eXaSkin is a medical device intended to be used as a bolus for the treatment of neoplastic skin lesions through radiotherapy. The product is designed to be applied to the irradiated area by qualified healthcare professionals to address anatomical irregularities and ensure precise delivery of the prescribed dose to the skin and underlying tissues during a radiation therapy treatment. eXaSkin was evaluated using 6 MV photons but has not been assessed for use with 10 MV photons, protons, or orthovoltage X-rays. eXaSkin Plus 150 and 225 is intended for general radiotherapy use. eXaSkin Plus 650 is intended for use on breast tissue.
Device Story
eXaSkin Plus is a moldable, two-component putty bolus used in radiotherapy to address anatomical irregularities and ensure precise dose delivery to skin and underlying tissues. Applied by healthcare professionals to the patient's skin, the putty hardens in approximately 3 minutes to conform to specific contours, including scalp, hands, neck, ears, and facial areas. It is non-reflective, enabling Surface Guided Radiotherapy (SGRT) without requiring additional optical surface imaging (OSI) of the bolus itself. The device eliminates build-up, does not shrink or break, and provides tissue-equivalent radiation attenuation. It benefits patients by improving surface dose accuracy and simplifying the clinical workflow for patient positioning and bolus alignment.
Clinical Evidence
No clinical or animal data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10, -23), physical property testing (ISO 4823), and dosimetric validation (6 MV photon beam attenuation, dose distribution accuracy within 3mm/3% agreement).
Technological Characteristics
Two-component moldable putty (putty and catalyzer). Hardens to Shore A 20-77 depending on time/component. Tissue-equivalent material. Non-reflective surface for SGRT compatibility. Biocompatible (ISO 10993). Manual application; no energy source or software. Dimensions customizable to patient anatomy.
Indications for Use
Indicated for patients undergoing radiotherapy for neoplastic skin lesions. eXaSkin Plus 150/225 for general use; eXaSkin Plus 650 for breast tissue. Contraindicated for use on open wounds. Evaluated for 6 MV photons; not assessed for 10 MV photons, protons, or orthovoltage X-rays.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 22, 2026
Anatomical Geometry S.L.
% Tyler Ting
Quality Director
Rook Quality Systems
1155 Mount Vernon Hwy
Suite 800
Dunwoody, Georgia 30338
Re: K252784
Trade/Device Name: eXaSkin Plus
Regulation Number: 21 CFR 892.5050
Regulation Name: Medical Charged-Particle Radiation Therapy System
Regulatory Class: Class II
Product Code: MUJ, IXI
Dated: April 23, 2026
Received: April 23, 2026
Dear Tyler Ting:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252784 – Tyler Ting
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K252784 – Tyler Ting
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lora D. Weidner, Ph.D.
Assistant Director
Radiation Therapy Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252784 | |
| Device Name eXaSkin Plus | |
| Indications for Use (Describe) eXaSkin is a medical device intended to be used as a bolus for the treatment of neoplastic skin lesions through radiotherapy. The product is designed to be applied to the irradiated area by qualified healthcare professionals to address anatomical irregularities and ensure precise delivery of the prescribed dose to the skin and underlying tissues during a radiation therapy treatment. eXaSkin was evaluated using 6 MV photons but has not been assessed for use with 10 MV photons, protons, or orthovoltage X-rays. eXaSkin Plus 150 and 225 is intended for general radiotherapy use. eXaSkin Plus 650 is intended for use on breast tissue. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary - K252784
Page 1 of 8
# Submitter
Anatomical Geometry S.L.
Parque Industrial y de Servicios Aljarafe. C/ Formación N°6,
CP 41927. Mairena del Aljarafe, Sevilla
Contact Person: Eduardo Suarez
Phone: +34 610 075 760
Email: esuarez@anatge.com
Date Prepared: 06 MAY 2025
# Device
Trade Name: eXaSkin Plus
Common Name: Radiotherapy Bolus
Classification Name: System, Planning, Radiation Therapy Treatment
Regulatory Class: II
Product Code(s): MUJ, IXI
Regulation Number: 892.5050, 892.5710
Classification Panel: Radiology
# Predicate Device
AnatGe submits the following information to demonstrate that the subject device eXaSkin Plus, is substantially equivalent to the following legally marketed predicate device:
| 510(k) Number | Device Name | Product Code |
| --- | --- | --- |
| K211511 | .decimal Bolus | MUJ, IXI |
# Reference Device
AnatGe submits the following information for the reference device utilized in the submission of the subject device:
| 510(k) Number | Device Name | Product Code |
| --- | --- | --- |
| K812893 | SUPERFLAB | IYP |
# Device Description
The eXaSkin Plus is a highly-adaptable bolus in a paste form that allows for a customizable bolus in 3 minutes. The eXaSkin Plus allows for precision dosage on the skin during photon beam therapy in clinical situations and is usable and specially recommended for highly-critical regions such as the scalp, hands,
{5}
510(k) Summary - K252784
neck, outer ear, facial contours, and genitals. The eXaSkin Plus eliminates build-up with less bolus thickness, does not dry, shrink, or break, and does not require complex 3D systems to achieve expected results.
The device enables the use of Surface Guided Radiotherapy (SGRT) technology. Unlike other SGRT-compatible boluses, eXaSkin is non-reflective and can be consistently repositioned at the same location on the patient's surface for each treatment session. As a result, it does not require an additional optical surface imaging (OSI) acquisition of the bolus itself. This allows for simultaneous patient positioning and bolus alignment in a single OSI step.
# Indications for Use
eXaSkin is a medical device intended to be used as a bolus for the treatment of neoplastic skin lesions through radiotherapy. The product is designed to be applied to the irradiated area by qualified healthcare professionals to address anatomical irregularities and ensure precise delivery of the prescribed dose to the skin and underlying tissues during a radiation therapy treatment.
eXaSkin was evaluated using 6 MV photons but has not been assessed for use with 10 MV photons, protons, or orthovoltage X-rays.
eXaSkin Plus 150 and 225 is intended for general radiotherapy use.
eXaSkin Plus 650 is intended for use on breast tissue.
# Comparison of Technological Characteristics with the Predicate and Reference Devices
The subject device has the same or similar intended use, indications, principle of operation, technological and material characteristics as the predicate device, .decimal Bolus (K211511), and reference device, Superflab (K812983). Comparisons of the predicate, reference, and subject devices can be found in Table 1:
Table 1 Comparison of the Subject, Primary Predicate, and Reference Devices
| | AnatGe eXaSkin Plus (Subject) | .decimal Bolus (Predicate) K211511 | Superflab (Reference) K812893 | Discussion |
| --- | --- | --- | --- | --- |
| Regulation | 892.5710, 892.5050 | 892.5710, 892.5050 | 892.1370 | Same as .decimal Bolus |
| Classification | Radiation therapy beam-shaping block,Medical charged-particle radiation therapy system. | Radiation therapy beam-shaping block,Medical charged-particle radiation therapy system. | Nuclear anthropomorphic phantom. | Same as .decimal Bolus |
{6}
510(k) Summary - K252784
Page 3 of 8
| Product Code | IXI, MUJ | MUJ, IXI | IYP | Same as .decimal Bolus |
| --- | --- | --- | --- | --- |
| Class | II | II | I | Same as .decimal Bolus |
| OTC or Rx | Rx | Rx | Rx | Same |
| Indications for Use | eXaSkin is a medical device intended to be used as a bolus for the treatment of neoplastic skin lesions through radiotherapy. The product is designed to be applied to the irradiated area by qualified healthcare professionals to address anatomical irregularities and ensure precise delivery of the prescribed dose to the skin and underlying tissues during a radiation therapy treatment. eXaSkin was evaluated using 6 MV photons but has not been assessed for use with 10 MV photons, protons, or orthovoltage X-rays. eXaSkin Plus 150 and 225 is intended for general radiotherapy use. eXaSkin Plus 650 is intended for use on breast tissue. | The decimal Bolus product is a solid piece of material (rigid or rubber-like) that will be placed on the skin of a patient with the intended use and primary purpose of helping control the dose received by that patient when undergoing radiation therapy treatment. decimal Bolus devices are designed by radiation therapy professionals to a unique shape that is specific to each patient being treated. The device is intended to modify the dose delivered during a radiation therapy treatment. As this product is a simple general purpose bolus device, the intended patient population and indications for use are quite broad. The most common indications for use are for the treatment of patients receiving radiation therapy, which encompasses a wide range of potential disease types and locations. As such, these devices will be required to have a wide range of potential shapes, sizes, and material properties and each device must be | This product is single-use only, and single patient use only. Do not Reprocess and dispose of this product after use. This product is to be used to increase the targeted radiation dose during photon and electron treatment by providing scattering of the beam and build-up of the radiation dose at the skin surface. | Different The .decimal Bolus is made of a solid material that is custom designed to be a unique shape specific to the patient being treated, while the eXaSkin is a putty designed to harden in a unique shape specific to the patient that it is applied to. This difference in material does not constitute a new intended use because the different materials used in each device serve the same intended purpose. Both the eXaSkin and .decimal Bolus are intended to control the radiation dose delivered to patients during radiation therapy treatments via a material that will be fit to the patient's skin. The difference does not raise different questions of substantial equivalence. Biocompatibility testing was performed, supporting substantial equivalence of skin-safe material. Data from bench testing for the eXaSkin supports substantial equivalence. Similar to the eXaSkin moldable putty, the Superflab is made of a moldable, elastic material that can conform to the contours of the patient. Testing was performed comparing the performance of the |
{7}
510(k) Summary - K252784
Page 4 of 8
| | | tested and approved by the radiation therapy professional prior to use on a patient. | | eXaSkin and Superflab (see Performance Testing section for more details), and performance was found to be substantially equivalent. |
| --- | --- | --- | --- | --- |
| Materials | Putty and Catalyzer | 100% Silicone | Synthetic oil gel including PVC Resin 01 and Plasticizer 34 | Different The eXaSkin, predicate, and reference device are all constructed from skin-safe materials and are used to control the dose delivered to patients undergoing radiation therapy treatments via fitting to the contours of a patient's skin. Unlike the predicate, the eXaSkin is not custom-made solid material, but a moldable putty. The difference in material does not raise concerns regarding substantial equivalence and the eXaSkin has been proven biocompatible via biocompatibility testing. |
| Material Characteristics | Skin-safe, moldable material | Skin-safe, solid piece of rubber-like, flexible material | Skin-safe, moldable material | Different The eXaSkin, predicate, |
{8}
510(k) Summary - K252784
| | | | | and reference device are all constructed from skinsafe materials and are used to control the dose delivered to patients undergoing radiation therapy treatments via fitting to the contours of a patient's skin. Unlike the predicate, the eXaSkin is not custom-made solid material, but a moldable putty. This difference in material characteristics does not raise different questions of substantial equivalence of the eXaSkin has been proven through literature and testing. |
| --- | --- | --- | --- | --- |
| Patient Contact | Direct skin | Direct skin | Direct skin | The eXaSkin, predicate, and reference device are all placed directly on patient skin, but the eXaSkin has a contraindication for being used on open wounds. This difference does not raise different question related to the substantial equivalence of the device. |
| Use Environment | Radiotherapy clinic | Radiotherapy clinic | Radiotherapy clinic | Same |
| Patient Population | Patients undergoing radiation therapy treatment | Patients undergoing radiation therapy treatment | Patients undergoing radiation therapy treatment | Same |
# Biocompatibility
Biocompatibility testing was conducted on representative final manufactured samples in accordance with ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Test results are summarized in Table 2:
Table 2 Summary of Biocompatibility Testing Conducted on the eXaSkin Plus
{9}
510(k) Summary - K252784
Page 6 of 8
| Biocompatibility Testing Requirements | Test Method | Material | Acceptance Criteria | Test Result |
| --- | --- | --- | --- | --- |
| ISO 10993-5 – Tests for in vitro cytotoxicity | Direct Contact | eXaSkin Plus | Not cytotoxic | Not cytotoxic |
| ISO 10993-10 – Tests for skin sensitization | Assessment of sensitizing properties on albino guinea pigs, Maximisation test according to Magnusson and Kligman | eXaSkin Plus | Non-skin sensitizer | Non-skin sensitizer |
| ISO 10993-23 - Tests for irritation | Assessment of acute local tolerance and general tolerance on white rabbits | eXaSkin Plus | Not an irritant | Not an irritant |
## Performance Data
### Bench
Non-clinical validation testing was performed to demonstrate substantial equivalence of the device. The following tests were performed:
| Test | Objective | Reference | Pass/Fail Criteria | Results |
| --- | --- | --- | --- | --- |
| Mixing Time | Evaluate the time required for two or more components to mix homogeneously in a mixture, process, or product. | ISO 4823: 2021 | 30 s | Pass |
| Setting time at 35°C | Evaluate the time it takes for a material, typically in paste or liquid form, to transition from a fluid or plastic state to a solid or rigid state. | 2100-AM QC-023 | Catalyzer: 190 - 320 s Putty: 210 - 75s | Pass |
| Hardness after 8 minutes | Evaluate the hardness of a material after 8 minutes (early curing stage). | 2100-AM QC-013 | Catalyzer: 20 - 34 Shore A Putty: 60 - 75 Shore A | Pass |
| Hardness after 1 hour | Evaluate the hardness of a material after 1 hour. | 2100-AM QC-013 | Catalyzer: 24 - 35 Shore A Putty: 65 - 77 Shore A | Pass |
{10}
510(k) Summary - K252784
Page 7 of 8
| Consistency | Evaluate the material’s ability to maintain a specific viscosity or fluidity under certain conditions. | 2100-AM QC-011 | Catalyzer: 33-44 mm Putty: ≤ 35 mm | Pass |
| --- | --- | --- | --- | --- |
| Mixed Material | Ensure that the mixture of materials has reached the expected homogeneity and to assess how the characteristics of the combined material behave under specific conditions. | 2100-AM QC-004 | Homogeneous and free of extraneous matter | Pass |
| Visual aspect | Evaluates the external appearance of the material to ensure it meets the quality standards in terms of aspect. | 2100-AM QC-004 | Homogeneous and free of extraneous matter | Pass |
| Color | evaluates the external appearance of the material to ensure it meets the quality standards in terms of color. | 2100-AM QC-004 | Purple | Pass |
| Odor | evaluates the external appearance of the material to ensure it meets the quality standards in terms of odor. | 2100-AM QC-005 | Catalyzer: Mint Putty: Vanilla | Pass |
| Working Time | Evaluates the time during which the material maintains its workability properties before it begins to change or harden significantly. | ISO 4823: 2021 | 75 s | Pass |
| Elastic recovery | Evaluates the material’s ability to recover its original shape after being deformed under load and then released. | ISO 4823: 2021 | ≥ 96.5% | Pass |
| Strain-in-compression | Evaluates deformation (or change in dimensions of the material) when a compressive force is applied until a specific point. | ISO 4823: 2021 | 0.8 - 20% | Pass |
| Detail reproduction | Assess the material’s ability to accurately reproduce fine details or characteristics of a mold or reference surface. | ISO 4823: 2021 | 75 μm | Pass |
| Compatibility with gypsum | Evaluates the interaction between the material and the gypsum used in impression or manufacturing processes. | ISO 4823: 2021 | 75 μm | Pass |
| Linear | Evaluate the material’s dimensional | ISO 4823: 2021 | ≤ 1.5% | Pass |
{11}
510(k) Summary - K252784
Page 8 of 8
| dimensional change 24h | stability after a 24-hour period. It measures whether the material undergoes expansion or contraction in its linear dimensions (length, width, or height) during the first 24 hours after its curing or formation process. | | | |
| --- | --- | --- | --- | --- |
| Material Properties | Evaluate eXaSkin, a novel high-density bolus alternative to commercial tissue-equivalent Superflab, for 6MV photon-beam radiotherapy. | N/A | Improved adaptation to phantom's surface over Superflab Full dosage achieved (agreement within 1% of calculations) | Pass |
| Dose Distribution and Radiation Attenuation | Determine dosimetric accuracy for calculating photon dose distribution in the presence of eXaSkin bolus. | N/A | Agreement in dose calculation - 3mm/3% Improved surface dose | Pass |
| Usability | Evaluate eXaSkin to commercial tissue-equivalent Superflab, for 6MV photon-beam radiotherapy. | N/A | N/A | Pass |
## Animal
No animal or clinical testing was performed in support of this submission.
## Clinical
No animal or clinical testing was performed in support of this submission.
## Substantial Equivalence Conclusion
Based on the comparison of indication of use, material characteristics, technological features, and principle of operation, and conclusions drawn from performance testing to demonstrate substantial equivalence (SE), the AnatGe eXaSkin Plus is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.