K252862 · Foshan Dahao Medical Technology Co., Ltd. · ITI · Dec 8, 2025 · Physical Medicine
Device Facts
Record ID
K252862
Device Name
Electric Wheelchair (DH01126)
Applicant
Foshan Dahao Medical Technology Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Dec 8, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Electric wheelchair (DH01126) provides mobility for disabled or elderly users. Input: user manipulation of joystick controller. Transformation: controller signals 24V DC brushless motors to drive rear wheels; electromagnetic brakes provide stopping. Output: powered movement. Used in indoor/flat outdoor environments; operated by patient. Output affects speed/direction; joystick release triggers automatic braking. Benefits: provides independent mobility for users with limited walking capacity. Foldable frame for storage/transport.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), electromagnetic compatibility (IEC 60601-1-2), and performance testing (ISO 7176 series).
Technological Characteristics
Carbon fiber frame; 24V DC brushless motors; lithium-ion battery; electromagnetic brakes. Dimensions: 860mm x 600mm. Connectivity: off-board charger. Software: embedded controller for joystick-based speed/direction control. Sterilization: N/A. Standards: ISO 7176 series, IEC 60601-1-2, ISO 10993.
Indications for Use
Indicated for disabled or elderly persons with mobility difficulties limited to a seated position requiring indoor and outdoor transportation.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 8, 2025
Foshan Dahao Medical Technology Co., Ltd.
% Sandra Jiang
Regulatory Affairs Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave
Shanghai, 200122
China
Re: K252862
Trade/Device Name: Electric Wheelchair (DH01126)
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: September 9, 2025
Received: September 9, 2025
Dear Sandra Jiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252862 - Sandra Jiang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252862 - Sandra Jiang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Electric Wheelchair
Page 12 of 45
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252862 | ? |
| Please provide the device trade name(s). | | ? |
| Electric Wheelchair (DH01126) | | |
| Please provide your Indications for Use below. | | ? |
| It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a
disabled or elderly person limited to a seated position. | | |
| Please select the types of uses (select one or both, as
applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building2, Building3,4,5,6,7, No.9 of Fanye
Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
K252862 510(K) Summary
(As requirement by 21 CFR 807.92)
Date prepared: 21st, August, 2025
A. Applicant:
Name: FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Address: Building 1, 2nd Floor of Building 2, 3rd Floor of Building2, Building3,4,5,6,7, No.9 of
Fanye Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
Contact: Ms. Bao Yi Xie
Tel: +86 13760674304
Email: 709111267@qq.com
Submission Correspondent:
Primary contact: Ms. Sandra Jiang
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-58817802
Email: yuan.jiang@sungoglobal.com
Secondary contact: Ms. Eva Li
Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China
Tel: +86-21-68828050
Email: zxfda@sungoglobal.com
B. Device:
Trade Name: Electric Wheelchair
Device Model: DH01126
Regulatory Information
Classification Name: Wheelchair, Powered
Classification: Class II
Product code: ITI
Regulation Number: 21 CFR 890.3860
Review Panel: Physical Medicine
C. Predicate device:
510(K) Number: K241529
Manufacturer: Zhejiang Nysin Medical Co., Ltd.
Device Name: Powered Wheelchair (NXN20-209)
K252862
{5}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building2, Building3, 4, 5, 6, 7, No.9 of Fanye Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
## D. Device Description:
The subject Electric Wheelchair is a wheeled personal mobility device that incorporates a seat-support system for a person with a disability or a person without the full capacity to walk designed to be propelled by power from electric motors. The electronic control of speed and direction can be performed by the occupant with the help of controlling joystick. The device can be quickly folded and disassembled, which makes it convenient to be stored or placed at the trunk of vehicles while traveling.
The subject Electric Wheelchair are intended to provide mobility to a disabled or elderly limited to a seated position. It is of indoor and outdoor type, suitable for the use indoor and flat path near buildings, but not on grass, gravel roads, large slopes or motorway, neither on muddy, rugged, soft, narrow, icy road, bad roads such as dangerous roads without guardrails or waterways.
The DH01126 Electric Wheelchair consists of two parts, the wheelchair main body and the electrical part. The main body includes a main foldable frame, two armrests, a backrest, a seat cushion, two rear driving wheels and two front wheels. The electrical part is composed of two 24V DC motors, two brakes, lithium battery, a controller and an off-board charger.
The device is powered by Li-ion battery, which can be recharged by an off-board battery charger that can be plugged into an AC socket outlet when the device is not in use.
The patient can activate the controller handle (joystick) to control the speed and direction of the wheelchair. The further the joystick is pushed from its central position, the faster the wheelchair moves, when it is released, it will automatically reset and brake.
The DC brushless motor and brake system are fixed on the rear wheels.
The max loading of both models is 100KG. Only for one person sit.
## E. Indication for Use
The proposed Electric Wheelchair is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
## F. Comparison with the predicate device
Table 1 - General Information Comparison
| Item | Proposed Device | Predicate Device | Discussion/Conclusion |
| --- | --- | --- | --- |
| Product Code | ITI | ITI | Same |
| Regulation No. | 21 CFR 890.3860 | 21 CFR 890.3860 | Same |
K252862
{6}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1,2nd Floor of Building 2,3rd Floor of Building2, Building3,4,5,6,7,No.9 of Fanye
Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
| Device classification | Class II | Class II | Same |
| --- | --- | --- | --- |
| Product Name | Electric Wheelchair | Powered Wheelchair | Same |
| Intended Use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | Same |
| Use Environment | Indoor and outdoor use | Indoor and outdoor use | Same |
| Patient Population | The device is suitable for disabled people with mobility difficulties and elderly people. | The device is suitable for disabled people with mobility difficulties and elderly people. | Same |
| Product Structure | Consist of two foldable armrests, a backrest, a seat cushion, a safety belt, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, a piece of Li-ion battery pack, an off-board battery charger, a control panel, and an electric motor controller. | Consist of two foldable armrests, a backrest, a seat cushion, a safety belt, a foldable frame, two rear driving wheels with hub motor/electromagnetic brake assemblies, two pivoting casters, two Li-ion batteries, an off-board battery charger, a control panel, and an electric motor controller. | Same |
| Driving System | Direct drive on the rear wheels | Direct drive on the rear wheels | Same |
| Number of wheels | 4 | 4 | Same |
| Motor | Brushless DC motor, 24V/150W, 2pcs | Brushless motor, 24V/200W, 2pcs | Analysis: Minor differences in the specification of electric parts will not impact the |
| Battery | Lithium-ion battery 24V10Ah*1 | Lithium-ion battery 24V12Ah*2 | |
| Battery Charger | Off-board charger: Input: 100-240VAC, | Off-board charger: Input: 100-240 VAC | |
K252862
{7}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building2, Building3, 4, 5, 6, 7, No.9 of Fanye
Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
| | 50/60Hz, 2A
Output: 29.4V, 2.0A | Output: DC 24V, 3A | safety and effectiveness of the substantial equivalence. |
| --- | --- | --- | --- |
| Main Frame Material | Carbon Fiber | Carbon Fiber | Same |
Table 2 - Technical Parameters Comparison
| Item | Subject Device | Predicate Device | Discussion/Conclusion |
| --- | --- | --- | --- |
| Overall length | 860mm | 930mm | Analysis:
Dimensions and mass of the subject electric wheelchair have been tested according to ISO 7176-5. Such minor differences of the parameters won’t will not affect the safety or effectiveness of the device |
| Overall width | 600mm | 585mm | |
| Storage length | 620mm | 740mm | |
| Storage width | 260mm | 265mm | |
| Storage height | 700mm | 720mm | |
| Weight | 14kg | 13.75kg | |
| Front Wheels | 6.5 inch PU tire | 7-inch PU solid tires | Similar |
| Rear Wheels | 12-inch PU solid tires | 12-inch PU solid tires | Same |
| Cruising Range | 12.4 km | 16.5 km | Analysis:
Cruising range of the subject electric wheelchair have been tested according to ISO 7176-4. Such minor differences of the parameters won’t will not affect the safety or effectiveness of the device. |
| Obstacle Climbing (mm) | 20 | 25 | Analysis:
The obstacle climbing ability of the subject wheelchair has been tested according to ISO 7176-10. Such difference will not affect the |
K252862
{8}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building2, Building3, 4, 5, 6, 7, No.9 of Fanye
Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
| | | | safety and effectiveness of the device. |
| --- | --- | --- | --- |
| Static stability forward | 23.1° | 18.9° | Analysis: The static stability of both proposed device and predicate devices is similar. Both devices have been evaluated according to ISO 7176-1 standard, therefore, such minor difference will not affect the safety and effectiveness of the device. |
| Static stability rearward | 15.9° | 18.2° | |
| Static stability sideways | Left: 23.2° Right: 23.0° | 20.7° | |
| Max. Loading (kg) | 100kg | 110kg | Analysis: The max. loading of the proposed device have been evaluated according to ISO 7176 standard, therefore, such minor difference will not affect the safety and effectiveness of the device. |
| Min. Turning radium | 825mm | 870mm | Analysis: Both devices have been evaluated according to ISO 7176-5 standard, therefore, such minor difference will not affect the safety and effectiveness of the device. |
| Min. braking distance | 0.6m | 1.76m | Analysis: Both devices have been evaluated according to ISO 7176-3 standard, therefore, such minor difference will not affect the safety and effectiveness of the device. |
| Max Speed Forwards | 1.3m/s | 1.66m/s | Analysis: |
K252862
{9}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building2, Building3, 4, 5, 6, 7, No.9 of Fanye
Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R.China.
| | | | The specification of both proposed device and predicate device is similar.
Both devices have been evaluated according to ISO 7176-6 standard, therefore, such minor difference will not affect the safety and effectiveness of the device. |
| --- | --- | --- | --- |
| Speed control method | Joystick control | Joystick control | Same |
Table 3 - Safety Comparison
| Item | Proposed Device | Predicate Device | Discussion/Conclusion |
| --- | --- | --- | --- |
| Biocompatibility | All user directly contacting materials are compliance with ISO10993-5, ISO10993-10 and ISO10993-23 requirements. | Comply with FDA Guidance | Same |
| EMC | IEC 60601-1-2& ISO7176-21 | IEC 60601-1-2& ISO7176-21 | Same |
| Performance | ISO7176 series | ISO7176 series | Same |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | Same |
## G. Summary of substantial equivalence discussion
The proposed DH01126 Electric Wheelchair complies with the requirements of ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2008, ISO 7176-13:1989, ISO 7176-14:2008, ISO 7176-15:1996, ISO 7176-16/ISO 16840-10, ISO 7176-21:2009, ISO 7176-22:2014, ISO 7176-25:2013, IEC 60601-1-2:2020, IEC TR 60601-4-2:2016, ISO 10993-1:2018, ISO10993-5:2009, ISO 10993-10:2021 and ISO 10993-23:2021.
The instructions for use, design and technological characteristics of the subject Electric Wheelchair are similar to the predicate device. The design principles of the controller and driving system are the same, and both meet the requirements of the ISO 7176-14:2008. Software validation has been carried out on both control systems. Brake system and speed
K252862
{10}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building 2, Building 3, 4, 5, 6, 7, No. 9 of Fanye Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R. China.
control are designed in the same way as well, and both meet the requirements of the ISO 7176-3:2012. Although technical parameter, such as device dimensions, weight, static and dynamic stability, maximum obstacle climbing etc., as listed in Table 2 are slightly different, such minor differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns because both devices been tested according to ISO 7176 series standard and the test results showed that the devices comply with those requirements.
The biocompatibility of the predicate device and subject device meet the requirements of the ISO 10993-5:2009, ISO 10993-10:2021 and ISO 10993-23:2021. The flame retardant test of the seat cushion/back cushion and armrest of both subject device and predicate device is carried out according to the ISO 7176-16 and ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level.
In conclusion, the technological characteristics, features, specifications, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
## H. Summary of Non-clinical Tests
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for Irritation
- ISO 7176-1:2014, Wheelchairs - Part 1: Determination of static stability
- ISO 7176-2:2017, Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs
- ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes
- ISO 7176-4:2008 Wheelchairs - Part 4: Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- ISO 7176-5:2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
- ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
- ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
K252862
{11}
FOSHAN DAHAO MEDICAL TECHNOLOGY CO., LTD.
Building 1, 2nd Floor of Building 2, 3rd Floor of Building 2, Building 3, 4, 5, 6, 7, No. 9 of Fanye Road, Leping Town, Sanshui District, FoshanCity, Guangdong Province, P.R. China.
- ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs
- ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- ISO 7176-11:2012 Wheelchairs - Part 11: Test dummies
- ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
- ISO 7176-14:2008 Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
- ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
- ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters.
- ISO 7176-22: 2014 Wheelchairs - Part 22: Set-up procedures
- ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchair
- ISO 16840-10:2021/AMD-1 2024 Wheelchair seating - Part 10: Resistance to ignition of postural support device - Requirements and test method
- IEC 60601-1-2:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance
- IEC 60601-4-2 Medical electrical equipment - Part 4-2: Guidance and interpretation for electromagnetic immunity of medical electrical equipment and medical electrical systems
I. Biocompatibility of Patient-contacting material
Biocompatibility of patient-contacting material are evaluated in accordance with ISO 10993-1: 2018. Although there are risks related to cytotoxicity, sensitization and irritation, the risk level is very low as the patient-contacting parts of the power wheelchair are made from materials in common use for other consumer products with a similar nature of contact, in addition, the product quality is controlled and managed by design, manufacturing, quality control, safety instructions, or warning information, hence the biological safety of the power wheelchair is acceptable without further biocompatibility testing on some parts contacted with the user during operation procedure of the product.
J. Summary of Clinical Testing
No clinical or animal study is available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate device.
K. Conclusion
The differences between DH01126 Electric Wheelchair and its predicate devices do not introduce a new intended use and do not raise new issues of safety and effectiveness. Verification and Validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. From the results of nonclinical testing described, it can be concluded that the subject device electric wheelchair is substantially equivalent to the legally marketed predicate device.
K252862