Biopsy Forceps

K252722 · Fehling Instruments GmbH · DWZ · Sep 30, 2025 · Cardiovascular

Device Facts

Record IDK252722
Device NameBiopsy Forceps
ApplicantFehling Instruments GmbH
Product CodeDWZ · Cardiovascular
Decision DateSep 30, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4075
Device ClassClass 2

Indications for Use

FEHLING Biopsy Forceps are used to obtain endomyocardial biopsy specimens from the right and left ventricle via percutaneous arterial or venous approach.

Device Story

Biopsy Forceps designed for percutaneous access to right or left heart ventricles to obtain diagnostic tissue samples. Device comprises three main components: ergonomic actuating handle, flexible shaft, and surgical stainless steel cutting jaws. Physician operates handle to steer device and activate distal jaws to excise tissue. Provided sterile for single use. Benefits patient by enabling minimally invasive cardiac tissue sampling for diagnostic purposes.

Clinical Evidence

No clinical testing was required. No additional non-clinical testing was required as the modifications did not necessitate further testing.

Technological Characteristics

Components include an actuating handle, flexible shaft, and surgical stainless steel cutting jaws. Device is a mechanical, non-powered, single-use, sterile instrument. Principle of operation is manual mechanical actuation of distal jaws for tissue excision.

Indications for Use

Indicated for patients requiring endomyocardial biopsy specimens from the right or left ventricle via percutaneous arterial or venous approach.

Regulatory Classification

Identification

An endomyocardial biopsy device is a device used in a catheterization procedure to remove samples of tissue from the inner wall of the heart.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 30, 2025 FEHLING Instruments GmbH Imad Ziate Quality Management Hanauer Landstr. 7A Karlstein, Bavaria 63791 Germany Re: K252722 Trade/Device Name: Biopsy Forceps Regulation Number: 21 CFR 870.4075 Regulation Name: Endomyocardial Biopsy Device Regulatory Class: Class II Product Code: DWZ Dated: August 28, 2025 Received: August 28, 2025 Dear Imad Ziate: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252722 - Imad Ziate Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252722 - Imad Ziate Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Samuel G. Raben -S Samuel Raben, PhD Assistant Director (Acting) DHT2B: Division of Circulatory Support, Structural, and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252722 | | | Device Name Biopsy Forceps | | | Indications for Use (Describe) FEHLING Biopsy Forceps are used to obtain endomyocardial biopsy specimens from the right and left ventricle via percutaneous arterial or venous approach. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K252722 Page 1 of 2 | 510(k) #: K252722 | 510(k) Summary | Prepared on: 2025-09-15 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | FEHLING Instruments GmbH | | | Applicant Address | Hanauer Landstr. 7A Karlstein Bavaria 63791 Germany | | | Applicant Contact Telephone | +4906188957429 | | | Applicant Contact | Mr. Imad Ziate | | | Applicant Contact Email | imad_ziate@fehling-instruments.de | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Biopsy Forceps | | | Common Name | Endomyocardial biopsy device | | | Classification Name | Device, Biopsy, Endomyocardial | | | Regulation Number | 870.4075 | | | Product Code(s) | DWZ | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K170726 | Biopsy Forceps | DWZ | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The Biopsy Forceps are designed to allow percutaneous access to the right or left ventricles of the heart to obtain diagnostic tissue samples. The forceps consist of three main components: 1. an actuating handle ergonomically designed for comfortable use, 2. a flexible shaft, 3. and surgical stainless steel cutting jaws. At the distal end of the forceps is a pair of stainless steel jaws used to obtain the tissue samples. At the proximately end of the forceps is the actuation handle used to activate the jaws and steer the device. The product is provided single for use and sterile. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | FEHLING Biopsy Forceps are used to obtain endomyocardial biopsy specimens from the right and left ventricle via percutaneous arterial or venous approach. | | | | Indications for Use Comparison | | 21 CFR 807.92(a)(5) | | The indications for use are the same as the predicate device. | | | | Technological Comparison | | 21 CFR 807.92(a)(6) | {5} K252722 Page 2 of 2 The design and principle of operation of the device are identical to those of the predicate device. The material of components was changed to similar materials. All other materials and technological aspects are identical. **Non-Clinical and/or Clinical Tests Summary & Conclusions** 21 CFR 807.92(b) There was no clinical testing required to support the medical device. No additional non-clinical testing was required as the modifications to the device did not require any testing.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...