← Product Code [IYN](/productcode/IYN) · K252709

# Wireless Probe Type Ultrasound Scanner (K252709)

_Xuzhou Yongkang Electronic Science Technology Co., Ltd. · IYN · Apr 2, 2026 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252709

## Device Facts

- **Applicant:** Xuzhou Yongkang Electronic Science Technology Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Apr 2, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, Pediatric

## Indications for Use

Diagnostic ultrasound imaging and fluid flow analysis of the human body.

## Device Story

Portable, software-controlled ultrasound system; acquires/displays real-time ultrasound data via wireless transducers. Transducers use Wi-Fi to communicate with mobile/PC app; Bluetooth used for device discovery. System features 32-channel digital beamformer with 96 or 128 array elements; advanced signal processing enhances image quality. Used in hospitals, clinics, physician offices, and point-of-care settings by trained healthcare professionals. App provides interface for mode control, image acquisition, measurement, and storage. Output displayed on smartphone, tablet, or PC; assists clinicians in diagnostic imaging and fluid flow analysis. Benefits include portable, real-time diagnostic visualization.

## Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical performance testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), acoustic output (IEC 60601-2-37, IEC 62359), software validation (IEC 62304), and cybersecurity assessment.

## Technological Characteristics

Portable wireless transducers; 32-channel digital beamformer; 96 or 128 array elements. Wi-Fi/Bluetooth connectivity. Powered by rechargeable Li-ion battery. Compatible with Windows 11 Pro, Android 10+, iOS 17.6.1+. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, IEC 62133-2, ISO 10993-1, IEC 62304, ISO 14971.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- NUSONO Handheld Ultrasound Scanner (NUSONO-C35); NUSONO Handheld Ultrasound Scanner (NUSONO-L75); NUSONO Handheld Ultrasound Scanner (NUSONO-P25) ([K242681](/device/K242681.md))

## Reference Devices

- Leltek Ultrasound Imaging System (Model: LU700 Series) ([K222365](/device/K222365.md))
- Clarius Ultrasound Scanner ([K192107](/device/K192107.md))
- SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SonoFinder Wireless Probe Type Ultrasound Scanner (Model: SF14L25) ([K220169](/device/K220169.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

April 2, 2026

Xuzhou Yongkang Electronic Science Technology Co., Ltd.
% Rachel Yu
Official Correspondent
ZHIHE Info-Tech Co., Ltd.
Room 616, Building 1, No. 1 Huayun Road, Industrial Park
Suzhou, Jiangsu 215134
CHINA

Re: K252709
Trade/Device Name: Wireless Probe Type Ultrasound Scanner
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX
Dated: August 25, 2025
Received: August 27, 2025

Dear Rachel Yu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}

K252709 - Rachel Yu
Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-

{2}

K252709 - Rachel Yu
Page 3

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Kaitlyn Mcdonald -S
for
Yanna Kang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K252709  |   |
|  Device Name Wireless Probe Type Ultrasound Scanner  |   |
|  Indications for Use (Describe)  |   |
|  The Wireless Probe Type Ultrasound Scanner is a portable and software-based ultrasound imaging system, indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: PIU1A,PEU1A,PIU1C,PEU1C,PAU2C: Small Organ, Musculo-skeletal(Conventional and Superficial), Peripheral vessels PIU1B,PEU1B: Fetal, Abdominal, Pediatric, Ob/GYN, Urology; PIU1D,PEU1D: Fetal, Abdominal, Pediatric, Ob/GYN, Urology, Cardiac Adult, Cardiac Pediatric; PIU2A,PEU2A, PAU2A: Small Organ, Musculo-skeletal(Conventional and Superficial), Peripheral vessels, Fetal, Abdominal, Pediatric, Ob/GYN, Urology; PIU2B,PEU2B, PAU2B: Small Organ, Musculo-skeletal(Conventional and Superficial), Peripheral vessels, Fetal, Abdominal, Pediatric, Ob/GYN, Urology, Cardiac Adult, Cardiac Pediatric;  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

510(k) Summary K252709

I Submitter

Device Submitter: Xuzhou Yongkang Electronic Science Technology Co., Ltd.
Address: 2nd &amp; 3rd Floor, Building 5, Block B, Huaihai Biomedical Industry Park, No. 5 Tengzhai North Road, Xuzhou High-tech Industrial Development Zone, Xuzhou, Jiangsu 221000, P.R. China
Contact Person: Xiao Wan
Title: +86 150 6214 9600
E-mail: wanxiao@yonker.cn
Date Prepared: March 10, 2026

II Device

Trade/Device Name: Wireless Probe Type Ultrasound Scanner
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic pulsed doppler imaging system
Regulation Class: Class II
Product code: IYN, ITX, IYO
Review Panel: Radiology

III Predicate Device

|  Items | Device Name | 510(k) Number  |
| --- | --- | --- |
|  Primary Predicate | NUSONO Handheld Ultrasound Scanner (NUSONO-C35);
NUSONO Handheld Ultrasound Scanner (NUSONO-L75);
NUSONO Handheld Ultrasound Scanner (NUSONO-P25) | K242681  |
|  Reference | Leltek Ultrasound Imaging System (Model: LU700 Series) | K222365  |
|  Reference | Clarius Ultrasound Scanner | K192107  |
|  Reference | SonoMe Wireless Probe Type Ultrasound Scanner (Model: 5C, 5CB, H5C, 10L, 14L, 10LB, H10L, H5C10L) and SonoFinder Wireless Probe Type Ultrasound Scanner (Model: SF14L25) | K220169  |

IV Description of the Device:

The Wireless Probe Type Ultrasound Scanner is a portable, software-controlled diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through an off-the-shelf (OTS) Windows 11 Pro, Android 10 or higher, and iOS 17.6.1 or higher based mobile or PC device. The system consists of a series of wireless transducers employing Wi-Fi-based technology to communicate with the Wireless Probe Type Ultrasound Scanner App on standard smartphone, tablet, or PC devices via direct Wi-Fi connection, with Bluetooth supported for device discovery. The application provides the interface for mode/setting control and image display, acquisition, measurement, and storage functions. The 32-channel digital beamformer with 96 or 128 array elements (depending on the probe model) and advanced image signal processing technology maximize the performance of all imaging

1 / 6

{5}

transducer elements to enhance diagnostic utility and confidence.

This transportable system is intended for use in environments where healthcare is provided by trained healthcare professionals. Intended users will be qualified and appropriately trained healthcare professionals with ultrasound experience.

The Wireless Probe Type Ultrasound Scanner product includes:

a) Scanner models:
- PIU1A; PIU1B; PIU1C; PIU1D; PIU2A; PIU2B
- PEU1A; PEU1B; PEU1C; PEU1D; PEU2A; PEU2B
- PAU2A; PAU2B; PAU2C

b) Components:
- Software:
- Wireless Probe Type Ultrasound Scanner App for Windows 11 Pro and compatible
- Wireless Probe Type Ultrasound Scanner App for Android 10 or higher
- Wireless Probe Type Ultrasound Scanner App for iOS 17.6.1 or higher
- Wireless charging dock
- USB Type-C charging cable (integrated with charging dock)
- Power adapter

## V Intended Use of the Device

Diagnostic ultrasound imaging and fluid flow analysis of the human body.

## Indication for Use:

The Wireless Probe Type Ultrasound Scanner is a portable and software-based ultrasound imaging system, indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:

PIU1A,PEU1A,PIU1C,PEU1C,PAU2C: Small Organ, Musculo-skeletal(Conventional and Superficial), Peripheral vessels

PIU1B,PEU1B: Fetal, Abdominal, Pediatric, Ob/GYN, Urology;

PIU1D,PEU1D: Fetal, Abdominal, Pediatric, Ob/GYN, Urology, Cardiac Adult, Cardiac Pediatric;

PIU2A,PEU2A, PAU2A: Small Organ, Musculo-skeletal(Conventional and Superficial), Peripheral vessels, Fetal, Abdominal, Pediatric, Ob/GYN, Urology;

PIU2B,PEU2B, PAU2B: Small Organ, Musculo-skeletal(Conventional and Superficial), Peripheral vessels, Fetal, Abdominal, Pediatric, Ob/GYN, Urology, Cardiac Adult, Cardiac Pediatric

The Wireless Probe Type Ultrasound Scanner provides diagnostic ultrasound imaging in B mode, M mode, Color Doppler mode, PW mode and combine mode (B+M,B+PW,B+CF,B+PDI,B+DPDI,B+PW+CF), intended for use in environments where healthcare is provided by trained healthcare professionals. The environments where the system can be used include physician offices, clinics, hospitals, and clinical point of care for diagnosis of patients except environments where intensity of electromagnetic disturbances is high.

2 / 6

{6}

VI Determination of SE

The subject device, "Wireless Probe Type Ultrasound Scanner," is a Track 3 system that utilizes the same fundamental scientific technology as the primary predicate device, NUSONO Handheld Ultrasound Scanner (NUSONO-C35); NUSONO Handheld Ultrasound Scanner (NUSONO-L75); NUSONO Handheld Ultrasound Scanner (NUSONO-P25)(K242681), along with the reference devices: Leltek Ultrasound Imaging System (Model: LU700 Series (K222365), Clarius Ultrasound Scanner (K192107), and SonoMe Wireless Probe Type Ultrasound Scanner/H5C10L (K220169). The Wireless Probe Type Ultrasound Scanner shares identical indications for use with the primary predicate device and at least one model of the reference devices. A detailed comparison between the predicate device and the subject device is provided below:

|  Items | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- |
|  510k number | Current submission | K242681 | K222365 | K192107 | K220169  |
|  Device name | Wireless Probe Type Ultrasound Scanner | NUSONO Handheld Ultrasound Scanner (NUSONO-C35); NUSONO Handheld Ultrasound Scanner (NUSONO-L75); NUSONO Handheld Ultrasound Scanner (NUSONO-P25) | Leltek Ultrasound Imaging System (Model: LU700 Series) | Clarius Ultrasound Scanner | SonoMe Wireless Probe Type Ultrasound Scanner / H5C10L  |
|  Product code | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX | IYN, IYO, ITX  |
|  Intended use | Diagnostic ultrasound imaging and fluid flow analysis of the human body. | Diagnostic ultrasound imaging and fluid flow analysis of the human body. | Diagnostic ultrasound imaging or fluid flow analysis of the human body. | Diagnostic ultrasound imaging and fluid flow analysis. | SonoMe is indicated for examining the adult, pregnant woman, and children. This products is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices. An appropriately trained healthcare professionals can have operator qualifications. The device use settings are intended in hospital clinic, and medical office settings. The general clinical applications include fetal/obstetrics(OB), gynecology(GYN), abdominal, small organ and peripheral vessel imaging. The modes of operation are B-mode, PDI(Power Doppler Imaging) mode, PW(Pulse Wave Spectral Doppler) mode and Harmonic mode.  |
|  Indications for use | - Fetal
- Abdominal
- Pediatric
- Small organ
- Musculoskeletal (conventional) | - Fetal
- Abdominal
- Pediatric
- Small organ
- Musculoskeletal (conventional) | - Ophthalmic
- Fetal
- Abdominal
- Pediatric
- Small organ
- Neonatal cephalic
- Musculoskeletal (conventional) | - Fetal
- Abdominal
- Intra-operative
- Pediatric
- Adult cephalic
- | - Fetal
- Abdominal
- Small organ
-  |

3 / 6

{7}

|  Items | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- |
|   | - Musculoskeletal (superficial)
- Urology
-Obstetrics(OB)/ Gynecology(GYN)
- Cardiac adult
- Cardiac pediatric
- Peripheral vessel
-
-
- | - Musculoskeletal (superficial)
- Urology
- Gynecology
- Cardiac adult
- Cardiac pediatric
- Peripheral vessel
- Carotid
- Lung
- | - Musculoskeletal (superficial)
-
-
- Cardiac adult
- Cardiac pediatric
- Peripheral vessel
- Carotid
- Pulmonary
- interventional
guidance(includes free
handneedle/catheter) | Musculoskeletal(conventional)
-
- Urology
- Gynecology
- Cardiac adult
- Cardiac pediatric
- Peripheral vessel
-
- | -
-
- Obstetrics(OB) , Gynecology(GYN)
-
-
- Peripheral vessel
-
-  |
|  Modes of operation | - B mode
- M mode
- Color Doppler
-
- PW
- Combined mode (B+M, B+PW, B+
Color Doppler, B+PDI, B+DPDI,
B+PW+ Color Doppler mode) | - B mode
- M mode
- Color doppler (CD)
- Power doppler(PD)
-
-
- Combined mode (B+M, B+CD,
B+PD) | - B mode
- M mode
- Color doppler (CD)
- Power doppler(PD)
- PWD
-
- Combined mode (B+M,
B+CD,B+PWD) | - B mode
- M mode
- Color doppler (CD)
- Power doppler(PD)
- PWD
-
- Combined mode
(B+M, B+CD,B+PD,
B+PWD) | - B mode
- B/M mode
- Color
- PDI
- PW
-  |
|  Wireless capability | Wi-Fi
Bluetooth | Wi-Fi | Wi-Fi | Wi-Fi
Bluetooth | Wireless  |
|  Transducer types | - Convex(PIU1B, PEU1B, PIU1D,
PEU1D, )
- Linear (PIU1A, PEU1A, PIU1C,
PEU1C, PAU2C)
-
- Convex & Linear (PIU2A, PEU2A,
PIU2B, PEU2B, PAU2A, PAU2B) | - Convex (NUSONO-C35)
- Linear (NUSONO-L75)
- Phased array (NUSONO-P25)
- | -
- Linear(LU710L)
- Phased array(LU710PA)
- | - Convex(C3 HD)
-
- | -
-
- Convex & Linear  |
|  Power source | Rechargeable battery (Li-ion) | Rechargeable battery (Li-ion) | Rechargeable battery (Li-ion) | Rechargeable battery (Li-ion) | Rechargeable battery  |
|  Display | iOS or Android mobile device and
Windows | iOS or Android mobile device | iOS or Android mobile device
and Windows | iOS or Android mobile device | iOS / Android /Windows  |
|  510(k) Track | Track 3 | Track 3 | Track 3 | Track 3 | Track 3  |
|  Standards | - IEC 60601-1 (2020)
- IEC 60601-1-2(2020)
- IEC 60601-1-6(2020)
-
- IEC 60601-2-37(2015)
- IEC TS 60601-4-2(2024)
-
-
- IEC 62359(2017)
- IEC 62133-2 (2022)
- IEC 62366-1 (2020) | - IEC 60601-1 (2020)
- IEC 60601-1-2(2020)
- IEC 60601-1-6(2020)
-
- IEC 60601-2-37(2015)
- IEC 60601-4-2(2016)
- UD 2-2004 (R2009)
-
-
- IEC 62133-2 (2017)
- IEC 62366-1 (2020) | - ANSI/AAMI ES60601-1 (2012)
- IEC 60601-1-2 (2014)
- IEC 60601-1-6 (2013)
-
- IEC 60601-2-37(2015)
-
- UD 2-2004 (R2009)
- UD 3-2004 (R2009)
-
- IEC 62133 (2012)
- IEC 62366 (2014) | - ANSI/AAMI ES60601-1 (2012)
-
- IEC 60601-1-2 (2014)
- IEC 60601-1-6 (2013)
-
- IEC 60601-1-12 (2014)
- IEC 60601-2-37 (2015)
-
- UD 2-2004 (R2009)
-
- IEC 62133 (2012)
- | IEC 60601-1
IEC 60601-1-2
IEC 60601-2-37  |

{8}

|  Items | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device  |
| --- | --- | --- | --- | --- | --- |
|   | - ISO 10993-1 (2018) | - ISO 10993-1 (2018) | - ISO 10993-1 (2009) | - |   |
|   | - | - ISO 10993-5 (2009) | - ISO 10993-5 (2009) | - ISO 10993-5 (2009) |   |
|   | - | - ISO 10993-10 (2021) | - ISO 10993-10 (2010) | - ISO 10993-10 (2010) |   |
|   | - | - | - | - ISO 10993-11 (2017) |   |
|   | - | - ISO 10993-23 (2021) | - | - |   |
|   | - IEC 62304 (2015) | - IEC 62304 (2015) | - IEC 62304 (2006) | - IEC 62304 (2006) |   |
|   | - ISO 15223-1 (2021) | - ISO 15223-1 (2021) | - ISO 15223-1 (2016) | - ISO 15223-1 (2012) |   |
|   | - ISO 14971 (2019) | - ISO 14971 (2019) | - ISO 14971 (2019) | - ISO 14971 (2007) |   |
|   | - ISO 13485 (2016) | - ISO 13485 (2016 | - ISO 13485 (2016) | - |   |

The subject device provides additional combined modes (e.g., B+PW+Color Doppler, B+PDI, B+DPDI) that are not fully covered in all predicate devices. However, these combined modes are constructed solely from individual imaging modes (B-mode, M-mode, Color Doppler, Power Doppler, PW Doppler) that are already present in the predicates. The combination does not introduce new diagnostic functionality but instead facilitates user workflow by enabling simultaneous display. Since the underlying acoustic output and image generation remain unchanged from the individual modes, these differences do not raise new questions of safety or effectiveness, and the subject device is therefore considered substantially equivalent with respect to modes of operation.

The subject device conforms to IEC 62359:2017 for the measurement and calculation of acoustic output indices (MI, TI), whereas the predicate devices list compliance with UD 2-2004 (R2009) and UD 3-2004 (R2009). IEC 62359:2017 is internationally recognized and encompasses the same technical principles as UD 2/UD 3, providing equivalent requirements for acoustic output characterization and display. Therefore the difference in standards does not raise new questions of safety or effectiveness.

As a result, it is reasonable to conclude that subject device (Wireless Probe Type Ultrasound Scanner) is substantial equivalent with predicate device.

5 / 6

{9}

VI Non-clinical Performance Test

The subject device was tested/analyzed according to the following standards in order to ensure its effectiveness and safety:

|  Test | Title/Test Method Summary | Result  |
| --- | --- | --- |
|  Thermal, Electrical, and Mechanical Safety  |   |   |
|  Electrical Safety | IEC 60601-1 | Pass  |
|  Electromagnetic Compatibility | IEC 60601-1-2
ANSI AAMI IEC 60601-1-2
IEC TS 60601-4-2 | Pass  |
|  Battery Safety | IEC 62133-2 | Pass  |
|  Wireless Coexistence | ANSI IEEE USEMCSC C63.27 | Pass  |
|  FCC Certification | FCC 15, FCC 18 | Certified  |
|  Acoustic Output | IEC 60601-2-37
IEC 62359 | Pass, Track 3  |
|  Software Verification and Validation  |   |   |
|  Software Verification and Validation | IEC 62304 | Pass  |
|  Cybersecurity  |   |   |
|  Cybersecurity Assessment and testing | Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions | Pass  |
|  Non-Clinical Verification Testing  |   |   |
|  Measurement Accuracy | Internal test methods | Pass  |

VII Clinical Performance Test

The Wireless Probe Type Ultrasound Scanner did not require clinical testing to establish substantial equivalence.

VIII Conclusion

The Wireless Probe Type Ultrasound Scanner is substantially equivalent to the predicate device. The Wireless Probe Type Ultrasound Scanner functions in a manner similar to and is intended for the same use as the predicate device. Based on the predicate device comparison of indications for use, acoustic output and general safety and effectiveness information, as well as the non-clinical performance test results, it is concluded that this device is as safe and effective as the predicate device for its intended use and performance, and is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K252709](https://fda.innolitics.com/device/K252709)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
