Rusch SoftSimplastic Foley Catheters

K252537 · Teleflex Medical Sdn. Bhd. · EZL · Jan 20, 2026 · Gastroenterology, Urology

Device Facts

Record IDK252537
Device NameRusch SoftSimplastic Foley Catheters
ApplicantTeleflex Medical Sdn. Bhd.
Product CodeEZL · Gastroenterology, Urology
Decision DateJan 20, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
AttributesTherapeutic

Indications for Use

2 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively or for patients with conditions requiring transurethral urine drainage. 3 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively, for patients with conditions requiring transurethral urine drainage and for patients requiring bladder irrigation.

Device Story

Rüsch SoftSimplastic Foley Catheters are sterile, single-use, short-term (up to 30 days) indwelling balloon catheters. Device consists of a transparent PVC shaft (2 or 3 lumen) with a distal latex retaining balloon, proximal funnel, inflation valve, and radiopaque contrast stripe. Available in 14-24 Fr sizes with Couvelaire or Coudé/Mercier tips. Used in clinical settings for transurethral bladder drainage or irrigation. Operated by healthcare professionals. Device is inserted transurethrally; balloon is inflated to retain catheter in bladder. Output is continuous urine drainage or irrigation fluid flow. Subject device is MR Conditional, allowing use in MRI environments under specified conditions.

Clinical Evidence

Bench testing only. No clinical data provided. MR safety and compatibility testing performed per ASTM F2052-21, ASTM F2213-17, ASTM F2182-19e2, ASTM F2119-07, and ASTM F2503-23 to establish MR Conditional status.

Technological Characteristics

Materials: PVC shaft, latex balloon. Configuration: 2 or 3 lumen, 42 cm length, 14-24 Fr. Radiopaque stripe included. Sterilization: Ethylene Oxide. MR Safety: MR Conditional per ASTM F2052, F2213, F2182, F2119, F2503. Single-use, short-term indwelling (up to 30 days).

Indications for Use

Indicated for adult male and female patients requiring routine transurethral bladder drainage (postoperative or due to medical conditions) or bladder irrigation (3-way only). Contraindicated for patients with latex allergy, insurmountable urethral obstructions, urethral injuries, urethral inflammation, or false passage.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION January 20, 2026 Teleflex Medical Sdn. Bhd. Yuvaanesswary Nallappan Regulatory Affairs Manager Lot PT 2577, Jalan Perusahaan 4 Kamunting, Perak 34000 MALAYSIA Re: K252537 Trade/Device Name: Rusch SoftSimplastic Foley Catheters Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and accessories Regulatory Class: II Product Code: EZL Dated: August 12, 2025 Received: December 19, 2025 Dear Yuvaanesswary Nallappan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K252537 - Yuvaanesswary Nallappan Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {3} K252537 - Yuvaanesswary Nallappan Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JESSICA K. NGUYEN -S Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252537 | ? | | Please provide the device trade name(s). | | ? | | Rusch SoftSimplastic Foley Catheters | | | | Please provide your Indications for Use below. | | ? | | 2 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively or for patients with conditions requiring transurethral urine drainage. | | | | 3 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively, for patients with conditions requiring transurethral urine drainage and for patients requiring bladder irrigation. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} K252537 Page 1 of 5 510(k) Summary Rüsch SoftSimplastic Foley Catheters # 510(k) Summary A. Name, Address, Phone and Fax Number of Applicant Teleflex Medical Sdn. Bhd. Lot PT 2577, Jalan Perusahaan 4, 34600 Kamunting, Perak, Malaysia. B. Contact Person Yuvaanesswary Nallappan Regulatory Affairs Manager C. Date Prepared January 19, 2026 D. Device Name Trade Name: Rüsch SoftSimplastic Foley Catheters Common Name: Catheter, Retention Type, Balloon Product Code: EZL Regulation Number: CFR 876.5130 Classification: II Classification Panel: Gastroenterology/Urology E. Predicate Device This submission demonstrates substantial equivalence to the predicate device, Teleflex Medical Rüsch SoftSimplastic Foley Catheter, K212077. F. Device Description The RÜSCH SoftSimplastic balloon catheters are made of transparent PVC. They have a 2 lumen or 3 lumen shaft with proximal funnel, inflation valve and a distal retaining balloon made of latex. Balloon capacity in ml and shaft size in French gauge (Fr.), Charrière (Ch.) or millimeters (mm) are indicated on the funnel of each individual catheter. They have a radiopaque contrast stripe along the shaft and are supplied sterile. The following tip types are available: Couvelaire and Coudé/Mercier. Sizes range from 14 to 24 Ch./Fr. The catheters have a length of approximately 42 cm. The RÜSCH SoftSimplastic balloon catheters are made of PVC uncoated. These catheters are only for short term usage. Maximum duration of use: Up to 30 days. Teleflex Medical, Inc. CONFIDENTIAL {6} K252537 Page 2 of 5 510(k) Summary Rüsch SoftSimplastic Foley Catheters ## G. Indications for Use 2 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively or for patients with conditions requiring transurethral urine drainage. 3 Way SoftSimplastic Catheters: Indicated where routine transurethral drainage of the bladder is required either postoperatively, for patients with conditions requiring transurethral urine drainage and for patients requiring bladder irrigation. ## H. Contraindications - Latex allergy - Insurmountable urethral obstructions - Urethral injuries - Urethral inflammation - False passage ## I. Substantial Equivalence The subject Rüsch SoftSimplastic Foley Catheter is substantially equivalent to the predicate device with respect to intended use, technology and construction. Table 1 summarizes the comparison between the subject and predicate device. Table 1: Comparison of Predicate Device vs. Subject Device Teleflex Medical, Inc. CONFIDENTIAL {7} K252537 Page 3 of 5 510(k) Summary Rüsch SoftSimplastic Foley Catheters | Features | Teleflex Medical Rüsch SoftSimplastic Foley Catheter (Predicate Device - K212077) | Teleflex Medical Rüsch SoftSimplastic Foley Catheter (Subject Device) | Similarities and Differences | | --- | --- | --- | --- | | Classification Name | Catheter, Retention Type, Balloon | Catheter, Retention Type, Balloon | Same | | Regulation Number | 876.5130 | 876.5130 | Same | | FDA Product Code | EZL | EZL | Same | | Class | II | II | Same | | Indications for Use | 2 Way SoftSimplastic Catheters: Indicated where routine drainage of the bladder is required either postoperatively or for patients with conditions requiring urine drainage. 3 Way SoftSimplastic Catheters: Indicated where routine drainage of the bladder is required either postoperatively, for patients with conditions requiring urine drainage and for patients requiring bladder irrigation. | 2 Way SoftSimplastic Catheters: Indicated where routine drainage of the bladder is required either postoperatively or for patients with conditions requiring urine drainage. 3 Way SoftSimplastic Catheters: Indicated where routine drainage of the bladder is required either postoperatively, for patients with conditions requiring urine drainage and for patients requiring bladder irrigation. | Same | | Contraindications | • Latex allergy • Insurmountable urethral obstructions • Urethral injuries • Urethral inflammation • False passage | • Latex allergy • Insurmountable urethral obstructions • Urethral injuries • Urethral inflammation • False passage | Same | | Single Use | Yes | Yes | Same | | Population | Adult, Male and Female | Adult, Male and Female | Same | | Size Range | Male/Female 14-24 Fr. | Male/Female 14-24 Fr. | Same | Teleflex Medical, Inc. CONFIDENTIAL {8} K252537 Page 4 of 5 510(k) Summary Rüsch SoftSimplastic Foley Catheters | Features | Teleflex Medical Rüsch SoftSimplastic Foley Catheter (Predicate Device - K212077) | Teleflex Medical Rüsch SoftSimplastic Foley Catheter (Subject Device) | Similarities and Differences | | --- | --- | --- | --- | | Indwelling Time | Maximum indwell time less than 30 days | Maximum indwell time less than 30 days | Same | | Single Use | Yes | Yes | Same | | Lumens | 2-way and 3-way | 2-way and 3-way | Same | | Nominal Balloon Inflation Volume | 30ml and 75ml | 30ml and 75ml | Same | | Shaft Material | PVC | PVC | Same | | Coated or uncoated | Uncoated | Uncoated | Same | | Eyelets | Yes | Yes | Same | | Primary Packaging | Paper and film peel back | Paper and film peel back | Same | | Sterile | Yes | Yes | Same | | Sterilization Method | Ethylene Oxide | Ethylene Oxide | Same | | Biocompatibility | Complies with ISO 10993-1 | Complies with ISO 10993-1 | Same | | MR Condition | Safety in MRI Not Evaluated | MR Conditional | Different | # J. Comparison of Technological Characteristics with the Predicate Device The Rüsch SoftSimplastic Foley Catheters have the same fundamental technological characteristics as the predicate device, Teleflex Medical Rüsch SoftSimplastic Foley Catheter (K212077). Both subject and predicate devices are sterile, single-use, short-term indwelling balloon catheters designed for transurethral bladder drainage. Both subject and predicate devices are based on the same technological principles and design features, including: - A flexible PVC shaft available in 2-way and 3-way lumen configurations. - A distal balloon made of latex to retain the catheter within the bladder. - A proximal funnel and inflation valve for urine drainage and balloon inflation. - A radiopaque stripe along the shaft for X-ray visibility. Teleflex Medical, Inc. CONFIDENTIAL {9} K252537 Page 5 of 5 510(k) Summary Rüsch SoftSimplastic Foley Catheters - Ethylene oxide sterilization and single-use packaging. - Intended for short-term use (up to 30 days) in adult male and female patients. The following minor differences exist between the subject and predicate devices: - The subject device has undergone magnetic resonance (MR) safety and compatibility testing and is now labeled as MR Conditional in accordance with ASTM F2052, ASTM F2213, ASTM F2182, ASTM F2119, and ASTM F2503. - The predicate device did not include MR labeling and was not previously evaluated for MR safety. The devices otherwise share identical materials, configuration, functional principles, and performance characteristics. # K. Summary of Non-Clinical Testing MR Safety and Compatibility testing was performed per the standards listed below. - ASTM F2052-21 “Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment”2 - ASTM F2213-17 “Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment” - ASTM F2182-19e2 “Standard Test Method for Measurement of Radio Frequency Induced Heating near Passive Implants during Magnetic Resonance Imaging” - ASTM F2119-07 (2013) “Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants” - ASTM F2503-23 “Standard Practice for Marking Devices and Other items for Safety in the Magnetic Resonance Environment” The device was determined to be MR Conditional when used under the specific conditions defined in the device labeling. # L. Conclusion The Rüsch SoftSimplastic Foley Catheters have the same intended use and technological characteristics as the predicate device. Test results demonstrate that the subject devices meet their intended use and performs as well as the legally marketed predicate device. The subject device is substantially equivalent to the predicate device. Teleflex Medical, Inc. CONFIDENTIAL
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...