← Product Code [CDZ](/productcode/CDZ) · K252728

# IDS-iSYS Total Testosterone (K252728)

_Immunodiagnostic Systems Limited · CDZ · Apr 15, 2026 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252728

## Device Facts

- **Applicant:** Immunodiagnostic Systems Limited
- **Product Code:** [CDZ](/productcode/CDZ.md)
- **Decision Date:** Apr 15, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1680
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The IDS-iSYS Total Testosterone assay is an in vitro diagnostic device intended for the quantitative determination of testosterone in human serum or plasma on the IDS system. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## Device Story

In vitro diagnostic assay for quantitative testosterone measurement in human serum or plasma. Operates on IDS-iSYS Multi-Discipline Automated System. Uses competitive immunoassay principle; biotinylated sheep monoclonal anti-testosterone antibody binds sample testosterone; acridinium ester-labeled testosterone conjugate competes for binding sites. Chemiluminescence detection quantifies analyte concentration. Used in clinical laboratories by trained personnel. Output provides numerical testosterone levels (ng/dL) to assist clinicians in diagnosing/monitoring androgen-related endocrine disorders. Benefits include standardized, automated assessment of hormonal status.

## Clinical Evidence

Bench testing only. Performance evaluated per CLSI guidelines. LoB 4 ng/dL, LoD 8 ng/dL, LoQ 14 ng/dL. Linearity 5-1578 ng/dL. Precision (repeatability/reproducibility) CVs generally <12%. Method comparison (N=125) against commercial CLIA showed Passing-Bablok slope 0.96, r=1.00. Matrix equivalence confirmed for serum, SST, K2/K3 EDTA, and Heparin plasma. Interference testing performed for endogenous substances and biotin; biotin interference observed at concentrations >350 ng/mL.

## Technological Characteristics

Competitive immunoassay. Reagents: magnetic particles (streptavidin-coated), biotinylated sheep monoclonal anti-testosterone antibody, acridinium ester-labeled testosterone conjugate. Detection: Chemiluminescence. Form factor: Reagent cartridge for IDS-iSYS automated system. Connectivity: Integrated with IDS-iSYS platform. Preservatives: Sodium azide, ProClin 300. Standards: CLSI EP17-A (limits), EP05-A3 (precision), EP06-Ed2 (linearity), EP07-A3/EP07-ED3 (interference), EP09c (method comparison), C28-A3 (reference intervals).

## Regulatory Identification

A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## Predicate Devices

- Elecsys Testosterone II ([K211685](/device/K211685.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K252728

B Applicant

Immunodiagnostic Systems Limited

C Proprietary and Established Names

IDS-iSYS Total Testosterone

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  CDZ | Class I, reserved | 21 CFR 862.1680 - Testosterone Test System | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

New device

B Measurand:

Testosterone

C Type of Test:

Quantitative Immunoassay

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K252728 - Page 2 of 12

# III Intended Use/Indications for Use:

## A Intended Use(s):

See Indications for Use below.

## B Indication(s) for Use:

The IDS-iSYS Total Testosterone assay is an in vitro diagnostic device intended for the quantitative determination of testosterone in human serum or plasma on the IDS system.

Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.

## C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

## D Special Instrument Requirements:

IDS-iSYS Multi-Discipline Automated Analyzer (K091849)

# IV Device/System Characteristics:

## A Device Description:

The IDS-iSYS Total Testosterone reagent cartridge contains the following reagents:

MP3: Magnetic particles coated with Streptavidin in a PBS Pluronic buffer with sodium azide as preservative (&lt;0.1 %), 1 bottle, 2.5 mL

CONJ: Testosterone linked to mouse protein labelled with an acridinium ester derivative, in a phosphate buffer with ProClin® 300 as preservative (&lt;0.0015%), 1 bottle, 3.5 mL

Ab-BIOT: Anti-Testosterone sheep monoclonal antibody labelled with biotin, in a phosphate buffer containing bovine and sheep protein with ProClin® 300 as preservative (&lt;0.0015%), 1 bottle, 11.5 mL

BUF: MES buffer containing BSA (0.1%), 4-Cloro-m-cresolo (0.1 mg/mL), Tween 20 (0.055%) and ProClin® 300 (0.0025%), 1 bottle, 11.5 mL

{2}

K252728 - Page 3 of 12

B Principle of Operation:
The IDS-iSYS Total Testosterone test system uses a competitive chemiluminescent immunoassay format. Patient samples or calibrators (30 µL) are first incubated with biotinylated anti-testosterone monoclonal antibody, an acridinium-labeled testosterone conjugate and streptavidin labeled magnetic particles. The magnetic particles are captured using a magnet and a wash step is performed to remove any unbound analyte. Trigger reagents are added and the chemiluminescent signal produced by the acridinium label is inversely proportional to the testosterone concentration in the sample.

V Substantial Equivalence Information:

A Predicate Device Name(s):
Elecsys Testosterone II

B Predicate 510(k) Number(s):
K211685

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K252728 | K211685  |
| --- | --- | --- |
|  Device Trade Name | IDS-iSYS Total Testosterone | Elecsys Testosterone II  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Quantitative determination of Testosterone | Same  |
|  Test Principle | Competitive immunoassay | Same  |
|  General Device Characteristic Differences |  |   |
|  Detection Method | Chemiluminescence | Electrochemiluminescence  |
|  Measuring Range | 14 – 1500 ng/dL | 2.50 – 1500 ng/dL  |
|  Expected Range of Values (ng/dL) | Females, 21 - 49 years <14 – 51
Females, ≥ 50 years <14 – 48
Males, 21 - 49 years 213 – 818
Males, ≥ 50 years 180 - 711 | Females, 20 – 49 8.4 – 48.1
Females, ≥ 50 years 2.9 – 40.8
Males, 20 – 49 249 – 836
Males, ≥ 50 years 193 - 740  |

{3}

VI Standards/Guidance Documents Referenced:

Clinical &amp; Laboratory Standards Institute (CLSI) EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Third Edition

CLSI EP07-A3; Interference testing in Clinical Chemistry; Approved Guideline – 3rd Edition

CLSI-EP06; Evaluation of the Linearity of Quantitative Measurement Procedures, 2nd Edition

CLSI-EP09c; Measurement Procedure Comparison and Bias Estimation Using Patient Samples, 3rd Edition

CLSI-EP17-A2; Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, 2nd Edition

CLSI-E28-A3c; Defining Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, 3rd Edition

CLSI-EP37; Supplemental Tables for Interference Testing in Clinical Chemistry, 1st Edition

CLSI-EP25; Evaluation of Stability of In Vitro Medical Laboratory Test Reagents, 2nd Edition

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Precision and repeatability were evaluated according to CLSI guideline EP05-A3.

Repeatability

A precision study was conducted to estimate repeatability and within-laboratory precision. Eight human serum samples with testosterone concentrations spanning the analytical measuring interval were assayed in duplicate in two runs per day over 20 days using one reagent lot on one IDS-iSYS Multi-Discipline Automated Analyzer. A total of 80 replicates per sample were measured. Repeatability and within laboratory precision were calculated using a two-way nested ANOVA according to CLSI EP05-A3. The results are provided in the table below:

|  Sample ID | N | Mean Conc. (ng/dL) | Repeatability |   | Within Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV | SD | CV  |
|  S1 | 80 | 42.5 | 2.2 | 5.2% | 4.3 | 10.2%  |
|  S2 | 80 | 73.1 | 3.5 | 4.8% | 5.5 | 7.5%  |
|  S3 | 80 | 125 | 3.9 | 3.1% | 7.3 | 5.9%  |
|  S4 | 80 | 180 | 5.2 | 2.9% | 9.6 | 5.3%  |
|  S5 | 80 | 402 | 7.6 | 1.9% | 16.4 | 4.1%  |
|  S6 | 80 | 588 | 16.5 | 2.8% | 28.2 | 4.8%  |

K252728 - Page 4 of 12

{4}

|  Sample ID | N | Mean Conc. (ng/dL) | Repeatability |   | Within Laboratory  |   |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV | SD | CV  |
|  S7 | 80 | 1146 | 24.2 | 2.1% | 52.7 | 4.6%  |
|  S8 | 80 | 1228 | 27.1 | 2.2% | 45.9 | 3.7%  |

## Reproducibility – Site to Site

A reproducibility study was conducted in which eight (8) human serum samples with testosterone concentrations spanning the analytical measuring interval were tested using one reagent lot on three instruments (IDS-iSYS Multi-Discipline Automated Analyzer) at 3 sites by 3 operators (one operator per instrument/site). Each sample was tested in replicates of 5 per run, 1 run per day for 5 days for a total of 75 replicates per sample. The results are provided in the table below:

|  Sample ID | N | Mean Conc. (ng/dL) | Repeatability |   | Between Site |   | Reproducibility  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV | SD | CV | SD | CV  |
|  S1 | 75 | 43.8 | 3.2 | 7.3% | 0.0 | 0.0% | 4.0 | 9.2%  |
|  S2 | 75 | 79.3 | 4.4 | 5.5% | 4.6 | 5.8% | 8.8 | 11.1%  |
|  S3 | 75 | 140.5 | 5.4 | 3.8% | 6.7 | 4.7% | 14.2 | 10.1%  |
|  S4 | 75 | 198.3 | 8.0 | 4.1% | 9.5 | 4.8% | 16.6 | 8.4%  |
|  S5 | 75 | 442.9 | 9.7 | 2.2% | 22.7 | 5.1% | 33.1 | 7.5%  |
|  S6 | 75 | 614.4 | 23.4 | 3.8% | 36.3 | 5.9% | 52.3 | 8.5%  |
|  S7 | 75 | 1167.6 | 53.6 | 4.6% | 28.8 | 2.5% | 77.5 | 6.6%  |
|  S8 | 75 | 1258.7 | 37.8 | 3.0% | 52.4 | 4.2% | 85.4 | 6.8%  |

## Reproducibility – Lot to Lot

A reproducibility study was conducted in which eight (8) human serum samples with testosterone concentrations spanning the analytical measuring interval were tested by one operator using three reagent lots on one IDS-iSYS Multi-Discipline Automated Analyzer. Each sample was tested in replicates of 5 per run, 1 run per day for 5 days for a total of 75 replicates per sample. The results are provided in the table below:

|  Sample ID | N | Mean Conc. (ng/dL) | Repeatability |   | Between Lot |   | Reproducibility  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  SD | CV | SD | CV | SD | CV  |
|  S1 | 75 | 43.4 | 2.8 | 6.4% | 2.3 | 5.4% | 4.3 | 10.0%  |
|  S2 | 75 | 73.6 | 4.0 | 5.4% | 3.9 | 5.3% | 7.5 | 10.2%  |
|  S3 | 75 | 136.8 | 6.4 | 4.7% | 6.3 | 4.6% | 12.8 | 9.3%  |
|  S4 | 75 | 191.4 | 5.5 | 2.9% | 7.5 | 3.9% | 12.4 | 6.5%  |
|  S5 | 75 | 438.2 | 13.2 | 3.0% | 23.4 | 5.3% | 31.1 | 7.1%  |
|  S6 | 75 | 576.8 | 14.3 | 2.5% | 12.7 | 2.2% | 27.1 | 4.7%  |
|  S7 | 75 | 1137.8 | 40.0 | 3.5% | 29.5 | 2.6% | 57.1 | 5.0%  |
|  S8 | 75 | 1206.5 | 44.0 | 3.7% | 31.4 | 2.6% | 67.1 | 5.6%  |

K252728 - Page 5 of 12

{5}

K252728 - Page 6 of 12

2. Linearity:

A study was performed based on the CLSI guideline EP06-Ed2. Two dilution series spanning a concentration range from 8-1,500 ng/dL were prepared by mixing high and low serum samples pools. Samples were assayed in replicates of seven or nine. Linearity was evaluated using linear regression analysis. The deviation from linearity did not exceed -6.5% for samples with total testosterone concentrations within the claimed measuring range.

The combined linear regression for two dilution series was as follows:
Y = 0.97x - 1.44

3. Analytical Specificity/Interference:

Interference and cross-reactivity studies were conducted following the CLSI EP7-Ed3 guideline.

Endogenous Interference

Serum samples with testosterone concentrations of 30, 75, 300 and 1200 ng/dL were spiked with potentially interfering substances. The samples were assayed, and the total testosterone concentrations of the spiked samples were compared to control samples without interferent. No significant interference (≤±10% bias) was observed when the interfering substances were tested at the following concentrations:

|  Interferent | Highest Concentration tested without interference  |
| --- | --- |
|  Hemoglobin | 1000 mg/dL  |
|  Triglycerides | 1500 mg/dL  |
|  Bilirubin, conjugated | 40 mg/dL  |
|  Bilirubin, unconjugated | 40 mg/dL  |

The sponsor includes the following limitation in the labeling:

- Heterophilic antibodies in human serum can react with reagent immunoglobulins, interfering with in vitro immunoassays⁴. Patients routinely exposed to animals or to animal serum products can be prone to this interference and anomalous values may be observed.

4. Boscato, LM. and Stuart, MC., 'Heterophilic antibodies: a problem for all immunoassays'. Clin Chem, 34, 1988, pp 27–33

Biotin

To evaluate the candidate device’s susceptibility to biotin interference, biotin was spiked into serum samples containing different concentrations of testosterone (approximately 40, 70, 300, and 1100 ng/dL). Samples were assayed in multiple replicates. The sponsor defined no significant interference as ≤10% bias. The results are summarized below.

{6}

K252728 - Page 7 of 12

|  % Bias for Samples Containing Various Concentration of Biotin  |   |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Testosterone Concentration (ng/dL) | Biotin Concentration (ng/mL)  |   |   |   |   |   |   |
|   |  250 | 350 | 450 | 500 | 750 | 1000 | 3500  |
|  40 | -4% | 5% | 86% | >100% | >100% | >100% | >100%  |
|  70 | -1% | 4% | 52% | NT | NT | NT | >100%  |
|  300 | 2% | 10% | 50% | NT | NT | NT | NT  |
|  1100 | -3% | -1% | 14% | >100% | >100% | >100% | >100%  |

NT = Not Tested

The sponsor includes the following limitation in the labeling:

- Specimens that contain biotin at a concentration of 350 ng/mL demonstrate a less than or equal to ±10% change in results. Biotin concentrations greater than this may lead to falsely elevated results for patient samples. The recommended adult daily dietary intake for biotin is 30 µg/day. Over the counter dietary supplements promoted for use in hair, skin and nail health may contain 5-10 mg of biotin. Pharmacokinetic studies in healthy adults have shown that ingesting 5 mg of biotin can result in serum levels as high as 73 ng/mL. In rare cases, subjects are prescribed up to 300 mg of biotin per day for therapeutic applications, resulting in serum biotin levels as high as 1,160 ng/mL.

## Exogenous Interferences

The effect on quantitation of analyte in the presence of drugs was determined by comparing values obtained from samples spiked with 17 common pharmaceutical compounds with the reference sample (unspiked). All samples used were native human serum pools. Samples (with testosterone concentrations near 60 ng/dL and 500 ng/dL) were divided into aliquots and spiked with the common drug interferents. The reference sample without drug was spiked with the respective amount of solvent. The definition of significant interference was set to ≤10% bias compared to the reference sample.

|  Drug | Highest Drug Concentration tested without Interference  |
| --- | --- |
|  Acetaminophen | 15.6 mg/dL  |
|  Acetylcysteine | 15.0 mg/dL  |
|  Acetylsalicylic Acid | 3.0 mg/dL  |
|  Ampicillin | 7.5 mg/dL  |
|  Ascorbic acid | 5.25 mg/dL  |
|  Cefoxitin | 75.0 mg/dL  |
|  Cyclosporine | 0.18 mg/dL  |
|  Doxycycline | 1.8 mg/dL  |
|  Heparin | 330.0 IU/dL  |
|  Ibuprofen | 21.9 mg/dL  |
|  Itraconazole | 3.0 mg/dL  |
|  Levodopa | 0.75 mg/dL  |
|  Methyldopa | 2.25 mg/dL  |
|  Metronidazole | 12.3 mg/dL  |
|  Phenylbutazone | 10.7 mg/dL  |

{7}

Significant interference was observed for Testosterone undecanoate and Nandrolone at the concentrations tested above. The sponsor includes the following limitation in the labeling:

- A significant interference with Nandrolone and Testosterone undecanoate was identified. The IDS Total Testosterone assay does not distinguish between endogenous testosterone and testosterone derived from supplementation therapy. Do not test samples from individuals undergoing treatment with Nandrolone or Testosterone undecanoate using the IDS Total Testosterone method.”

## Cross Reactivity

A cross-reactivity study was performed to evaluate the following substances. Aliquots from pools of human serum with testosterone concentrations of 60 ng/dL and 500 ng/dL were spiked with potentially cross-reactive substances and measured in the presence or absence of the potential cross-reactants and cross reactivity was calculated using the following equation:

$$
\% \text{ Cross Reactivity} = 100 \times (\text{Average "spike" concentration} - \text{Average "blank concentration}) / \text{Spike concentration of cross reactant}
$$

|  Cross Reactant | Cross Reactant Concentration | % Cross Reactivity  |
| --- | --- | --- |
|  *11-Ketotestosterone | 200 ng/mL | 4.9%  |
|  11-Ketotestosterone | 100 ng/mL | 6.1%  |
|  *11-β-Hydroxy testosterone | 50 ng/mL | 24.3%  |
|  11-β-Hydroxy testosterone | 25 ng/mL | 27.4%  |
|  19-Norethisterone | 40 ng/mL | 0.80%  |
|  5α-Androstane-3β,17β-diol | 1000 ng/mL | 0.10%  |
|  5-α-Androstene-3β,17β-diol | 1000 ng/mL | 0.10%  |
|  Androstenedione | 100 ng/mL | 2.30%  |
|  Cortisol | 5000 ng/mL | n.d  |
|  Cortisone | 5000 ng/mL | n.d  |
|  Danazol | 1000 ng/mL | n.d  |
|  Dexamethasone | 2000 ng/mL | n.d  |
|  DHEA | 5000 ng/mL | n.d  |
|  DHEA-S | 50000 ng/mL | n.d  |
|  Dihydrotestosterone | 500 ng/mL | 0.40%  |
|  Estradiol | 5000 ng/mL | 0.10%  |
|  Estrone | 5000 ng/mL | n.d  |
|  Ethisterone | 1000 ng/mL | 0.20%  |
|  Norgestrel | 1000 ng/mL | 0.10%  |

K252728 - Page 8 of 12

{8}

|  Cross Reactant | Cross Reactant Concentration | % Cross Reactivity  |
| --- | --- | --- |
|  Prednisolone | 5000 ng/mL | n.d  |
|  Prednisone | 5000 ng/mL | n.d  |
|  Progesterone | 5000 ng/mL | n.d  |
|  Testosterone propionate | 100 ng/mL | n.d  |

*Serum with testosterone level ~20 ng/dL

Interference was observed for 11-Ketotestosterone and 11-β-Hydroxy testosterone at the concentrations tested above. The sponsor includes the following limitation in the labeling:

- A strong interaction with 11-Ketotestosterone and 11β-Hydroxy testosterone was found using the IDS Total Testosterone method. n.d. = not detectable; cross-reactivity is &lt; 0.1%.

4. Detection Limit and Assay Reportable Range:

**Limit of Blank (LoB)**

For determination of LoB four analyte free samples were measured in 5 replicates, once per day for 3 days, with 2 different reagent lots resulting in 15 replicates per sample for a total of 60 replicates per reagent lot on one IDS-iSYS analyzer. LoB was calculated according to the parametric function as described in CLSI EP17-A2.

The LoB claim in the labeling is 4 ng/dL.

**Limit of Detection (LoD)**

For determination of LoD, seven serum samples with low-analyte concentrations were measured in replicates of 5, with one reagent lot over 5 days, resulting in 25 replicates per sample for a total of 175 replicates per reagent lot on one IDS-iSYS Analyzer. LoD was calculated according to CLSI EP17-A2.

The LoD claim in the labeling is 8 ng/dL.

**Limit of Quantitation (LoQ)**

For the determination of LoQ, seven serum samples with low-analyte concentrations were measured in replicates of 5, with one reagent lot over 5 days, resulting in 25 replicates per sample for a total of 175 replicates per reagent lot on one IDS-iSYS Analyzer. The LoQ was defined as the concentration of analyte which has imprecision less than 20% CV.

The LoQ claim in the labeling is 14.0 ng/dL.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

The device is traceable to the certified reference testosterone material from the National Measurement Institute of Australia (NMIA), product code M914c, which is a certified reference material reviewed for compliance with ISO 15194.

K252728 - Page 9 of 12

{9}

# Sample stability

The sponsor provided information to support the following sample stability labeling claims using serum and  $\mathrm{K}_2\mathrm{EDTA}$  plasma samples:

|  Storage Condition | Duration  |
| --- | --- |
|  Room temperature (20±3°C) | 72 hours (3 days)  |
|  5±3°C (2 - 8°C) | 22 days  |
|  -20°C or below | 35 days (5 weeks)  |
|  Freeze/Thaw cycle | 3  |
|  On-board the system* | 6 hours  |

*Continuous on-board stability

# 6. Assay Cut-Off:

Not applicable

# B. Comparison Studies:

# 1. Method Comparison with Predicate Device:

A method comparison study was performed comparing the IDS-iSYS Testosterone assay to the predicate device, using a protocol based on CLSI EP09c-A3. A total of 125 native human serum samples with testosterone concentrations ranging from 14 to  $1490\mathrm{ng / dL}$  (as measured by the predicate device) were evaluated with the candidate and predicate devices in singlicate. The Passing-Bablok regression analysis results between the candidate device (dependent variable, y) and the comparator device (x, comparator), are shown below:

|  N | Concentration Range (ng/dL)* | Slope | Slope 95% CI | Intercept | Intercept 95% CI | Correlation Coefficient (r)  |
| --- | --- | --- | --- | --- | --- | --- |
|  125 | 14-1449 | 0.96 | 0.93 – 0.98 | 1.06 | -2.01 – 4.29 | 1.00  |

*As measured by the candidate device

# 2. Matrix Comparison:

A matrix comparison study was conducted to assess the equivalence between serum (serum without additives, serum gel separator tubes (SST)) and plasma  $\mathrm{(K_2EDTA}$ ,  $\mathrm{K}_3$  EDTA, Sodium Heparin and Lithium Heparin) sample matrices when using the IDS Total Testosterone assay. A total of 50 matched serum, serum gel separator tubes (SST) and plasma  $\mathrm{(K_2EDTA}$ , Lithium Heparin, and Sodium Heparin) and 39 matched serum and plasma  $\mathrm{(K_3EDTA}$  samples with concentrations ranging from 14 to  $1500\mathrm{ng / dL}$  were used. The samples were tested in singlicate by one operator on one instrument using one reagent lot. Data was assessed by Passing-Bablok regression analysis. The results are summarized below.

K252728 - Page 10 of 12

{10}

|  Tube type | N | Slope | Intercept | Correlation coefficient r  |
| --- | --- | --- | --- | --- |
|  SST | 50 | 0.99 | 0.42 | 1.00  |
|  K2 EDTA plasma | 50 | 0.99 | -1.25 | 1.00  |
|  K3 EDTA plasma | 39 | 0.99 | -0.49 | 1.00  |
|  Li Heparin plasma | 50 | 0.98 | 2.36 | 1.00  |
|  Na Heparin plasma | 50 | 0.99 | 1.94 | 1.00  |

The results demonstrate equivalency between serum, serum gel separator tubes (SST) and plasma (K2 EDTA, K3 EDTA, Lithium Heparin, and Sodium Heparin) sample matrices.

# B Clinical Studies:

1. Clinical Sensitivity:

Not Applicable

2. Clinical Specificity:

Not Applicable

3. Clinical Cut-Off:

Not Applicable

4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not Applicable

# C Expected Values/Reference Range:

A reference interval study was performed for the total Testosterone assay in accordance with the CLSI EP28-A3c guideline. A total of 776 adult serum samples were collected in the USA from apparently healthy individuals. The sample groups tested consisted of:

497 males between 21 and 78 years of age
279 females between 21 and 78 years of age

The samples were categorized into two age groups: 21 to 49, and  $\geq 50$  years. The data were analyzed applying a nonparametric method with  $90\%$  confidence interval using the 5th and 95th percentiles of measured data. The resulting reference interval is summarized in the following table:

K252728 - Page 11 of 12

{11}

|  Males | 21 to 49 years | ≥ 50 years  |
| --- | --- | --- |
|  N of subjects | 266 | 231  |
|  Median (ng/dL) | 459 | 402  |
|  Observed Range (ng/dL) (5th to 95th percentile) | 213-818 | 180-711  |
|  Females | 21 to 49 years | ≥ 50 years  |
| --- | --- | --- |
|  N of subjects | 119 | 108  |
|  Median (ng/dL) | 23 | 22  |
|  Observed Range (ng/dL) (5th to 95th percentile) | <14-51 | <14-48  |

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K252728 - Page 12 of 12

---

**Source:** [https://fda.innolitics.com/device/K252728](https://fda.innolitics.com/device/K252728)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
