MAASTRO Rectal Applicator Set

K252717 · Varian Medical Systems · JAQ · May 20, 2026 · Radiology

Device Facts

Record IDK252717
Device NameMAASTRO Rectal Applicator Set
ApplicantVarian Medical Systems
Product CodeJAQ · Radiology
Decision DateMay 20, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The MAASTRO Rectal Applicator Set is intended for use for cancer treatment of the rectum using HDR brachytherapy.

Device Story

MAASTRO Rectal Applicator Set enables HDR brachytherapy for rectal cancer. System includes proctoscope, dilator, illumination adapter, targeting window applicator, and fixation screw. Physician inserts proctoscope and applicator through anus; places applicator in close contact with tumor. Applicator contains multiple channels for HDR 192Ir source delivery; design allows uniform dose delivery to tumor while shielding healthy tissue. Used in brachytherapy treatment rooms. Compatible with GammaMedplus and BRAVOS afterloader systems. Device provides mechanical positioning control and adjustment of treatment site. Benefits include targeted radiation delivery to rectal tumors. Device is MR Unsafe and not CT compatible.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including mechanical stability (20N pull force), dose profile verification, biocompatibility (ISO 10993), and human factors validation (IEC 62366). Literature cited supports dose prescription and fractionation guidance for contact brachytherapy.

Technological Characteristics

Materials: Polyamide (PA) 2200, Stainless Steel, PEEK. Components: Proctoscope, dilator, illumination adapter, targeting window applicator. Energy: HDR 192Ir source (via afterloader). Connectivity: Compatible with GammaMedplus and BRAVOS afterloaders. Sterilization: Steam sterilization or high-level disinfection. MR Unsafe. Not CT compatible. Complies with ISO 10993, ISO 14971, IEC 60601-2-17, and ASTM F2503-23.

Indications for Use

Indicated for cancer treatment of the rectum using HDR brachytherapy in adult patients (18 years and older) as determined by a prescribing physician.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION May 20, 2026 Varian Medical Systems Lynn Allman Senior Director of Regulatory Affairs 3100 Hansen Way Palo Alto, California 94304 Re: K252717 Trade/Device Name: MAASTRO Rectal Applicator Set Regulation Number: 21 CFR 892.5700 Regulation Name: Remote Controlled Radionuclide Applicator System Regulatory Class: Class II Product Code: JAQ Dated: August 27, 2025 Received: August 28, 2025 Dear Lynn Allman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252717 - Lynn Allman Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252717 - Lynn Allman Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JULIE SULLIVAN -S Julie Sullivan, Ph.D. Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252717 | | | Device Name MAASTRO Rectal Applicator Set | | | Indications for Use (Describe) The MAASTRO Rectal Applicator Set is intended for use for cancer treatment of the rectum using HDR brachytherapy. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Varian A Siemens Healthineers Company # Premarket Notification - 510(k) Summary Traditional 510(k) Submission for MAASTRO Rectal Applicator Set I. Submitter's Name Varian Medical Systems 3100 Hansen Way Palo Alto, CA 94304 Contact Name: Lynn, Allman, PhD., Senior Director Regulatory Affairs Phone: (650) 424-5369 E-mail: submissions.support@varian.com Date Prepared: May 19th, 2026 II. Device Information Proprietary Name: MAASTRO Rectal Applicator Set Classification Name: Remote Controlled Radionucleotide Applicator System Regulation Number: §892.5700 Product Code: JAQ III. Predicate Device Universal Cylinder Applicator Family (K193240) IV. Device Description The MAASTRO Rectal Applicator Set enables HDR treatment of rectal cancer. This involves inserting a proctoscope and applicator through the anus and placing it in close contact with the tumor. The dilator enables smooth insertion. The Illumination Adapter serves as interface to an external illumination device. The Targeting Window Applicator allows positioning control and adjustment of the treatment position. All this can be fixed to the proctoscope using the fixation screw. The Applicator for Proctoscope and Dilator for Proctoscope components are designed for single use. Key performance characteristics are as follows: - Suitable for treating body sites allowing insertion of a catheter. 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set Page 1 of 9 {5} Varian A Siemens Healthineers Company - Not CT Compatible - MR Compatibility: Unsafe - The Proctoscope is placed eye-guided allowing placement close to the tumor - The cylindrical Applicator containing channels is inserted into proctoscope up to the tumor - The channels within the Applicator are aligned to allow uniform high dose delivery to the tumor - Healthy tissue is spared from radiation by shielding V. Intended Use The MAASTRO Rectal Applicator Set is intended for use for cancer treatment of the rectum using HDR brachytherapy. VI. Comparison of Technological Characteristics with the Predicate Device Table 1: Comparison of Subject Device to Predicate Device | Feature and/or Specification of new/modified device | Predicate Device: Universal Cylinder Applicator Family (K193240) *specifically, the universal multi-channel cylinder set | Subject Device: MAASTRO Rectal Applicator Set | Comparison | | --- | --- | --- | --- | | Intended use | The Universal Multi-Channel Cylinder Set is intended for use when performing HDR or PDR brachytherapy within a CT or MRI environment. | The MAASTRO Rectal Applicator Set is intended for use for cancer treatment of the rectum using HDR brachytherapy. | Use only in the rectum for cancer treatment by means of HDR brachytherapy Please note, the subject device is being compared specifically to the multi-channel cylinder set from the predicate device. | | Indications for Use | The Universal Multi-Channel Cylinder Set are indicated for use for cancer treatment of the vagina, vaginal stump, cervix, uterus, endometrium, and rectum using HDR or PDR brachytherapy. | The MAASTRO Rectal Applicator Set is indicated for use for cancer treatment of the rectum using HDR brachytherapy | Use only in the rectum for cancer treatment by means of HDR brachytherapy | 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set {6} Varian A Siemens Healthineers Company | Feature and/or Specification of new/modified device | Predicate Device: Universal Cylinder Applicator Family (K193240) *specifically, the universal multi-channel cylinder set | Subject Device: MAASTRO Rectal Applicator Set | Comparison | | --- | --- | --- | --- | | Patient Target Group | N/A | The patient target group includes adult patients of 18 years and above whose cancer can be treated using HDR brachytherapy, as determined by the prescribing physician. | Clarification that the intended patient population is limited to adults only. | | Compatible Afterloader | GammaMedplus Afterloader Series VariSource 200 and iX Afterloader BRAVOS Afterloader System | GammaMedplus Afterloader Series BRAVOS Afterloader System | The subject device is not compatible with the VariSource 200 and iX Afterloader due to product being end of service at the end of 2025. | | Design | Components: * Please note, the subject device is being compared specifically to the multi-channel cylinder set from the predicate device. | Components: | | 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set Page 3 of 9 {7} Varian A Siemens Healthineers Company | Feature and/or Specification of new/modified device | Predicate Device: Universal Cylinder Applicator Family (K193240) *specifically, the universal multi-channel cylinder set | Subject Device: MAASTRO Rectal Applicator Set | Comparison | | --- | --- | --- | --- | | | Universal Multi-channel Cylinder Set: Diameters 25, 30, 35, 40 mm Rigid Guide Tube 250 mm length Clamping segment | MAASTRO rectal cylindrical applicator: diameters 20 and 24 mm Proctoscope: diameters 26 and 30 mm (requesting clearance with this submission) Dilator: diameters 20 and 24 mm Illumination adapters: for proctoscope diameter 26 mm, and proctoscope diameter 30 mm Targeting Window applicator for proctoscope diameter 26 mm and proctoscope diameter 30 mm. Fixation Screw | The MAASTRO Rectal Applicator Set is a new device that is indicated specifically for treatment of cancer in the rectum. The differences in dimension of the cylindrical applicator of the MAASTRO Rectal Applicator Set are due to the treatment design for the rectum. The proctoscope is specifically needed to access to the rectum, but functions as a ‘guide’ to the treatment area in conjunction with the dilator, illumination adapters, and targeting window applicator similar to the rigid guide tube and clamping segment. Similar to the Multi-Channel Cylinder Set, the rectal cylindrical applicator includes multiple channels which are inserted into a proctoscope through the anus and placing it in close contact with the tumor. | | Materials | PPSU (Polyphenylsulfone) PEEK (Polyether ether ketone) Titanium | Polyamide (PA) 2200 Stainless Steel PEEK (Polyether ether ketone) | While the MAASTRO Rectal Applicator Set is a new device and is comprised of different biocompatible material-, the materials used in the subject device have been tested to and comply with the same biological standards as the predicate device. Furthermore, the materials of the subject device are from common polymers and metals with well-established biocompatibility profiles. | 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set Page 4 of 9 {8} Varian A Siemens Healthineers Company | Feature and/or Specification of new/modified device | Predicate Device: Universal Cylinder Applicator Family (K193240) *specifically, the universal multi-channel cylinder set | Subject Device: MAASTRO Rectal Applicator Set | Comparison | | --- | --- | --- | --- | | Packing | Individual | Individual | Similar | | Sterility | Provided non-sterile | Provided non-sterile | Similar | | Sterilization method (Bronchial catheter and Monofil) | Steam Sterilization 132°C for 4 min 134°C for 3 min 134°C for 5 min | Steam Sterilization 132°C for 4 min 134°C for 3 min 134°C for 5 min Or High-Level Disinfection | Added high level disinfection for consistency with other, Varian applicator sets | | Reprocessing | Universal Multi-Channel Cylinder: 100 cycles Rigid Guide Tube, 250 mm length: 500 cycles Clamping segment: 500 cycles | Applicator: Single Use Proctoscope: 200 cycles Dilator: Single Use Illumination adapters: 200 cycles Targeting Window applicator: 200 cycles | Based on materials and testing, the MAASTRO Rectal Applicator Set is comprised of single use and reusable components. | | Biocompatibility | Fully biocompatible | Fully biocompatible | Similar | | Anatomical sites | Vagina, vaginal stump, cervix, endometrium, uterus, and rectum | Rectum | The subject device is indicated for use only in the rectum | 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set Page 5 of 9 {9} Varian A Siemens Healthineers Company | Feature and/or Specification of new/modified device | Predicate Device: Universal Cylinder Applicator Family (K193240) *specifically, the universal multi-channel cylinder set | Subject Device: MAASTRO Rectal Applicator Set | Comparison | | --- | --- | --- | --- | | Compatibility with the environment and other devices | CT compatible MR Conditional | Not CT compatible MR Unsafe | The MAASTRO Rectal Applicator Set is partially made of metallic and conductive materials and is, therefore, unsafe in a magnetic environment | | Where used | Brachytherapy treatment room | Brachytherapy treatment room | Similar | ## VII. Summary of Performance Testing (Non-Clinical Testing) The following performance data was provided in support of the substantial equivalence determination. ### Dose Distribution: The following literature provides guidance on dose prescription and fractionation: - Jean-Pierre Gerard et al, paper, Targeted Radiotherapy Using Contact X-ray Brachytherapy 50 kV: This paper provides guidance on appropriate tumor staging for contact brachytherapy. - Murillo Bellezzo et al, paper, Advanced design, simulation, and dosimetry of a novel rectal applicator for contact brachytherapy with a conventional HDR 192Ir source: This paper explains how the MAASTRO applicator design uses HDR 192Ir sources to deliver a dose distribution similar to those of contact X-ray brachytherapy devices. The paper also discusses the depth dose curves that this device can achieve. ### Biocompatibility Testing: The compatibility of the skin-contact component material in the finished product meets the biocompatibility requirements. The Biological Evaluation Tests are in compliance with the standards of ISO 10993-1 and its applicable parts, "Biological Evaluation of Medical Device – Part 1: Evaluation and Testing Within a Risk Management Process". Biocompatibility testing conducted included: Cytotoxicity, Irritation, and Sensitization. 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set {10} Varian A Siemens Healthineers Company ## Cleaning, Disinfection and Sterilization Testing: Cleaning, disinfection and sterilization testing was conducted to demonstrate the device components may be sterilized effectively (as appropriate) and the device can be used and sterilized for the specified number of times. Human factors validation study was conducted according to the standard IEC 62366 to verify that the device performs well as intended for the intended users, uses, and use environments. ## Mechanical and Acoustic Testing: MAASTRO Rectal Applicator Set Applicator Set has undergone formal design verification and design validation testing. Design verification and design validation testing demonstrates that the device performs as intended. - Mechanical Stability Testing: The applicator was tested to withstand mechanical loads which occur during normal application without functional damage. The device maintained structural integrity and functional performance following repeated use and reprocessing cycles and application of 20N pull force. - Dose Profile: Testing demonstrated that the applicator provides a suitable dose distribution within the target region, including acceptable range of surface dose and dose fall-off. Evaluation of dose distribution confirmed that shielding features and applicator design effectively limit radiation exposure to surrounding tissue, consistent with design requirements. Design verification and design validation testing was performed according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System Standard, and ISO 14971 Risk Management Standard. ## Magnetic Resonance Testing (MR): MR Compatibility testing was completed, and documentation was provided to FDA in support of the substantial equivalence determination. The MAASTRO Rectal Applicator Set is MR Unsafe. This testing was conducted in accordance with ASTM F2503-23 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment. ## Electromagnetic Compatibility (EMC) Testing: The electromagnetic compatibility of the MAASTRO Rectal Applicator Set was assessed by testing the device to the requirements of IEC 60601-1 and IEC 60601-2-17. The results of the testing demonstrate that the subject device is compatible in an electromagnetic setting. 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set {11} Varian A Siemens Healthineers Company Use of Consensus Standards: The following list of FDA-recognized, voluntary consensus standards were utilized in the design and evaluation of the subject device's safety and efficacy. | Standard Number | Standard Title | | --- | --- | | ANSI/AAMI/ISO 14971:2019 | Medical devices – Applications of risk management to medical devices | | EN ISO 15223-1:2021 | Medical devices – Symbols to be used with the information to be supplied by the manufacturer – Part 1: General requirements | | EN ISO 20417: 2021 | Information supplied by the manufacturer of medical devices | | IEC 62366-1:2015+A1:2020 | Application of Usability Engineering to Medical Devices | | ANSI/AAMI ST98:2022 | Cleaning Validation of health care products –cleaning process for medical devices | | ANSI / AAMI ST79:2017(R)2022 | Comprehensive guide to steam sterilization and sterility assurance in health care facilities | | ISO 17664-1:2021 | Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices –Part 1: Critical and semi-critical medical devices | | EN ISO 11607-1:2020 | Packaging for terminally sterilized medical devices –Part 1: Requirements for materials, sterile barrier systems and packaging systems | | EN ISO 17665-1:2006 | Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices | | ISTA 3A (2018) | General Simulation Performance Tests for packaged-products weighing 150 lb. (70 kg) or less (standard, small, flat or elongated) | | EN ISO 10993-1:2020 | Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. | | EN ISO 10993-5:2009 | Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity | | EN ISO 10993-10:2021 | Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization | | ISO 10993-11:2017 | Biological evaluation of medical devices – Part 11: Tests for systemic toxicity | | EN ISO 10993-12:2021 | Biological evaluation of medical devices –Part 12: Sample preparation and reference materials | | ISO 10993-18:2020/ AMD 1:2022 | Biological evaluation of medical devices -- Part 18: Chemical characterization of medical device materials within a risk management process | | EN ISO 10993-23:2021 | Biological evaluation of medical devices – Part 23: Tests for irritation | 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set {12} Varian A Siemens Healthineers Company | IEC 60601-2-17:2013 | Medical electrical equipment - Part 2-17: Particular requirements for the basic safety and essential performance of automatically-controlled brachytherapy after loading equipment | | --- | --- | | ASTM F2503 –23 | Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment | | ASTM D4332-22 | Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing | | EN ISO 10993-2:2022 | Biological evaluation of medical devices. –Part 2: Animal welfare requirements | | EN ISO 10993-17:2023 | Biological evaluation of medical devices - Part 17: Establishment of allowable limits for leachable substances | ## VIII. Determination of Substantial Equivalence to the Predicate Varian believes the major technological characteristics are substantially equivalent to the predicate device and the differences do not raise new questions of safety and effectiveness compared to the predicate. The results of verification and validation as well as conformance to relevant safety standards demonstrate that the device meets the safety and performance criteria. Varian considers the MAASTRO Rectal Applicator Set to be as safe and effective as the predicate and therefore substantially equivalent to the predicate device. 510(k) Summary Traditional 510(k) Application MAASTRO Rectal Applicator Set
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...