The Zio ECG Utilization Software (ZEUS) Platform is intended to process, store, and analyze electrocardiogram (ECG) recordings captured from a compatible ECG monitor, including automatically detected arrhythmia events and patient-triggered events. The ZEUS Platform also allows for review of the analyzed data and, subsequently, report generation. Interpretation of the analyzed ECG data and the associated diagnosis is the responsibility of the prescribing or treating healthcare professional. The ZEUS Platform is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre- syncope, syncope, fatigue, or anxiety. The ZEUS Platform is not intended for detection or notification of hemodynamically unstable or life- threatening arrhythmias or cardiac events requiring urgent medical response. It is not intended for patients at elevated risk of serious cardiovascular events that would require immediate intervention.
Device Story
ZEUS Platform is SaMD processing/analyzing ECG recordings from compatible Zio monitors; identifies/annotates beats, rhythms, and artifacts. Utilizes deep-learned AI (ECGDLv2/v3) for analysis of transmissions and post-wear recordings. Enables Qualified Cardiac Technicians (QCTs) to review/edit algorithm-generated findings; facilitates final report generation for clinicians. Used in clinical settings to aid arrhythmia diagnosis/management. Output provides preliminary findings for physician review; supports clinical decision-making regarding cardiac rhythm management. Benefits patients by providing diagnostic data for symptomatic/asymptomatic cardiac events.
Clinical Evidence
No clinical testing was performed in support of this premarket notification. Evidence is based on nonclinical design verification and validation testing.
Technological Characteristics
SaMD; deep-learned AI-based ECG analysis (ECGDLv2/v3). Compatible with Zio XT, Zio monitor, and Zio AT devices. Conforms to ANSI AAMI ISO 14971:2019, ANSI AAMI IEC 62304:2006/A1:2016, ANSI AAMI IEC 60601-2-47:2012/(R)2016, and ANSI AAMI EC57:2012.
Indications for Use
Indicated for patients 18+ years, asymptomatic or symptomatic (palpitations, dyspnea, dizziness, light-headedness, pre-syncope, syncope, fatigue, anxiety). Not for hemodynamically unstable/life-threatening arrhythmias or patients requiring urgent/immediate intervention.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
Zio® ECG Utilization Software (ZEUS) System (K222389)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2026
iRhythm Technologies, Inc.
Brittany Davis
Staff Regulatory Affairs Specialist
699 8th Street
Suite 600
San Francisco, California 94103
Re: K252859
Trade/Device Name: ZEUS Platform (FG0501US)
Regulation Number: 21 CFR 870.1425
Regulation Name: Programmable Diagnostic Computer
Regulatory Class: Class II
Product Code: DQK
Dated: May 4, 2026
Received: May 4, 2026
Dear Brittany Davis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252859 - Brittany Davis
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K252859 - Brittany Davis
Page 3
Sincerely,
JENNIFER W. SHIH - S
Jennifer Kozen
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252859 | ? |
| Please provide the device trade name(s). | | ? |
| ZEUS Platform (FG0501US) | | |
| Please provide your Indications for Use below. | | ? |
| The Zio ECG Utilization Software (ZEUS) Platform is intended to process, store, and analyze electrocardiogram (ECG) recordings captured from a compatible ECG monitor, including automatically detected arrhythmia events and patient-triggered events. The ZEUS Platform also allows for review of the analyzed data and, subsequently, report generation. Interpretation of the analyzed ECG data and the associated diagnosis is the responsibility of the prescribing or treating healthcare professional. The ZEUS Platform is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre- syncope, syncope, fatigue, or anxiety. The ZEUS Platform is not intended for detection or notification of hemodynamically unstable or life- threatening arrhythmias or cardiac events requiring urgent medical response. It is not intended for patients at elevated risk of serious cardiovascular events that would require immediate intervention. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
iRhythm Technologies, Inc.
Traditional 510(k): 510(k) Summary
ZEUS Platform: K252859
# I. General Information
Applicant: iRhythm Technologies, Inc.
699 8th Street, Suite 600
San Francisco, CA 94103 USA
Phone: 415-632-5700
Fax: 415-632-5701
Contact Person: Brittany Davis
Staff Regulatory Affairs Specialist
Phone: 415-757-3012
Email: brittany.davis@irhythmtech.com
Date Prepared: May 4, 2026
# II. Device Information
Trade Name: ZEUS Platform
Generic/Common Name: Programmable diagnostic computer
Classification Names: Programmable diagnostic computer [21CFR§870.1425]
Regulatory Class: Class II
Product Codes: DQK, Computer, Diagnostic, Programmable
# III. Predicate Devices
The following predicate device has been selected:
- iRhythm Technologies, Inc. Zio® ECG Utilization Software (ZEUS) System [K222389]
# IV. Indications for Use
The Indications for Use statement for the ZEUS Platform is as follows:
The Zio ECG Utilization Software (ZEUS) Platform is intended to process, store, and analyze electrocardiogram (ECG) recordings captured from a compatible ECG monitor, including automatically detected arrhythmia events and patient-triggered events.
The ZEUS Platform also allows for review of the analyzed data and, subsequently, report generation.
Page 1 of 6
{6}
iRhythm Technologies, Inc.
Traditional 510(k): 510(k) Summary
ZEUS Platform: K252859
Interpretation of the analyzed ECG data and the associated diagnosis is the responsibility of the prescribing or treating healthcare professional.
The ZEUS Platform is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety.
The ZEUS Platform is not intended for detection or notification of hemodynamically unstable or life-threatening arrhythmias or cardiac events requiring urgent medical response. It is not intended for patients at elevated risk of serious cardiovascular events that would require immediate intervention.
## V. Device Description
ZEUS Platform is a Software as a Medical Device (SaMD) consisting of a collection of software modules intended to process, store and analyze electrocardiogram (ECG) recordings captured from compatible Zio ECG monitoring devices, including automatically detected arrhythmia events and patient-triggered events. ZEUS Platform also enables Qualified Cardiac Technicians (QCTs) to review and edit ECG data after algorithm analysis and facilitate report generation. Reports curated with preliminary findings are intended for use by clinicians as an aid to arrhythmia diagnosis and management.
ZEUS Platform utilizes a deep-learned, artificial intelligence-based ECG analysis software module present in both the predicate and subject devices. It is intended to analyze both ECG transmissions during wear and full ECG recordings post-wear captured by all compatible Zio ECG monitoring devices. As a part of the analysis, the ECGDL will identify and annotate beats, beat types, ectopic runs, and rhythms/artifact present in the ECG data.
## VI. Comparison of Technological Characteristics with Predicate Devices (Substantial Equivalence)
The subject ZEUS Platform has the same intended use as the predicate device. The differences in the technological characteristics between the subject and predicate device do not raise any issues of safety or efficacy as the fundamental scientific technology and intended use is unchanged. Thus, the ZEUS Platform is considered substantially equivalent to the predicate device.
A comparison table outlining the differences and similarities between the subject device and the predicate device is provided in Table 1.
Table 1. Substantial Equivalence Summary Table
Page 2 of 6
{7}
iRhythm Technologies, Inc.
Traditional 510(k): 510(k) Summary
ZEUS Platform: K252859
| Category | Subject Device | Predicate Device (K222389) |
| --- | --- | --- |
| Device Name | ZEUS Platform | ZEUS System |
| Manufacturer | iRhythm Technologies, Inc. | Identical |
| Classification | Class II | Class II (Special controls) |
| Product Code | • Classification Product Code: DQK | • Classification Product Code: DQK • Subsequent Product Codes: DSI, DXH |
| Review Panel | Cardiovascular | Identical |
| Prescription/OTC | Prescription | Identical |
| Indications for use statement | The Zio ECG Utilization Software (ZEUS) Platform is intended to process, store, and analyze electrocardiogram (ECG) recordings captured from a compatible ECG monitor, including automatically detected arrhythmia events and patient-triggered events. The ZEUS Platform also allows for review of the analyzed data and, subsequently, report generation. Interpretation of the analyzed ECG data and the associated diagnosis is the responsibility of the prescribing or treating healthcare professional. The ZEUS Platform is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from symptoms such as palpitations, shortness of breath, dizziness, light-headedness, presyncope, syncope, fatigue, or anxiety. The ZEUS Platform is not intended for detection or notification of hemodynamically unstable or life-threatening arrhythmias or cardiac events requiring urgent medical | The device is intended to capture, analyze and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring. After wear, ECG data from compatible monitoring devices is processed and analyzed by the ZEUS System. A final report is generated on the beat-to-beat information from the entire ECG recording. For the Zio AT service, the ZEUS System supports the capture and analysis of automatically-detected arrhythmia events, as well as the analysis of uploaded patient-triggered events. The ZEUS System is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety and patients who are asymptomatic. The reports are provided for review by the intended user to render a diagnosis based on clinical judgment and experience. It is not intended for use on critical care patients. |
Page 3 of 6
{8}
iRhythm Technologies, Inc.
Traditional 510(k): 510(k) Summary
ZEUS Platform: K252859
| Category | Subject Device | Predicate Device (K222389) |
| --- | --- | --- |
| | response. It is not intended for patients at elevated risk of serious cardiovascular events that would require immediate intervention. | |
| Principle functionality | • ECG analysis • Post-wear reporting • Enabling QCT review | • ECG analysis • Post-wear reporting • Enabling QCT review • During wear arrhythmia detection and reporting |
| Interoperability | • Zio XT device • Zio monitor device • Zio AT device | • Zio XT device • Zio monitor device • Zio AT device |
| Rhythm detection algorithm | • ECGDLv2 (transmissions) • ECGDLv3 (post-wear) | • ECGDLv2 (transmissions and post-wear) • AutoTrigger Engine |
| Beat detection | ECGDL • QRS Complex • Supraventricular Beat (S Beat) • Ventricular Beat (V Beat) • Supraventricular Ectopic (SVE) Couplet • Supraventricular Ectopic (SVE) Triplet • Ventricular Ectopic (VE) Couplet • Ventricular Ectopic (VE) Triplet | • Identical |
| Rhythm detection | ECGDL • Pause ≥3 seconds • Ventricular fibrillation • Atrial fibrillation/flutter • Complete heart block • Second degree AV block-type II • Sinus rhythm • Supraventricular tachycardia • Ventricular bigeminy | • Identical |
Page 4 of 6
{9}
iRhythm Technologies, Inc.
Traditional 510(k): 510(k) Summary
ZEUS Platform: K252859
| Category | Subject Device | Predicate Device (K222389) |
| --- | --- | --- |
| | • Ventricular tachycardia • Ventricular trigeminy • Second degree AV block-type I (Wenckebach) • Ectopic atrial rhythm • Junctional rhythm • Idioventricular rhythm | |
# VII. Performance Data
Safety and performance of the subject ZEUS Platform has been evaluated and verified in accordance with design specifications and to support a determination of substantial equivalence to the predicate device.
The design verification and validation testing performed on the subject device demonstrates that the subject ZEUS Platform is in conformance with FDA-recognized consensus standards and FDA guidance documents as highlighted in Table 2.
Table 2: FDA-Recognized Consensus Standards & Guidance Document Summary
| FDA# | Body | Number/Version | Title |
| --- | --- | --- | --- |
| 5-125 | ANSI AAMI ISO | 14971: 2019 | Medical devices — Application of risk management to medical devices |
| 13-79 | ANSI AAMI IEC | 62304:2006/A1:2016 | Medical device software — Software lifecycle processes |
| 3-155 | ANSI AAMI IEC | 60601-2-47: 2012/(R)2016 | Medical electrical equipment — Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems |
| 3-118 | ANSI AAMI | EC57:2012 | Testing and Reporting Performance Results of Cardiac Rhythm and ST Segment Measurement Algorithms |
| 5-134 | ISO | 15223-1 Fourth edition 2021-07 | Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements |
| 5-129 | ANSI AAMI IEC | 62366-1: 2015+AMD1:2020 (Consolidated Text) | Medical devices Part 1: Application of usability engineering to medical devices, including Amendment 1 |
| 5-132 | IEC | 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
Page 5 of 6
{10}
iRhythm Technologies, Inc.
Traditional 510(k): 510(k) Summary
ZEUS Platform: K252859
| FDA# | Body | Number/Version | Title |
| --- | --- | --- | --- |
| N/A | FDA | Guidance for Industry and FDA Staff | The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (issued July 28, 2014) |
| N/A | FDA | Guidance for Industry and FDA Staff | Content of Premarket Submissions for Device Software Functions (issued June 14, 2023) |
| N/A | FDA | Guidance for Industry and FDA Staff | General Principles of Software Validation (issued January 11, 2002) |
| N/A | FDA | Guidance for Industry and FDA Staff | Multiple Function Device Products: Policy and Considerations (issued July 29, 2020) |
| N/A | FDA | Guidance for Industry and FDA Staff | Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (issued June 27, 2025) |
| N/A | FDA | Guidance for Industry and FDA Staff | Design Considerations and Premarket Submission Recommendations for Interoperable Medical Devices (issued September 6, 2017) |
The nonclinical verification and performance test results established that the subject device meets its design requirements and intended use, that the design differences with the predicate device do not raise new questions of safety and efficacy. During development, potential hazards were evaluated and controlled by risk management activities, including risk analysis, risk mitigation, verification and benefit-risk analysis. The verification and validation testing demonstrate that the device meets all predetermined specifications.
## VIII. Clinical Testing in Support of Substantial Equivalence Determination
No clinical testing was performed in support of this premarket notification.
## IX. Conclusion
The results confirm by evaluation and provision of objective evidence that the design outputs met the design input requirements. The results of the nonclinical testing performed demonstrate that the subject ZEUS Platform meets the requirements of established conformance standards and performance specifications necessary for its intended use and does not raise new questions of safety or effectiveness as compared to the predicate device. The subject ZEUS Platform is substantially equivalent to its predicate device.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.