← Product Code [PLF](/productcode/PLF) · K252733

# ATLAS™ Expandable Osteotomy Wedge System (K252733)

_MiRus, LLC · PLF · Feb 19, 2026 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252733

## Device Facts

- **Applicant:** MiRus, LLC
- **Product Code:** [PLF](/productcode/PLF.md)
- **Decision Date:** Feb 19, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ATLAS™ Expandable Osteotomy Wedge System is an expandable osteotomy wedge fusion device used for internal bone fixation for bone fractures, fusions, or osteotomies in the foot and ankle. The ATLAS™ Expandable Osteotomy Wedge System is intended for use with supplemental fixation for use in the foot or ankle. The ATLAS™ Expandable Osteotomy Wedge System is not intended for use in the spine.

## Device Story

ATLAS™ Expandable Osteotomy Wedge System is an implantable device for internal bone fixation in foot and ankle procedures; used for fractures, fusions, or osteotomies. Device requires supplemental fixation. Implants are expandable to accommodate patient anatomy; provided sterile for single use. Instrumentation provided non-sterile for steam sterilization by end user. Surgeon operates device during orthopedic surgery to stabilize bone segments. Output is mechanical stabilization of bone; clinical benefit is improved fixation and fusion outcomes.

## Clinical Evidence

Bench testing only. No clinical data provided. Mechanical performance assessed via static and dynamic compression (ASTM F2077-18), static and dynamic compression shear (ASTM F2077-18), subsidence (ASTM F2267-04), and expulsion testing (ASTM Draft Standard F-04.25.02.02).

## Technological Characteristics

Implants: Titanium-6 Aluminum-4 Vanadium ELI (ASTM F136). Expansion mechanism: Molybdenum-47.5 Rhenium Alloy (ASTM F3273). Instrumentation: Stainless Steel (ASTM F899). Expandable wedge design. Sterile-packed implants; non-sterile instruments requiring steam sterilization.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- TITAN 3-D™ Wedge System ([K162241](/device/K162241.md))
- Restor3d Utility Wedge ([K201314](/device/K201314.md))
- IO™ Expandable Lumbar Interbody Fusion System ([K210800](/device/K210800.md))

## Reference Devices

- IO™ Expandable Lumbar Interbody Fusion System ([K210800](/device/K210800.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

February 19, 2026

MiRus, LLC
Anuradha Nagulapati
Senior Regulatory Affairs Engineer
1755 W. Oak Parkway
Suite 100
Marietta, Georgia 30062

Re: K252733
Trade/Device Name: ATLAS™ Expandable Osteotomy Wedge System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: PLF
Dated: January 20, 2026
Received: January 20, 2026

Dear Anuradha Nagulapati:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252733 - Anuradha Nagulapati
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252733 - Anuradha Nagulapati
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CHRISTOPHER FERREIRA -S

Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

|  Submission Number (if known)  |
| --- |
|  K252733  |
|  Device Name  |
|  ATLAS™ Expandable Osteotomy Wedge System  |
|  Indications for Use (Describe)  |
|  The ATLAS™ Expandable Osteotomy Wedge System is an expandable osteotomy wedge fusion device used for internal bone fixation for bone fractures, fusions, or osteotomies in the foot and ankle.  |
|  The ATLAS™ Expandable Osteotomy Wedge System is intended for use with supplemental fixation for use in the foot or ankle.  |
|  The ATLAS™ Expandable Osteotomy Wedge System is not intended for use in the spine.  |

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

&gt; "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}

K252733 - Page 1 of 2

|  510(k) #: K252733 | 510(k) Summary | Prepared on: 2026-01-20  |
| --- | --- | --- |
|  Contact Details |   | 21 CFR 807.92(a)(1)  |
|  Applicant Name | MiRus, LLC |   |
|  Applicant Address | 1755 W. Oak Parkway Suite 100 Marietta GA 30062 United States |   |
|  Applicant Contact Telephone | 470-428-8684 |   |
|  Applicant Contact | Ms. Anuradha Nagulapati |   |
|  Applicant Contact Email | anagulapati@mirusmed.com |   |
|  Device Name |   | 21 CFR 807.92(a)(2)  |
|  Device Trade Name | ATLAS™ Expandable Osteotomy Wedge System |   |
|  Common Name | Single/multiple component metallic bone fixation appliances and accessories |   |
|  Classification Name | Bone Wedge |   |
|  Regulation Number | 888.3030 |   |
|  Product Code(s) | PLF |   |
|  Legally Marketed Predicate Devices |   | 21 CFR 807.92(a)(3)  |
|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
|  K162241 | TITAN 3-D™ Wedge System | PLF  |
|  K201314 | Restor3d Utility Wedge | PLF  |
|  K210800 | IO™ Expandable Lumbar Interbody Fusion System | MAX  |
|  Device Description Summary |   | 21 CFR 807.92(a)(4)  |
|  The ATLAS™ Expandable Osteotomy Wedge System is an expandable osteotomy wedge fusion device used for internal bone fixation for bone fractures, fusions, or osteotomies in the foot and ankle.
The ATLAS™ Expandable Osteotomy Wedge System consists of implants manufactured from Titanium-6 Aluminum-4 Vanadium ELI per ASTM F136. Expansion mechanism components are manufactured from Molybdenum-47.5 Rhenium Alloy (MoRe) per ASTM F3273. Instrumentation are manufactured from Stainless Steel per ASTM F899.
The ATLAS™ Expandable Osteotomy Wedge System implants are offered in multiple configurations and different sizes to accommodate various patient anatomical requirements. The implants (osteotomy wedges) will be provided sterile packed and are intended for single use only. The instruments will be provided non-sterile and will be steam sterilized by the end user.  |   |   |
|  Intended Use/Indications for Use |   | 21 CFR 807.92(a)(5)  |
|  The ATLAS™ Expandable Osteotomy Wedge System is an expandable osteotomy wedge fusion device used for internal bone fixation for bone fractures, fusions, or osteotomies in the foot and ankle.  |   |   |

{5}

K252733 - Page 2 of 2

The ATLAS™ Expandable Osteotomy Wedge System is intended for use with supplemental fixation for use in the foot or ankle.

The ATLAS™ Expandable Osteotomy Wedge System is not intended for use in the spine.

## Indications for Use Comparison
21 CFR 807.92(a)(5)

The indications for the use are identical for the subject device and predicate devices.

## Technological Comparison
21 CFR 807.92(a)(6)

The ATLAS™ Expandable Osteotomy Wedge System has the same intended use, indications for use, labeling, and technological characteristics as the predicate systems, including the same design features, geometries, sizes, and materials. The predicate device components are manufactured from titanium alloy. The ATLAS™ Expandable Osteotomy Wedge System components are manufactured from titanium alloy (ASTM F136) and molybdenum-rhenium alloy (ASTM F3273), which are the same materials as the IO™ Expandable Lumbar Interbody Fusion System (Reference device – K210800).

Performance data demonstrate that the ATLAS™ Expandable Osteotomy Wedge System is substantially equivalent to other legally marketed predicate systems.

## Non-Clinical and/or Clinical Tests Summary &amp; Conclusions
21 CFR 807.92(b)

The mechanical performance profile of the ATLAS™ Expandable Osteotomy Wedge System was assessed through static and dynamic testing in accordance with the following test methods:

- Static and dynamic compression testing (ASTM F2077-18)
- Static and dynamic compression shear testing (ASTM F2077-18)
- Subsidence testing (ASTM F2267-04)
- Expulsion testing (ASTM Draft Standard F-04.25.02.02)

The ATLAS™ Expandable Osteotomy Wedge System has the same intended use, indications for use, labeling, and technological characteristics as the predicate systems, including the same design features, geometries, sizes, and materials. The predicate device components are manufactured from titanium alloy. The ATLAS™ Expandable Osteotomy Wedge System components are manufactured from titanium alloy (ASTM F136) and molybdenum-rhenium alloy (ASTM F3273), which are the same materials as the IO™ Expandable Lumbar Interbody Fusion System (Reference device – K210800).

Performance data demonstrate that the ATLAS™ Expandable Osteotomy Wedge System is substantially equivalent to other legally marketed predicate systems.

---

**Source:** [https://fda.innolitics.com/device/K252733](https://fda.innolitics.com/device/K252733)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
