Halo Sterile Single-use Radial Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Halo Sterile Single-use Radial Fibers are intended for use with any cleared surgical laser with an SMA 905 connector.
Device Story
Surgical laser fiber; radial-typed; conical distal end for radial laser emission. Input: laser energy from external diode laser system (500-2200nm, max 30W). Operation: fiber transmits laser energy to target tissue in contact or non-contact mode; facilitates incision, excision, vaporization, ablation, hemostasis, or coagulation. Used in clinical surgical settings by physicians. Output: laser light delivered to tissue; enables precise tissue interaction. Benefits: provides controlled thermal effect for endovascular and general surgical procedures. Single-use; EO sterilized.
Clinical Evidence
No clinical data. Bench testing only. Performance testing evaluated packaging, dimensions, transmission power/efficiency, tensile strength, and bending radius. Biocompatibility testing performed per ISO 10993 (cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity, thrombosis, hemolysis, and genotoxicity).
Technological Characteristics
Silica glass radial fiber; SMA 905 connector; 0.40mm/0.60mm core diameters; 0.9mm/1.3mm outer diameters. EO gas sterilization. Biocompatibility per ISO 10993. Compatible with 500-2200nm diode lasers up to 30W.
Indications for Use
Indicated for adult patients in all surgical specialties requiring soft tissue incision, excision, vaporization, ablation, hemostasis, or coagulation, specifically endovascular blood vessel coagulation, using compatible laser systems (500nm-2200nm) with SMA 905 connectors.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
Micro-Energy Medical Technology Co., Ltd.
Danning Zhu
Registered Engineer
Section B4301, B4701 and B4901, No. 13, Baolan Road
Kengzi Street, Pingshan District
Shenzhen, 518118
China
Re: K252745
Trade/Device Name: Halo Sterile Single-use Radial Fiber (Halo-R-0.40-2.5/Halo-R-0.60-2.5/Novo-R-0.60-1.8/Novo-R-0.40-1.8)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: August 29, 2025
Received: August 29, 2025
Dear Danning Zhu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252745 - Danning Zhu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252745 - Danning Zhu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
YAN FU-S
Digitally signed by YAN FU
Date: 2025.12.12 12:03:49
-05'00"
for Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252745 | ? |
| Please provide the device trade name(s). | | ? |
| Halo Sterile Single-use Radial Fiber (Halo-R-0.40-2.5/Halo-R-0.60-2.5/Novo-R-0.60-1.8/Novo-R-0.40-1.8) | | |
| Please provide your Indications for Use below. | | ? |
| Halo Sterile Single-use Radial Fibers are indicated for use in all surgical specialties in which compatible
laser systems with operational wavelengths between 500nm - 2200nm have
received regulatory clearance. Halo Sterile Single-use Radial Fibers are intended for use with any cleared
surgical laser with an SMA 905 connector. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) Summary: K252745
Date Prepared: December 12, 2025
This summary of 510(k) information is submitted as required by requirements of 21 CFR §807.92.
1. Submitter's Information
Name of Sponsor: Micro-Energy Medical Technology Co., Ltd.
Address: Section B4301, B4701 and B4901 No.13, Baolan Road, Kengzi Street, Pingshan District, Shenzhen, China
Contact Name: Elly Xu
Telephone No.: +86-755-28336782
Fax No.: +86-755-28336782
Email Address.: sha.x@micro-energy.com
2. Correspondent's Information
Company Name: Micro-Energy Medical Technology Co., Ltd.
Correspondent Name: Danning Zhu
Telephone No.: +86-755-28339524
Email Address: danning.z@micro-energy.com
3. Trade Name, Common Name, Classification
| Trade Name: | Halo Sterile Single-use Radial Fiber |
| --- | --- |
| Common Name: | Sterile Single-use Radial Fiber |
| Model: | Halo-R-0.40-2.5/Halo-R-0.60-2.5/Novo-R-0.60-1.8/Novo-R-0.40-1.8 |
| Classification Name: | Powered Laser Surgical Instrument |
| Product Code: | GEX |
| Classification Panel: | General & Plastic Surgery |
| Device Class: | II |
| Classification Regulation: | 21 CFR 878.4810 |
{5}
2
# 4. Identification of Predicate Device(s)
The identified predicates within this submission are as follows:
| Predicate Device : | 510(k) number:K241643
Trade name:WONTECH Surgical Optic Fibers
(BA400/BA400R/BA600/BA600R)
Product code:GEX |
| --- | --- |
# 5. Description of the Device
The Halo Sterile Single - use Radial Fibers (Model: Halo - R - 0.40 - 2.5/Halo - R - 0.60 - 2.5/Novo-R-0.60-1.8/Novo-R-0.40-1.8), as surgical laser fibers, are suitable for use in combination with commercially available diode laser systems with a wavelength ranging from 500nm to 2200nm and equipped with SMA905 connectors. The output power should not exceed 30 W. This device is single use only and sterilized by EO gas. The Halo Sterile Single - use Radial Fibers are intended for general surgical operations such as incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissues in either contact or non - contact mode. In particular, they are designed for endovascular coagulation of blood vessels. The Fiber is radial-typed fiber. The shape of distal end of fiber is conical to radiate laser radially.
# 6. Intended Use/Indication for Use
Halo Sterile Single-use Radial Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Halo Sterile Single-use Radial Fibers are intended for use with any cleared surgical laser with an SMA 905 connector.
# 7. Technological Characteristics
The Halo Sterile Single - use Radial Fibers (Model: Halo-R-0.40-2.5/Halo-R-0.60-2.5/Novo-R-0.60-1.8/Novo-R-0.40-1.8) uses the same technology that is utilized in the predicate device. A comparison of technological characteristics is provided in the following table:
{6}
| SUBSTANTIAL EQUIVALENCE COMPARISON TABLE | | | |
| --- | --- | --- | --- |
| Item | Subject device | Predicate device (K241643) | Comparison |
| Product Code | GEX | GEX | Same |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Class | II | II | Same |
| Device Name | Halo Sterile Single-use Radial Fiber | WONTECH Surgical Optic Fibers (BA400/BA400R/BA600/BA600R) | NA |
| Device Model | Halo-R-0.40-2.5/Halo-R-0.60-2.5/Novo-R-0.60-1.8/Novo-R-0.40-1.8 | WONTECH Surgical Optic Fibers | NA |
| Manufacturer | Micro-Energy Medical Technology Co., Ltd. | WON TECH Co., Ltd. | NA |
3
{7}
| Intended Use/Indication for Use | Halo Sterile Single-use Radial Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Halo Sterile Single-use Radial Fibers are intended for use with any cleared surgical laser with an SMA 905 connector. | WONTECH Surgical Optic Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths Between 500nm - 2200nm have received regulatory clearance. WONTECH Surgical Optic Fibers are intended for use with any cleared surgical laser with an SMA 905 connector. | same |
| --- | --- | --- | --- |
| Targer Population | Adult | Adult | Same |
| Wavelength Range | 500nm-2200mm | 532nm-2100mm | Same |
| Fiber Core Diameter | 0.40mm (Halo-R-0.40-2.5)
0.60mm (Halo-R-0.60-2.5)
0.40mm (Novo-R-0.40-1.8)
0.60mm (Novo-R-0.60-1.8) | 0.40mm
0.60mm | Same |
| Outer Diameter | Halo-R-0.40-2.5:0.9mm
Halo-R-0.60-2.5:1.3mm
Novo-R-0.40-1.8:0.9mm
Novo-R-0.60-1.8:1.3mm | Diameter 0.40mm:0.73mm
Diameter 0.60mm:0.75mm | Similar Note01 |
| Maximum Output Power | 1-30Watts | 1-30Watts | Same |
{8}
| Fiber Core Material | Silica Glass | Silica Glass | Same |
| --- | --- | --- | --- |
| Connectors | SMA 905 | SMA 905 | Same |
| Single Use | Yes | Yes | Same |
| Sterility | EO gas sterilized | EO gas sterilized | Same |
| Biocompatibility | ISO10993 | ISO10993 | Same |
As seen in the comparison tables, The Halo Sterile Single-use Radial Fibers that are the subject of this premarket notification use same or similar technology as that of the WONTECH Surgical Optic Fibers of the K241643 510(k). The minor differences only exist in the outer diameter. Operators can select the products with appropriate outer diameters for the surgery according to the actual situation. The design is intended to meet the claimed requirements of the intended use. Such differences will not give rise to any issues regarding safety and effectiveness. And performance testing demonstrates that the Halo Sterile Single-use Radial Fibers can be used safely and effectively for the proposed indications for use. Moreover, both the subject devices and the predicate devices meet the same standard ISO10993. The Halo Sterile Single-use Radial Fibers are considered to be substantially equivalent to the predicate K241643.
{9}
8. Non-Clinical Test Summary
1) Sterilization
The proposed device met all design specifications. Standard written in the sterilization test report are:
1) ISO 11135:2014 Sterilization of health care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
1) ISO 11737-1:2018 Sterilization of health care products - Microbiological methods - Part 1:
2) Determination of a population of microorganisms on products
4) ISO 11737-2:2019 Sterilization of health care products - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
5) ISO 11138-1:2017 Sterilization of health care products - Biological indicators - Part 1: General requirements
6) ISO 11138-2:2017 sterilization of health care products - Biological indicators - Part 2: Biological indicators for ethylene oxide sterilization processes
7) ISO 11138-7:2019 Sterilization of health care products - Biological indicators - Part 7: Guidance for the selection, use and interpretation of results
8) ISO 10993-7: 2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals
9) ASTM F88/F88M-21 Standard Test Method for Seal Strength of Flexible Barrier Materials
10) ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
11) ASTM F2096-11(2019) Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
12) ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
13) ANSI/AAMI ST72:2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
14) ISO 2859-1:1999 Sampling procedures for inspection by attributes - Part 1: Sampling schemes indexed by acceptance quality limit (AQL) for lot-by-lot inspection
2) Software Validation
There is no software available for this device.
3) Biocompatibility
The biocompatibility evaluation for Halo Sterile Single-use Radial Fiber was conducted in accordance with the International Standard ISO 10993 - 1:2020, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recommended by FDA. The biocompatible testing included the following tests:
- Cytotoxicity – (ISO 10993 - 5: 2009)
- Skin Sensitization – (ISO 10993 - 10:2021)
- Intracutaneous Injection Irritation – (ISO 10993 - 23:2021)
- Material-mediated Pyrogens – (ISO 10993-11:2017)
- Acute Systemic Toxicity – (ISO 10993 - 11:2017)
- Thrombosis Test in Vivo – (ISO 10993-4:2017)
- Partial Thromboplastin Time (PTT) – (ISO 10993-4: 2017)
{10}
- In Vitro Hemolysis Study– (ISO 10993-4: 2017)
- Complement activity– (ISO 10993-4: 2017)
- Bacterial Reverse Mutation– (ISO 10993-3: 2014)
- In vitro Mammalian Chromosome Aberration– (ISO 10993-3: 2014)
## 4) Performance testing
Performance testing was conducted on Halo Sterile Single-use Radial Fiber. Technical parameters about Packaging Performance, dimension, Maximum Transmission Power, Transmission Efficiency, Transmission Efficiency, Tensile Strength, Minimum Bending Working Radius of the Fiber were evaluated in the performance testing. All of the tested parameters met the predefined acceptance criteria.
## 9. Brief discussion of clinical tests
No clinical studies were considered necessary and performed.
## 10. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Micro-Energy Medical Technology Co., Ltd concludes that:
- The indications for use of Halo Sterile Single-use Radial Fiber (model:Halo-0.40-2.5/Halo-0.60-2.5//Novo-R-0.40-1.8/Novo-R-0.60-1.8) are the same as those of the predicate device. The technological characteristics are the same or similar to those of the predicate device.
- Halo Sterile Single-use Radial Fiber (model:Halo-0.40-2.5/Halo-0.60-2.5/Novo-R-0.40-1.8/Novo-R-0.60-1.8) is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.