name: Fule Metallic Lockable Intramedullary Nail Systems
Regulation Number: 21 CFR 888.3020
Regulation Name: Intramedullary fixation rod
Regulatory Class: Class II
Product Code: HSB, HWC
Dated: May 20, 2026
Received: May 20, 2026
Dear Xiaoliang Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253155 - Xiaoliang Chen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K253155 - Xiaoliang Chen
Page 3
Sincerely,
**FARZANA SHARMIN -S**
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K253155
Device Name
Fule Metallic Lockable Intramedullary Nail Systems
Indications for Use (Describe)
The Intramedullary Nail System of The Fule Metallic Lockable Intramedullary Nail is intended to provide temporary stabilization of various types of fractures, malunions and nonunion of the tibia. The nails are inserted using an open or closed technique and can be statically, dynamically and compressed locked. It is indicated for long bone fracture fixation, specifically tibial fracture fixation, which may include the following:
- Open and closed tibial fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures, and tumor resections
- Nonunion and malunion.
The Fule Proximal Femoral Nail Systems (PFNA System and SISF System) of the Fule Metallic Lockable Intramedullary Nail are intended for the treatment of stable and unstable fractures of the proximal femur, including pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations thereof.
- Short nails (PFNA:170–240 mm / SISF: 180–240 mm):
- Pertrochanteric fractures
- Intertrochanteric fractures
- Basal neck fractures
- Combinations of these fracture types
- Long nails (PFNA:300–420 mm):
- Subtrochanteric fractures
- Pertrochanteric fractures associated with shaft fractures
- Pathologic fractures (including prophylactic use) in both trochanteric and diaphyseal regions
- Long subtrochanteric fractures
- Proximal or distal non-unions
- Malunions
- Revisions
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K253155 - Page 1 of 8
## 510(k) Summary
**510(k) Number:** K253155
**Date Prepared:** 2026-06-17
### 1. Submitter / Applicant Information
**Applicant Name:**
BEIJING FULE SCIENCE & TECHNOLOGY DEVELOPMENT CO., LTD.
**Applicant Address:**
No. 50, Mafang West Industry Zone, Pinggu District, Beijing 101101, China
**Contact Person:**
Mr. Xiaoliang Chen
**Telephone:**
+86 18210731904
**Email:**
xiao.liang.c.fule@gmail.com
### 2. Device Information
**Trade Name:**
Fule Metallic Lockable Intramedullary Nail Systems
**Common Name:**
Intramedullary fixation rod
**Classification Name:**
Rod, Fixation, Intramedullary and Accessories
**Regulation Number:**
21 CFR 888.3020
**Product Code(s):**
HSB, HWC
**Device Classification:**
Class II
{5}
K253155 - Page 2 of 8
### 3. Legally Marketed Predicate Devices
The subject device claims substantial equivalence to the primary predicate device identified below. Other legally marketed devices are identified as additional predicate devices to support approach-specific configurations and indications.
| Predicate Status | 510(k) Number | Predicate Device | Product Code |
| --- | --- | --- | --- |
| **Primary predicate** | K200880 | T2 Tibial Nailing System, T2 Femoral Nail System, T2 Supracondylar Nail System, T2 Recon Nail System, T2 Greater Trochanter Nail (GTN),T2 Ankle Arthrodesis Nail,T2 Arthrodesis Nail System | HSB |
| **Additional predicate** | K182783 | DePuy Synthes Trauma Orthopedic Nail Implants / Proximal Femoral Nail System | HRS / HSB as applicable |
### 4. Device Description
The Fule Metallic Lockable Intramedullary Nail Systems are metallic orthopedic internal fixation devices intended to provide temporary stabilization of long bone fractures by insertion of an intramedullary nail into the medullary canal. The system consists of intramedullary nails, proximal fixation components, compression nails/blades, locking heads, interlocking screws, and associated reusable surgical instruments. The devices are intended for use in skeletally mature patients.
The product family includes two device groups:
#### 1. Fule Intramedullary Nail System
This system is intended for tibial fracture fixation. It includes intramedullary nails and associated locking screws designed to provide temporary stabilization of tibial fractures, malunions, nonunions, and related long bone conditions.
#### 2. Fule Proximal Femoral Nail Systems, including PFNA and SISF sub-systems
These systems are intended for proximal femoral fracture fixation. The systems include short nail configurations and long nail configurations,
{6}
K253155 - Page 3 of 8
depending on the fracture indication. The proximal femoral systems include femoral nails, anti-rotation compression nails/blades or screw-type cephalomedullary elements, locking heads, and interlocking screws.
The subject implants are manufactured from titanium alloy material suitable for long-term orthopedic implantation. The devices are supplied non-sterile and are intended to be sterilized before use according to the validated sterilization instructions in the labeling.
Reusable surgical instruments associated with the system are supplied non-sterile and are packaged in protective polyethylene packaging. The packaging is intended for physical protection during shipping, storage, and handling, and is not intended to provide or maintain sterility.
### 5. Intended Use / Indications for Use
The Intramedullary Nail System of The Fule Metallic Lockable Intramedullary Nail is intended to provide temporary stabilization of various types of fractures, malunions and nonunion of the tibia. The nails are inserted using an open or closed technique and can be statically, dynamically and compressed locked. It is indicated for long bone fracture fixation, specifically tibial fracture fixation, which may include the following:
- Open and closed tibial fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures, and tumor resections
- Nonunion and malunion.
* The Fule Proximal Femoral Nail Systems (PFNA System and SISF System) of the Fule Metallic Lockable Intramedullary Nail are intended for the treatment of stable and unstable fractures of the proximal femur, including pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations thereof.
- Short nails (PFNA:170–240 mm / SISF: 180–240 mm):
- Pertrochanteric fractures
- Intertrochanteric fractures
{7}
K253155 - Page 4 of 8
- Basal neck fractures
- Combinations of these fracture types
- Long nails (PFNA:300–420 mm):
- Subtrochanteric fractures
- Pertrochanteric fractures associated with shaft fractures
- Pathologic fractures (including prophylactic use) in both trochanteric and diaphyseal regions
- Long subtrochanteric fractures
- Proximal or distal non-unions
- Malunions
- Revisions
### 6. Technological Characteristics and Comparison to Predicate Devices
The subject devices and predicate devices have the same general intended use: temporary stabilization and fixation of bone fractures using metallic intramedullary nail constructs.
The Fule Intramedullary Nail System and the T2 Tibial Nailing System predicate are both intramedullary nail systems intended for tibial fracture fixation. Both systems use metallic intramedullary nails inserted into the medullary canal and locking screws to stabilize fractured bone segments during healing.
The Fule Proximal Femoral Nail Systems, including the PFNA and SISF sub-systems, and the K182783 predicate device are proximal femoral intramedullary nail systems intended for the treatment of stable and unstable proximal femoral fractures. Both systems use a metallic femoral nail, proximal femoral head/neck fixation components, and distal/proximal locking components to stabilize fractures during healing. The PFNA substantial equivalence comparison states that the subject and predicate devices have the same intended use and similar technological characteristics, including intramedullary nail-based fixation, titanium alloy construction, comparable dimensions, proximal femoral head/neck fixation
{8}
K253155 - Page 5 of 8
components, and locking screw/bolt functions.
Differences between the subject and predicate devices include dimensional ranges, titanium alloy grade, component geometry, and sub-system configurations. These differences do not introduce a new intended use, a new anatomical site, a new patient population, a new surgical approach, or a new mode of action. The differences were evaluated through dimensional comparison, material conformity evaluation, biocompatibility evaluation, and nonclinical mechanical performance testing.
## 7. Nonclinical Performance Testing
Nonclinical bench testing was performed using final or representative finished devices to evaluate the mechanical safety, structural integrity, fixation strength, and functional performance of the subject devices under loading modes relevant to intramedullary fixation.
The following nonclinical performance tests were conducted or relied upon:
- Static and dynamic four-point bending testing of the intramedullary nail in accordance with ASTM F1264-24
- Static torsion testing of the intramedullary nail in accordance with ASTM F1264-24
- Static and dynamic bending testing of locking screws in accordance with ASTM F1264-24
- Torsional properties, axial pullout strength, and driving torque testing of metallic interlocking screws in accordance with ASTM F543-23
- Static and dynamic axial compression bending testing of the compression nail in accordance with ASTM F384-17
- Cutout resistance testing of the compression nail using a device-specific method relevant to cephalomedullary fixation performance
The submitted nonclinical testing supports that the subject devices have adequate mechanical performance for their intended use and do not raise new or different questions of safety or effectiveness compared with the predicate devices. The
{9}
K253155 - Page 6 of 8
submitted supporting information identifies ASTM F1264-24, ASTM F543-23, ASTM F384-17, cutout resistance testing, instrument biocompatibility bridging, and non-sterile packaging documentation as supporting evidence.
## 8. Biocompatibility
The subject implants are manufactured from titanium alloy materials commonly used for long-term orthopedic implantation. Biocompatibility and material information were evaluated to support the biological safety of the patient-contacting implant materials and final finished device.
For the reusable surgical instruments, patient-contacting components are medical-grade stainless steel. The biocompatibility of the reusable surgical instruments was supported through a bridging assessment to the FDA-cleared Fule Spinal Fixation System instruments under K252730. The bridging rationale states that the patient-contacting stainless steel components are equivalent with respect to material formulation, raw material specification, processing, machining, surface finishing, cleaning, packaging configuration, and sterilization status, and that no new chemicals, additives, fillers, plasticizers, lubricants, coatings, or mold-release agents are introduced.
Based on the material information, manufacturing information, cleaning and sterilization status, and biocompatibility evaluation, the subject devices do not present new or increased biocompatibility risks compared with the predicate devices.
## 9. Sterilization, Cleaning, and Packaging
The subject implants and reusable surgical instruments are supplied non-sterile. The implants are single-use devices and must be sterilized before implantation according to the validated sterilization cycle specified in the labeling. The reusable instruments must be cleaned and steam sterilized prior to first use and after each use.
The reusable instruments are packaged in protective polyethylene self-sealing pouches with product identification labeling. The instrument packaging prominently identifies the non-sterile status and includes instructions to clean and sterilize before use. The packaging system is intended only to protect the instruments during transportation, storage, and handling and is not a sterile barrier system.
{10}
K253155 - Page 7 of 8
A product adoption rationale was provided for the cleaning validation of the reusable surgical instruments. The assessment compares the subject instruments with predicate instruments previously reviewed under K252730 and concludes that the subject instruments do not represent a new or worse cleaning challenge compared with the predicate representative instruments. The comparison addresses materials, manufacturing processes, sterilization parameters, hollow structures, assembled components, crevices, material complexity, surface area, and overall cleanability.
#### **10. Clinical Testing**
Clinical testing was not required to support substantial equivalence. The determination of substantial equivalence is supported by comparison to legally marketed predicate devices, technological comparison, material evaluation, biocompatibility evaluation, sterilization and reprocessing information, packaging information, and nonclinical performance testing.
#### **11. Predetermined Change Control Plan**
A Predetermined Change Control Plan is not included in this 510(k). Therefore, no PCCP information is applicable to this 510(k) Summary.
#### **12. Substantial Equivalence Conclusion**
The Fule Metallic Lockable Intramedullary Nail Systems and the identified predicate devices have the same intended therapeutic and surgical use, similar indications for use, similar anatomical applications, similar technological characteristics, and similar operating principles.
Any differences between the subject and predicate devices, including dimensional ranges, component geometry, sub-system configurations, and titanium alloy grade, do not introduce a new intended use, new indication, new anatomical site, new patient population, new surgical approach, or new mode of action. The differences were evaluated through material assessment, dimensional comparison, biocompatibility evaluation, sterilization and reprocessing information, packaging information, and nonclinical performance testing.
Performance testing demonstrated that the subject devices have acceptable mechanical performance for their intended use. Biocompatibility and material
{11}
K253155 - Page 8 of 8
information support the suitability of the subject device materials for their intended patient-contacting use. Cleaning, sterilization, and packaging information support the labeled non-sterile supply configuration and required processing before use.
Therefore, the Fule Metallic Lockable Intramedullary Nail Systems are substantially equivalent to the identified predicate device, and do not raise different questions of safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.