K252621 · El Global Trade, Ltd. · NFO · May 8, 2026 · Neurology
Device Facts
Record ID
K252621
Device Name
DeepSkin (DEP100)
Applicant
El Global Trade, Ltd.
Product Code
NFO · Neurology
Decision Date
May 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DEP100 device is an over-the-counter home-use device intended for the stimulation of healthy facial skin.
Device Story
DEP100 is a handheld, battery-powered, non-invasive microcurrent stimulator for OTC home cosmetic use. Device consists of user interface, microcontroller (PLC), power module, and electrodes. User applies conductive gel to face; device delivers microcurrents via active tip electrodes to stimulate skin. Operation is non-invasive; intended for 5-10 minute treatment sessions. Device features include automatic no-load trip, automatic shut-off, and low battery indicator. Designed for single-user, home environment. Safety and efficacy supported by bench testing, electrical safety/EMC compliance, and human factors validation study confirming user ability to self-select and operate safely.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing, including electrical safety, EMC, software verification/validation, and a human factors/usability study (n=18) demonstrating 100% success rate for critical tasks.
Indicated for facial stimulation for over-the-counter cosmetic use in healthy adults on intact skin.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Facial Spa device by Nu Skin Enterprises Inc. (K122711)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 8, 2026
El Global Trade Ltd.
Stephanie Khoury
RA/QA Associate
Gibore Israel 13
Netanya, 42407
Israel
Re: K252621
Trade/Device Name: DeepSkin (DEP100)
Regulation Number: 21 CFR 882.5890
Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief
Regulatory Class: Class II
Product Code: NFO
Dated: April 9, 2026
Received: April 9, 2026
Dear Stephanie Khoury:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252621 - Stephanie Khoury
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252621 - Stephanie Khoury
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
Submission Number (if known)
K252621
Device Name
Microcurrent Device (model number: DEP100)
Indications for Use (Describe)
The Microcurrent Device (Model: DEP100) is intended for facial stimulation for over-the-counter cosmetic use.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K252621 510(K) Summary EL Global Trade Ltd. For DEP100
1. DATE PREPARED: April 9th, 2026
2. 510(K) OWNER NAME
EL Global Trade Ltd.
13th Gibore Israel St., Netanya 4250413, Israel
Tel: +972-9-7889069 Fax: +972-9-7734831
Contact person name:
Sivan Fishman, VP Global Compliance
Phone: +972-54-5223677
E-mail: sivanf@sensica.com
3. DEVICE NAME
Common/Usual Name: Microcurrent Device
Model Name: DEP100
Classification: Class II device under the following classification names:
| Classification Name | Product Code | Regulation Number | Panel |
| --- | --- | --- | --- |
| Transcutaneous electrical nerve stimulator for pain relief | NFO | 882.5890 | Neurology |
4. PREDICATE DEVICES
Facial Spa device by Nu Skin Enterprises Inc.
Cleared under 510(k) number K122711 on September 17th, 2013.
5. DEVICE DESCRIPTION
Page 1
{5}
K252621 510(K) Summary EL Global Trade Ltd. For DEP100
The DEP100 device is an OTC, home-use hand-held battery-powered device used with FDA cleared conductive gel to stimulate the face with microcurrents. The device is intended for facial stimulation for OTC cosmetic use. It is a non-invasive, non-ablative device consisting of:
- User Interface
- Programmable logic controller (PLC, microcontroller) embedded in PCBA
- Microcurrent power module
- Power Supply
- Microcurrent electrodes
The PLC (on the PCBA) is specially configured software that, combined with hardware circuits, provides the operational and safety function of the system. The Microcurrent power module provides microcurrent to the active tip electrodes. This device is supplied non-sterile.
## 6. INDICATIONS FOR USE:
The DEP100 device is indicated for facial stimulation for over-the-counter cosmetic use.
## 7. INTENDED USE:
The DEP100 device is an over-the-counter home-use device intended for the stimulation of healthy facial skin.
## 8. NON-CLINICAL (BENCH) PERFORMANCE DATA
The following performance data (bench tests) were provided in support of the performance, safety, and efficacy of the DEP100 device as well as the substantial equivalence determination.
### Safety Bench Tests and Verification & Validation (V&V) Summary
The device was internally tested for V&V as per pre-determined device requirements and specifications.
Page 2
{6}
Page 3
K252621 510(K) Summary
EL Global Trade Ltd. For DEP100
Electrical safety and electromagnetic compatibility (EMC)
{7}
Page 4
K252621 510(K) Summary
EL Global Trade Ltd. For DEP100
Electrical safety and EMC testing were conducted on the DEP100 device. The device complies with:
- IEC 60601-1:2005/EN 60601-1:2006, General safety standard: safety requirements for medical electrical systems.
- IEC 60601-1-2:2014/EN 60601-1-2:2015, Medical electrical equipment Part 1-2 - General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility – Requirements and Tests.
- IEC 60601-1-11:2015, Medical electrical equipment Part 1-11 – Requirements for the medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-2-10: 2012, Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff. “Guidance for the Content of Premarket Submissions for Device Software Functions.” The document level for this software was considered “Basic”.
{8}
K252621 510(K) Summary EL Global Trade Ltd. For DEP100
## Human Factors Validation Testing
The human factors and usability validation of the DEP100 device are based on:
1. IEC 60601-1-6:2010 test report
2. A self-selection and human factors validation study
A Self-Selection and Human Factors validation study was performed to demonstrate that users can safely and effectively self-select, prepare and perform treatment with the DEP100 device.
The Self-Selection study proved the self-selection of the users, reaching El Global’s goal. 18 subjects who successfully self-selected themselves as eligible for treatment with the device, performed treatment with the DEP100 device using the device labeling and IFU. The users participating reached a validation rate of 100% success rate for all critical tasks, with no critical issues raised during use. These results demonstrated that the design of the DEP100 device is safe for use by the intended lay users.
## 9. PERFORMANCE TESTING – ANIMAL
No animal testing was performed with the subject device.
## 10. CLEANING, STERILIZATION, SHELF LIFE AND BIOCOMPATIBILITY
The DEP100 device is a non-sterile, reusable device, intended for a single user. The device cleaning instructions are based on the cleaning instructions of the predicate device due to the fact that both devices are made of the same materials and used similarly.
The shelf-life expectancy of the device is 5 years, similarly to the predicate device.
The biocompatibility evaluation for the DEP100 device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 “Use of International Standard ISO-10993, ‘Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,’” September 4, 2020, and International Standard ISO 10993-1 “Biological Evaluation of Medical Devices – Part 1:
{9}
Page 6
K252621 510(K) Summary
EL Global Trade Ltd. For DEP100
Evaluation and Testing Within a Risk Management Process,” as recognized by FDA.
In the case of a device with such a classification, the prerequisite information for the risk assessment is the physical and/or chemical information (in accordance with ISO 10993-18). Additional endpoints required for this classification in case of missing or insufficient information are Cytotoxicity, Sensitization and Irritation.
Components of the DEP100 device were tested in accordance with REACH and RoHS standards in order to confirm that no dangerous or toxic materials are present in the device composition. Additionally, the materials used in the manufacture of the device are materials commonly used in the medical device industry. These materials are also used in the manufacturing process of El Global Trade Ltd.’s devices, for which Cytotoxicity, Irritation, and Sensitization tests were performed.
11. CLINICAL PERFORMANCE DATA
No clinical testing was gathered for the subject device, as it is substantially equivalent to the predicate device, and thus requires no additional clinical data to prove its safety and efficacy.
12. SUBSTANTIAL EQUIVALENCE
The indications for use and technological characteristics of the DEP100 device are substantially equivalent to the indications for use and technological characteristics of the predicate device, as can be seen in the technology comparison table below.
The design and components of both devices (i.e., power supply, Micro-current generator, and controller) are similar. The performance specifications of the DEP100 (i.e., Micro-current and electrical power) are substantially equivalent to those of the predicate device. The safety features and compliance with safety standards of both are similar. The body contact materials
{10}
Page 7
# K252621 510(K) Summary EL Global Trade Ltd. For DEP100
are similar. The minor differences in the technological characteristics do not raise new safety or effectiveness concerns and are demonstrated to be substantially equivalent through relevant performance tests. Furthermore, the DEP100 device has qualified with varied performance tests, including software validation testing, electrical and mechanical safety testing according to IEC 60601-1, electromagnetic compatibility testing according to IEC 60601-1-2, compatibility as medical electrical equipment for home healthcare environment according to 60601-1-11, and basic safety and essential performance of nerve and muscle stimulators according to IEC 60601-2-10. The performance tests demonstrated that the device specifications meet the system requirements and do not raise new safety or effectiveness concerns.
| Feature | DEP100 (EL Global Trade LTD) Subject Device | Facial Spa (Nu Skin Enterprises, Inc.) [K122711] Predicate Device |
| --- | --- | --- |
| Device Class | Class II | Class II |
| Product Code | NFO | NFO |
| Regulation Number | 21 CFR 882.5890 | 21 CFR 882.5890 |
| Device | Stimulator, Transcutaneous Electrical, Aesthetic Purposes | Stimulator, Transcutaneous Electrical, Aesthetic Purposes |
| Regulation Description | Transcutaneous Electrical Nerve Stimulator | Transcutaneous Electrical Nerve Stimulator |
| Review Panel | Neurology | Neurology |
| Indication for Use | Over-the-counter cosmetic use | Over-the-counter cosmetic use |
| Intended Use | The device is intended for facial stimulation for over-the-counter cosmetic use. | The device is intended for facial stimulation for over-the-counter cosmetic use. |
| Intended User Population | OTC device. General population – healthy adults, treatment on intact skin (according to contraindications). | OTC device. General population – healthy adults, treatment on intact skin (according to contraindications). |
| Intended Treated / Application Site | Face | Face |
| Principle of Operation | Non-invasive, one hand operation, based on micro-current. | Non-invasive, one hand operation, based on micro-current. |
{11}
K252621 510(K) Summary EL Global Trade Ltd. For DEP100
| Feature | DEP100 (EL Global Trade LTD) Subject Device | Facial Spa (Nu Skin Enterprises, Inc.) [K122711] Predicate Device |
| --- | --- | --- |
| | To be used with prior applied conductive gel. | To be used with prior applied conductive gel. |
| Treatment Duration | 5-10 minute treatment depending on the area. | 5-10 minute treatment depending on the area. |
| Technology | Micro-current device | Micro-current device |
| Waveform | Direct current | Direct current |
| Maximum Average Output Voltage | ~200 mV @ 500 Ω ~800 mV @ 2 KΩ ~4 V @ 10 KΩ | ~200 mV @ 500 Ω ~800 mV @ 2 KΩ ~4 V @ 10 KΩ |
| Maximum Output Current | ~400 μA @ 500Ω ~400 μA @ 2kΩ ~400 μA @ 10kΩ | ~400 μA @ 500Ω ~400 μA @ 2kΩ ~400 μA @ 10kΩ |
| Charge Per Phase | N/A – no phases. The current flows in one direction. | N/A – no phases. The current flows in one direction. |
| Average DC when the device is ON, but no pulse is applied | 0 | 0 |
| Number of Output Channels | 1 | 1 |
| Software/Microprocessor Control | Yes | Yes |
| Automatic No-Load Trip | Yes | Yes |
| Automatic Shut-Off | Yes | Yes |
| User Over-ride Control | Yes | Yes |
| Low Battery Display Indicator | Yes | Yes |
| Voltage/Current Level Indicator | No | No |
| Device Main Components | Hand-held handpiece, two treatment surfaces, power adaptor, conductive gel in cartridge. | Hand-held handpiece, two treatment surfaces, power adaptor, conductive gel. |
| Power Source | Battery operated – Rechargeable Li-ion battery | Battery operated – two AAA Alkaline batteries |
| Main unit weight | 120gr (4 oz) | 120 gr (4 oz) |
| Main unit size | 19.3 × 6.4 × 4.9 [cm³] | With large conductor: 14.3 × 3.12 × 6.75 [cm³] With small conductor: 13.6 × 3.12 × 6.75 [cm³] |
Page 8
{12}
K252621 510(K) Summary EL Global Trade Ltd. For DEP100
| Feature | DEP100 (EL Global Trade LTD) Subject Device | Facial Spa (Nu Skin Enterprises, Inc.) [K122711] Predicate Device |
| --- | --- | --- |
| **Electrodes size** | Large conductor: ~22 cm² Small conductor: ~2 cm² | Large conductor: ~8.3 cm² Small conductor: ~2.4 cm² |
| **Maximum Average Current Density** (@ 500 Ω) | Large conductor: ~18 µA/cm² Small conductor: ~200 µA/cm² | Large conductor: ~48 µA/cm² Small conductor: ~160 µA/cm² |
| **Contact (electrodes) material** | Chrome plated ABS | Chrome plated ABS |
| **Safety/ EMC** | IEC 60601-1:2005 (Ed. 3.1), Medical Electrical Equipment - Part 1: General safety standard safety requirements for medical electrical systems. IEC 60601-1-2: 2014 (Ed. 4), Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility – Requirements And Tests. IEC 60601-2-10:2012 (Ed. 2), Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. IEC 60601-1-11:2015 (Ed. 2), Medical electrical equipment Part 1-11 – Requirements for the medical electrical equipment and medical electrical systems used in the home healthcare environment. | IEC 60601-1:1998 (Ed. 2), Medical Electrical Equipment - Part 1: General safety standard safety requirements for medical electrical systems. IEC 60601-1-2: 2012 (Ed. 3), Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility – Requirements And Tests. IEC 60601-2-10:1987 (Ed. 1), Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. |
| **Biocompatibility** | All parts that are in contact with user comply with the requirements of ISO-10993-1. | All parts that are in contact with user comply with the requirements of ISO-10993-1. |
Page 9
{13}
Page 10
# K252621 510(K) Summary
EL Global Trade Ltd. For DEP100
| Feature | DEP100 (EL Global Trade LTD) Subject Device | Facial Spa (Nu Skin Enterprises, Inc.) [K122711] Predicate Device |
| --- | --- | --- |
| Software | The software was verified and validated according to the FDA guidance. | IEC 60601-1-4:1996 (Ed. 1.1), Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems |
| Compliance with 21 CFR 898 | Yes | Yes |
{14}
K252621 510(K) Summary EL Global Trade Ltd. For DEP100
The DEP100 device is substantially equivalent to the predicate device, as the devices share a common intended use, and technological differences between the DEP100 device and the predicate do not raise new questions of safety or effectiveness.
## 13. CONCLUSIONS
Due to these identical clauses and high similarities, the DEP100 device is at least as safe and as effective as the predicate device, since they share the same intended use, technological characteristics, features, specifications, materials, and anatomical site for use. The differences between the DEP100 device and the predicate device do not raise new safety or effectiveness issues or questions, as detailed above. Based on performance testing and comparison to predicate devices, the DEP100 device is substantially equivalent to the previously cleared Facial Spa predicate device.
Page 11
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.