K252844 · Tyto Care , Ltd. · PHZ · Jun 2, 2026 · Anesthesiology
Device Facts
Record ID
K252844
Device Name
Tyto Insights for Wheeze Detection (with PCCP)
Applicant
Tyto Care , Ltd.
Product Code
PHZ · Anesthesiology
Decision Date
Jun 2, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1900
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, PCCP, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K252844 · Jun 2, 2026
Tyto Insights for Wheeze Detection (with PCCP)
Tyto Care , Ltd.
Retrospective datasets of lung sound recordings acquired via real-world use of the Tyto Stethoscope
The PCCP requires retrospective analysis of real-world clinical recordings to validate future algorithm enhancements, technical performance modifications, and device input expansions.
PCCP; Retrospective study; Real-world use data; Algorithm validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective validation of algorithm modifications; Retrospective study; Follow-up/Duration: Not applicable; Study Period: Future (post-clearance)
Adults and pediatrics (2 years and older) representative of the intended patient population; Sample Size: Not specified (to be determined per future modification)
Most recently validated and released version of the device
The Tyto Insights for Wheeze Detection (with PCCP) is an over-the-counter artificial intelligence (AI) enabled decision support software system used in the evaluation of lung sounds in adults and pediatrics (2 years and older). It automatically analyzes the acoustic signal of the lung as recorded by the FDA cleared compatible Tyto Stethoscope and identifies recordings where a specific abnormal lung sound suggestive of “Wheeze” is suspected. It is not intended to detect other abnormal or normal lung sounds. A licensed health care professional’s advice is required to understand the meaning of the Tyto Insights for Wheeze Detection result. Healthcare providers should consider the device result in conjunction with recording and other relevant patient data.
Device Story
Tyto Insights for Wheeze Detection is a cloud-hosted, AI-enabled decision support software. It receives lung sound audio files (6-12 seconds) recorded by an FDA-cleared Tyto Stethoscope (K181612). The system comprises an Application Server, Algorithm Server, and Web Server. The AI algorithm, based on a neural network, analyzes the acoustic signal to detect wheeze. Results (positive/negative/unable to analyze) are transmitted via API to a telehealth server, then displayed on clinician and patient interfaces. Used in clinical or home environments by professional or lay users (18-65 years). The output assists healthcare providers in clinical assessment; providers must interpret results alongside other patient data. The device includes a Predetermined Change Control Plan (PCCP) for future algorithm updates, performance enhancements, and input device expansion.
Clinical Evidence
Clinical validation previously conducted for the predicate device (K232237) remains applicable. Performance metrics: AUC 0.96 (95% CI: 0.95–0.98), Sensitivity 0.55 (95% CI: 0.43–0.65), Specificity 0.99 (95% CI: 0.99–1.00). No new clinical data was required for this submission as the minor software update did not impact performance.
Technological Characteristics
Cloud-hosted software system; neural network-based AI algorithm; processes 6-12 second audio segments from compatible electronic stethoscopes. Connectivity via IP network. Standards: ISO 14971 (risk management), IEC 62304 (software lifecycle), IEC 62366-1 (usability).
Indications for Use
Indicated for adults and pediatrics (2 years and older) for the evaluation of lung sounds to identify suspected wheeze. Not intended to detect other abnormal or normal lung sounds. Requires professional interpretation.
Regulatory Classification
Identification
A diagnostic pulmonary-function interpretation calculator is a device that interprets pulmonary study data to determine clinical significance of pulmonary-function values.
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 2, 2026
Tyto Care Ltd.
Stella Raizelman Perry
RA&QA Director
14 Beni Gaon Street
Netanya,
Israel
Re: K252844
Trade/Device Name: Tyto Insights for Wheeze Detection (with PCCP)
Regulation Number: 21 CFR 868.1900
Regulation Name: Diagnostic Pulmonary-Function Interpretation Calculator
Regulatory Class: Class II
Product Code: PHZ
Dated: April 29, 2026
Received: April 29, 2026
Dear Stella Raizelman Perry:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K252844 - Stella Raizelman Perry
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{3}
K252844 - Stella Raizelman Perry
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Rachana Visaria
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K252844
Device Name
Tyto Insights for Wheeze Detection (with PCCP)
Indications for Use (Describe)
The Tyto Insights for Wheeze Detection (with PCCP) is an over-the-counter artificial intelligence (AI) enabled decision support software system used in the evaluation of lung sounds in adults and pediatrics (2 years and older). It automatically analyzes the acoustic signal of the lung as recorded by the FDA cleared compatible Tyto Stethoscope and identifies recordings where a specific abnormal lung sound suggestive of "Wheeze" is suspected. It is not intended to detect other abnormal or normal lung sounds. A licensed health care professional's advice is required to understand the meaning of the Tyto Insights for Wheeze Detection result. Healthcare providers should consider the device result in conjunction with recording and other relevant patient data.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}

## 510(k) Summary
| **Submission Number** | K252844 |
| --- | --- |
| **Submitter Name and Address:** | Tyto Care Ltd. 14 Beni Gaon Street Netanya, Israel, 4250803 |
| **Contact Person:** | Stella Raizelman Perry RA & QA Director Email: stellar@tytocare.com Phone Number: +972 72-2210750 Fax Number: +972 72-2210752 |
| **Establishment Registration Number:** | 3012678246 |
| **Date Prepared:** | April 28, 2026 |
| **Device Trade Name(s):** | Tyto Insights for Wheeze Detection (with PCCP) |
| **Device Common Name:** | Tyto Insights for Wheeze Detection (with PCCP) |
| **Classification:** | **Name:** Diagnostic pulmonary-function interpretation calculator **Product code:** PHZ **Regulation No:** 21 CFR 868.1900 **Class:** II **Panel:** Anesthesiology |
### **Predicate Device(s):**
| **Device name** | **510(k) No.** | **Date of Clearance** |
| --- | --- | --- |
| Tyto Insights for Wheeze Detection | K232237 | December 13, 2023 |
### **Reference Device(s):**
| **Device name** | **510(k) No.** | **Date of Clearance** |
| --- | --- | --- |
| Tyto Insights for Rhonchi Detection | K243567 | April 07, 2025 |
The reference device is cited to demonstrate consistency of Predetermined Change Control Plan (PCCP) elements with the subject device.
1
{6}
tytocare™
# Intended use / indication for use statement
The Tyto Insights for Wheeze Detection (with PCCP) is an over-the-counter artificial intelligence (AI) enabled decision support software system used in the evaluation of lung sounds in adults and pediatrics (2 years and older). It automatically analyzes the acoustic signal of the lung as recorded by the FDA cleared compatible Tyto Stethoscope and identifies recordings where a specific abnormal lung sound suggestive of “Wheeze” is suspected. It is not intended to detect other abnormal or normal lung sounds. A licensed health care professional’s advice is required to understand the meaning of the Tyto Insights for Wheeze Detection result. Healthcare providers should consider the device result in conjunction with recording and other relevant patient data.
# Device description
The Tyto Insights for Wheeze Detection is a web-based (AI) enabled software system designed to aid in the clinical assessment of lungs auscultation sound data by analyzing recorded lung sounds to determine whether a Wheeze is detected within the recorded sound data.
The Tyto Insights for Wheeze Detection Software is intended to process recordings from the FDA-cleared compatible Tyto Stethoscope (Tyto Stethoscope, K181612). The acquisition of the acoustic data (recordings) is carried out by a professional user in a clinical environment or by a lay- user in a non-medical environment, in compliance with the labeling of the Tyto Stethoscope. The system is composed of the following sub-systems:
1. The Tyto Insights for Wheeze Detection Application Server (APS) communicates with the Tyto Insights for Wheeze Detection Algorithm Server (ALS) and implements an application programming interface (API) for communication with the telehealth server.
2. The Tyto Insights for Wheeze Detection Algorithm Server (ALS) receives an audio file as input and returns an analysis result of positive or negative regarding whether a wheeze was detected as output.
3. The Tyto Insights for Wheeze Detection Web Server (WBS) provides a graphic indication whether a wheeze is detected in the recording. It can be utilized both in patient and clinician side.
All the software subsystems (servers and storage) are hosted in the cloud and communicate through IP network.
2
{7}
[LOGO]
The recordings from the compatible Tyto Stethoscope (K181612) are sent by the third-party point of care app to the clinician app through the telehealth server. The telehealth server sends the set of lung sound recordings to the Tyto Insights for Wheeze Detection web server using its dedicated API. The AI enabled algorithm runs automatically and returns a response for each audio file with the indication of wheeze to the telehealth server which sends a response to both the clinician side and the patient side.
3
{8}

### Substantial Equivalence to Predicate Devices
The following table compares Tyto Insights for Wheeze Detection (with PCCP) to the predicate and reference devices.
**Table 1.** Substantial Equivalence Summary
| | Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| **Device Name** | Tyto Insights for Wheeze Detection (with PCCP) | Tyto Insights for Wheeze Detection | Tyto Insights for Rhonchi Detection | NA |
| **Device Manufacturer** | Tyto Care Ltd. | Tyto Care Ltd. | Tyto Care Ltd. | Same |
| **510(k) Number** | K252844 | K232237 | K243567 | NA |
| **Device Class** | Class II | Class II | Class II | Same |
| **Review Panel** | Anesthesiology | Anesthesiology | Anesthesiology | Same |
| **Product code** | PHZ | PHZ | PHZ | Same |
| **Regulation number** | 21 CFR 868.1900 | 21 CFR 868.1900 | 21 CFR 868.1900 | Same |
| **Device Classification Name** | Abnormal breath sound device | Abnormal breath sound device | Abnormal breath sound device | Same |
4
{9}
# tytocare™
| | Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| Intended use and indication for use | The “Tyto Insights for Wheeze Detection” is an over-the-counter artificial intelligence (AI) enabled decision support software system used in the evaluation of lung sounds in adults and pediatrics (2 years and older). It automatically analyzes the acoustic signal of the lung as recorded by the FDA cleared compatible Tyto Stethoscope and identifies recordings where a specific abnormal lung sound suggestive of “Wheeze” is suspected. It is not intended to detect other abnormal or normal lung sounds. A licensed health care professional’s advice is required to understand the meaning of the Tyto Insights for Wheeze Detection result. Healthcare providers should consider the device result in conjunction with recording and other relevant patient data. | The “Tyto Insights for Wheeze Detection” is an over-the-counter artificial intelligence (AI) enabled decision support software system used in the evaluation of lung sounds in adults and pediatrics (2 years and older). It automatically analyzes the acoustic signal of the lung as recorded by the FDA cleared compatible Tyto Stethoscope and identifies recordings where a specific abnormal lung sound suggestive of “Wheeze” is suspected. It is not intended to detect other abnormal or normal lung sounds. A licensed health care professional’s advice is required to understand the meaning of the Tyto Insights for Wheeze Detection result. Healthcare providers should consider the device result in conjunction with recording and other relevant patient data. | The Tyto Insights for Rhonchi Detection is a prescription-use artificial intelligence (AI) enabled decision support software system used in the evaluation of lung sounds in adults and pediatrics (2 years and older). It automatically analyzes the acoustic signal of the lung as recorded by the FDA 510k cleared compatible Tyto Stethoscope and identifies recordings where a specific abnormal lung sound suggestive of “Rhonchi” is suspected. It is not intended to detect other abnormal or normal lung sounds. A licensed health care professional’s advice is required to understand the meaning of the Tyto Insights for Rhonchi Detection result. Healthcare providers should consider the device result in conjunction with recording | Same as the predicate device. |
5
{10}
# tytocare™
| | Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | | | and other relevant patient data. | |
| Type of use | Over-The-Counter Use | Over-The-Counter Use | Prescription use | Same as the predicate device |
| Intended users | Professional users and lay users (18-65 years old). | Professional users and lay users (18-65 years old). | Professional users and lay users (18-65 years old). | Same |
| Intended patient population | Patients of 2 years and older | Patients of 2 years and older | Patients of 2 years and older | Same |
| Intended environment | Non-clinical (home) and clinical | Non-clinical (home) and clinical | Non-clinical (home) and clinical | Same |
| Input | Lung sounds recorded by compatible Tyto Stethoscope (K181612) | Lung sounds recorded by compatible Tyto Stethoscope (K181612) | Lung sounds recorded by compatible Tyto Stethoscope (K181612) | Same |
| Signal length | Recordings are processed in segments of up to 12 seconds while signals shorter than 6 seconds will not be processed. | Recordings are processed in segments of up to 12 seconds while signals shorter than 6 seconds will not be Processed. | Recordings are processed in segments of up to 12 seconds while signals shorter than 6 seconds will not be Processed. | Same |
| Algorithm | Machine learning based (Neural network) | Machine learning based (Neural network) | Machine learning based (Neural network) | Same as the predicate device. |
| Output | • Positive (wheeze suspected), • Negative (Wheeze not suspected), • The Tyto Insights for Wheeze Detection was | • Positive (wheeze suspected), • Negative (Wheeze not suspected), • The Tyto Insights for Wheeze Detection was | • Positive (Rhonchi suspected), • Negative (Rhonchi not suspected), • The Tyto Insights for Rhonchi Detection was | Same as the predicate device. |
6
{11}
# tytocare™
| | Device | Predicate Device | Reference Device | Summary |
| --- | --- | --- | --- | --- |
| | not able to analyze the recording | not able to analyze the recording | not able to analyze the recording | |
| **Performance** | AUC :0.96 (0.95– 0.98) Estimate two-sided 95% CI): Sensitivity: 0.55 [0.43 – 0.65] Specificity: 0.99 [0.99 – 1.00] | AUC :0.96 (0.95– 0.98) Estimate two-sided 95% CI): Sensitivity: 0.55 [0.43 – 0.65] Specificity: 0.99 [0.99 – 1.00] | AUC: 0.96 (0.92–0.98) Estimate two-sided 95% CI): Sensitivity: 0.60 (0.50-0.69) Specificity: 0.99 (0.97 - 1.00) | Same as the predicate. |
| **User interface** | Web view | Web view | Web view | Same |
| **Predetermined Change Control plan (PCCP)** | Includes PCCP (submitted for FDA review in this 510(k)) | Not included | Includes FDA authorized PCCP | The predicate device didn’t include PCCP. The types of modifications, data management practices, re-training practices, performance evaluation methods, and update procedures for the subject device are consistent with those of the reference device. |
7
{12}
tytocare™
The intended use of the subject device is identical to that of the predicate device. The purpose of this 510(k) submission is to obtain FDA’s review and authorization of a Predetermined Change Control Plan (PCCP) for the subject device. The only additional difference from the predicate device is a minor software update in which third-party libraries were updated to address known vulnerabilities, consistent with Tyto Care’s software development lifecycle and cybersecurity procedures. The subject device maintains the same technological characteristics as the predicate device. This minor software update does not alter these characteristics and does not raise new questions of safety or effectiveness.
Standards Conformance:
- ANSI AAMI ISO 14971:2019, Medical devices - Application of risk management to medical devices
- ANSI AAMI IEC 62304:2006/A1:2016, Medical device software - Software life cycle processes
- ISO 15223-1 Fourth edition 2021-07, Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements.
- ANSI AAMI IEC 62366-1:2015+AMD1:2020 (Consolidated Text) Medical devices Part 1: Application of usability engineering to medical device.
Performance evaluation:
- SW verification and validation - The software including both custom developed software and OTS software, have been verified and validated and have been demonstrated to be safe and effective for its intended use. The software documentation level is basic per Content of Premarket Submissions for Device Software Functions, Guidance for Industry and Food and Drug Administration Staff, dated June 14, 2023. All required items related to software as required by FDA guidance for Basic Documentation Level have been included in this submission.
- Cybersecurity- all the applicable information to reflect effective cybersecurity management and to address the FDA’s recommendations described in Cybersecurity in Medical Devices: Refuse to Accept Policy for Cyber Devices and Related Systems Under Section 524B of the FD&C Act, issue date March 2023 , Cybersecurity in
8
{13}
tytocare™
Medical Devices: Quality System Considerations and Content of Premarket Submissions, issue date February 2026 and in the other FDA's applicable policies have been included in this submission.
- Performance evaluation - The minor software update did not impact device performance. The clinical validation previously conducted for the predicate device remains applicable to the subject device. Accordingly, no additional performance validation was required.
- Human factors validation - The minor software update did not introduce any modifications to the user interface, did not create new critical tasks, and did not affect existing critical tasks. Therefore, no additional human factors validation was necessary. The evaluations conducted for the predicate device remain fully applicable to the subject device.
### Predetermined Change Control Plan (PCCP)
Predetermined Change Control Plan (PCCP) that specifies anticipated modifications to the Tyto Insights for Wheeze detection device is included in this submission. The same type of software modifications, data management practices, re-training practices, performance evaluation methods, and update procedures to establish substantial equivalence as were applied for the Tyto Insights for Rhonchi Detection (K243567), the reference device, were followed.
The table below includes a description of the software modifications that can be made to the algorithms in the device subject to the authorized PCCP as well as a description of the test methods that will be used to support substantial equivalence determination.
| Detailed list of changes | Test Methods | Acceptance criteria to establish substantial equivalence |
| --- | --- | --- |
| 1. Modifications related to performance specifications enhancement: Re-training of the ML model with additional data to improve the device clinical performance, as measured by sensitivity and specificity, compared to the most recently validated and released version of the device, while maintaining the same type and range of input signal. | • Software verification and validation • Clinical performance Validation. Retrospective study on dataset of the compatible Tyto Stethoscope recordings that are representative of the intended recorded patient population (data will be acquired by the real-world use of the Tyto Stethoscope). | • Software verification and validation meet the requirements • Clinical performance Validation: **Stand alone:** Accuracy of device – Sensitivity, Specificity, PPV, NPV **Co-Primary endpoints:** • Sensitivity of the modified device calculated on the new validation dataset. Specificity of the modified device calculated on the new validation dataset. |
9
{14}
| | | **Success criteria:** The modification will be considered successful only if the lower bound of the 95% confidence interval for both sensitivity and specificity is not lower than the corresponding lower confidence bounds of the most recently validated and released version of the device **Secondary Endpoint** - The PPV and NPV of the newly trained algorithm compared to the PPV and the NPV of the most recently validated and released version of the device - Performance of the modified device (including sensitivity and specificity) will also be descriptively evaluated across the following clinically relevant subgroups: age group (2–18 years vs. >18 years), gender, and user type (lay user vs. healthcare professional) and stethoscope model (e.g., due to prior modifications). |
| --- | --- | --- |
| 2. Modifications related to technical performance specifications. Modification of data preprocessing methodologies /data augmentation methodologies and hyper-parameters tuning to improve the technical performance characteristics (including running time, memory consumption and CPU utilization). These changes are implemented without modifications to the core model architecture, intended use, or output of the device. | - Software verification and validation - Verification testing will be conducted for the improved computational parameters - Clinical performance Validation. Retrospective study on dataset of the compatible Tyto Stethoscope recordings that are representative of the intended recorded patient population (data will be acquired by the real-world use of the Tyto Stethoscope). | - Software verification and validation meet the requirements - Verification testing meets the requirements - Clinical performance Validation: **Stand-alone:** Accuracy of device – Sensitivity, Specificity, PPV, NPV **Co-Primary endpoints:** - Sensitivity of the modified device calculated on the new validation dataset. Specificity of the modified device calculated on the new validation dataset. **Success criteria:** The modification will be considered successful only if the lower bound of the 95% confidence interval for both sensitivity and specificity is not lower than the |
10
{15}
tytocare™
| | | corresponding lower confidence bounds of the most recently validated and released version of the device. Meeting both endpoints is required for the modification to be declared successfully. **Secondary Endpoint** - The PPV and NPV of the modified algorithm will be compared to the PPV and the NPV of the most recently validated and released version of the device. - Performance of the modified device will be descriptively evaluated across the following clinically relevant subgroups: age group (2–18 years vs. >18 years), gender, and user type (lay user vs. healthcare professional) and stethoscope model (e.g., due to prior modifications). |
| --- | --- | --- |
| 3. Modifications related to device input expansion Expanding the algorithm to include new sources of the same signal type (different model of FDA compatible Stethoscope with equivalent audio acquisition specifications). Modification is limited to expanding to electronic stethoscopes that have FDA 510k clearance (for over-the-counter use) at the time that the proposed modification is made. | - Software verification and validation - Clinical performance Validation. Retrospective study on dataset of the compatible Stethoscope recordings that are representative of the intended recorded patient population (data will be acquired by the real-world use of the applicable Stethoscope models). | - Software verification and validation meet the requirements - Clinical performance Validation: **Stand-alone:** Accuracy of device – Sensitivity, Specificity, PPV, NPV Co-Primary endpoints: - Sensitivity of the modified device calculated on the new validation dataset. - Specificity of the modified device calculated on the new validation dataset. **Success Criteria:** The modification will be considered successful only if the lower bound of the 95% confidence interval for both sensitivity and specificity is not lower than the corresponding lower confidence bounds of the most recently validated and released version of the device. Meeting both endpoints is required for the modification to |
11
{16}
tytocare™
| | | be declared successfully. **Secondary Endpoint** - The PPV and NPV of the newly trained algorithm compared to the most recently validated and released version of the device calculated on new validation dataset. - Performance of the modified device (including sensitivity and specificity) will also be descriptively evaluated across the following clinically relevant subgroups: age group (2–18 years vs. >18 years), gender, user type (lay user vs. healthcare professional), and input device. Subgroup analysis by input device will include performance evaluation for each compatible stethoscope model. |
| --- | --- | --- |
## Conclusion
Tyto Insights for Wheeze Detection (with PCCP) has the same intended use, indication for use and technological characteristics as the predicate device.
The proposed modifications under the PCCP and the minor software differences do not raise new questions of safety and effectiveness. Thus, we conclude that the Tyto Insights for Wheeze Detection (with PCCP) is substantially equivalent, i.e., as safe and effective as the predicate device.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.