K252752 · Cytrellis Biosystems, Inc. · QAI · Oct 30, 2025 · General, Plastic Surgery
Device Facts
Record ID
K252752
Device Name
ellacor System with Micro-Coring Technology
Applicant
Cytrellis Biosystems, Inc.
Product Code
QAI · General, Plastic Surgery
Decision Date
Oct 30, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4430
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ellacor System with Micro-Coring Technology is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. It is further indicated for Fractional Skin Resurfacing.
Device Story
System uses hollow coring needles to rapidly excise full-thickness micro-columns of dermal and epidermal tissue; needles controlled by console and handpiece to precise depths/positions. System offers adjustable treatment densities and needle depths via LCD touch display. Handpiece uses vacuum-assisted disposable for alignment. Mechanical removal of skin columns (up to 7.9% surface area) triggers natural healing response via skin elastin, closing holes without visible scarring; reduces wrinkle appearance. Used by medical professionals in clinical settings. Output is mechanical skin excision; provider uses device to perform fractional skin resurfacing. Benefits include wrinkle reduction without surgical procedure.
Clinical Evidence
Single-center, prospective histology study (n=3) in patients undergoing abdominoplasty. Subjects treated at 4mm depth across 3 areas; non-treated control area included. Safety assessments at 7, 14, 21, 30 days; histology at day 30. Results showed no evidence of trauma, significant inflammation, or hemorrhage; normal dermis/adipose structures observed. Complete epithelialization by day 3.
Technological Characteristics
Powered microneedle device; hollow needles (single or triple configuration) with 15-degree tip angle. Max penetration 4.0mm core depth/5mm tip depth. Frequency 8-12 Hz. Vacuum-assisted handpiece. RFID-enabled disposable cartridges. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Biocompatibility per ISO 10993-1. Sterilization per ISO 11137.
Indications for Use
Indicated for treatment of moderate and severe wrinkles in mid and lower face in adults aged 22+ with Fitzpatrick skin types I-IV; also indicated for fractional skin resurfacing.
Regulatory Classification
Identification
A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.
Special Controls
The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate. Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Accuracy of needle penetration depth and puncture rate; (ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and (iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use. Performance data must demonstrate the sterility of the patient-contacting components of the device. Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. Software verification, validation, and hazard analysis must be performed for all software components of the device. The patient-contacting components of the device must be demonstrated to be biocompatible. Performance data must validate the cleaning and disinfection instructions for reusable components of the device. Labeling must include the following: (i) Information on how to operate the device and its components and the typical course of treatment; (ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; (iii) Validated methods and instructions for reprocessing of any reusable components: (iv) Disposal instructions; and (v) Shelf life. Patient labeling must be provided and must include: (i) Information on how the device operates and the typical course of treatment; (ii) The probable risks and benefits associated with use of the device; and (iii) Post-operative care instructions.
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.
(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Accuracy of needle penetration depth and puncture rate;
(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and
(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.
(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.
(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.
(7) The patient-contacting components of the device must be demonstrated to be biocompatible.
(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.
(9) Labeling must include the following:
(i) Information on how to operate the device and its components and the typical course of treatment;
(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;
(iii) Validated methods and instructions for reprocessing of any reusable components;
(iv) Disposal instructions; and
(v) A shelf life.
(10) Patient labeling must be provided and must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) The probable risks and benefits associated with use of the device; and
(iii) Postoperative care instructions.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 30, 2025
Cytrellis Biosystems, Inc.
Robert Hinz
Vice President, Operations
34 Commerce Way
Woburn, Massachusetts 01801
Re: K252752
Trade/Device Name: ellacor System with Micro-Coring Technology
Regulation Number: 21 CFR 878.4430
Regulation Name: Microneedling Device For Aesthetic Use
Regulatory Class: Class II
Product Code: QAI
Dated: August 26, 2025
Received: August 29, 2025
Dear Robert Hinz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252752 - Robert Hinz
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252752 - Robert Hinz
Page 3
Sincerely,
Wilmarie Flores -S
Digitally signed by Wilmarie Flores -S
For Jodie Giordano, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252752 | |
| Device Name ellacor Sytem with Micro-Coring Technology | |
| Indications for Use (Describe) The ellacor System with Micro-Coring Technology is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. It is further indicated for Fractional Skin Resurfacing. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
CYTRELLIS
510(k) Summary K252752
Submitter:
Cytrellis BioSystems, Inc.
34 Commerce Way
Woburn, MA 01801
Contact Person:
Robert Hinz
V.P., Operation
Phone: 781.689.3149
Email: rhinz@cytrellis.com
Date Prepared:
October 30, 2025
Device Trade Name:
ellacor System with Micro-Coring Technology
Common Name:
Powered Microneedle Device
Classification Name:
Microneedling device for aesthetic use
Regulation Number:
21 CFR 878.4430 (Product Code: QAI)
Predicate Device:
Cytrellis® Dermal Micro-Coring™ System Device
K202517
## Device Description:
The ellacor System with Micro-Coring Technology removes unwanted skin by rapidly excising full thickness micro-columns of the dermal and epidermal tissue via specially designed hollow coring needles. The needles are controlled to precise coring depths and positions by the device console and handpiece. The system provides a range of skin removal percentages (treatment densities) and needle depths settings to customize the user needs that is controlled by the LCD touch display on the console. The treatment is applied in 1cm² areas of the skin at a time by the handpiece, which has a vacuum assisted disposable in order to help with alignment between treatment areas. The ellacor System with Micro-Coring Technology mechanically removes full thickness sections of skin without evidence of scarring or need for surgical procedure. As much as 7.9% of the surface area of the skin is removed in columns that are the size of the inner diameter of the needle which is too small to cause visible scarring. The holes quickly close because of the natural elastin of the skin as they start to heal. This wound response, combined with the mechanical removal of the skin's surface area, leads to a reduction in the appearance of wrinkles.
## Intended Use and Indications for Use:
The ellacor System with Micro-Coring Technology is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. It is further indicated for Fractional Skin Resurfacing.
Page 1 of 6
{5}
CYTRELLIS
510(k) Summary K252752
Comparison of Technological Characteristics with the Predicate Device:
| Characteristics | ellacor® System with Micro-Coring Technology (Proposed – K pending) | Cytrellis Dermal Micro-Coring System (K202517) (Predicate) | AI.ME System (K221011) (Reference) | Comparison to Predicate/Reference and Discussion |
| --- | --- | --- | --- | --- |
| Product Code | QAI | QAI | QAI | Identical |
| Classification Name | Microneedling device for aesthetic use | Microneedling device for aesthetic use | Microneedling device for aesthetic use | Identical |
| Device Classification | Class II | Class II | Class II | Identical |
| Intended Location of Use | Mid and Lower Face Skin Resurfacing | Mid and Lower Face | Any where | The proposed device’s intended location of use is the same as that of the predicate and reference devices. |
| Intended Use and Indications for Use | The ellacor® System with Micro-Coring Technology is intended for the treatment of moderate and severe wrinkles in the mid and lower face.
The ellacor® System with Micro-Coring Technology is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in | The Cytrellis® Dermal Micro-Coring™ System is intended for the treatment of moderate and severe wrinkles in the mid and lower face.
The Cytrellis Dermal Micro-Coring System is indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower | The AI.ME system is indicated for fractional skin resurfacing. | The proposed device’s intended use and indications for use are the same as that of the predicate and reference devices. |
{6}
CYTRELLIS
510(k) Summary K252752
| Characteristics | ellacor® System with Micro-Coring Technology (Proposed – K pending) | Cytrellis Dermal Micro-Coring System (K202517) (Predicate) | ALME System (K221011) (Reference) | Comparison to Predicate/Reference and Discussion |
| --- | --- | --- | --- | --- |
| | adults aged 22 years or older with Fitzpatrick skin types I-IV. It is further indicated for Fractional Skin Resurfacing. | face in adults aged 22 years or older with Fitzpatrick skin types I-IV | | |
| Geometry | Single (1) or three (3) needles arranged in a straight line. Needles are ground to 15-degree tip angle resulting in three (3) cutting surfaces and three (3) tips. | Single (1) or three (3) needles arranged in a straight line. Needles are ground to 10-degree tip angle resulting in two (2) cutting surfaces and two (2) tips. | 6 hollow needles, hexagon arrangement | The proposed device’s needle geometry is within the predicate, and it is supported in the safety study and does not raise new concerns for safety and effectiveness. |
| Maximum needle penetration (maximum needle length) | Up to 4.0mm core depth setting or 5mm needle tip depth | Up to 4.0mm core depth setting or 5mm needle tip depth | Fixed 3mm core depth | Identical |
| Needle protrusion setting | 0-4.0mm core depth setting or 1- 5mm needle tip depth | 0-4.0mm core depth setting or 1- 5mm needle tip depth | Fixed 3mm core depth | Identical |
| Frequency | 8-12 Hz | 8-12 Hz | Unknown | Identical |
## Summary of performance testing:
To demonstrate safety and effectiveness and support substantial equivalence, the ellacor System with Micro-Coring Technology has undergone non-clinical performance testing in line with recognized standards.
Page 3 of 6
{7}
CYTRELLIS
510(k) Summary K252752
| (1) The technical specifications and needle characteristics have been identified, including needle length, geometry, maximum penetration depth, and puncture rate. |
| --- |
| (2) Non-clinical performance data demonstrates that the device performs as intended under anticipated conditions of use. The following performance characteristics have been tested: |
| (i) | Accuracy of needle penetration depth and puncture rate | Accuracy of needle penetration depth and puncture rate was tested in a suitable skin substrate model and measured using the Keyence Laser System |
| --- | --- | --- |
| (ii) | Safety features built into the device to protect against cross-contamination & fluid ingress protection | The device design prevents cross contamination including fluid ingress protection due to the needle cartridge design. Design elements include RFID to prevent needle cartridge re-use and tortuous path for fluid ingress. Testing was performed under worst case scenarios. |
| (iii) | Identification of the max safe needle penetration depth for the device & for the labeled indications for use | Maximum safe needle penetration depth was identified in a suitable skin substrate model and measured using the Keyence Laser System. Note: Do not treat areas where critical structures may be less than 4mm below the skin surface. |
| (3) Performance data must demonstrate the sterility of the patient-contacting components of the device. | Performance data demonstrates the sterility of the patient-contacting components of the device according to ISO 11137-1:2006, ISO 11137-2:2013 and ISO/TS 13004:2013 | |
| (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life. | Performance data supports the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life according to ASTM F1980 -16, ISO 11607-1, ASTM F88-15, ASTM F1140 / F1140M-13, ASTM F2096-11, ASTM F1886 / F1886M-16 | |
| (5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device. | Performance data demonstrates the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device according to IEC 60601-1:2012 3.1 edition and IEC 60601-1-2:2014 | |
| (6) Software verification, validation, and hazard analysis must be performed for all software components of the device. | Software verification, validation, and hazard analysis were performed for all software components of the device according to in accordance with FDA’s Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices | |
Page 4 of 6
{8}
CYTRELLIS
510(k) Summary K252752
| (7) The patient-contacting components of the device must be demonstrated to be biocompatible. | | The patient-contacting components of the device were demonstrated to be biocompatible including evaluation of cytotoxicity, irritation, and sensitization, acute systemic toxicity and material-mediated pyrogenicity and hemocompatibility per ISO 10993-1 |
| --- | --- | --- |
| (8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device. | | A cleaning and disinfection validation was performed for reusable components of the device per AAMI TIR30:2011/(R)2016 and AAMI TIR12:2010 |
| (9) Labeling includes the following: | | |
| (i) | Information on how to operate the device and its components and the typical course of treatment; | |
| (ii) | A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate; | |
| (iii) | Validated methods and instructions for reprocessing of any reusable components; | |
| (iv) | Disposal instructions; and | |
| (v) | A shelf life | |
| (10) Patient labeling includes: | | |
| (i) | Information on how the device operates and the typical course of treatment; | |
| (ii) | The probable risks and benefits associated with use of the device; | |
| (iii) | Postoperative care instructions. | |
## Summary of Clinical testing:
A single center, prospective histology study was conducted evaluate the safety and histological effect on tissue using the ellacor Micro-Coring procedure in patients undergoing abdominoplasty. The study included three subjects treated a single time at 4mm at a minimum of 3 treatment areas to evaluate safety of the varying treatment depths. A non-treated control area was also designated. Cohort 1 subjects were followed for safety assessments at 7, 14, 21 and 30 days post treatment with histology evaluations for all treatment areas and the control area occurring 30 days after treatment (Day 30).
There is no evidence of trauma, significant inflammation, or hemorrhage. Normal dermis and adipose tissue structures are seen in all histology samples. The results of this study establish the safety of the ellacor Micro-Coring procedure at densities up to 10% and at a depth of up to 4mm, showing complete epithelialization by day-3.
## Conclusion:
As described in the comparison tables above, the ellacor® System with Micro-Coring Technology subject device has same intended use and indications for use, similar technological characteristics, and same principles of operation as its predicate and reference device. The technological differences between ellacor
Page 5 of 6
{9}
CYTRELLIS
510(k) Summary K252752
and the predicate/reference device do not raise any new issues of safety and effectiveness. The ellacor® System with Micro-Coring Technology is substantially equivalent to the previously cleared device.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.