Hyperdrive 35 Guidewire

K253146 · Rapid-Medical , Ltd. · DQX · Sep 18, 2026 · Cardiovascular

Device Facts

Record IDK253146
Device NameHyperdrive 35 Guidewire
ApplicantRapid-Medical , Ltd.
Product CodeDQX · Cardiovascular
Decision DateSep 18, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Hyperdrive 35 Guidewire is intended to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions.

Device Story

Hyperdrive 35 Guidewire is a 0.035" diameter steerable guidewire; used to navigate peripheral vasculature during diagnostic/interventional procedures. Device features stainless steel hypotube, Nitinol braided coil, and handle assembly. Physician operates handle to rotate inner core wire, enabling distal tip deflection (straight/curved). Hydrophilic coating reduces friction during navigation. Fluoro-safe markers assist fluoroscopy-guided insertion. Supplied sterile for single use. Output is physical guidance of catheters to target sites; aids clinical decision-making by facilitating access to challenging anatomy. Benefits include improved steerability and navigation in peripheral vessels.

Clinical Evidence

No clinical studies conducted. Evidence consists of bench testing (visual/dimensional, tip flexibility/deflection, simulated use, torqueability, kink resistance, fracture/flexing, tensile strength, durability, lubricity, radiopacity, corrosion) and GLP animal study (porcine model, n=not specified, primary endpoint: thromboresistance/thrombus formation). Biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, hemocompatibility, pyrogenicity, systemic toxicity).

Technological Characteristics

0.035" diameter; stainless steel hypotube; Nitinol braided coil; Nitinol core wire; hydrophilic coating. ETO sterilization (ISO 11135). Manual handle-actuated tip deflection. Corrosion testing per ISO 11070:2014(E). Biocompatibility per ISO 10993.

Indications for Use

Indicated for directing catheters to desired anatomical locations in the peripheral vasculature during diagnostic or interventional procedures. Not for neurovascular or coronary interventions.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} **FDA** U.S. FOOD & DRUG ADMINISTRATION September 18, 2026 Rapid-Medical, Ltd. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Dr. Suite #510k Saint Paul, Minnesota 55114 Re: K253146 Trade/Device Name: Hyperdrive 35 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: July 1, 2026 Received: July 1, 2026 Dear Prithul Bom: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K253146 - Prithul Bom Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K253146 - Prithul Bom Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, LYDIA S. GLAW -S Digitally signed by LYDIA S. GLAW -S Date: 2026.09.18 17:06:59 -04'00' Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K253146 Device Name Hyperdrive 35 Guidewire Indications for Use (Describe) The Hyperdrive 35 Guidewire is intended to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} Rapid Medical # **510(k) Summary K253146** # **Submission Sponsor** Rapid Medical Ltd. Carmel Building, P.O. Box 337 Yokneam, 2069205 Israel Company Phone No.: +972-72-250-3331 # **Contacts:** Ina Gutman, RA/QA Director Email: ina.gutman@rapid-medical.com Ronen Eckhouse, CEO Email: ronen@rapid-medical.com # **Date Prepared** August 19, 2026 # **Device Identification** **Trade/Proprietary Name:** Hyperdrive 35 Guidewire **Common/Usual Name:** Guidewire **Classification** **Name:** Guide, Wire, Catheter **Regulation Number:** 21 CFR 870.1330 **Product Codes:** DQX **Device Class:** II **Classification Panel:** Cardiovascular # **Legally Marketed Predicate Device** Predicate Device: Radifocus Glidewire (K152740) # **Reference Device** Reference Device: Drivewire 24 Guidewire (K233791) # **Indications for Use Statement** The Hyperdrive 35 Guidewire is intended to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions Page 1 of 9 {6} **Rapid Medical** ## Device Description The Hyperdrive 35 Guidewire is a 0.035” diameter steerable guidewire with a deflectable tip to aid in accessing vasculature. The guidewire is supplied sterile (ETO sterilization) and is for single use only. The Hyperdrive 35 Guidewire is comprised of a stainless steel hypotube that is cut along its length to provide flexibility and tip deflection ability through control of the handle, an inner Nitinol braided flexible coil, an inner core wire, and a handle. The inner core wire runs inside the hypotube from the distal end to the handle. The distal end of the inner core wire is flattened, looped around and joined to the tip of the distal section of the hypotube, forming a deflectable tip. The hypotube is marked with fluoro-safe markers to provide visual clues to the user to initiate fluoroscopy guided insertion. In order to actuate the tip deflection in two directions, curved or straight, the Hyperdrive 35 guidewire handle contains a tube assembly section. The handle is assembled to the proximal end of the core wire and controls the movement of the distal tip by rotating it counterclockwise. This rotation activates the inner moveable core wire, allowing the bending of the distal tip in two directions counterclockwise will curve the wire, clockwise will straighten the wire. The handle assembly has landmarks to identify the location where the tip is straight. The Hyperdrive 35 Guidewire has a hydrophilic coating along its distal side in order to reduce the friction of the guidewire while navigating. The Hyperdrive 35 Guidewire is provided with a “Torquer” accessory to facilitate use of the guidewire and is not intended to have patient contact. ## Comparison of Technological Characteristics The table below compares the indications for use, principles of operation, technological characteristics, and materials of the Hyperdrive 35 Guidewire with that of the predicate device (the Radifocus Glidewire) and the reference device (Drivewire 24 Guidewire). The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. Differences in the subject device compared to the predicate device are the subject device (and the reference device) have a handle that controls tip deflection, and differences in dimensions and materials. The differences in technological characteristics of the Hyperdrive 35 Guidewire do not raise any new questions of safety or effectiveness. Page 2 of 9 {7} Rapid Medical | | Hyperdrive 35 Guidewire (Subject device) | Radifocus Glidewire (Predicate device) | Drivewire 24 Guidewire (Reference device) | | --- | --- | --- | --- | | 510(k) Number | TBD | K152740 | K233791 | | Regulation | 21 CFR 870.1330 | 21 CFR 870.1330 | 21 CFR 870.1330 | | Indications for Use | The Hyperdrive 35 Guidewire is intended to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions. | The Glidewire is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions | The Drivewire 24 Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in the coronary arteries. | | Anatomical Location | General intravascular use, including peripheral vasculature but not neuro and coronary arteries | General intravascular use, including peripheral vasculature but not neuro and coronary arteries | General intravascular use, including the neuro and peripheral vasculature but not coronary arteries | | Overall Length | 150, 180cm | 260, 300, 350, 400, and 450cm | 204 cm | | Proximal Diameter | 0.035" | 0.035" | 0.024" | | Distal Diameter | 0.035" | 0.035" | 0.024" | | Core Wire Material | Nitinol | Nitinol | Nitinol | | Proximal section Material | Stainless Steel | Polyurethane containing Tungsten | Stainless Steel | | Coil Length | 20 cm | NA | 40 cm | | Coil Material | Nitinol and Nitinol DFT 30% tantalum | NA | Nitinol and Nitinol DFT 30% tantalum | | Shapeable Tip Length | 1.6cm | NA | 1.6cm | | Coating | Hydrophilic coating | Hydrophilic polymer | Hydrophilic coating | | Sterilization | Sterile | Sterile | Sterile | Page 3 of 9 {8} **Rapid Medical** | Sterilization Method | Ethylene oxide | Ethylene oxide | Ethylene oxide | | --- | --- | --- | --- | | Single Use | Yes | Yes | Yes | | Packaging | Placed into a Dispenser hoop, Tyvek pouch, and Carton box | Placed into a Dispenser hoop, Tyvek pouch, and Carton box | Placed into a Dispenser hoop, Tyvek pouch, and Carton box | Page 4 of 9 {9} **Rapid Medical** ## Non-Clinical Testing In order to demonstrate the safety and performance of the subject device and to support substantial equivalence to the predicate device, Rapid Medical Ltd. completed a number of non-clinical performance tests. ### Bench Tests The subject device passed all non-clinical performance bench testing in accordance with internal requirements, national standards and international standards as shown in the table below to support substantial equivalence of the device. | Performance Bench Testing | | | | --- | --- | --- | | Test | Test Method Summary | Results | | Visual and Dimensional Verification | The overall guidewire length, diameter, bend diameter, and coating length were measured. A visual inspection confirmed there was no damage to the device. | Dimensional and visual inspection results meet acceptance criteria. | | Tip Flexibility | The force required to flex the distal tip was measured using a loading cell. | Tip flexibility was shown to be comparable to the predicate device. | | Tip Deflection Force | The force that the guidewire tip applies on the vessel wall was measured using a loading cell. | The force applied by the tip is equal to or less than the predicate device's. | | Simulated Use - Delivery and Retrieval Force | Forces for the delivery and retrieval of the guidewire through a simulated use vascular model were measured. | The delivery and retrieval forces were less than the predicate device's. | | Simulated Use - Performance and Compatibility | Guidewire performance and compatibility were assessed in a simulated use model. | All devices met acceptance criteria in demonstrating the intended use of the guidewire in challenging simulated use conditions. | Page 5 of 9 {10} **Rapid Medical** | Performance Bench Testing | | | | --- | --- | --- | | Test | Test Method Summary | Results | | Usability Evaluation | Physicians evaluated the guidewire in clinically relevant simulated use models. | The device met the acceptance criteria. | | Torqueability | The ability of the guidewire to translate torque from the proximal end to the distal end was measured. | All devices met acceptance criteria in having the required initial rotation to the point when the tip starts to respond be equivalent to or less than the predicate device. | | Kink Resistance | Kink resistance was tested along the guidewire by wrapping the device around mandrels of smaller radii until failure. | All devices met the acceptance criteria for kink resistance which represent clinical use scenarios. | | Fracture Test | The guidewire was wound around a cylinder and examined for fractures or other damage. | All devices met the acceptance criteria of having no fractures, loosening, or failures. | | Flexing Test | The distal portion of the guidewire was subjected to repeated bending and examined for damage. | All devices met the acceptance criteria of having no fractures, loosening, or failures. | | Tensile Force | The tensile strength of the guidewire joints was measured. | All joints met peak tensile force acceptance criteria established by delivery and retrieval force testing. | | Tip Mechanism Durability | In a simulated use model, the tip deflection mechanism was cycled to assess durability. | All devices met the acceptance criteria of withstanding a predetermined number of handle actuations. | | Torque Strength | With the distal tip constrained, the number of turns to failure was measured. | All devices met the acceptance criteria of having the number of turns to failure be no less than a predetermined value. | | Torquer Performance | The torque device was tested for performance (functional, tensile and torque) with the guidewire. | All devices met the acceptance criteria of having no damage to the handle or shaft, and equivalent or higher tensile force and measured torque force at slipping to predicate the device. | | Particulate | Particulates were measured under simulated use conditions. | All devices have particulates equivalent to or less than the predicate device. | Page 6 of 9 {11} Rapid Medical | Performance Bench Testing | | | | --- | --- | --- | | Test | Test Method Summary | Results | | Lubricity | A pinch test using a pushability test system was performed on the device through cycles to measure the friction properties of the hydrophilic coating | All devices met the acceptance criteria of having a maximal force and an average force comparable to the predicate device. | | Coating integrity | Visual inspection of the hydrophilic coating following simulated use was performed. | All devices met the acceptance criteria of having no coating separation after simulated use testing. | | Radiopacity | Images were evaluated during animal study. | All devices were visible under fluoroscopy. | | Corrosion | The device was tested according to Annex B of ISO 11070: 2014(E). | The test sample did not have evidence of corrosion. | ### Biocompatibility Biocompatibility assessment of the reference device supports the biocompatibility of the Hyperdrive 35 Guidewire Device in accordance with ISO–10993 and the testing on the reference device consisted of the following tests: | Test | Test Description | Results | | --- | --- | --- | | Cytotoxicity | Cytotoxicity study using ISO Elution Method | The test article extract showed no evidence of causing cell lysis or toxicity. The test met the requirements as the grade was less than a grade 2 (mild reactivity). | | Irritation (Intracutaneous Reactivity) | ISO Intracutaneous Irritation Study using extracts | The test article met the requirements, with no differences between each test article extract’s overall mean score and the corresponding control extract’s mean score. | | Sensitization | ISO Guinea Pig Maximization Sensitization Test | Test articles showed no evidence of causing delayed dermal contact sensitization and were not considered sensitizers. | | Hemocompatibility – Hemolysis | ASTM Hemolysis Study – Extract and Direct Contact | The test articles, both in direct and indirect contact with blood, were non-hemolytic. | | Hemocompatibility – Complement Activation | SC5b-9 Complement Activation Assay | The test article was not considered a potential activator of the complement system. | Page 7 of 9 {12} Rapid Medical | Pyrogenicity | USP Rabbit Pyrogen Study, Material Mediated | The total rise of rabbit temperatures during the observation period was within acceptable USP limits. The test article was considered nonpyrogenic. | | --- | --- | --- | | Acute Systemic Toxicity | ISO Acute Systemic Toxicity in Mice | There was no mortality or evidence of systemic toxicity from extracts injected into mice. | | In Vivo Thrombogenicity | In vivo thrombogenicity in a porcine model | There was no evidence of thrombus formation. | The biocompatibility assessment confirmed that the Hyperdrive 35 Guidewire met biological safety requirements per the ISO 10993 standard. ### Animal Study Rapid Medical Ltd. completed an acute, Good Laboratory Practices (GLP) study in domestic swine to support the substantial equivalence of the Hyperdrive 35 Guidewire to the predicate device. Devices were navigated to the left and right renal arteries and allowed to dwell for 10 minutes, removed, examined for thrombus, scored, and photographed; this procedure was repeated three times for a total dwell time of 30 minutes. Activated clotting times were maintained within clinically relevant limits (250-350 seconds) throughout the procedure. After termination, the renal arteries, medulla, and capsule were examined. There was no evidence of thrombus on devices, in renal arteries or in downstream organs (kidneys). Therefore, the study results support the thromboresistance of the Hyperdrive 35 Guidewire as substantially equivalent to that of the Radifocus Glidewire. ### Sterilization and Shelf Life The device and its accessories underwent sterilization validation studies in accordance with ISO 11135: 2014/AC: 2014 (“Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices”) requirements to demonstrate that the Ethylene Oxide (EtO) sterilization process will reliably sterilize the product loads to a sterility assurance level (SAL) of 10⁻⁶ or less, according to the overkill half cycle approach and relevant standards. The Hyperdrive 35 Guidewire device and package met all shelf life and package integrity test acceptance criteria, supporting a shelf-life of one (1) year. ### Clinical Testing No clinical studies were conducted, as substantial equivalence of the Hyperdrive 35 Guidewire to the predicate device is established through the non-clinical bench, biocompatibility, and animal testing. Page 8 of 9 {13} Rapid Medical ### Conclusion of Substantial Equivalence The Hyperdrive 35 Guidewire has the same intended use and similar indications for use and technological characteristics compared to the predicate Radifocus Glidewire. The differences do not raise new or different questions regarding the safety and effectiveness of the device. The non-clinical bench, biocompatibility, shelf-life, and animal testing with passing results discussed above further support the substantial equivalence of the Hyperdrive 35 Guidewire to the predicate device. Page 9 of 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...