Standalone validation study: 97 independent imaging series from 97 unique male patients
4 (board-certified radiologists)
Indications for Use
The product is intended for automatic segmentation, visualization, and volumetric quantification of the prostate on 3D MRI data (in DICOM format). The software automates the currently manual process of identifying, labeling and volumetric calculation of segmented prostate structures on 3D MRI images. The calculated volumetric data, in addition to the image data at hand, can assist trained medical professionals in the interpretation of prostate MR studies. Diagnosis should not be made solely based on the analysis performed using mdprostate. The application of the software refers to patients with presumed changes in the prostate tissue and is intended to be used for men of age between 35 and 99 years.
Device Story
Software processes 3D MRI data (DICOM) to perform automatic prostate segmentation and volumetric quantification; calculates PSA density based on user-provided PSA values; allows manual lesion evaluation (location, size, volume, ADC values, PI-RADS score). Used in hospitals, imaging centers, or image processing labs by trained medical professionals. Operates on off-the-shelf hardware or cloud. Outputs include annotated DICOM images, electronic reports, and RTStruct files for ultrasound systems to facilitate MR-US fusion biopsy. Assists clinicians in interpreting prostate MR studies; does not provide definitive diagnosis. Benefits include automated, standardized prostate measurement and improved workflow for targeted biopsy planning.
Clinical Evidence
Bench testing only. No clinical trials conducted. Standalone validation study evaluated 97 independent MRI series from 97 male patients across diverse demographics and scanner manufacturers (Siemens, GE, Philips). Reference standard established by a panel of four radiologists. Primary endpoint (mean Dice Similarity Coefficient) was 0.90 (acceptance >0.88). Secondary endpoint (mean 95th percentile Hausdorff distance) was 3.70 mm. Stratified analysis across age, ethnicity, scanner type, field strength (1.5T/3.0T), and disease status (PI-RADS 1-5) showed stable performance with variance <0.03.
Technological Characteristics
Software-only device; operates on off-the-shelf hardware or cloud. DICOM compatible. Deep learning-based automatic segmentation. Inputs: Axial T2W MRI, DWI, ADC maps (optional DCE). Outputs: Annotated DICOM, electronic reports, RTStruct. Compliant with ISO 14971 (risk management), IEC 62366-1 (usability), IEC 62304 (software lifecycle), and NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for men aged 35-99 with presumed prostate tissue changes. Used to assist trained medical professionals in interpreting prostate MR studies via automatic segmentation, visualization, and volumetric quantification.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 7, 2026
mediaire GmbH
Alexander Wegener
Director Regulatory Affairs and Quality Management
Ritterstrasse 16-18
Berlin, 10969
Germany
Re: K252571
Trade/Device Name: mdprostate
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: July 7, 2026
Received: July 7, 2026
Dear Wegener Alexander:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252571 - Wegener Alexander
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252571 - Wegener Alexander
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252571
Device Name
mdprostate
Indications for Use (Describe)
The product is intended for automatic segmentation, visualization, and volumetric quantification of the prostate on 3D MRI data (in DICOM format). The software automates the currently manual process of identifying, labeling and volumetric calculation of segmented prostate structures on 3D MRI images.
The calculated volumetric data, in addition to the image data at hand, can assist trained medical professionals in the interpretation of prostate MR studies.
Diagnosis should not be made solely based on the analysis performed using mdprostate.
The application of the software refers to patients with presumed changes in the prostate tissue and is intended to be used for men of age between 35 and 99 years.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 1 of 9
# 510(k) SUMMARY FOR MDPROSTATE
# K252571
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92.
# 1. Submitter
Submitter's Name: mediaire GmbH
Submitter's Address: Ritterstrasse 16-18
10969 Berlin, Germany
Date 2026-08-04
# 2. Contact Person
Contact Person: Alexander Wegener
Contact Person Title: Director Regulatory Affairs and Quality Management
Application Correspondent: Alexander Wegener
Telephone: +49 30 286 490 67
Email: a.wegener@mediaire.de
# 3. Device Name and Classification
Product Name: mdprostate
Trade Name mdprostate
Classification Name: Automated Radiological Image Processing Software
Common Name: Medical Image Processing Software
Product Code: QIH
CFR Section: 21 CFR 892.2050
Recommended Classification: Class II
Classification Panel: Radiology
510(k) Number: K252571
{5}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 2 of 9
### 4. Predicate Device
| Product Name: | AI-Rad Companion Prostate MR |
| --- | --- |
| Propriety Trade Name | AI-Rad Companion Prostate MR |
| 510(k) Number | K193283 |
| Clearance Date: | July 2, 2020 |
| Classification Name: | Picture Archiving and Communication System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.2050 |
| Secondary CFR Section: | 21 CFR §892.1000 |
| Device Class: | Class II |
| Product Code | QIH |
| Secondary Product Code | LNH |
### 5. Intended Use
The product is intended for automatic segmentation, visualization, and volumetric quantification of the prostate on 3D MRI data (in DICOM format). The software automates the currently manual process of identifying, labeling and volumetric calculation of segmented prostate structures on 3D MRI images.
The calculated volumetric data, in addition to the image data at hand, can assist trained medical professionals in the interpretation of prostate MR studies.
Diagnosis should not be made solely based on the analysis performed using mdprostate.
The application of the software refers to patients with presumed changes in the prostate tissue and is intended to be used for men of age between 35 and 99 years.
{6}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 3 of 9
## 6. Device Description
The product is intended for automatic segmentation, visualization and volumetric quantification of the prostate on 3D MRI data (in DICOM format). The software automates the currently manual process of identifying, labeling and volumetric calculation of segmented prostate structures on 3D MRI images.
The primary features of mdprostate include:
- Automatic prostate segmentation and volume estimation
- Calculation of the PSA density, based on the input of the PSA value of the patient by the user
- The user can manually add his/her evaluation of lesions (and their characteristics) in a suitable user interface. The characteristics provided include location, size, volume and ADC values and other characteristics (like PI-RADS score) for each lesion.
- Export in a suitable format for reading and archiving in PACS, as well as in a second format that can be imported by ultrasound systems (e.g. RTStruct), allowing the urologist to perform targeted MR-US fusion biopsy
## 7. Technological Characteristics
The subject device mdprostate is substantially equivalent to the predicate device with regards to software, programming languages, operating system, and fundamental technology. Mediaire made some device modifications and implemented enhancements to the existing predicate device, AI-Rad Companion Prostate MR (K193283).
While these enhancements and improvements offer additional image viewing and evaluation capabilities compared to the predicate device, the conclusions from all verification and validation data suggest that these modifications do not adversely affect the safety and effectiveness of the predicate device.
Table 1: Substantial Equivalence Comparison Table
| Description | Subject Device | Predicate Device | Comparison Results |
| --- | --- | --- | --- |
| Device Trade Name | mdprostate | AI-Rad Companion Prostate MR | |
| Environment of use | mdprostate is used by trained professionals in hospitals, imaging centers or in image processing labs. | | EQUIVALENT |
| Physical characteristics | Software package Operates on off-the-shelf hardware (multiple vendors) | Software package Operates on the cloud | ENHANCED |
{7}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 4 of 9
| Table 1: Substantial Equivalence Comparison Table | | | |
| --- | --- | --- | --- |
| Description | Subject Device | Predicate Device | Comparison Results |
| Device Trade Name | mdprostate | AI-Rad Companion Prostate MR | |
| | or on the cloud | | |
| DICOM compatible | YES | YES | SAME |
| Design and incorporate technology | Automatic segmentation by deep machine learning | Automatic segmentation by deep machine learning | EQUIVALENT |
| Data Source | Axial T2W MRI scans and corresponding DWI and ADC maps. Can optionally evaluate DCEs. | 2D TSE axial or 3D TSE / SPACE MR image data and DWI image | EQUIVALENT |
| Output | • electronic reports with information about the prostate in DICOM format • annotated DICOM images • format that can be imported by ultrasound systems (e.g. RTStruct) | • electronic reports with information about the prostate • annotated DICOM images • format that can be imported by ultrasound systems (e.g. RTStruct) | SAME |
| Testing | • Production Risk Assessment • Software Verification tests • Software validation tests | • Production Risk Assessment • Software Verification tests • Software validation tests | SAME |
### 7.1. Substantial Equivalent Discussion
The technical characteristics are equivalent for the subject (mdprostate) and predicate (AI-RAD companion Prostate MR) device. Both applications use the same environment, are DICOM compatible and create similar outputs. Both devices are pure software packages and operate in the cloud. Additionally, mdprostate can be installed and operated on-premise. The technologies in both devices include deep learning algorithms. Those may differ in their configurations and/or tuning parameters, but are in general equivalent. The data used for training those algorithms differ, but both include T2-weighted and DWI MR images. Additionally, mdprostate uses the respective ADC map as an additional input.
{8}
**mediaire** | DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
**Summary K252571**
Page 5 of 9
## 8. Performance Testing
Non-clinical testing was performed to evaluate the functionality of **mdprostate**. This included software validation and bench testing, which were conducted to assess the device's performance claims and to support the determination of substantial equivalence to the predicate device. **mdprostate** was further tested for conformity with multiple applicable industry standards. The results of the non-clinical performance testing demonstrate that **mdprostate** complies with the FDA guidance document "Content of Premarket Submissions for Device Software Functions" (June 14, 2023), as well as with the voluntary FDA-recognized consensus standards listed in Table 2 below.
| Recognition Number | Title of Standard | Reference Number and Date | Standards Development Organization |
| --- | --- | --- | --- |
| 5-125 | Medical devices - Application of risk management to medical devices | 14971:2019 | ISO |
| 5-129 | Medical devices - Part 1: Application of usability engineering to medical devices | 62366-1:2020 | IEC |
| 5-134 | Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements | 15223-1:2021 | ISO |
| 13-79 | Medical device software - Software life cycle processes | 62304:2006/A1:2016 | IEC |
| 12-261 | Information Technology –Digital Compression and coding of continuous -tone still images: Requirements and Guidelines [including: Technical Corrigendum 1 (2005)] | 10918-1:1994 | ISO IEC |
| 12-363 | Digital Imaging and Communications in Medicine (DICOM) Set | PS 3.1 – 3.20 (2024) | NEMA |
{9}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 6 of 9
### 8.1 Verification and Validation
Development of mdprostate was made in compliance with 21 CFR 820.30 (Design Controls). Risk management was completed in accordance with ISO 14971, Application of Risk Management to Medical Devices to assess the potential risk and harms of the device. Based on the risk analysis, appropriate verification and validation was completed and results demonstrated that the predetermined acceptance criteria were met. Software was developed and validated in accordance with ISO 62304, Software Life Cycle Processes for Medical Device Software.
Software documentation consistent for software with a basic documentation level, in accordance with the FDA guidance document "Content of Premarket Submissions for Device Software Functions" (June 14, 2023), was completed and placed in the Design History File. Software verification and validation testing demonstrated that the software performs as intended and in accordance with specifications. The potential risks of mdprostate have been identified and evaluated, and the risks were determined to be acceptable, or have been addressed with risk control measures.
### 8.2 Standalone Performance Validation
A standalone validation study was conducted to assess the segmentation performance of mdprostate on bi-parametric MRI. The study evaluated 97 independent imaging series from 97 unique male patients, with a strict 1:1 relationship between patients and samples (no duplicate scans, follow-ups, or repeat imaging sessions).
Reference Standard: A consensus ground truth was established for each case through a hierarchical panel of three independent board-certified radiologists who manually outlined prostate boundaries slice-by-slice; disagreements were adjudicated by a fourth senior radiologist.
Independence of Test Data: The validation dataset was subject to a strict data-lock protocol. None of the 97 cases were used at any stage of algorithm development, architecture selection, hyperparameter tuning, or model training, and the data originated from institutions independent of those that supplied the training data, ensuring an unbiased assessment of real-world generalizability.
Imaging Equipment and Protocol: Data were acquired across multiple institutions using scanners from three major manufacturers (Siemens Healthineers, GE Healthcare and Philips Healthcare), at both 1.5T and 3.0T field strengths, using high-resolution T2-weighted turbo spin-echo (TSE) axial sequences compliant with PI-RADS v2.1.
Demographics and Clinical Subgroups: The cohort was 100% male, consistent with the intended use. Participants were between 35 and 99 years of age and represented diverse ethnic groups, including White Americans, Black or African Americans, Asians, and Other (Hispanics, Native Hawaiians, Pacific Islanders, and American Indians and Alaska Natives). Performance was additionally stratified by prostate volume and by PI-RADS v2.1 category to evaluate robustness against clinically challenging confounders.
Suspected malignant lesions (PI-RADS 3–5) represent a clinically challenging confounder for segmentation algorithms, as they can distort local tissue architecture, alter signal intensity, and
{10}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 7 of 9
obscure peripheral zone boundaries. Nevertheless, stratified analysis across all evaluated subgroups, including demographic characteristics, prostate volume, lesion status, and MRI field strength, demonstrated stable performance. Mean Dice scores ranged from 0.89 to 0.91, while mean HD95 values ranged from 3.07 to 3.84, with a variance in Dice Similarity Coefficient of less than 0.03 across all stratifications. These findings suggest no meaningful performance disparities and support the robustness of mdprostate across diverse patient characteristics, clinical presentations, imaging conditions, and prostate volumes.
Results: The pre-defined primary endpoint, mean Dice Similarity Coefficient (DSC), had an acceptance criterion of >0.88; mdprostate achieved a mean DSC of 0.90. As a secondary endpoint, the 95th percentile Hausdorff distance was evaluated; mdprostate achieved a mean 95th percentile Hausdorff distance of 3.70 mm (2.47 pixels), favorably comparable to the literature benchmark. All pre-specified performance endpoints were met.
To demonstrate that mdprostate maintains consistent segmentation accuracy and does not suffer from algorithmic degradation when applied to diverse clinical subgroups, we performed a stratified performance analysis. Table 3 reports the mean Dice Similarity Coefficient (DICE) and 95th percentile Hausdorff Distance (HD95) for each category.
Table 3: Subgroup Performance Analysis
| Variable | Subgroup Category | Patients (N) | Mean DICE | Mean HD95 [mm] |
| --- | --- | --- | --- | --- |
| Age Group (Years) | < 55 | 4 | 0.89 [0.86 | 0.91] | 3.13 [2.67 | 3.57] |
| 55 - 70 | 66 | 0.91 [0.904 | 0.917] | 3.74 [3.47 | 4.00] |
| > 70 | 27 | 0.92 [0.909 | 0.927] | 3.63 [3.22 | 4.19] |
| Ethnicity | White American | 58 | 0.91 [0.904 | 0.918] | 3.84 [3.55 | 4.16] |
| Black or African American | 19 | 0.91 [0.898 | 0.924] | 3.65 [3.19 | 4.24] |
| Asian American | 10 | 0.92 [0.899 | 0.934] | 3.07 [2.61 | 3.49] |
| Other* | 10 | 0.91 [0.894 | 0.925] | 3.49 [3.10 | 3.98] |
| Scanner Manufacturer | Siemens Healthineers | 49 | 0.91 [0.903 | 0.920] | 3.80 [3.45 | 4.20] |
| GE Healthcare | 38 | 0.90 [0.882 | 0.914] | 3.59 [3.26 | 3.95] |
{11}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 8 of 9
Table 3: Subgroup Performance Analysis
| Variable | Subgroup Category | Patients (N) | Mean DICE | Mean HD95 [mm] |
| --- | --- | --- | --- | --- |
| | Philips Healthcare | 10 | 0.92 [0.907 | 0.922] | 3.58 [3.17 | 4.14] |
| Field Strength | 1.5 Tesla | 28 | 0.91 [0.901 | 0.921] | 3.70 [3.27 | 4.16] |
| 3.0 Tesla | 69 | 0.91 [0.905 | 0.918] | 3.67 [3.42 | 3.98] |
| Prostate Volume | Small (<50 ml) | 33 | 0.90 [0.884 | 0.907] | 3.53 [3.19 | 3.92] |
| Medium (50–100 ml) | 43 | 0.92 [0.910 | 0.922] | 3.74 [3.43 | 4.04] |
| Large (>100 ml) | 21 | 0.93 [0.920 | 0.937] | 3.81 [3.31 | 4.61] |
| Disease Presence | PI-RADS 1-2 (Benign/ No Lesion) | 47 | 0.91 [0.91 | 0.92] | 3.77 [3.43 | 4.12] |
| PI-RADS 3-5 (Suspected Cancer) | 50 | 0.91 [0.90 | 0.92] | 3.59 [3.31 | 3.93] |
* incl. Hispanic or Latino, Native Hawaiian or other Pacific Islander, American Indian or Alaska Native
### 9. Cybersecurity
mediaire has established and maintains a cybersecurity risk management process in accordance with the FDA guidance "Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions" (February 2, 2026), and the applicable statutory requirements of Section 524B of the Federal Food, Drug, and Cosmetic Act. As part of this process, which is embedded within the company's Quality Management System (QMS), mediaire implements administrative and technical controls designed to prevent unauthorized access to, unauthorized modification of, misuse of, or denial of use of the device, as well as to safeguard information stored on, accessed by, or transmitted from the device to external recipients against unauthorized use. Key elements of this process include the maintenance of a Software Bill of Materials (SBOM), a structured vulnerability monitoring and remediation procedure, and mechanisms to support the ongoing deployment of security patches and updates over the device's total product lifecycle.
{12}
mediaire
DIGITAL INNOVATION
IN RADIOLOGY
mediaire GmbH • Ritterstrasse 16-18 • 10969 Berlin • Germany
Trade Register: Berlin HRB 195423B • VAT ID No.: DE318215731
Summary K252571
Page 9 of 9
## 10. Clinical Tests
Clinical testing was not conducted to evaluate the performance and functionality of the modifications introduced in mdprostate. Instead, verification and validation activities were performed to confirm that the modifications and enhancements meet their design specifications and perform as intended for their intended use. The results of these verification and validation activities support the safety and effectiveness of the subject device and the determination of substantial equivalence to the predicate device. No animal testing was performed on the subject device.
## 11. Safety and Effectiveness
The device labeling includes instructions for use, along with all necessary warnings and precautions, to ensure the device is used safely and effectively.
Safety is further ensured through a risk management process compliant with ISO 14971, under which potential hazards are systematically identified and mitigated through risk analysis conducted early in the design phase and maintained continuously throughout product development. Identified risks are controlled through a combination of measures implemented during software development, verification and validation testing, and product labeling.
In addition, the device is intended for use by healthcare professionals who are experienced in the post-processing and interpretation of magnetic resonance images.
## 12. Substantial Equivalence and Conclusion
Based on thorough verification and validation testing using applicable industry standards, mediaire GmbH concludes that mdprostate is substantially equivalent to AI-Rad Companion Prostate MR cleared under K193283.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.