Lux HD 2530 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 2530 Detector is not intended for mammography or dental applications.
Device Story
Lux HD 2530 is a digital flat panel detector for general radiography. It uses a CsI scintillator to convert X-rays to visible light, which is then transformed into electronic signals by a TFT/diode capacitor array. The device transmits digital image data to a PC via wired (Gigabit Ethernet) or wireless (IEEE 802.11a/b/g/n/ac) connection. Operated by clinicians in clinical settings, the detector integrates with a DR software platform for image acquisition, enhancement, and viewing. It features Automatic Exposure Detection (AED) to trigger image acquisition upon X-ray detection and Detector Exposure Control (DEC) to signal the high-voltage generator to terminate exposure once a target dose is reached. The output assists physicians in diagnosing disease or injury, potentially reducing patient radiation dose through precise exposure control.
Clinical Evidence
Bench testing only. Performance evaluated via electrical safety, EMC, biological evaluation, wireless coexistence, and cybersecurity testing. No clinical data provided.
Technological Characteristics
Flat panel detector with CsI scintillator, TFT readout, 16-bit ADC. Active area 25cm x 30cm, 100μm pixel size, 5.0 lp/mm spatial resolution. Connectivity: Gigabit Ethernet, IEEE 802.11a/b/g/n/ac. IP67 rating. Carbon fiber construction. Air-cooled. Powered via 10-pin port.
Indications for Use
Indicated for digital radiographic diagnosis of human anatomy in adult and pediatric patients. Not for mammography or dental applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 9, 2025
iRay Imaging Technology (Haining) Limited
c/o Junjie Qian
No. 2, Caohejing RD.
Haining, Jiaxing, Zhejiang 314499
CHINA
Re: K252911
Trade/Device Name: Lux HD 2530 Detector (Lux HD 2530)
Regulation Number: 21 CFR 892.1680
Regulation Name: Stationary X-Ray System
Regulatory Class: Class II
Product Code: MQB
Dated: August 1, 2025
Received: September 12, 2025
Dear Junjie Qian:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252911 - Junjie Qian
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K252911 - Junjie Qian
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252911 | |
| Device Name Lux HD 2530 detector (Lux HD 2530) | |
| Indications for Use (Describe) | |
| Lux HD 2530 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 2530 Detector is not intended for mammography or dental applications. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
iRay Imaging Technology (Haining) Limited [510(k)] Application
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
(As Required by 21 CFR 807.92)
K252911
1. Date Prepared [21 CFR 807.92(a)(1)]
August 8, 2025
2. Submitter's Information [21 CFR 807.92(a)(1)]
Company Name: iRay Imaging Technology (Haining) Limited
Company Address: No. 2, Caohejing RD., Haining 314499, Jiaxing, Zhejiang, China
Contact Person: Junjie Qian
Phone: 0573-87399739
Email: Junjie.qian@iraygroup.com
3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Name: Lux HD 2530 Detector (Lux HD 2530)
Common Name: Solid State X-Ray Imager
Model Name: Lux HD 2530
Classification Name: Stationary X-Ray System
Product Code: MQB
Regulation Number: 21 CFR 892.1680
Device Class: Class II
1 of 10
{5}
iRay Imaging Technology (Haining) Limited [510(k)] Application
# 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)]
The identification predicates within this submission are as follows:
Manufacturer: iRay Imaging Technology (Haining) Limited
Trade Name: Lux HD 35 Detector (Lux HD 35)
Lux HD 43 Detector (Lux HD 43)
Model Name: Lux HD 35
Lux HD 43
Product Code: MQB
Classification Name: Stationary X-Ray System
Regulation Number: 21 CFR 892.1680
Device Class: Class II
FDA 510 (k) #: K243556
# 5. Description of the Device [21 CFR 807.92(a)(4)]
Lux HD 2530 Detector is digital flat panel detector. They support the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 25cm × 30cm (Lux HD 2530 Detector). The differences between two models are overall change in the dimensions of the image receptor
The sensor plate of Lux HD 2530 Detector is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface.
The major function of the Lux HD 2530 Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging.
2 of 10
{6}
iRay Imaging Technology (Haining) Limited [510(k)] Application
The Digital Radiographic Imaging Acquisition Software Platform - DR is part of the system, it is used to acquire, enhance, view image from Lux HD 2530 Detector. Based on the risks and intended use, documentation level of the software is basic
## 6. Intended Use [21 CFR 807.92(a)(5)]
### 6.1. Indications for use
Lux HD 2530 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 2530 Detector is not intended for mammography or dental applications.
### 6.2. Suitable patient
Lux HD 2530 Detector is suitable for providing digital X-ray imaging for DR system to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients, but not intended for mammography or dental applications.
### 6.3. Processing of input and output
When Lux HD 2530 Detector work continuously, it can automatically distinguish X-ray and output an imaging for diagnosis of disease, injury, or of any applicable health problem.
3 of 10
{7}
iRay Imaging Technology (Haining) Limited [510(k)] Application
7. Technological Characteristic [21 CFR 807.92(a)(6)]
| Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector |
| --- | --- | --- |
| 510(K) Number | K243556 | K252911 |
| Intended Use | The Lux HD detectors are indicated for digital imaging solution designed for providing general radiographic system in all general-purpose diagnostic procedures. | Same |
| Indications for Use | Lux HD Detectors are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. The devices are not intended for mammography or dental applications. | same |
| Classification Name | Stationary X-ray system | Same |
| Product Code | MQB | Same |
{8}
iRay Imaging Technology (Haining) Limited [510(k)] Application
| Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector |
| --- | --- | --- |
| Regulation Number | 21 CFR 892.1680 | Same |
| Panel | Radiology | Same |
| Classification | II | Same |
| X-Ray Absorber (Scintillator) | CsI | Same |
| Installation Type | Wireless or Wired | Same |
| Readout Mechanism | Thin Film Transistor | Same |
| Image Matrix Size | Lux HD 35 Detector: 3500 ×4300pixels Lux HD 43 Detector: 4267 × 4267 pixels | 2500 × 3000pixels |
| Pixel Size | 100μm | Same |
| ADC Digitization | 16 bit | Same |
| Effective Imaging Area | Lux HD 35 Detector: 350mm ×430mm Lux HD 43 Detector: 426.7 mm × 426.7mm | 250mm × 300mm |
| Spatial Resolution | 5.0 lp/mm | Same |
| Detective Quantum Efficiency (DQE) | 0.54 at 1 lp/mm (RQA5, 2.5μGy) | Same |
5 of 10
{9}
iRay Imaging Technology (Haining) Limited [510(k)] Application
| Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector |
| --- | --- | --- |
| Power Consumption | Max. 30W | Same |
| Communications | a) Wired: Gigabit Ethernet (1000BASE-T) b) Wireless: IEEE 802.11a/b/g/n/ac (2.4 GHz / 5 GHz) | Same |
| Imaging protect Plate | Carbon Fiber Plate | Same |
| Cooling | Air cooling | Same |
| Dimensions | Lux HD 35 Detector: 384mm×460mm×15mm Lux HD 43 Detector: 460mm×460mm×15mm | 281.5mm×332.3mm×15 mm |
| Detector IP grade | IP67 | Same |
| Power input port | 10 pin port | Same |
| Frame material | Carbon fiber | Same |
| Surface pressure | Uniform load: 300 kg over the whole area of the surface; Local load: 100 kg on an area 4 cm diameter of center | Same |
| Operation | Temperature: +5 ~ +35°C Humidity: 5 ~ 90% (Non-Condensing) | Same |
{10}
iRay Imaging Technology (Haining) Limited [510(k)] Application
| Item | Predicate Device: Lux HD 43 DetectorLux HD 35 Detector | Proposed Device: Lux HD 2530 Detector |
| --- | --- | --- |
| | Atmospheric pressure: 70 ~ 106 kPaAltitude: Max. 3000 meters | |
| Storage and Transportation: (detector) | Temperature: -10 to + 60°C.(Excluding battery)Relative Humidity: 5 to 95%, non-condensing.Atmosphere: 700mBar~1060mBarAltitude: Max. 3000 meter | Same |
| Software | Digital Radiographic Imaging Acquisition Software Platform - DRThe software used for getting Digital X-ray radiography images from the flat panel detectors.The software is used to handle the DICOM protocol (DICOM 3.0).The software is responsible for the DR equipment management, acquisition and processing functions, to | Same |
{11}
iRay Imaging Technology (Haining) Limited [510(k)] Application
| Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector |
| --- | --- | --- |
| | provide patient registration, scanning, image processing and other functions. | |
# 8. Minimum requirement for connection
Lux HD 2530 Detector connected via wireless and wired communication.
Operating System: Windows 11
CPU: Intel Core i3- 8100 3.6GHz 4C 65W
Memory: 16GB (2x8GB) DDR4 2666 DIMM
Hard Disk: 1TB
# 9. X-ray exposure mode
The AED (Automatic Exposure Detection) trigger module is a unit can connect X-ray signal in the Lux HD 2530 Detector. Once there is X-ray generator exposure exist, the AED trigger module will detect the X-ray radiation and output signal to the detector. Until the exposure finished, the detector will receive a signal which represent the end of exposure from the inner trigger module and begin to acquire the image.
The DEC (Detector Exposure Control) trigger module is a unit can cutoff HVG output when patients need specified dose, which avoid extra dose absorbed by patients. Once the dose detected by the detector reaches the target value, the detector will send a cut-off signal to the high-voltage system. This signal indicates that no further exposure from HVG is required.
8 of 10
{12}
iRay Imaging Technology (Haining) Limited [510(k)] Application
# 10. Non-clinical study
1) Electrical Safety and EMC testing:
Electrical, mechanical, environmental safety according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC 60601-1-2. All test results are meet the standard requirements.
2) Biological Evaluation:
The materials of the detector which contact operators' or patients' skin have been evaluated with the FDA guidance "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". And the evaluation result assured the biological safety.
3) Non-clinical Considerations:
The non-clinical studies have been performed and the results have shown that sections of the non-clinical consideration mentioned in the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices' are substantially equivalent to the non-clinical consideration of predicate device on the Market (Lux HD 35/43 detector, K243556).
4) Clinical Consideration:
Intended use, fundamental scientific technology, regulatory requirement, non-clinical performance, labeling, quality-assurance program keep the same with those of predicate device. Software keep the same with that of reference device except "image forwarding, image printing" function and literal name. There is no any negative change about clinical performance from predicate device.
5) Wireless testing
Wireless functionality and wireless coexistence testing in accordance with ANSI IEEE C63.27-2017 was performed. All test results are meet the standard requirements.
9 of 10
{13}
iRay Imaging Technology (Haining) Limited [510(k)] Application
6) Cybersecurity testing
Cybersecurity threat modeling, risk assessment, and controls and testing were performed to comply with requirements specified in section 524B(b)(2) of the Federal Food, Drug, and Cosmetics Act to provide a reasonable assurance that the subject device with its wireless capabilities are cybersecure.
11. Conclusion [21 CFR 807.92(b)(3)]
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, applicant concludes that Lux HD 2530 Detector is substantially equivalent to predicate device with regards to safety and effectiveness.
10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.