Training: 80% of 2852 X-ray, 2073 CT, 209 MRI datasets. Development: 10% of datasets. Testing: 10% of datasets.
—
External validation dataset: 87 unique datasets.
—
Indications for Use
PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients’ imaging studies. Experience in usage and a clinical assessment are necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers. This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical specialties within surgery, and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population.
Device Story
PeekMed web is a cloud-based, multi-platform decision support software for orthopedic pre-operative planning. It ingests medical images (X-ray, CT, MRI) to perform 2D/3D anatomical reconstruction, bone segmentation, landmarking, and classification. The system allows surgeons to overlay digital templates of prosthetic materials onto patient images for surgical simulation. Operated by healthcare professionals in clinical environments, the device provides automated measurements and planning outputs that require clinician review and approval. The software assists in surgical planning, potentially improving efficiency and accuracy. It does not interact with the patient body and is not life-supporting.
Clinical Evidence
Bench testing only. ML models were trained/validated on 2852 X-rays, 2073 CTs, and 209 MRIs. Performance metrics met acceptance criteria: Segmentation (DICE ≥90%, HD-95 ≤8), Landmarking (MRE ≤7mm), Classification (Accuracy ≥90%), Detection (MAP ≥90%), and Reconstruction (DICE ≥90%). External validation used independent datasets to ensure generalizability across demographics, equipment, and institutions.
Technological Characteristics
Cloud-based software; distributed architecture. Functions: 2D/3D image processing, bone segmentation, landmarking, classification, detection, and 3D reconstruction. ML-based algorithms. Connectivity: Internet required. Platform: Web browser-based. No patient contact.
Indications for Use
Indicated for adult patients with previously diagnosed injury or disability requiring orthopedic pre-surgical planning in the musculoskeletal system (hip, knee, upper limb, foot).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
Peek Health, S.A.
Sara Silva
Chief Quality Officer (CQO)
Centro de Negócios Ideia Atlântico, Rua Padres Carmelitas
Braga, 4719-005
Portugal
Re: K252856
Trade/Device Name: PeekMed web
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: September 9, 2025
Received: November 12, 2025
Dear Sara Silva:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252856 - Sara Silva
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K252856 - Sara Silva
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D
Assistant Director
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K252856
Device Name
PeekMed web
Indications for Use (Describe)
PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients’ imaging studies. Experience in usage and a clinical assessment are necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.
This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical specialties within surgery, and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{5}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
K252856
PeekMed
# 510(k) summary
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
## 1. Submitter
Peek Health, S.A.
Centro de Negócios Ideia Atlântico
Rua Padres Carmelitas
4719-005 Braga
Portugal
Contact Person: Sara Silva
Chief Quality Officer (CQO)
Email: regulatory@peekmed.com
Office number: +351 253 128 941
Date Summary Prepared: December 18, 2025
## 2. Device
### 2.1 PeekMed web
Trade Name: PeekMed web
Common or Usual Name: Medical image management and processing system
Classification Name: System, Image Processing, Radiological
(21 C.F.R. § 892.2050)
Regulatory Class: Class II
Product Code: QIH, LLZ
510(k) Summary - PeekMed web
{6}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
# 3. Legally Marketed Predicate Device
## 3.1 PeekMed web
| 510(k) | Product Name | Clearance Date |
| --- | --- | --- |
| K251096 | PeekMed web | July 14, 2025 |
# 4. Device Description Summary
PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment are necessary for the proper use of the system in the revision and approval of the output of the planning.
The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.
As the PeekMed web is capable of representing medical images in a 2D or 3D environment, performing relevant measurements on those images, and also capable of adding templates, it can then provide a total overview of the surgery. Being software, it does not interact with any part of the body of the user and/or patient.
# 5. Intended Use/Indications for Use
PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment are necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers.
This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical specialties within surgery, and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population.
510(k) Summary - PeekMed web
{7}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
## 5.1. Contraindications
No contraindications specific to this device.
## 5.2 Indications for Use Comparison
There are NO differences between the indications for use of this device and its predicate.
## 6. Technological Comparison to Predicate
The PeekMed web was compared to its respective predicate device in intended use, indications for use, design, function, and technology, and it was demonstrated that they are substantially equivalent. Any technological differences within this 510(k), between the subject device and the predicate device, do not impact substantial equivalence, or safety and effectiveness.
The subject device and its predicate are both medical software that allow healthcare professionals to perform orthopedic pre-surgical planning efficiently in the musculoskeletal system of adults in a healthcare environment, therefore sharing the same intended use, intended user, and intended patient population. To properly and fully use both devices, clinical judgment and experience are mandatory.
Compared to the predicate device, the subject device includes updates to existing ML variants and the introduction of new ML variants, such as the segmentation and landmarking for the knee region. In addition, certain previously cleared variants have undergone updates to improve performance while maintaining the same intended use. These modifications do not alter the overall intended use of the device.
Both devices have the same workflows, use requirements (e.g., internet connection, output validation), and planning features (e.g., model representation, digital overlap of prosthetic material, possible 2D and 3D environments). Both devices generate a final report of the planning, which consists of the selected images with templates, measurements, and textual information describing the patient and/or the surgical procedure to be performed.
The technological characteristics of the PeekMed web are substantially equivalent to those of the predicate device PeekMed web, K251096.
510(k) Summary - PeekMed web
{8}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
Table 1: Summary of Predicate and Subject Device Characteristics to Demonstrate Substantial Equivalence
| Characteristic | PeekMed web
Predicate device K251096 | PeekMed web
Subject device | Substantially Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| Product Code | QIH, LLZ | QIH, LLZ | Yes | --- |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Yes | --- |
| Regulation Name | Medical Image Management And Processing System | Medical Image Management And Processing System | Yes | --- |
| Intended use/Indications for use | PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment are necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers. This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical | PeekMed web is a system designed to help healthcare professionals carry out pre-operative planning for several surgical procedures, based on their imported patients' imaging studies. Experience in usage and a clinical assessment are necessary for the proper use of the system in the revision and approval of the output of the planning. The multi-platform system works with a database of digital representations related to surgical materials supplied by their manufacturers. This medical device consists of a decision support tool for qualified healthcare professionals to quickly and efficiently perform the pre-operative planning for several surgical procedures, using medical imaging with the additional capability of planning the 2D or 3D environment. The system is designed for the medical | Yes | --- |
510(k) Summary - PeekMed web
{9}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
| Characteristic | PeekMed web
Predicate device K251096 | PeekMed web
Subject device | Substantially
Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| | specialties within surgery, and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population. | specialties within surgery, and no specific use environment is mandatory, whereas the typical use environment is a room with a computer. The patient target group is adult patients who have an injury or disability diagnosed previously. There are no other considerations for the intended patient population. | | |
| Contraindications | No contraindications specific to this device. | No contraindications specific to this device | Yes | --- |
| Clinical purpose | The PeekMed web allows the surgeon to perform orthopedic pre-surgical planning efficiently in the musculoskeletal system (e.g., Hip procedures, Knee procedures) | The PeekMed web allows the surgeon to perform orthopedic pre-surgical planning efficiently in the musculoskeletal system (e.g., Hip procedures, Knee procedures) | Yes | --- |
| Anatomical regions | The PeekMed web allows the surgeon to perform pre-surgical planning efficiently in the following anatomical regions:
- Hip
- Knee
- Upper limb
- Foot | The PeekMed web allows the surgeon to perform pre-surgical planning efficiently in the following anatomical regions:
- Hip
- Knee
- Upper limb
- Foot | Yes | --- |
| Patient Population | Adults | Adults | Yes | --- |
| End users | Healthcare Professionals | Healthcare Professionals | Yes | --- |
| Device availability | Software is cloud-based (not installable) and can be displayed on any personal device or workstation that can run a web browser | Software is cloud-based (not installable) and can be displayed on any personal device or workstation that can run a web browser | Yes | --- |
510(k) Summary - PeekMed web
Page 5
{10}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
| Characteristic | PeekMed web
Predicate device K251098 | PeekMed web
Subject device | Substantially
Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| Software Architecture | Distributed system (cloud-based). This distributed system is a combination of software modules placed on servers that are able to communicate with each other. | Distributed system (cloud-based). This distributed system is a combination of software modules placed on servers that are able to communicate with each other. | Yes | --- |
| Workflow | The workflow is as follows: Import case images, configure images, identify the case, pre-surgical planning, and export the case. | The workflow is as follows: Import case images, configure images, identify the case, pre-surgical planning, and export the case. | Yes | --- |
| Internet connection | Required | Required | Yes | --- |
| Images source | Receives medical images from various sources | Receives medical images from various sources | Yes | --- |
| Data processing | The software processes data to provide an overlap and dimensioning of digital representations of the prosthetic material | The software processes data to provide an overlap and dimensioning of digital representations of the prosthetic material | Yes | --- |
| Digital overlap of templates | Allows the overlap of models and the intersection of the models | Allows the overlap of models and the intersection of the models | Yes | --- |
| Interactive model positioning | Yes | Yes | Yes | --- |
| Interactive model dimensioning | Yes | Yes | Yes | --- |
| Model rotation | Yes | Yes | Yes | --- |
| Support for digital prosthetic materials provided | Yes | Yes | Yes | --- |
510(k) Summary - PeekMed web
Page 6
{11}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
| Characteristic | PeekMed web
Predicate device K251098 | PeekMed web
Subject device | Substantially
Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| by the manufacturers | | | | |
| Pre-surgical planning | Yes | Yes | Yes | --- |
| Type of pre-surgical planning | Automatic or Manual | Automatic or Manual | Yes | --- |
| Contact with the patient | No | No | Yes | --- |
| Control of life supporting devices | No | No | Yes | --- |
| Human intervention for image interpretation | Yes | Yes | Yes | --- |
| Ability to add additional modules when available | Yes | Yes | Yes | --- |
| Automatic bone segmentation | Yes
- Hip (X-ray and CT scan)
- Knee (X-ray, CT scan, and MRI)
- Upper limb (CT scan)
- Foot (X-ray and CT scan) | Yes
- Hip (X-ray and CT scan)
- Knee (X-ray - AP & LAT view, CT scan, and MRI)
- Upper limb (CT scan)
- Foot (X-ray and CT scan) | Yes | The subject device includes new ML model variants for the segmentation of the Knee in a Lateral view.
Both devices allow planning for the knee region with X-rays in an AP view, but the subject device also offers a new image perspective (Lateral view). |
510(k) Summary - PeekMed web
Page 7
{12}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
| Characteristic | PeekMed web
Predicate device K251098 | PeekMed web
Subject device | Substantially
Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| | | | | This does not constitute an intended purpose update, nor does it raise questions of safety and performance, since the development, verification, validation, and deployment processes are the same for both devices.
Compared to the predicate device, the subject device includes updates to existing ML variants and the introduction of new ML variants, such as the segmentation and landmarking for the knee region. In addition, certain previously cleared variants have undergone updates to improve performance while maintaining the same intended use. These modifications do not alter the overall intended use of the device |
| Type of
landmarking | Automatic or Manual
- Hip (X-ray and CT scan)
- Knee (X-ray, CT scan, and MRI) | Automatic or Manual
- Hip (X-ray and CT scan)
- Knee (X-ray - AP & LAT view, CT scan, | Yes | The subject device includes new ML model variants for landmarking for the Knee in a Lateral view. |
510(k) Summary - PeekMed web
Page 8
{13}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
| Characteristic | PeekMed web
Predicate device K251098 | PeekMed web
Subject device | Substantially
Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| | - Upper limb (CT scan)
- Foot (X-ray and CT scan) | and MRI)
- Upper limb (CT scan)
- Foot (X-ray and CT scan) | | Both devices allow planning for the knee region with X-rays in an AP view, but the subject device also offers a new image perspective (Lateral view). This does not constitute an intended purpose update, nor does it raise questions of safety and performance, since the development, verification, validation, and deployment processes are the same for both devices.
Compared to the predicate device, the subject device includes updates to existing ML variants and the introduction of new ML variants, such as the segmentation and landmarking for the knee region. In addition, certain previously cleared variants have undergone updates to improve performance while maintaining the same intended use. These modifications do not alter the overall intended use of the device |
510(k) Summary - PeekMed web
Page 9
{14}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
| Characteristic | PeekMed web
Predicate device K251098 | PeekMed web
Subject device | Substantially
Equivalent? | Justification and rationale |
| --- | --- | --- | --- | --- |
| Bone reconstruction | Yes – Reconstruction of 3D anatomical representation from 3D imaging modalities (e.g., CT, MRI). | Yes
Reconstruction of 3D anatomical representation from 3D imaging modalities (e.g., CT, MRI) and additionally from 2D imaging modalities (X-ray). | Yes | Both devices provide reconstruction capabilities for generating an anatomical representation. The subject device extends the input modalities to include 2D X-rays, enabling reconstruction of 3D models where only planar images are available. This does not alter the intended use or indications for use, as the function remains within pre-surgical planning for the same anatomical regions. The reconstructed models can be integrated into planning and analysis workflows. The capability is decision-support only, with outputs subject to review and approval by healthcare professionals. The development, verification, validation, and deployment processes follow the same methodology, and no new safety or effectiveness concerns are raised. |
PeekMed web shares the same intended use, indications for use, end users, and has the patient population, overall technical and functional capabilities, and therefore is substantially equivalent to the predicate device. The subject device has the same design and function as the predicate device for the modes of operation and use. The subject device includes new ML model variants for segmentation and landmarking for the Knee in a Lateral view, a new ML model type (Reconstruction), and its ML model variants.
510(k) Summary - PeekMed web
Page 10
{15}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
# 7. Performance Data
Nonclinical performance testing performed on the subject device, PeekMed web, supports substantial equivalence to the predicate device. The following testing was performed on the subject device:
A. Verification activities were conducted to ensure that features were implemented following the requirements and covering the acceptance criteria.
B. ML models incorporated into the PeekMed web were also developed, trained, tested, and externally validated for their performance according to the internal procedures.
- A dedicated validation dataset containing different data from the ML development dataset was used. Specifically, the validation dataset was not a sampling of the development dataset, had never been used for the algorithm training or for tuning the algorithm, and leakage between development and validation data sets did not occur.
# Training, development, testing and external validation data information
ML models were developed with datasets from multiple sites in a total of 2852 X-ray datasets, 2073 CT scans, and 209 MRIs. We trained the ML models with 80% of the dataset, developed with 10%, and tested with the remaining 10%. External validation is performed by sample size with a total unique dataset, for Segmentation ML model: 672; Landmarking ML model: 561; Classification ML model: 367; and Detection ML model: 198; and Reconstruction ML model: 87. This comprehensive dataset was designed to cover the intended use population while ensuring a variety of data, maintaining diverse patient characteristics.
# Subgroup definition (generalizability)
Datasets were divided according to the subgroups listed below:
- Demographics
- Patient Sex
- Patient Age
- Equipment and Protocols for Image Collection
- Institution Name
- Manufacturer
- Manufacturer Model Name
510(k) Summary - PeekMed web
{16}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
# Acceptance criteria
The acceptance criteria for each ML model are shown in the following table:
| ML model | Acceptance Criteria |
| --- | --- |
| Segmentation | DICE is no less than 90%
HD-95 is no more than 8
STD DICE is between +/- 10%
Precision is more than 85%
Recall is more than 90% |
| Landmarking | MRE is no more than 7mm
STD MRE is between +/- 5mm |
| Classification | Accuracy is no less than 90%.
Precision is no less than 85%
Recall is no less than 90%
F1 score is no less than 90% |
| Detection | MAP is no less than 90%.
Precision is no less than 85%
Recall is no less than 90% |
| Reconstruction | DICE is no less than 90%
HD-95 is no more than 8
STD DICE is between +/- 10%
Precision is more than 85%
Recall is more than 90% |
# Reference Standard ("Truthing" Process)
Comparison of the efficacy results of the Segmentation ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.
Comparison of the efficacy results of the Landmarking ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.
510(k) Summary - PeekMed web
{17}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
Comparison of the efficacy results of the Classification ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.
Comparison of the efficacy results of the Detection ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance, demonstrating the substantial equivalence of the subject device to its predicate.
Comparison of the efficacy results of the Reconstruction ML model using the testing and external validation datasets against the predefined ground truth met the acceptance criteria for ML model performance.
# Independence of Training and External Validation Data
External validation datasets were collected independently of the development data to prevent bias, ensuring the reliability of the results. For the external validation, a fully independent dataset, labeled by a separate team, was employed to provide an accurate assessment of the model's performance, over the whole population and for each sub-group mentioned before, to prove that it generalizes well to unseen, real-world data. All the testing and external validation performed indicate acceptable performance of the ML models for their intended population.
C. Validation tests were performed internally before the release to the market by qualified personnel, in an environment simulating the real end-user environment. This validation follows a pre-defined test script document, according to the tests defined in the test cases management software.
Nonclinical performance testing allowed us to understand that there are no related problems in the subject device. Furthermore, these tests will be repeated and updated when appropriate to ensure that the software is always properly validated, making it possible to understand in which version the problems arise and in which they are solved. Consequently, any problem that may appear in a given PeekMed web version will be identified and can be solved in subsequent versions, as all steps are
510(k) Summary - PeekMed web
Page 13
{18}
Peek Health, S.A.
Traditional 510(k) Premarket Submission
PeekMed web
PeekMed
traceable. All anatomical areas were tested, as well as other main areas of the software, such as the planning final report, and saved planning, ML models, among others.
After these successful validation tests, it is possible to deem the subject device, PeekMed web, as substantially equivalent to its predicate device.
# 9. Conclusion
Based on the information provided in this 510(k) submission, it was determined that the subject device, PeekMed web, is substantially equivalent to the legally marketed predicate device concerning indications for use, intended use, design, technology, and performance.
510(k) Summary - PeekMed web
Page 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.