# PRIMED Surgical and Procedure Masks (K252964)

_Primed Medical Products, Inc. · FXX · Nov 4, 2025 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252964

## Device Facts

- **Applicant:** Primed Medical Products, Inc.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Nov 4, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

PRIMED Surgical masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings and during surgical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials. PRIMED Procedure Masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials.

## Device Story

Single-use, non-sterile, disposable surgical and procedure masks; constructed with melt-blown polypropylene filter layer between non-woven inner and outer layers; available in various models (ear loops or head ties); includes features like anti-fog tape and visors. Used by healthcare personnel in clinical and surgical settings to provide barrier protection against microorganisms, body fluids, and particulates. Performance levels (ASTM Level 1 or 3) determined by material basis weight and layer count. Masks are donned by the user to reduce cross-contamination risk between patient and provider.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-23 standards, including Bacterial Filtration Efficiency (ASTM F2101), Sub-micron Particulate Filtration (ASTM F3502), Differential Pressure (EN 14683), Synthetic Blood Penetration (ASTM F1862), and Flammability (16 CFR 1610). Biocompatibility confirmed per ISO 10993-1, ISO 10993-10, and ISO 10993-23.

## Technological Characteristics

Materials: Melt-blown polypropylene filter, spunbond/SSS/wetlaid cellulose/ES inner/outer layers, spandex/polyester ear loops, annealed aluminum/plastic-coated nosepiece. Design: Pleated/pleated omega, ultrasonic welding. Energy: None. Connectivity: None. Sterilization: Non-sterile. Standards: ASTM F2100:2023, ASTM F2101, ASTM F3502, EN 14683, ASTM F1862, 16 CFR 1610, ISO 10993.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- PRIMAGARD Surgical Masks ([K240286](/device/K240286.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 4, 2025

PRIMED Medical Products Inc.
Mitra Fard
Senior Manager, Regulatory Affairs
200, 2003-91 St SW
Edmonton, AB T6X0W8
Canada

Re: K252964
Trade/Device Name: PRIMED Surgical and Procedure Masks
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Regulatory Class: Class II
Product Code: FXX
Dated: September 2, 2025
Received: September 17, 2025

Dear Mitra Fard:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252964 - Mitra Fard
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K252964 - Mitra Fard
Page 3

Sincerely,

ALLAN GUAN -S

For Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252964  |   |
|  Device Name PRIMED Surgical Masks  |   |
|  Indications for Use (Describe) PRIMED Surgical masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings and during surgical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252964  |   |
|  Device Name PRIMED Procedure Masks  |   |
|  Indications for Use (Describe) PRIMED Procedure Masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials.  |   |
|  Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{5}

PRIMED Surgical and Procedure Masks
K252964 - 510(K) Summary

|  Summary Date: | October 31, 2025  |
| --- | --- |
|  Manufacturer: | PRIMED Medical Products Inc.  |
|  Address: | 200, 2003-91 St. SW, Edmonton, AB, Canada, T6X 0W8  |
|  Phone Number | +1.877.877.4633  |
|  Fax Number | +1.780.497.7670  |
|  Email | mitra.fard@primed.com  |
|  Contact person: | Mitra Fard  |
|  Common name of the device | Mask, Surgical  |
|  Trade or Proprietary Name: | PRIMED Surgical Masks and PRIMED Procedure Masks  |
|  Models | PG4-1001: ASTM Level 1, Procedure Ear Loop, Yellow Mask,
PG4-1223: ASTM Level 3, Procedure Ear Loops, Black Mask,
PG4-2001: ASTM Level 1, Surgical Ties, Blue Mask,
PG4-4053: ASTM Level 3, Anti-fog Tape, Surgical Ties, Green Mask,
PG4-5033: ASTM Level 3 Soft Mask, Anti-fog, Anti-Glare, Surgical Ties, Visor White,
PG4-5553: ASTM Level 3 Visor Anti-Fog Anti-Glare Surgical Ties Indigo Mask  |
|  Regulation Description | Surgical Apparel  |
|  Review Panel | General and Plastic Surgery  |
|  Regulation number | 21 CFR 878.4040  |
|  Device Class: | 2  |
|  Product Code: | FXX  |
|  Predicate device(s) | PRIMAGARD Surgical Masks, # K240286  |

## Device Description:

PRIMED Surgical masks and PRIMED Procedure masks are single use, non-sterile masks composed of a melt-blown filter layer placed between layers of non-woven inner and SS outer layer(s) material. PRIMED Medical Products Inc. offers various models in two main categories: Procedure masks with ear loops and Surgical masks with head ties. The level of protection varies based on the basis weight of the inner layer and the number of outer layers used in the mask's construction, in accordance with ASTM F2100:2023 specifications.

## Indication for Use:

PRIMED Surgical masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings and during surgical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials.

PRIMED Procedure Masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials.

Page 1 of 3

{6}

PRIMED Surgical and Procedure Masks

K252964 - 510(K) Summary

# Technical comparison:

Here below is the summary of the technological characteristics and the performance of the proposed device compared to the predicate device.

Table 1: Technological Characteristics Comparison Table

|  Elements of Comparison | Proposed device | Predicate | comments  |
| --- | --- | --- | --- |
|  Trade name | PRIMED Surgical Masks and PRIMED Procedure Masks | PRIMED Surgical Masks and PRIMED Procedure Masks | -  |
|  510k number | K252964 | K240286 | -  |
|  Manufacturer | PRIMED Medical Products Inc. | PRIMED Medical Products Inc. | -  |
|  Regulation number | 21 CFR 878.4040 | 21 CFR 878.4040 | Same  |
|  Product code | FXX | FXX | Same  |
|  Common name | Mask, Surgical | Mask, Surgical | Same  |
|  Indications For Use | PRIMED Surgical masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings and during surgical procedures to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials. | PRIMED Surgical Masks and PRIMED Procedure Masks are single use, non-sterile products intended to be worn by operating room and health care personnel in health care setting and during surgical procedures to protect both the patient and the operating room personnel from transfer of microorganisms, body fluids and particulate materials. | Similar  |
|   |  PRIMED Procedure Masks are single use, non-sterile, disposable products intended to be worn by healthcare personnel in healthcare settings to protect both the patient and healthcare personnel from transfer of microorganisms, body fluid and particulate materials.  |   |   |
|  ASTM F2100 level | Level 1 and Level 3 | Level 1 and Level 3 | Similar  |
|  Shelf life | PG4-1001: 5 yearsPG4-1223: 3 yearsPG4-2001: 5 yearsPG4-4053: 5 yearsPG4-5033: 5 yearsPG4-5553: 5 years | N/A | Different  |
|  Biocompatibility | Under the conditions of the testing, not a sensitizer, not an irritant, and not cytotoxic | Under the conditions of the testing, not a sensitizer, not an irritant, and not cytotoxic | Same  |
|  Single use vs. reusable | Single Use | Single Use | Same  |
|  Sterile vs. non-sterile | Non-sterile | Non-sterile | Same  |
|  Outer layer | Spunbond material | Spunbond material | Same  |
|  Filter layer | Meltblown PP | Meltblown PP  |   |
|  Inner layer | SSS materialWetlaid CelluloseES material | SSS materialWetlaid CelluloseES material  |   |
|  Ear loop | Spandex Polyester Ear loops | Spandex Polyester Ear loops  |   |
|  Head Tie | SSS Head tie | SSS Head tie  |   |
|  Nose piece | Annealed AluminiumPlastic coated nosepiece | Annealed AluminiumPlastic coated nosepiece  |   |
|  Color | White, Yellow, Green, Blue, Black, Indigo colour | White, Yellow, Green, Blue, Black, Indigo colour  |   |
|  Design | Pleated designPleated Omega design | Pleated designPleated Omega design | Same  |
|  Welding | Ultrasonic welding pattern | Ultrasonic welding pattern | Same  |
|  Accessories | Anti-fog, visor | Anti-fog, visor | Same  |

{7}

PRIMED Surgical and Procedure Masks
K252964 - 510(K) Summary

Summary of non-clinical testing:

Table 2: Summary of the Non-Clinical Tests:

|  ASTM F2100-23  |   |   |   |
| --- | --- | --- | --- |
|  Test Performed | Test method | Acceptance Criteria | Results (Pass/Fail)  |
|  Bacterial Filtration Efficiency | ASTM F2101 | Level 1: ≥ 95%, Level 3: ≥ 98% | Pass  |
|  Sub-micron Particulate Filtration | ASTM F3502 | Level 1: ≥ 80%, Level 3: ≥ 85% | Pass  |
|  Differential Pressure | EN 14683 | Level 1: < 5 mmH₂O/cm², Level 3: < 6 mmH₂O/cm² | Pass  |
|  Resistance to Penetration by synthetic blood | ASTM F1862 | Level 1: 80 mmHg, Level 3: 160 mmHg | Pass  |
|  Flammability | 16 CFR 1610 | Class I | Pass  |
|  Cytotoxicity | ISO 10993-1 | Under the conditions of the testing, non-Cytotoxic | Pass  |
|  Sensitization | ISO 10993-10 | Under the conditions of the testing, non sensitizing | Pass  |
|  Irritation | ISO 10993-23 | Under the conditions of the testing, not an irritant | Pass  |

Summary of clinical testing:

N/A

Conclusion:

Based on the non-clinical testing performed, it can be concluded that the subject device, PRIMED Surgical masks and PRIMED Procedure masks are as safe, as effective and perform as well as or better than the legally marketed predicate device, K240286.

Page 3 of 3

---

**Source:** [https://fda.innolitics.com/device/K252964](https://fda.innolitics.com/device/K252964)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).