DYNE PORT

K252933 · Dyne Medical Group, Inc. · ODC · Jun 12, 2026 · Gastroenterology, Urology

Device Facts

Record IDK252933
Device NameDYNE PORT
ApplicantDyne Medical Group, Inc.
Product CodeODC · Gastroenterology, Urology
Decision DateJun 12, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope.

Device Story

DYNE PORT is a single-use, sterile endoscope accessory; available in straight (SP-100) and Y-port (YP-100) configurations. Device attaches to endoscope accessory port via Luer connector. Primary components include adjustment cap, main body, and Luer connector; YP-100 includes position adjustment lever for fine depth control. Device secures inserted instruments and prevents irrigation fluid leakage during endoscopic procedures. Used in healthcare facilities/hospitals by clinicians. Benefits include maintaining fluid pressure and preventing backflow during instrument manipulation.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (cytotoxicity, skin sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogen), sterilization validation, leakage testing, device compatibility/integrity, and accelerated aging.

Technological Characteristics

Materials: Polycarbonate, Polypropylene, TPE, Silicone. Principle: Mechanical seal/adapter. Energy: None. Form factor: Straight or Y-shaped adapter with Luer connector. Sterilization: Ethylene Oxide (SAL 10^-6).

Indications for Use

Indicated for use as an endoscope channel accessory to prevent fluid backflow around instruments inserted into ureteroscopes, cystoscopes, or other endoscopes.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION June 12, 2026 DYNE Medical Group Inc % Jonghyeon Kim Chief Consultant GMS Consulting #B-612, 66 Chengcho-ro Gyeonggi-do, Goyang-si 10543 KOREA Re: K252933 Trade/Device Name: DYNE PORT Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODC Dated: May 12, 2026 Received: May 12, 2026 Dear Jonghyeon Kim: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252933 - Jonghyeon Kim Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {2} K252933 - Jonghyeon Kim Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MARK J. ANTONINO -S Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. # Indications for Use 510(k) Number (if known) K252933 Device Name DYNE PORT Indications for Use (Describe) This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 441-6740 EF {4} DYNE MEDICAL GROUP INC K252933 Page 1 of 4 ## 510(k) Summary [As Required by 21 CFR 807.92] 1. Date Prepared [21 CFR 807.92(a)(a)] Aug 29, 2025 2. Submitter's Information [21 CFR 807.92(a)(1)] - Name of Manufacturer: DYNE Medical Group Inc. - Address: #1001-1013, #1015-1016, #1024-1025, 10F, 165, Gimpohangang 10-ro 133beon-gil - Contact Name: Young-Hoon Moon / Quality Manager - Telephone No.: +82-31-8049-9013 - Email Address: yhmoon@dynemedical.com 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] - Trade Name: DYNE PORT - Model Name: SP-100 / YP-100 - Classification: | **Common Name** | Endoscope Channel Accessory | | --- | --- | | **Classification Name** | Endoscope and Accessories | | **Classification Number** | 21 CFR 876.1500 | | **Product Code** | ODC | | **Device Class** | II | | **Classification Panel** | Gastroenterology/Urology | 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] The identified predicate devices within this submission are shown as follow; Predicate Device - 510(k) Number: K173105 - Applicant: Cook Incorporated - Classification Name: Endoscope Channel Accessory - Trade Name: Tuohy-Borst Adapter 510(k) Summary 1 / 4 DYNE PORT {5} DYNE MEDICAL GROUP INC K252933 Page 2 of 4 ### 5. Description of the Device [21 CFR 807.92(a)(4)] The DYNE PORT is a single-use, sterile accessory for endoscopes, available in two configurations: a straight type (SP-100) and a Y-port type (YP-100). The device features a Luer connector on its distal end for attachment to an endoscope's accessory port. The DYNE PORT is constructed from its primary components: an Adjustment Cap, a main body, and a Luer Connector. The main body is designated as the Single Port (SP-100) or the Dual Port (YP-100). The device is designed to secure the position of an instrument inserted through the device, and simultaneously prevent the leakage of irrigation fluid. The YP-100 is an alternate design that incorporates a Y-shaped body with a side port for additional access. A distinguishing feature of the YP-100 is a Position Adjustment Lever, which allows for fine control of the instrument's depth after it has been secured by the Adjustment Cap. The DYNE PORT is sterilized using Ethylene Oxide (EO) gas and is intended for single use only. ### 6. Indications for use [21 CFR 807.92(a)(5)] This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope. ### 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)] The table below presents comparisons between the subject device (DYNE PORT) and the legally marketed predicate devices (K173105): Table 1. Comparison of Proposed Device to Predicate Device | | Proposed Device | Predicate Device | Remark | | --- | --- | --- | --- | | K Number | | K173105 | - | | Manufacturer | DYNE Medical Group Inc. | Cook Incorporated | - | | Trade Name | DYNE PORT | Tuohy-Borst Adapter | - | | Model Name | SP-100, YP-100 | TBA-6, TBAYR-6 | - | | Product code | ODC; Endoscope Channel Accessory | ODC; Endoscope Channel Accessory | Same | | Classification Regulation Number | 21 CFR §876.1500 | 21 CFR 876.1500 | Same | | Device class | II | II | Same | | Usage | Single-Use | Single-Use | Same | 510(k) Summary 2 / 4 DYNE PORT {6} DYNE MEDICAL GROUP INC K252933 Page 3 of 4 | Environment of Use | healthcare facility/hospital | healthcare facility/hospital | Same | | --- | --- | --- | --- | | Intended Use | This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope. | This device is intended to resist the backflow of fluid around an instrument inserted through the working channel of a ureteroscope, cystoscope, or other endoscope. | Same | | Accepts Instrumentation Fr (Working channel Inner diameter) | 9.0 FR (3.0 mm) | 3.0-6.0 FR | Different | | Energy Used/Delivered | None | None | | | Sterilization | Ethylene Oxide (SAL: 10^{-6}) | Ethylene Oxide | Same | | Shelf life | 3 years | 3 years | Same | | Biologics | Performed for the items below: • Cytotoxicity • Skin sensitization • Intracutaneous reactivity • Acute systemic toxicity • Pyrogen | Performed | Same | | Duration and type of contact | Limited (< 24 hours) | Limited (< 24 hours) | Same | | Material | Polycarbonate, Polypropylene, TPE, Silicone | ACETAL, polycarbonate, polyethylene | Different | | Label and Labeling | Meet FDA's Requirements | Meet FDA's Requirements | | The comparison table shows that the subject device, the DYNE PORT, has the same Indications for Use as the predicate device, Cook Incorporated's Tuohy-Borst Adapter. Both devices share the same intended use, and also share the same device class, single-use designation, and duration of patient contact. Although the devices have some different technological characteristics, primarily in accepted instrumentation size, shelf life, and the materials of construction. Characteristics of the subject devices that differ from the predicate devices are supported by testing and analysis. 510(k) Summary 3 / 4 DYNE PORT {7} DYNE MEDICAL GROUP INC K252933 Page 4 of 4 ### 8. Performance Data The following testing was performed in order to demonstrate that the proposed DYNE PORT met applicable design and performance requirements. - Biocompatibility - Sterilization - Leakage - Device Compatibility and Integrity - Accelerated Age ### 9. Conclusion [21 CFR 807.92(b)(3)] In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification DYNE Medical Group Inc., concludes that the DYNE PORT is substantially equivalent in safety and effectiveness to the predicate devices as described herein. 510(k) Summary 4 / 4 DYNE PORT
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...