K252776 · ZheJiang Decans Medical Devices Co., Ltd. · KWQ · May 14, 2026 · Orthopedic
Device Facts
Record ID
K252776
Device Name
ARIES Anterior Cervical Plate Systems
Applicant
ZheJiang Decans Medical Devices Co., Ltd.
Product Code
KWQ · Orthopedic
Decision Date
May 14, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ARIES Anterior Cervical Plate Systems is intended for anterior interbody screw fixation from C2-T1 in the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2) Trauma (including fractures) 3) Tumors 4) Deformity (defined as kyphosis, lordosis, or scoliosis) 5) Pseudarthrosis, and/or 6) Failed previous fusions
Device Story
System consists of titanium alloy plates and screws for anterior cervical spine fixation (C2-T1); includes screw anti-extraction locking mechanism to prevent withdrawal. Used in hospital/healthcare settings by surgeons for temporary stabilization during spinal fusion. Device provides mechanical support to spine; surgeon selects appropriate plate length and screws to secure vertebrae. Benefits include stabilization of spinal segments to facilitate fusion. Supplied non-sterile; requires moist heat sterilization before use.
Clinical Evidence
No clinical or animal study data provided. Substantial equivalence supported by non-clinical bench testing, including static compression bending, static torsion, and dynamic compression bending fatigue tests per ASTM F1717-21, and biocompatibility testing per ISO 10993 series.
Technological Characteristics
Material: Ti6Al4V (ISO 5832-3). Components: Titanium plates and screws with anti-extraction locking mechanism. Sterilization: Moist heat (ISO 17665:2024). Biocompatibility: ISO 10993-1 compliant. Mechanical testing: ASTM F1717-21.
Indications for Use
Indicated for patients requiring temporary anterior spinal stabilization during cervical fusion (C2-T1) due to degenerative disc disease, trauma (fractures), tumors, deformity (kyphosis, lordosis, scoliosis), pseudarthrosis, or failed previous fusions.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
ATLANTIS™ Anterior Cervical Plate System (K021461)
ATLANTIS™ Anterior Cervical Plate System (K130640)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 14, 2026
ZheJiang Decans Medical Devices Co., Ltd.
Peiwen Feng
No. 2836 Xincheng Avenue, Gaozhao Street, Xiuzhou District
Jiaxing City, Zhejiang Province 314031
China
Re: K252776
Trade/Device Name: ARIES Anterior Cervical Plate Systems
Regulation Number: 21 CFR 888.3060
Regulation Name: Spinal Intervertebral Body Fixation Orthosis
Regulatory Class: Class II
Product Code: KWQ
Dated: April 14, 2026
Received: April 15, 2026
Dear Peiwen Feng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252776 - Peiwen Feng
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252776 - Peiwen Feng
Page 3
Sincerely,
MAZIAR SHAH-MOHAMMADI -S
For: Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
K252776 - Page 1 of 1
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252776 | ? |
| Please provide the device trade name(s). | | ? |
| ARIES Anterior Cervical Plate Systems | | |
| Please provide your Indications for Use below. | | ? |
| The ARIES Anterior Cervical Plate Systems is intended for anterior interbody screw fixation from C2-T1 in the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with: 1) Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies) 2) Trauma (including fractures) 3) Tumors 4) Deformity (defined as kyphosis, lordosis, or scoliosis) 5) Pseudarthrosis, and/or 6) Failed previous fusions | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
K252776 - Page 1 of 5
510(k) Summary
ARIES Anterior Cervical Plate Systems
1. Submitter
ZheJiang Decans Medical Devices Co., Ltd.
No.2836 Xincheng Avenue, Gaozhao Street, Xiuzhou District, Jiaxing City, Zhejiang Province, 314031 P.R.China
Contact Person: Haifeng Liu, RA Manager
Telephone: +86 15210058659
Email: hfliu@decansmd.com
Correspondent: Peiwen Feng, RA
Telephone: +86 18758326652
Email: pwfeng@decansmd.com
Date of Preparation: August/20/2025
Prepared in accordance with 21CFR807.92
2. Identification of subject device
Trade name: ARIES Anterior Cervical Plate Systems
Regulation number: 21CFR888.3060
Regulation name: Spinal intervertebral body fixation orthosis
Regulation class: Class II
Product code: KWQ
Common name: appliance, fixation, spinal intervertebral body
Review panel: Orthopedic
Material: Ti6Al4V per ISO 5832-3
Patient contact: Bone and surrounding tissue
Contact duration: Permanent, >30 days
Sterilization method: Moist heat sterilization
Environment of Use: Healthcare facility/Hospital
Single Use: yes
Provided status: non-sterile
3. Identification of predicate device(s)
> Primary predicate device
K021461-ATLANTIS™ Anterior Cervical Plate System
Regulation number: 21CFR888.3060
Regulation name: Spinal intervertebral body fixation orthosis
Regulation class: Class II
{5}
K252776 - Page 2 of 5
Product code: KWQ
Common name: appliance, fixation, spinal intervertebral body
Review panel: Orthopedic
> Additional predicate device:
K130640-ATLANTIS™ Anterior Cervical Plate System
Regulation number: 21CFR888.3060
Regulation name: Spinal intervertebral body fixation orthosis
Regulation class: Class II
Product code: KWQ
Common name: appliance, fixation, spinal intervertebral body
Review panel: Orthopedic
K103491-SKYLINE® Anterior Cervical Plate System
Regulation number: 21CFR888.3060
Regulation name: Spinal intervertebral body fixation orthosis
Regulation class: Class II
Product code: KWQ
Common name: appliance, fixation, spinal intervertebral body
Review panel: Orthopedic
K231090-ZEVO™ Anterior Cervical Plate System
Regulation number: 21CFR888.3060
Regulation name: Spinal intervertebral body fixation orthosis
Regulation class: Class II
Product code: KWQ
Common name: appliance, fixation, spinal intervertebral body
Review panel: Orthopedic
## 4. Device description
The ARIES Anterior Cervical Plate Systems consist of titanium plates of various lengths and a range of screws. The screws are inserted into the cervical vertebrae through the holes on the fixation plates for fixation. The titanium plate includes a screw anti-extraction locking mechanism, which covers the head of the screw to reduce the possibility of screw withdrawal. The components of the device are made of titanium alloy(Ti6Al4V) per ISO 5832-3, and are supplied in non-sterile form.
{6}
K252776 - Page 3 of 5
## 5. Indication for use
The ARIES Anterior Cervical Plate Systems is intended for anterior interbody screw fixation from C2-T1 in the cervical spine. The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spinal fusions in patients with:
1) Degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies)
2) Trauma (including fractures)
3) Tumors
4) Deformity (defined as kyphosis, lordosis, or scoliosis)
5) Pseudarthrosis, and/or
6) Failed previous fusions
## 6. Substantial equivalence comparison
We did comparisons and identified similarities and differences of the subject device to the legally marketed predicate devices with respect to Indication for use, technological characteristics following the FDA guidance The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (issued on July 2014).
The subject device ARIES Anterior Cervical Plate Systems and predicate devices are equivalent in the indication for use and anatomical site, they are same in patient population, patient contact type, contact duration, component, mechanism of action, single use, SAL. Any technological similarities or differences between the subject and predicate devices are considered minor and do not raise any new issues of safety or effectiveness.
## 7. Non-clinical Test summary
### Standard performance test
Static Compression Bending Test per ASTM F1717-21
Static Torsion Test per ASTM F1717-21
Dynamic Compression Bending Fatigue Test per ASTM F1717-21
Performance criteria proposed in the FDA Guidance Spinal Plating Systems-Performance Criteria for Safety and Performance Based Pathway issued on September 20,2019, is used to support substantial equivalence, rather than a direct comparison of the performance of the subject device to that of the predicate device. Test results were satisfactory and met acceptance criteria.
### Sterilization efficacy validation
{7}
K252776 - Page 4 of 5
The recommended sterilization method for subject device is moist heat sterilization. The sterilization validation has been performed on an representative product in accordance with ISO 17665:2024. A sterility assurance level (SAL) of $10^{-6}$ has been demonstrated.
> Biocompatibility
The biocompatibility evaluation was performed per ISO10993-1, and biological tests were conducted on an representative products following applicable standards. Through test and evaluation, the subject is determined as biocompatible. Tests included:
- In vitro cytotoxicity Test per ISO 10993-5
- Skin sensitization Test per ISO 10993-10
- Intracutaneous Reactivity Test per ISO 10993-23
- Material mediated Pyrogen Test per ISO 10993-11
- Chemical Characterization per ISO 10993-18
- Toxicological risk assessment of extractable chemicals per ISO 10993-17
## 8. Performance-Clinical or animal
The subject of this premarket submission, did not require clinical study data or animal study data to support substantial equivalence.
## 9. Referenced FDA guidance
- Guidance for Industry and FDA Staff: Spinal System 510(k)s (issued on May, 2004)
- Format for Traditional and Abbreviated 510(k)s Guidance (issued on December, 2019)
- Electronic Submission Template for Medical Device 510(k) Submissions (issued on October, 2023)
- Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process" (issued on September, 2023)
- Chemical Analysis for Biocompatibility Assessment of Medical Device draft guidance (issued on September, 2024)
- The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (issued on July, 2014)
- Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions (issued on December 2019)
{8}
K252776 - Page 5 of 5
## 10. Conclusion
The subject device is substantially equivalent to the predicate devices. The identified similarities or differences does not raise any new issues of safety or effectiveness. ZheJiang Decans Medical Devices Co.,Ltd. believes that ARIES Anterior Cervical Plate Systems is as safe and effective, and perform as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.