← Product Code [OOE](/productcode/OOE) · K252682

# LenSx Laser System (8065000944) (K252682)

_Alcon Laboratories, Inc. · OOE · Sep 24, 2025 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252682

## Device Facts

- **Applicant:** Alcon Laboratories, Inc.
- **Product Code:** [OOE](/productcode/OOE.md)
- **Decision Date:** Sep 24, 2025
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 886.4390
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The LenSx Laser system is indicated for use: • In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate), anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery. • In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate) during Implantable Collamer Lens (ICL) surgery. • In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. • In the creation of corneal pockets for placement/insertion of a corneal inlay device; and for creation of corneal tunnels for the placement of corneal rings.

## Device Story

Ophthalmic surgical laser; uses focused femtosecond laser pulses for photodisruption; creates vapor bubbles to cut/separate tissue in lens capsule, crystalline lens, and cornea. System includes computer-guided delivery, video microscope, and Optical Coherence Tomography (OCT) for surgical field imaging. Operated by trained physicians in clinics/hospitals. Patient interface (sterile, disposable applanating lens and suction ring) docks to cornea to fix eye position. Physician uses real-time imaging to guide laser patterns for incisions, capsulotomy, or phacofragmentation. Benefits include precise, automated tissue disruption for various ophthalmic surgeries.

## Clinical Evidence

No clinical data; bench testing only (referenced from predicate K243896).

## Technological Characteristics

Femtosecond laser (1030 nm, 600-800 fs pulse duration); Kb:KYW gain medium. OCT (820-880 nm) and LED video microscope for imaging. Integrated suction device. Max pulse energy 15 μJ (cataract) / 2.6 μJ (corneal). Repetition rate 50-150 kHz. Sterile, disposable patient interface accessories. QNX 6.3.2 operating environment.

## Regulatory Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

## Predicate Devices

- LenSx Laser System ([K243896](/device/K243896.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

September 24, 2025

Alcon Laboratories, Inc.
Tammy Vu
Senior Manager, Regulatory Affairs
6201 South Freeway
Fort Worth, Texas 76134-2099

Re: K252682
Trade/Device Name: LenSx Laser System (8065000944)
Regulation Number: 21 CFR 886.4390
Regulation Name: Ophthalmic laser
Regulatory Class: Class II
Product Code: OOE, HNO, HQC
Dated: August 22, 2025
Received: August 25, 2025

Dear Tammy Vu:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252682 - Tammy Vu
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K252682 - Tammy Vu
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CLAUDINE H. KRAWCZYK -S

Claudine Krawczyk
Assistant Director
DHT1A: Division of Ophthalmic Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252682 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  LenSx Laser System (8065000944)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The LenSx Laser system is indicated for use: • In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate), anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery. • In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate) during Implantable Collamer Lens (ICL) surgery. • In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. • In the creation of corneal pockets for placement/insertion of a corneal inlay device; and for creation of corneal tunnels for the placement of corneal rings.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |

{4}

510(k)# K252682

510(k) Summary: LenSx Laser System

In accordance with 21 CFR 807.92, Alcon hereby provides the 510(k) summary for the LenSx Laser system.

# 1. SUBMITTER [per 807.92(a)(1)]

|  Applicant | Alcon Laboratories, Inc.
6201 South Freeway
Fort Worth, TX 76134-2099  |
| --- | --- |
|  Primary Correspondent | Tammy Vu
Senior Manager, Regulatory Affairs
Alcon Research LLC, on behalf of Alcon Laboratories, Inc.
20511 Lake Forest Drive
Lake Forest, CA 92630-7741
+1 (949) 505-7519
tammy.vu@alcon.com  |
|  Secondary Correspondent | Ophelia Biggs
Senior Director, Regulatory Affairs
Alcon Research LLC, on behalf of Alcon Laboratories, Inc.
20511 Lake Forest Drive
Lake Forest, CA 92630-7741
+1 (949) 505-7557
ophelia.biggs@alcon.com  |
|  Date Prepared | September 16, 2025  |

# 2. SUBJECT DEVICE [per 807.92(a)(2)]

|  Device Trade Name | LenSx Laser System (8065000944)  |
| --- | --- |
|  Regulation Number and Name (Product Code in Parentheses) | 21 CFR 886.4390, Ophthalmic Laser (OOE)  |
|  Associated Product Codes | HNO, HQC  |
|  Regulatory Class | II  |

{5}

510(k)# K252682

510(k) Summary: LenSx Laser System

# 3. PREDICATE DEVICE [per [807.92(a)(3)]

|  Predicate Device | LenSx Laser System (8065998162)  |
| --- | --- |
|  510(k) Number | K243896  |
|  Regulation Number and Name (Product Code in Parentheses) | 21 CFR 886.4390, Ophthalmic Laser (OOE)  |
|  Associated Product Codes | HNO, HQC  |
|  Regulatory Class | II  |

# 4. DEVICE DESCRIPTION

The LenSx Laser system is an ophthalmic surgical laser which uses focused femtosecond laser pulses to create vapor bubbles which disrupts/separates tissue (photodisruption) within the lens capsule, crystalline lens, and the cornea. A computer-guided delivery system places the laser pulses in a pattern to produce an incision/cut.

The laser pulses are delivered through a sterile, disposable applanating lens and suction ring that contacts the cornea and fixes the eye with respect to the laser delivery system.

The interface between the laser and patient is the Patient Interface that connects to the delivery system which is docked to the patient's cornea. Two models of the Patient Interface accessory are offered for use with the LenSx Laser: the LenSx Laser Patient Interface and the LenSx Laser SoftFit Patient Interface. Both models consist of a sterile, disposable applanating lens and suction ring assembly. The LenSx Laser SoftFit Patient Interface also comes with a soft contact lens that is positioned against the external surface of the Patient Interface glass. For cataract procedures, the LenSx Laser SoftFit Patient Interface is used. The LenSx Laser Patient Interface is used for corneal, flap, tunnel, and pocket incisions. Refer to the Instructions for Use supplied with the LenSx Laser Patient Interface for preparation and application.

The LenSx Laser system is for prescription use and should only be operated by a trained physician. The LenSx Laser system is intended to be used within a clinic(s)/hospital(s)/surgical practice network.

# 5. INDICATIONS FOR USE

The LenSx Laser system is indicated for use:

- In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate), anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the

{6}

510(k)# K252682

510(k) Summary: LenSx Laser System

same surgery.

- In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate) during Implantable Collamer Lens (ICL) surgery.
- In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea.
- In the creation of corneal pockets for placement/insertion of a corneal inlay device; and for creation of corneal tunnels for the placement of corneal rings.

# 6. TECHNOLOGICAL COMPARISON

The technological characteristics are the same to those of the predicate device previously cleared in K243896.

|   | PREDICATE DEVICE | SUBJECT DEVICE  |
| --- | --- | --- |
|  CHARACTERISTIC | LENSX LASER SYSTEM | LENSX LASER SYSTEM  |
|  Administrative  |   |   |
|  510(k) | K243896 | K252682  |
|  Regulation Number | 21 CFR 886.4390 | 21 CFR 886.4390  |
|  Regulatory Class | Class II | Class II  |
|  Product Code | OOE; HNO; HQC | OOE; HNO; HQC  |
|  Intended Use (Summarized) | Intended for creation of corneal cuts/incisions, pockets, tunnels, and flaps, and for anterior capsulotomy and phacofragmentation. | Same  |
|  Indications for Use | The LenSx Laser system is indicated for use: • In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate), anterior capsulotomy and laser phacofragmentation during cataract surgery. Each of these procedures may be performed either individually or consecutively during the same surgery. • In the creation of corneal cuts/incisions (single-plane, multi-plane, and arcuate) during Implantable Collamer Lens (ICL) surgery. • In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea. • In the creation of corneal pockets for placement/insertion of a corneal inlay device; and for creation of corneal tunnels for the placement of corneal rings. | Same  |

{7}

510(k)# K252682

510(k) Summary: LenSx Laser System

|  CHARACTERISTIC | PREDICATE DEVICE LENSX LASER SYSTEM | SUBJECT DEVICE LENSX LASER SYSTEM  |
| --- | --- | --- |
|  System Features  |   |   |
|  Operating Principle | Femtosecond laser photodisruption | Same  |
|  Mechanism of Action | Cutting and resection surfaces are created by scanned pattern of femtosecond laser micro-photodisruptions in tissue. | Same  |
|  Surgical Field Imaging | Video Microscope and Optical Coherence Tomography | Same  |
|  Suction Method | Integrated suction device | Same  |
|  Laser Mode of Operation | Pulsed | Same  |
|  Laser Gain Medium | Kb:KYW | Same  |
|  Laser Wavelength | 1030 nm | Same  |
|  Laser Pulse Duration | 600-800 fs | Same  |
|  Laser Pulse Repetition Rate | 50 kHz for cataract and ICL, 150 kHz for corneal flaps, pockets, and tunnels | Same  |
|  Maximum Pulse Energy | 15 μJ for cataract, 2.6 μJ for corneal flaps, pockets, and tunnels | Same  |
|  Maximum Average Power | 1 W during service 0.750 W for cataract 0.390 W for corneal flaps, pockets, and tunnels | Same  |
|  Maximum Surgical Diameter (Clear Aperture) | 12.5 mm | Same  |
|  Maximum Depth | 8000 μm for cataract, 190 μm for corneal flaps, 400 μm for corneal pockets, and corneal tunnels | Same  |
|  OCT Optical Source | Superluminescent Diode | Same  |
|  Mode of Operation | Spectral Domain | Same  |
|  OCT Wavelength | 820-880 nm | Same  |
|  OCT Maximum Power | 3.0 mW | Same  |
|  Illumination for Video Microscope (VM) | LED-Light Emitting Diode | Same  |
|  VM Illumination Wavelength | 500-650 nm | Same  |
|  VM Illumination Maximum Power | 0.330 mW | Same  |
|  Accessories | LenSx Laser Patient Interface LenSx Laser SoftFit Patient Interface | Same  |
|  Software Features  |   |   |
|  Software Operating Environment GUI & Instrument Host | QNX, version 6.3.2 | Same  |
|  Software Version | LenSx Laser System version 2.40 | Same  |

Page 4 of 5

{8}

510(k)# K252682

510(k) Summary: LenSx Laser System

## 7 NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY

The risk assessment for the subject device did not require any non-clinical or clinical testing as mitigation against any identified risks. Non-clinical testing previously performed in the predicate device are deemed applicable to the subject device and can be referenced in K243896.

## 8 CONCLUSION

The modified LenSx Laser system shares the same intended use, indications for use, technological characteristics, and risk profiles as the predicate device. Non-clinical testing previously performed and cleared in K243896 are deemed applicable to the subject device. The predicate device and the subject device have been developed and manufactured in compliance with 21 CFR 820 and ISO 14971. Therefore, the modified LenSx Laser system is deemed substantially equivalent to the currently cleared, LenSx Laser system.

Page 5 of 5

---

**Source:** [https://fda.innolitics.com/device/K252682](https://fda.innolitics.com/device/K252682)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
