← Product Code [DRG](/productcode/DRG) · K252984

# Perin Health System (PHD80060-2) (K252984)

_Perin Health Devices, LLC · DRG · Jan 22, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252984

## Device Facts

- **Applicant:** Perin Health Devices, LLC
- **Product Code:** [DRG](/productcode/DRG.md)
- **Decision Date:** Jan 22, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall). The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate with optional third-party devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological data falls outside set limits or manual trigger by the patient. The device is intended to provide physiological information for non-critical, adult population.

## Device Story

Wireless remote monitoring system for adult patients; captures spot-check physiological data via chest-worn patch. Inputs: auscultation sounds, 1-channel ECG, PPG, bioimpedance, skin temperature, accelerometer data. Patch communicates via encrypted BLE to mobile app; app relays data to cloud. Cloud processes signals to derive vital signs (HR, HRV, RR, SpO2, posture, activity). Healthcare providers review data, trends, and audio recordings via web-based portal; portal supports alert configuration, virtual appointments, and EHR integration. System notifies providers of threshold exceedances or patient-triggered events. Benefits: enables retrospective clinical review, tracking of physiological changes, and remote patient engagement in non-critical care settings.

## Clinical Evidence

Clinical validation included: SpO2 accuracy (12 subjects, 70-100% range, Arms 3.3%); Respiratory Rate accuracy (35 subjects, Arms 1.7 bpm vs. end-tidal CO2); ECG/HR/HRV performance (243 participants vs. Holter monitor, showing high agreement); and 360-hour wear-life evaluation (26 participants). All studies confirmed performance across diverse demographic and clinical subgroups.

## Technological Characteristics

Chest-worn patch; dry electrodes; thermistor; contact acoustic sensor; bioimpedance; PPG (530nm, 650nm, 940nm, 475nm); accelerometer. Connectivity: BLE 5.4, NFC pairing. Power: LiMnO2 battery. Dimensions: 80x45x14mm. Software: Class C, enhanced documentation. Standards: ISO 10993, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ANSI/AAMI/IEC 60601-2-27/47, ANSI/CTA-2056-A.

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- UbiqVue 2A Multiparameter System ([K242018](/device/K242018.md))
- Alio Medical Remote Monitoring System ([K211365](/device/K211365.md))

## Reference Devices

- EKO CORE 500 Digital Stethoscope ([K230111](/device/K230111.md))
- Biobeat Platform-2 and BB-613WP Patch ([K222010](/device/K222010.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 22, 2026

Perin Health Devices, LLC
% Chandler Thames
Quality/Regulatory Consultant
Rook Quality Systems
1155 Mount Vernon Hwy
Suite 800
Dunwoody, Georgia 30338

Re: K252984
Trade/Device Name: Perin Health System (PHD80060-2)
Regulation Number: 21 CFR 870.2910
Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver
Regulatory Class: Class II
Product Code: DRG, DQD, MHX, DQA, DXN, FLL, BZQ
Dated: December 19, 2025
Received: December 19, 2025

Dear Chandler Thames:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252984 - Chandler Thames
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252984 - Chandler Thames
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

JENNIFER W. SHIH -S

Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

# Indications for Use

Submission Number (if known)
K252984

Device Name
Perin Health System (PHD80060-2)

Indications for Use (Describe)
The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall).

The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate with optional third-party devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological data falls outside set limits or manual trigger by the patient.

The device is intended to provide physiological information for non-critical, adult population.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”

{4}

Perin Health
K252984 510(k) Summary

# 510(k) Summary for Perin Health System

Submitter
Perin Health Devices, LLC

Address
755 Research Parkway
Suite 530
Oklahoma City, OK 73104

Contact
Ian McLane, Chief Technology Officer
Phone: 818-697-8551
Email: ian.mclane@perinhealth.com

Date Prepared:
January 22, 2026

Device
Perin Health System

Trade Name:
Perin Health System

Common Device Name:
Transmitters and Receivers, Physiological Signal, Radio Frequency

|  Classification Product Code  |   |   |   |
| --- | --- | --- | --- |
|  Regulation Medical Specialty | Regulation Classification | Product Code | Description  |
|  Cardiovascular | 21 CFR 870.2910 Class II | DRG | Radiofrequency physiological signal transmitter and receiver  |
|  Subsequent Product Code  |   |   |   |
|  Cardiovascular | 21 CFR 870.1025 Class II | MHX | Monitor, physiological, patient (with arrhythmia detection or alarms)  |
|  Cardiovascular | 21 CFR 870.1875 Class II | DQD | Stethoscope, electronic  |
|  Cardiovascular | 21 CFR 870.2700 Class II | DQA | Oximeter  |
|  Cardiovascular | 21 CFR 870.1130 Class II | DXN | System, measurement, blood-pressure, non-invasive  |
|  General Hospital | 21 CFR 880.2910 Class II | FLL | Continuous measurement thermometer  |
|  Anesthesiology | 21 CFR 868.2375 Class II | BZQ | Monitor, breathing frequency  |

Perin Health System
Page 1 of 22

{5}

Perin Health
K252984 510(k) Summary

Primary Predicate Device UbiqVue 2A Multiparameter System
510(k) Number K242018

Secondary Predicate Device Alio Medical Remote Monitoring System
510(k) Number K211365
Predicate for Auscultation Sound Data

Reference Device EKO CORE 500 Digital Stethoscope
510(k) Number K230111
Reference for Auscultation Sounds of the Heart and Lungs

Reference Device Biobeat Platform-2 and BB-613WP Patch
510(k) Number K222010
Reference for Spot-Checking

## Product Description

The Perin Health System is a wireless remote patient monitoring platform that enables healthcare professionals to perform spot-checking and retrospective monitoring of physiological data from adult patients. The Perin Health System is designed for use in hospitals, clinics, long-term care facilities, physician offices, and home environments.

The Perin Health System comprises the following components:

1. Perin Health Patch wearable device
2. Perin Health Patient Mobile Application
3. Perin Health Cloud
4. Perin Health Provider Portal
5. Perin Health Inpatient Application

Perin Health System
Page 2 of 22

{6}

Perin Health
K252984 510(k) Summary

1. The Perin Health Patch is a chest-worn wearable device that performs scheduled spot-check measurements of multiple physiological parameters. Unlike continuous monitoring systems, the Perin Health Patch captures measurements at predetermined intervals configured by healthcare providers based on clinical need.

The device integrates six primary sensing modalities:

- Auscultation (heart and lung sounds)
- Electrocardiography (1-channel ECG)
- Pulse oximetry via photoplethysmography (PPG)
- Bioimpedance (BioZ) for respiratory monitoring
- Temperature sensing (skin)
- Motion and orientation detection via accelerometer

The combination of these modalities in a small, low-power wearable form allows for the spot-checking of primary vital signs:

- Heart rate and R-R intervals
- Heart rate variability (HRV) parameters
- ECG waveform data
- Auscultation sound data (heart and lung sounds)
- Respiratory rate
- Pulse (PPG) waveform
- Oxygen saturation (SpO2%)
- Skin temperature
- Fall detection events
- Body posture
- Activity level and step count

The device adheres to the patient's upper left chest at the second intercostal space with a medical-grade long-term wear adhesive. The adhesive is placed on the patient-

Perin Health System
Page 3 of 22

{7}

Perin Health
K252984 510(k) Summary

facing side of the wearable, with cutouts for the sensors to make direct contact with the skin. The wearable device is lightweight and semi-flexible, allowing for the device to conform to the natural curvature of the chest. It is water resistant, allowing for bathing and normal activities while the patient is wearing the system.

The wearable communicates to the receiving unit (mobile phone) via an encrypted Bluetooth Low Energy connection. Measurements, all notifications and control commands, and software updates are transmitted over the BLE connection. The wearable uses Near Field Communication (NFC) to facilitate the Bluetooth pairing process with the mobile phone by simply having to tap their phone to the device to initiate a Bluetooth connection. The wearable device also contains on-board memory that can store over two weeks of spot-check data. When measurements are taken and no receiving unit is present, the wearable can store recordings in the onboard memory. Recordings are stored in a stack, such that at the next connection possibility between the wearable and the receiving unit, the most recent data will be transmitted first followed by other measurements in reverse chronological order.

Other key features of the wearable include:

- Customizable recording schedule set by the healthcare provider in their care program
- Replaceable battery
- Patient-triggered recordings via double-tap
- Signal quality indicators for measurement validation and identification of noisy measurements

2. The Patient Mobile Application, available on iOS or Android platforms, is intended exclusively for use in home environments by patients under healthcare provider supervision. The application serves as a data relay and display interface, allowing the patient to complete key tasks, including onboarding, device setup, device communication, and patient-reported data.

The application serves as the primary interface between the Perin Health Patch and the cloud infrastructure, receiving spot-check measurements from the device and uploading them for provider review. The application establishes and manages secured BLE communication with the Health Patch. Given that the Health Patch operates on provider-configured recording schedules, the application manages data transfer in the background with minimal patient interaction required. When internet connectivity is

Perin Health System
Page 4 of 22

{8}

Perin Health
K252984 510(k) Summary

unavailable, the application stores measurements locally until transmission becomes possible. The system also manages firmware updates for the Perin Health Patch.

The application integrates with FDA-cleared third-party blood pressure cuff and scale using BLE and transfers the data to the Cloud System. Healthcare providers determine which patients require the additional third-party device monitoring as part of their individualized care programs. The system also allows users to optionally enter manual data for blood pressure and weight if no third-party device is connected.

Patients are able to review their historical measurement data taken throughout their monitoring program and their goals and thresholds set by their providers. The patient can view metrics assigned within their care program:

- Heart Rate and Heart Rate Variability
- Respiratory Rate
- Oxygen Saturation
- Step Count
- Temperature
- Blood Pressure
- Weight

Patients can also select audio segments captured by the device for playback (no visualization).

The application provides comprehensive patient engagement features. Patients can complete customized questionnaires with up to 20 questions in various formats, review educational content delivered through their care programs, and submit non-critical medical reports to their care team. The reporting feature includes anatomical body mapping for location-specific symptoms, severity scaling, and photo attachment capabilities. The application supports secure messaging with care providers, virtual appointment attendance with waiting room functionality, and comprehensive offline operation with automatic synchronization upon connectivity restoration.

3. The Perin Health Cloud infrastructure serves as the central hub for data management and processing. The cloud system receives encrypted spot-check data from relay systems and manages raw data processing (for Health Patch data only), storage, and retrieval of physiological measurements for retrospective clinical review.

Perin Health System
Page 5 of 22

{9}

Perin Health
K252984 510(k) Summary

Algorithms are run in the cloud to process measurements from the Health Patch and generate Signal Quality Index, Heart Rate, Heart Rate Variability, Respiratory Rate, Oxygen Saturation, and Posture.

The alert and notification system enables healthcare professionals to configure multi-level alerts based on clinical parameters, technical issues, or manual patient triggers. Clinical alerts are based on provider-configured thresholds that are set in during the enrollment of a patient in a care program. The system supports complex notification rules including threshold exceedances, percentage changes, trending patterns, and consecutive violations. Alerts are displayed to providers for the purpose of highlighting data during their retrospective review and are not intended to support real-time patient monitoring or urgent care provider action.

The cloud infrastructure includes comprehensive audit logging of all user actions, data access, and system events. The system provides API access for integration with electronic health records with HL7 v2.x, HL7 FHIR R4, and other standard protocols, enabling bidirectional data exchange with major EHR systems.

4. The web-based Provider Portal enables healthcare professionals to access and manage patient data and alert statuses remotely through any compatible web browser. Through the portal, providers can review spot-check measurements and historical trends, playback audio recordings of auscultation sounds captured by the Patch, configure individualized care programs, set measurement schedules and alert thresholds, and communicate with patients through various modalities.

Through the portal, providers can review spot-check measurements with customizable vital sign charts displaying trends over days, weeks, or months. Advanced visualization includes waveform analysis for ECG and PPG signals, audio playback for auscultation recordings, and comprehensive annotation tools. The portal displays signal quality indicators and out-of-range values with appropriate visual highlighting based on configured thresholds. The portal also displays patient severity levels (Low/Medium/High) based on the NEWS2 scoring methodology. Additional clinical measures, such blood pressure and weight, that are manually input into the EHR can be read into the Perin Health System and viewed in the Provider Portal using the EHR interface.

The system employs a structured care program architecture that ensures appropriate clinical oversight throughout the monitoring process. Healthcare organizations create standardized care program templates for common conditions. Individual providers can then select from these approved templates and customize them for specific patient

Perin Health System
Page 6 of 22

{10}

Perin Health
K252984 510(k) Summary

needs, prescribing the specific devices needed, measurement frequencies appropriate to the condition, and recording schedules tailored to clinical requirements.

The portal includes comprehensive communication capabilities supporting both patient and care team interactions. Providers can conduct virtual appointments with integrated video calling, AI-powered real-time transcription using AWS HealthScribe, and automated clinical note generation structured into standard sections. The messaging system supports secure text communication with file attachments, while the task management system enables care coordination across team members. Providers can create and deploy customized questionnaires with various response types and scoring algorithms, manage educational content delivery, and review patient-submitted reports with collaborative response capabilities.

Additional portal features include appointment scheduling with EHR integration, comprehensive alert management with acknowledgment workflows, administrative functions for user and device management, and organization hierarchy configuration. The portal provides detailed audit trails, performance analytics, and compliance reporting to support quality improvement initiatives.

5. The Perin Health Inpatient Module provides a monitoring dashboard for monitoring capabilities in healthcare facility environments through. The modules leverage the existing architectures for the Mobile Application and Provider Portal but offer unique interfaces for inpatient spot-check measurements.

The web-based monitoring dashboard, a page accessible through the Provider Portal, displays vital signs for up to 50 concurrent patients in a grid layout. Each patient card shows the latest values for heart rate, respiratory rate, oxygen saturation, temperature, and device status, with automatic sorting by alert priority and visual indicators for threshold violations. The dashboard refreshes every second, updating as new spot-check recordings are captured from patients across the unit.

The bedside Inpatient Application is built on top of the Android architecture of the Patient Mobile app and operates in kiosk mode. The Beside app only interfaces with the Perin Health Patch and relays information to the Cloud to provide clinicians with access to recent measurements in the Provider Portal Inpatient view. The application also maintains local data storage for backup operation and automatically synchronizes with the cloud upon connectivity restoration. Providers are unable to manually input clinical data (e.g., blood pressure measurements) directly into the bedside Inpatient Application but manual data input into the EHR can be read into and visualized in the Provider Portal over the EHR interface.

Perin Health System
Page 7 of 22

{11}

Perin Health
K252984 510(k) Summary

The Perin Health System supports monitoring in hospitals and out-of-hospital patient care settings where care is administered by healthcare professionals. Visual alarm indicators highlight parameter exceedances according to configured thresholds. High-priority alerts display prominently with appropriate color coding, though all clinical responses and acknowledgments must be performed through the Provider Portal to maintain proper documentation and workflow management.

The Perin Health System facilitates comprehensive spot-checking and retrospective monitoring across the continuum of care. Data flows from the wearable patch and third-party devices through the patient mobile application to the central cloud infrastructure, where processing algorithms derive clinical insights. Healthcare providers access this information through the web portal or inpatient displays for clinical review and analysis, enabling healthcare providers to track patient progress, adjust treatment plans based on measurements, and identify patients requiring intervention based on retrospective data trends.

## Indications for Use

The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall).

The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate with optional third-party devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological data falls outside set limits or manual trigger by the patient.

The device is intended to provide physiological information for non-critical, adult population.

Perin Health System
Page 8 of 22

{12}

Perin Health
K252984 510(k) Summary

Substantial Equivalence comparison (Subject Device and Predicate Devices)

|   | Perin Health System (Subject Device) | UbiqVue (Primary Predicate) – Used for ECG, Skin Temperature, Respiratory Rate, Oxygen Saturation | Alio (Secondary Predicate) – Used for Auscultation Sound Data  |
| --- | --- | --- | --- |
|  510k | K252984 | K242018 | K211365  |
|  Product code & regulation | DRG
DQD, DQA, FLL, BZQ, DXN, MHX
870.2910 | DRG
DQA, FLL, MHX
870.2910 | DRG
DQD
870.2910  |
|  Intended Use | The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall).

The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate with optional third-party | The UbiqVue 2A Multiparameter System is a wireless remote patient monitoring system intended for use by healthcare professionals for continuous collection of physiological data at home and in healthcare settings. This shall include electrocardiography (2-Channel ECG), heart Rate, SpO2%, respiration rate, pulse Rate, photoplethysmography waveform, skin temperature, body temperature, body posture, body motion, R-R Interval, heart Rate variability (HRV) and blood Pressure (optional).

Data is transmitted wirelessly near real time from UbiqVue 2A Wearable Biosensor and 3rd party device (for Blood Pressure only) to remote central server for display, storage, and analysis. The UbiqVue 2A Multiparameter System is intended for non-critical, adult population.

The UbiqVue 2A Multiparameter System shall include the ability to notify healthcare professionals through alerts when physiological parameters fall outside the set limits, manual trigger by the patient and to display multiple patient's physiological data for remote monitoring at home and | The Alio Medical Remote Monitoring System is a wireless remote monitoring system intended for use by healthcare professionals to intermittently collect physiological data in home use settings. The data includes skin temperature, auscultation sound data and heart rate. Data is transmitted wirelessly from the SmartPatch wearable sensor to a web-based portal for the healthcare provider's (HCP) review. The Alio Medical RMS is intended for use on general care patients who are 18 years of age or older. The SmartPatch sensor is indicated to measure skin temperature and pulse rate where clinically indicated. The SmartPatch sensor is indicated to record and transmit auscultation sound data where clinically indicated. The device is not intended for use in critical care or other high-acuity  |

Perin Health System
Page 9 of 22

{13}

Perin Health
K252984 510(k) Summary

|   | Perin Health System (Subject Device) | UbiqVue (Primary Predicate) – Used for ECG, Skin Temperature, Respiratory Rate, Oxygen Saturation | Alio (Secondary Predicate) – Used for Auscultation Sound Data  |
| --- | --- | --- | --- |
|   | devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological data falls outside set limits or manual trigger by the patient.

The device is intended to provide physiological information for non-critical, adult population. | with visual alarm for active monitoring at hospitals and out-of-hospital patient care settings (such as clinics, outpatient surgery facilities, long-term care facilities and physician offices) in which care is administered by healthcare professionals. | environments. The Alio Medical RMS is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring practices.  |
|  Intended Population | Non-critical, adult population, 22 years or older | Non-critical, adult population, 18 years or older | Non-critical, adult population, 18 years or older  |
|  Intended Use environment | Home & Healthcare settings | Home & Healthcare settings | Home setting only  |
|  Rx or OTC | Rx | Rx | Rx  |
|  Spot – Checking or Continuous | Spot-checking | Continuous | Intermittent  |
|  ECG | 1-channel | 2-channel | N/A  |
|  HR | 20-200 bpm ± 3 BPM or 5%, whichever is greater | 30-250BPM +- 3 BPM or 5% whichever is greater | 30-200 BPM  |
|  RR Range and Accuracy | 5-30 Breaths per Minutes ± 1 Breaths per Minute (Accuracy Root Mean | 5-60 Breaths per Minute | N/A  |

Perin Health System
Page 10 of 22

{14}

Perin Health
K252984 510(k) Summary

|   | Perin Health System (Subject Device) | UbiqVue (Primary Predicate) – Used for ECG, Skin Temperature, Respiratory Rate, Oxygen Saturation | Alio (Secondary Predicate) – Used for Auscultation Sound Data  |
| --- | --- | --- | --- |
|   | Square (Arms) based on bench testing)
± 3 Breaths per Minute (Accuracy Root Mean Square (Arms) based on clinical study)

Derived from Trans-thoracic Impedance (TTI) and ECG Derived Respiration (EDR) based on RS Amplitude | ≤ 1 Breath per minute (Mean Absolute Error based on Simulation study)
≤ 3 Breaths per minute (Mean Absolute Error based on Clinical study)

Trans-thoracic Impedance (TTI), ECG Derived Respiration (EDR) based on RS Amplitude & Respiratory Sinus arrhythmia (RSA) and Accelerometer data |   |
|  Skin Temperature | 15 C - 50°C
± 0.3°C
Resolution: 0.008°C
Time response: 30 minutes
Measurement mode: Direct
ISO 80601-2-56 | 15C – 43C//59F-109.4F
As per ASTM E1112-00 | 15 C - 50°C  |
|  Posture | Prone, supine, left lateral recumbent, right lateral recumbent, Fowler’s, Trendelenburg, upright, leaning forward (> 80% average sensitivity and specificity, compared to visual) | Supine, Prone, Lying- Left/Right, Reclining, Reclining-Left/Right, Leaning Forward, Upright, Upside down | N/A  |
|  SpO2% | 70% - 100%
± 3 %
PPG, Signal Strength Indicator
Chest Measurement site 530 nm (Green), 650 nm (Red), 940 nm (IR), Blue (475 nm) | 0 to 100%
± 3 % (100 to 70%), Less than 70% unspecified
PPG, Signal Strength Indicator
Chest Measurement Site Spectral Sensor | N/A  |
|  R-R interval | Sinus Heart Rate | Beat Classification
Sinus Heart Rate | N/A  |
|  Heart Rate Variability (HRV) | SDNN (short), RMSSD, TINN, pNN50, VLF Band, LF Band, HF Band | SDNN, SDANN, ASDNN, RMSSD, VLF Band, LF Band, HF Band | N/A  |
|  Body Motion | Active or sedentary (> 90% sensitivity and specificity) | No Body Motion, Mild Body Motion, Moderate Body Motion, | N/A  |

Perin Health System
Page 11 of 22

{15}

Perin Health
K252984 510(k) Summary

|   | Perin Health System (Subject Device) | UbiqVue (Primary Predicate) – Used for ECG, Skin Temperature, Respiratory Rate, Oxygen Saturation | Alio (Secondary Predicate) – Used for Auscultation Sound Data  |
| --- | --- | --- | --- |
|   |  | Severe Body Motion |   |
|  Fall Detection | Fall or no fall ( > 80% sensitivity and specificity) | N/A | N/A  |
|  Step Count | < 5% Absolute Error Compared to Manual Count for speeds of at least 2 miles per hour Compliance: ANSI/CTA-2056-A | N/A | N/A  |
|  Emitted light peak wavelength | 530 nm (Green), 650 nm (Red), 940 nm (IR), Blue (475 nm) | N/A | 530 nm (Green), 650 nm (Red), 940 nm (IR)  |
|  Sound Data | Heart Lungs | N/A | Heart Lungs
Bowel
Arteries
Veins  |
|  Sound Amplification | Yes | N/A | Yes  |
|  Record and Playback Sound | Yes (Playback via Provider Portal and Mobile Application) | N/A | Yes (Playback via Clinician Portal)  |
|  Sensor Type | Thermistor
Electrocardiogram (ECG)
Contact Acoustic sensor
Bioimpedance
PPG
Accelerometer | Thermistor
Electrocardiogram (ECG)
-
Bioimpedance
PPG
Accelerometer | Thermistor
-
Microphone
-
PPG
Accelerometer  |
|  Single Use | Multiple Use by Single User | Yes | Yes  |
|  Wear Life | 360 hours | 120 hours | 120 Hours  |
|  Dimension | 80 mm x 45 mm x 14 mm | 116mm x 91mm x17 mm | 150mm x 94mm x 36mm  |
|  Weight | 19 g | 35 g | Unknown  |
|  ECG Electrode distance | Single lead
60mm electrode distance | 90.0 mm
73.0 mm | N/A  |
|  Location on body | Upper chest, below the left collarbone | Upper Center Chest or Upper Left Chest | Unknown  |
|  Data can be transferred and temporarily stored | Yes | Yes | Unknown  |
|  Battery | LiMnO2 (Primary) | LiMnO2 (Primary) | Li-ion  |

Perin Health System
Page 12 of 22

{16}

Perin Health
K252984 510(k) Summary

|   | Perin Health System (Subject Device) | UbiqVue (Primary Predicate) – Used for ECG, Skin Temperature, Respiratory Rate, Oxygen Saturation | Alio (Secondary Predicate) – Used for Auscultation Sound Data  |
| --- | --- | --- | --- |
|  Applied part category | BF | CF | Unknown  |
|  Defibrillation proof | Yes | Yes | Unknown  |
|  HF Surgical equipment compatibility | No | Yes | Unknown  |
|  Communication protocol | BLE 5.4 | WLAN (802.11b), BLE 5.2 | BLE  |
|  Wireless Radio Frequency | 2.4 - 2.5 GHz | 2.4 - 2.4835 GHz | 2.4 - 2.5 GHz  |
|  Communication Security | AES-CTR 256 | WPA2-PSK /CCMP AES-CCM 128 | Unknown  |
|  Authentication from server | Yes | Yes | Unknown  |
|  Data transferred to server | Yes | Yes | Unknown  |
|  Data buffered if there is no connection with server | Yes | Yes | Unknown  |
|  GUI for patient admission (only for single patient relay) | Yes | Yes | Unknown  |
|  GUI for PPG signal quality (only for single patient relay) | Yes | Yes | Unknown  |
|  Secure Server | Data is stored for access by the Perin Health Mobile Application and Provider Portal. | Data is stored for access by Active Monitoring Portal & any 3rd party server (through API) | Unknown  |
|  Programmable Alert Notification & Configurable | Yes (Clinical, Technical, Manual) | Yes (Clinical, Technical, Manual) | Unknown  |
|  Early Warning Score (NEWS2) | Yes | Yes | Unknown  |
|  Multi-patient tile view | Yes | Yes | Unknown  |

Perin Health System
Page 13 of 22

{17}

Perin Health
K252984 510(k) Summary

|   | Perin Health System (Subject Device) | UbiqVue (Primary Predicate) – Used for ECG, Skin Temperature, Respiratory Rate, Oxygen Saturation | Alio (Secondary Predicate) – Used for Auscultation Sound Data  |
| --- | --- | --- | --- |
|  Single patient zoom view | Yes | Yes | Unknown  |
|  Visual alarm indication | Yes | Yes | Unknown  |

# Differences Discussion:

- The Subject device has an additional physiological monitoring capability via 3rd party device integration, weight data. However, the performance of this parameter is verified and validated, and the 3rd party device (Transtek TeleRPM GBS2012-B) is a Class I, 510(k) exempt device. Hence, this additional parameter does not affect the intended use of the Subject device.
- The Subject Device does not claim to measure body temperature as the Primary Predicate Device. This does not impact the indications for use nor substantial equivalence.
- The Subject device is intended to perform spot-checking monitoring compared to continuous monitoring of the Primary Predicate device, as this is a narrower scope of the Primary Predicate Device, this does not impact the indications for use nor substantial equivalence.

# Comparison with Reference Devices

The EKO Core 500 Digital Stethoscope is used as a reference for the auscultation sounds of the heart and lungs performed by the Perin Health System.

|   | Perin Health System (Subject Device) | EKO CORE 500 Digital Stethoscope (Reference Device)  |
| --- | --- | --- |
|  510k | K252984 | K230111  |
|  Product code & regulation | DRG
DQD, DQA, FLL, BZQ
870.2910 | DQD
DPS
870.1875  |
|  Intended Use | The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare | The CORE 500 Digital Stethoscope is intended to be used by clinicians to electronically amplify, The CORE  |

Perin Health System

{18}

Perin Health
K252984 510(k) Summary

|   | Perin Health System
(Subject Device) | EKO CORE 500 Digital Stethoscope
(Reference Device)  |
| --- | --- | --- |
|   | professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall).

The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate with optional third-party devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological | 500 Digital Stethoscope is intended to be used by clinicians to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or used under the care of a clinician). The data offered by the device is only significant when used in conjunction with clinician evaluation as well as consideration of other relevant patient data.  |

Perin Health System
Page 15 of 22

{19}

Perin Health
K252984 510(k) Summary

|   | Perin Health System
(Subject Device) | EKO CORE 500 Digital Stethoscope
(Reference Device)  |
| --- | --- | --- |
|   | data falls outside set limits or manual trigger by the patient.

The device is intended to provide physiological information for non-critical, adult population. |   |
|  Single Use | Multiple Use by Single User | Unknown  |
|  Wear Life | 360 hours | 60 hours  |
|  Intended Population | Non-critical, adult population, 22 years or older | Adults and pediatrics  |
|  Intended Use environment | Home & Healthcare settings | Healthcare  |
|  Rx or OTC | Rx | Rx  |
|  Sensor Type | Contact Acoustic Sensor | Contact Acoustic Sensor  |
|  Sound Amplification | Yes | Yes  |
|  Audio Frequency Range | 40 – 2000 Hz | 20 – 2000 Hz  |
|  Software Interface | Mobile Application
Web Portal | Mobile Application  |

Perin Health System
Page 16 of 22

{20}

Perin Health
K252984 510(k) Summary

The Biobeat Platform-2 and BB-613WP Patch (K222010) is used as a reference for spot-checking performed by the Perin Health System.

|   | Perin Health System (Subject Device) | Biobeat Platform-2 and BB-613WP Patch (Reference Device)  |
| --- | --- | --- |
|  510k | K252984 | K222010  |
|  Product code & regulation | DRG
DQD, DQA, FLL, BZQ, DXN, MHX
870.2910 | DQA
BZQ, DRG, DXG, DXN, FLL
870.2700  |
|  Intended Use | The Perin Health Platform is a wireless remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in healthcare and home settings for long-term monitoring. The Perin Health Patch can monitor auscultation data of heart and lung sounds, photoplethysmography waveforms (PPG), oxygen saturation (%SpO2), heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity detection (including step count), and posture (body position relative to gravity including fall).

The Perin Health System is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. In home-use environments, the Perin Health Platform is able to integrate | The Biobeat Platform-2 is a wireless noninvasive remote monitoring system intended for use by healthcare professionals for spot check collection of physiological data in home and healthcare settings. This can include, functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, blood pressure, respiration rate (RRp), hemodynamic parameters (stroke volume, cardiac output), and body temperature.

The Biobeat Platform-2 tracks changes in blood pressure based on Pulse Wave Transit Time (PWTT) which is obtained utilizing pulse measurements from the integrated SpO2 sensor, following a calibration process using an FDA-cleared oscillometric blood pressure monitor.  |

Perin Health System
Page 17 of 22

{21}

Perin Health
K252984 510(k) Summary

|   | Perin Health System (Subject Device) | Biobeat Platform-2 and BB-613WP Patch (Reference Device)  |
| --- | --- | --- |
|   | with optional third-party devices for blood pressure, and weight data collection via the mobile application. The mobile application transmits data from the Health Patch and third party devices to the cloud and web-based portal for storage, analysis, and review by healthcare professionals. The Perin Health Platform can include the ability to notify healthcare professionals when physiological data falls outside set limits or manual trigger by the patient.

The device is intended to provide physiological information for non-critical, adult population. | The Biobeat Platform-2 is intended for spot-checking and tracking changes of adult patients in hospitals, clinics, long-term care, and at home. The data from the Biobeat Platform-2 are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.  |
|  Intended Population | Non-critical, adult population, 22 years or older | Non-critical, adult population, 18 years or older  |
|  Intended Use environment | Home & Healthcare settings | Home & Healthcare settings  |
|  Rx or OTC | Rx | Rx  |
|  Spot – Checking or Continuous | Spot-checking | Spot-checking  |

## Discussion of technological differences:

- The Subject device has an additional physiological monitoring capability, weight data, via integration with a third-party device (Class I, 510(k) exempt, Transtek TeleRPM GBS2012-B). The weight data integration has been verified and validated to

Perin Health System
Page 18 of 22

{22}

Perin Health
K252984 510(k) Summary

perform as intended. Additionally, weight data is at the discretion of the treating healthcare provider, based on clinical need. Since this parameter is acquired through a low-risk technology, when deemed appropriate by a physician, for supplemental patient information, there is no introduction of additional risk.

- The Subject device is intended to perform spot-checking monitoring compared to continuous monitoring of the Primary Predicate device, as this is a narrower scope of the Primary Predicate Device, this does not add any additional risk.
- The Subject Device utilizes one ECG channel whereas the Primary Predicate Device utilizes two. As this is a narrower scope of the Primary Predicate Device, this does not add any additional risk.
- The Subject Device utilizes a contact acoustic sensor to measure auscultation data of the heart and lungs. The Secondary Predicate Device utilizes a microphone to measure auscultation data. However, the Reference Device for Auscultation Data also utilizes a contact acoustic sensor to perform this measurement, and thus, there is no introduction of additional risk.
- The Subject device also allows playback of auscultation sound data via the Provider Portal and the Mobile Application (patient). In contrast, the Secondary Predicate Device only allows playback via the clinician portal. The auscultation feature of the Perin Health System is solely indicated for recording and playback of body sounds to support clinical assessment, including interpretation and findings, by qualified healthcare professionals. However, patients using the Perin Health System are able to confirm they were able to take a recording via the Mobile Application playback option. Thus, this is a risk mitigation, ensuring patients can verify a successful recording, and it does not add additional risk.
- The predicate device, the UbiqVue Multiparameter 2A, utilizes wet (hydrogel) electrodes for biopotential signal acquisition, whereas the subject device employs dry electrodes. Although the electrode technology differs, the subject device has been demonstrated to perform equivalently to standard wet electrodes. Bench and performance testing, including verification against ANSI/AAMI EC12, confirmed that the dry electrodes meet the same safety and performance requirements as hydrogel electrodes. Accordingly, this difference in electrode technology does not raise new questions of safety or effectiveness.

Perin Health System
Page 19 of 22

{23}

Perin Health
K252984 510(k) Summary

# Summary of Performance Testing

Verification &amp; Validation activities were performed on Perin Health System to demonstrate substantial equivalence to the predicate device:

- Biocompatibility testing of In-vitro cytotoxicity, skin irritation and skin sensitization were conducted on Perin Health System, according to ISO 10993-1: 2018 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process. Tests were performed in accordance with ISO 10993-5:2009, ISO 10993-10:2021, and ISO 10993-23:2021.
- Electrical Safety and electromagnetic compatibility testing were conducted on the Perin Health System for compliance with IEC 60601-1(Ed.3.2), IEC 60601-1-2 (Ed. 4.1) and IEC 60601-1-11(Ed. 2.1).
- The Skin temperature accuracy performance of the Perin Health System was verified by using bench testing, as per ISO 80601-2-56:2017(E).
- ECG, Heart Rate, R-R Interval, and Heart Rate Variability performance testing was performed on the Perin Health System in compliance with ANSI/AAMI/IEC 60601-2-27:2011, ANSI/AAMI/IEC 60601-2-47:2012.
- Auscultation data of heart and lung sounds has been verified by using bench testing in accordance with acceptance criteria.
- Defibrillation proof testing was performed on the Perin Health System for compliance with IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020.
- Body Motion, Fall Detection, and Posture performance of the Perin Health System were verified by using bench testing as per the acceptance criteria.
- Step Count performance of the Perin Health System was verified via bench testing as per ANSI/CTA-2056-A.
- The Alert system &amp; Visual alarm display on the Perin Health System was verified for compliance with IEC 60601-1-8 (Ed. 2.2).
- Usability study was conducted on the Perin Health System for compliance with IEC 60601-1-6 (Ed. 3.2) and “USFDA Guidance- Applying Human Factors and Usability Engineering to Medical Devices.”
- Wireless performance &amp; coexistence testing conducted as per ANSI/IEEE C63.27:2017: American National Standard for Evaluation of Wireless coexistence on the Perin Health System.
- Software in the Perin Health System was designed, documented, verified &amp; validated as per the IEC 62304: Medical device Software - Software Life Cycle Process and USFDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Device. The Software for this device is determined as Class C and Enhanced Documentation Level.

Perin Health System
Page 20 of 22

{24}

Perin Health
K252984 510(k) Summary

- Software in the Perin Health System was securely designed, documented, verified &amp; validated as per the USFDA Guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions.
- Shelf-life testing was conducted and verified on the Perin Health Patch, as per the acceptance criteria.
- Packaging testing conducted on the packaged Perin Health Patch as per ISTA 6.

# Summary of Clinical Testing

Clinical validation activities were performed on Perin Health System to demonstrate substantial equivalence to the predicate device:

- The Perin Health System and Patch were verified &amp; validated for no motion accuracy in compliance with ISO 80601-2-61(Ed. 2.0) including clinical validation to determine the accuracy of SpO2% in induced hypoxia studies in the range of 70% to 100% SpO2 against arterial blood samples analyzed by a laboratory co-oximeter.
- 12 healthy adults (5 female, 7 male), ages 24-42, Monk scale assessment across 5 anatomical sites (1 B, 2 C, 2 D, 3 E, 2 G, 2 H).
- The overall measured Arms in the range of 70 to 100% SpO₂ was 3.3%, with Arms of 3.5 for 67% to &lt;80%, 3.1 for 80% to &lt;90%, and 3.3 for 90% to 100%.

- The Respiratory Rate accuracy performance of the Perin Health System was verified by using bench testing and through clinical validation against the gold standard, manually counted end-tidal CO₂.
- The demographics of the analyzed population were 17 males and 18 females (age: 24-72 years, weight: 111-390 lbs, height: 60-75 inches, BMI: 18-50). 1 American Indian/Alaska Native, 2 Asian, 8 Black/African-American, 23 White, and 1 other, with 8 participants of Hispanic/Latino ethnicity and 27 of Non-Hispanic/Non-Latino ethnicity were completed.
- After baseline data was collected at the participant’s natural respiratory rate, a range of stable respiratory rates (approx. 5, 10, 15, 20, 25, and 30 breaths per minute) were elicited from each volunteer.
- Subgroup analysis was conducted for sex, comorbidities (asthma, diabetes, COPD, smoker, hypertension, congestive heart failure), BMI, age, and skin tone.
- The accuracy root mean square (Arms) for 259 points was 1.7 breaths per minute, with all subgroups exhibiting Arms between 0.5 and 2.8.
- The mean absolute error (MAE) for 259 points was 0.8 breaths per minute, with all subgroups exhibiting MAE between 0.4 and 1.3.

- ECG, Heart Rate, R-R Interval, and Heart Rate Variability performance testing was performed on the Perin Health System through clinical validation against a standard Holter monitor.

Perin Health System
Page 21 of 22

{25}

Perin Health
K252984 510(k) Summary

- 243 participants were analyzed for ECG, HR, and HRV performance. Participants were enrolled from a diverse composition of gender (51.4% female/48.6% male), race (Asian 17.7%, Black/African American 15.2%, Hispanic or Latino 16.9%, White 39.9%, Other or Multiple 9.5%), age (22-90 with 32.1% ≥ 60 years), and a mix of healthy controls (50.78%) and patients with chronic conditions (hypertension, cardiovascular diseases, COPD, diabetes, and arrhythmias).
- The ECG performance validation demonstrated high levels of agreement between the Perin Health patch and the reference Holter monitor across all evaluated parameters (timing intervals, SNR, morphological features) and for all demographic and clinical subgroups.

- Wear-life performance of the Perin Health Patch was verified &amp; validated for compliance via internal clinical wear life evaluation to confirm sustained adhesion to the body for 360 hours.
- 26 participants were enrolled across 3 clinical sites, with a balanced sex distribution, 30% aged ≥ 60 years, 30% with BMI ≥ 30, and including healthy controls and patients with documented arrhythmias.
- The system demonstrated stable performance across all evaluated parameters (timing intervals, SNR, morphological features) and for all demographic and clinical subgroups.

# Conclusion

The Perin Health System is substantially equivalent in indications for use &amp; intended use to the legally marketed Predicate Device(s) &amp; Reference Device(s). Minor differences between the Perin Health System and the Predicate &amp; Reference devices have no significant effect on safety or effectiveness, as established through various performance tests. Thus, the proposed device/system is substantially equivalent to the legally marketed predicate &amp; reference devices.

Perin Health System
Page 22 of 22

---

**Source:** [https://fda.innolitics.com/device/K252984](https://fda.innolitics.com/device/K252984)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
