K252518 · Duopross Meditech Corporate · FMF · Dec 17, 2025 · General Hospital
Device Facts
Record ID
K252518
Device Name
DuoprossTM Smart Cap (Type I)
Applicant
Duopross Meditech Corporate
Product Code
FMF · General Hospital
Decision Date
Dec 17, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
Smart Caps are indicated for use as sterile tip caps with tamper-evident function for luer-lock syringes. The use time of the Smart cap should be ≤ 24 hours.
Device Story
Smart Cap is a sterile, single-use, tamper-evident tip cap for luer-lock syringes; designed to maintain sterility and prevent unauthorized access to medical fluids. Device assembly consists of an outer cover, inner cap, and inner ring. Upon removal, the clear outer shell separates from the cap, providing visual evidence of tampering. Used in clinical settings by healthcare providers. Benefits include reduced risk of accidental or intentional fluid compromise. Device is sterilized via ethylene oxide.
Clinical Evidence
Bench testing only. Testing included ISO 80369-7 connector compliance, biocompatibility (ISO 10993-1), sterilization validation (ISO 11135) achieving SAL 10^-6, EO/ECH residual testing (ISO 10993-7), bacterial endotoxin testing (LAL < 20 EU/device), and sterile barrier packaging integrity (ISO 11607, ASTM F1886, F1929, F88, D4169, F1980).
Technological Characteristics
Materials: Acrylonitrile Butadiene Styrene (ABS) and Polypropylene (PP). Connection: Female luer lock. Sterilization: Ethylene oxide (EO) to SAL 10^-6. Design: Tamper-evident outer shell/housing. Standards: ISO 80369-7, ISO 10993-1, ISO 10993-7, ISO 11135, ISO 11607-1/2, ASTM F1886, ASTM F1929, ASTM F88/F88M, ASTM D4169, ASTM F1980.
Indications for Use
Indicated for use as sterile tip caps with tamper-evident function for luer-lock syringes. For prescription use only. Use time ≤ 24 hours.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 17, 2025
Duopross Meditech Corporate
% Eva Li
Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
Room 1401, Dongfang Building, 1500# Century Ave.
Shanghai, 200122
China
Re: K252518
Trade/Device Name: DuoprossTM Smart Cap (Type I)
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston Syringe
Regulatory Class: Class II
Product Code: FMF
Dated: November 21, 2025
Received: November 21, 2025
Dear Eva Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252518 - Eva Li
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252518 - Eva Li
Page 3
Sincerely,
Shruti N. Mistry -S
Shruti Mistry
Assistant Director, Injection Devices
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252518 | |
| Device Name DuoprossTM Smart Cap (Type I) | |
| Indications for Use (Describe) Smart Caps are indicated for use as sterile tip caps with tamper-evident function for luer-lock syringes. The use time of the Smart cap should be ≤ 24 hours. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Duopross Meditech Corporate
154 Marine St, Farmingdale N.Y. 11735
K252518
# 510(K) Summary
## 1. Basic information
| Date prepared: | Dec 17, 2025 |
| --- | --- |
| Submitter: | Duopross Meditech Corporate |
| Address: | 154 Marine St, Farmingdale N.Y. 11735 |
| Manufacturing Company: | Duopross Meditech (Kunshan) Co., Ltd. |
| Manufacturing Address: | Taihong Road Yushan Town Wusong Jiang, Kunshan Jiangsu, CN 215300 |
| Contact person: | Alan Yang |
| Telephone Number: | +86 155 4242 0089 |
| Email Address: | hongkang.yang_alan@dpsmeditech.com |
Trade Name: DuoprossTM Smart Cap (Type I)
Device 510(k): K252518
Device Common Name: Tamper Evident Cap
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston syringe
Product Class: II
Product Panel: General Hospital
Product Code: FMF
Predicate Device Name: TrueCare Biomedix Tamper Evident Cap
Predicate Device 510(k): K210818
Device Common Name: Tamper Evident Cap
Regulation Number: 21 CFR 880.5860
Regulation Name: Piston syringe
Product Class: II
Product Panel: General Hospital
Product Code: FMF
## 2. Device Description
The Smart Cap is a tamper-evident syringe cap with a clear indication of cap removal to ensure sterility and reduce the risk of medical fluids being compromised accidentally or intentionally. Its assembly includes an outer cover, an inner cap and an inner ring. It is made of acrylonitrile butadiene styrene (ABS) and polypropylene (PP) and is sterilized by ethylene oxide.
## 3. Indication for use
Smart Caps are indicated for use as sterile tip caps with tamper-evident function for luer-lock syringes. The use time of the Smart cap should be ≤ 24 hours.
## 4. Comparison of Technological Characteristics and Basis for
{5}
Duopross Meditech Corporate
154 Marine St, Farmingdale N.Y. 11735
K252518
## Substantial Equivalence
| Item | Predicate Device | Subject device | Comparison |
| --- | --- | --- | --- |
| Device Name | TrueCare Biomedix Tamper Evident Cap | Duopross^{TM} Smart Cap | N/A |
| Indication for use | Tamper Evident Caps are indicated for use as a sterile tamper evident cap for IV syringes. | Smart Caps are indicated for use as sterile tip caps with tamper-evident function for luer-lock syringes. The use time of the Smart cap should be ≤ 24 hours. | Difference #1 |
| Fundamental Scientific Technology | Female luer lock cap with clear tamper evident outer shell/housing | Female luer lock cap with clear tamper evident outer shell/housing | same |
| Sterility | Sterile EO (10^-6) | Sterile EO (10^-6) | same |
| Number of Uses | Single Use, Rx only | Single Use, Rx only | same |
| Connection | Female luer lock | Female luer lock | same |
| Tamper Evidence Design Feature | Clear outer shell/housing that separates from the cap upon opening | Clear outer shell/housing that separates from the cap upon opening | same |
| Materials | Indirect patient contacting: PP 1024 polypropylene with red color additive
Non-patient contacting: Acrylonitrile Butadiene Styrene | Indirect patient contacting: Acrylonitrile Butadiene Styrene with blue color additive
Non-patient contacting: polypropylene | Difference #2 |
| Biocompatibility Contact / Duration | Indirect blood contact, limited duration. | Indirect blood contact, limited duration. | same |
Discussion of Technological differences:
Difference #1: The intended use of Smart Cap is similar to intended use of the predicate. Both are indicated for use as tamper evident caps for syringes. Both devices are Rx Only. This difference does not raise any new or different questions of safety or effectiveness when compared to the predicate device.
Difference #2: While the cap material is different, the subject device meets the requirements of ISO 10993-1. This difference does not raise any new or different questions of safety or effectiveness when compared to the predicate device.
## 5. Non-clinical bench testing
### Bench performance
The testing was conducted as part of the Design Control process for the TEC product family to keep current with the latest ISO standard, 80369. For this product line, the appli
{6}
Duopross Meditech Corporate
154 Marine St, Farmingdale N.Y. 11735
K252518
cable standards are ISO 80369-7, "Small-bore connectors for liquids and gases in healthcare applications – Part 7 Connectors for intravascular or hypodermic applications"
## Biocompatibility
In accordance with ISO 10993-1, the device is classified as Limited duration, externally communicating device, Blood Path Indirect (<24hours).
## Sterility
A sterilization validation was conducted to ensure a Sterility Assurance Level (SAL) of 10⁻⁶ using the overkill method in compliance with ISO 11135 Sterilization of health care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices.
EO and ECH residual testing was completed for the subject device per ISO 10993-7 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals and the devices met the limits of < 4mg/device and < 9mg/device, respectively.
The sterile subject devices were evaluated for the potential to produce a pyrogenic response. The Limulus Amebocyte Lysate (LAL) test was used to test for Bacterial Endotoxins as part of process validations, and the devices met the limits of < 20 EU/device. Additionally, material-mediated pyrogen testing conducted as part of the biocompatibility testing conducted in support of this submission.
The subject device is currently labeled with a 3-year shelf life based on shelf life verification testing. Packaging system characteristics and integrity testing was conducted in accordance with the following standards:
- ISO 11607-1 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ISO 11607-2 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1886 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM F1929 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM F88/F88M Standard Test Method for Seal Strength of Flexible Barrier Materials
- ASTM D4169 Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
## 6. Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Duopross™ Smart Cap is substantially equivalent to the TrueCare Biomedix Tamper Evident Cap under K210818 with respect to the
{7}
Duopross Meditech Corporate
154 Marine St, Farmingdale N.Y. 11735
K252518
indications for use, target populations, treatment method, and technological characteristics.
4 / 4
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.