← Product Code [GCJ](/productcode/GCJ) · K252697

# 3D Laparoscope (K252697)

_Guangdong Optomedic Technologies, Inc. · GCJ · Oct 29, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252697

## Device Facts

- **Applicant:** Guangdong Optomedic Technologies, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Oct 29, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Stellar Imaging System is intended to provide real-time visible and near-infrared imaging in minimally invasive procedures. Upon intravenous administration and use of an ICG agent consistent with its approved label, Stellar Imaging System is used to perform fluorescence imaging visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts. Upon interstitial administration and use of an ICG agent consistent with its approved label, the Stellar Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

## Device Story

Stellar Imaging System provides real-time 3D visible and near-infrared (NIR) fluorescence imaging for minimally invasive endoscopic surgery. System components include image processing unit, light source, camera head, and laparoscope. Device uses excitation light (785 or 805 nm) to excite intravenously or interstitially administered ICG dye in the surgical field; camera head captures resulting fluorescence emission. System displays parallel white light and fluorescence images to surgeon. Features include procedure recording, storage, and replay. Used in clinical settings by surgeons to visualize anatomy, blood flow, tissue perfusion, and lymphatic structures. Output assists intraoperative decision-making by providing enhanced anatomical and physiological visualization, potentially improving surgical precision and patient outcomes.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

System comprises image processing unit, light source, camera head, and laparoscope. Excitation light source (785/805 nm). Field of view: 80°. Max insertion width: 10.3 mm. Working length: 300 mm or 490 mm. Supports 3D visible and fluorescence imaging. Connectivity includes removable storage for recording/replay.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Stellar Imaging System ([K241530](/device/K241530.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 29, 2025

Guangdong OptoMedic Technologies, Inc.
Minghua Wu
Regulatory Affairs Engineer
Suite 503, Building A, Golden Valley Intellicreation
Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai
Foshan, Guangdong 528200
China

Re: K252697
Trade/Device Name: 3D Laparoscope
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: GCJ
Dated: September 10, 2025
Received: September 11, 2025

Dear Minghua Wu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252697 - Minghua Wu
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K252697 - Minghua Wu
Page 3

Sincerely,

Colin K. Chen -S

Digitally signed by Colin K. Chen -S
Date: 2025.10.29 21:29:38 -04'00'

Colin Kejing Chen
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252697 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Stellar Imaging System  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  Stellar Imaging System is intended to provide real-time visible and near-infrared imaging in minimally invasive procedures.
Upon intravenous administration and use of an ICG agent consistent with its approved label, Stellar Imaging System is used to perform fluorescence imaging visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts.
Upon interstitial administration and use of an ICG agent consistent with its approved label, the Stellar Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) #: K252697
510(k) Summary
Prepared on: 2025-09-11

# Contact Details
21 CFR 807.92(a)(1)

|  Applicant Name | Guangdong OptoMedic Technologies, Inc.  |
| --- | --- |
|  Applicant Address | Suite 503, Building A, Golden Valley Intellicreation Community, No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, 528200, Foshan, Guangdong, P.R. China Foshan 528200 China  |
|  Applicant Contact Telephone | 0757-8670-2920  |
|  Applicant Contact | Ms. Minghua Wu  |
|  Applicant Contact Email | wuminghua@optomedic.com  |

# Device Name
21 CFR 807.92(a)(2)

|  Device Trade Name | Stellar Imaging System  |
| --- | --- |
|  Common Name | Endoscope and accessories  |
|  Classification Name | laparoscope, general & plastic surgery  |
|  Regulation Number | 876.1500  |
|  Product Code(s) | GCJ  |

# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)

|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
| --- | --- | --- |
|  K241530 | Stellar Imaging System | GCJ  |

# Device Description Summary
21 CFR 807.92(a)(4)

Stellar Imaging System comprises Image Processing Unit, Light Source, Camera Head, Laparoscope and relevant accessories.
During surgical procedures, Stellar Imaging System is used to provide real-time 3D visible and fluorescence imaging in endoscopic surgery.
The Stellar Imaging System including the laparoscope is designed to work with an approved infrared dye (Indocyanine green (ICG). ICG may be excited at excitation at either 785 or 805 nm. The System provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye using a camera head. Fluorescence imaging with ICG permits the system to visualize blood flow and related tissue circulation, of lymphatic flow, etc.
The System allows the capture of normal (white) light image in parallel with the fluorescence image and displays both to the surgeon to provide a view of the anatomy. In addition, the System permits recording surgical procedures, storing them on removable storage devices, and replaying the procedures.

# Intended Use/Indications for Use
21 CFR 807.92(a)(5)

Stellar Imaging System is intended to provide real-time visible and near-infrared imaging in minimally invasive procedures.
Upon intravenous administration and use of an ICG agent consistent with its approved label, Stellar Imaging System is used to perform fluorescence imaging visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts.
Upon interstitial administration and use of an ICG agent consistent with its approved label, the Stellar Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

{5}

|  Indications for Use Comparison 21 CFR 807.92(a)(5)  |
| --- |
|  The indications for use are the same.  |
|  Technological Comparison 21 CFR 807.92(a)(6)  |
|  The differences between changed 21033SA,21033SC, new models 21050SA,21050SC and cleared 21033SA,21033SC are: 1. Field of view: The field of view of changed 21033SA,21033SC, new models 21050SA,21050SC is 80°, while cleared 21033SA,21033SC is 70°. 2. Maximum insertion portion width (mm): The maximum insertion portion width of changed 21033SA,21033SC, new models 21050SA,21050SC is 10.3, while cleared 21033SA,21033SC is 10.1. 3. Working length (mm): The working length of changed 21033SA,21033SC is 300, new models 21050SA,21050SC is 490, while cleared 21033SA,21033SC is 315.  |
|  Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)  |
|  The bench testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device to support a substantial equivalence determination.  |

---

**Source:** [https://fda.innolitics.com/device/K252697](https://fda.innolitics.com/device/K252697)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).