K252495 · Newpong Co., Ltd. · QPB · Apr 3, 2026 · General, Plastic Surgery
Device Facts
Record ID
K252495
Device Name
LSSA (NP-200)
Applicant
Newpong Co., Ltd.
Product Code
QPB · General, Plastic Surgery
Decision Date
Apr 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LSSA is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. LSSA is indicated for use in the following surgical specialists when the fragmentation and aspiration of adipose tissue is desired: - Plastic and Reconstruction Surgery - Orthopedic Surgery
Device Story
LSSA (NP-200) is an ultrasonic surgical system for fragmentation and emulsification of subcutaneous fatty tissue. System comprises a main control unit, two handpieces (B and F), foot switch, and various probes. Operation relies on piezoelectric effect: PZT ceramic crystals in handpieces convert electrical energy from control unit into mechanical ultrasonic vibrations (31-37.5 kHz). Vibrations transmitted to probe to emulsify adipose tissue. Used in clinical settings (OR) by plastic/reconstructive or orthopedic surgeons. Operator adjusts vibration amplitude/frequency via control unit; supports continuous and pulsed (50% duty cycle) modes. Output is mechanical ultrasonic energy for tissue fragmentation. Benefits include precise body contouring through targeted tissue emulsification.
Clinical Evidence
Bench testing only. Includes electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), ultrasound output power, and biocompatibility (ISO 10993-1, -5, -10, -11, -23).
Technological Characteristics
Ultrasonic surgical unit; piezoelectric PZT ceramic crystals; operating frequencies 31 kHz and 37.5 kHz; output power 0-40W; continuous and pulsed (50% duty cycle) modes; probe diameters 0.9mm to 3.7mm; electrical energy source; IEC 60601-1/1-2 compliant; ISO 10993 compliant materials.
Indications for Use
Indicated for fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring in patients undergoing procedures by plastic/reconstructive or orthopedic surgeons.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
4/03/2026
Newpong Co., Ltd.
Leo W Choi
President
4855 Shady Pines Dr.
Saint Cloud, Florida 34772
Re: K252495
Trade/Device Name: LSSA (NP-200)
Regulation Number: 21 CFR 878.5040
Regulation Name: Suction Lipoplasty System
Regulatory Class: Class II
Product Code: QPB
Dated: August 8, 2025
Received: August 8, 2025
Dear Leo W Choi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252495 - Leo W Choi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K252495 - Leo W Choi
Page 3
Sincerely,
**Alicia Hemphill -S**
Digitally signed by Alicia Hemphill -S
Date: 2026.04.03 10:46:01 -05'00"
Alicia L. Hemphill (Johnson), MS
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K252495 | ? |
| Please provide the device trade name(s). | | ? |
| LSSA (NP-200) | | |
| Please provide your Indications for Use below. | | ? |
| LSSA is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body
contouring. LSSA is indicated for use in the following surgical specialists when the fragmentation and
aspiration of adipose tissue is desired:
- Plastic and Reconstruction Surgery
- Orthopedic Surgery | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
K252495, Page | 1
# 510(k) Summary
1. Date of Preparation: March 05, 2026
2. Applicant
1) Company: NEWPONG Co., Ltd.
2) Address: #1212, #1216 SJ Technoville, 278, Beotkkot-ro, Geumcheon-gu, Seoul, Korea
3) Tel: +82-2-541-0145
4) Fax: +82-2-541-0147
5) Contact person: Leo W Choi, 786-815-9121
6) Contact person address: 4855 Shady Pines Dr. Saint Cloud, FL, 34772 USA
7) Submission date: March 05, 2026
8) Prior related submission: Not applicable
3. Subject Device Information
1) Trade name: LSSA (NP-200)
2) Common name: Suction lipoplasty system
3) Classification name: System, Suction, Lipoplasty For Removal
4) Product code: QPB
5) Regulation number: 878.5040
6) Class of device: Class II
7) Panel: General & Plastic Surgery
4. Predicate Devices
1) Trade name (Predicate Submission Number): LipoSaver™ (K212969)
2) Manufacturer: LHbiomed Co., Ltd.
5. Device description
LSSA (NP-200) of Newpong Co., Ltd. is an ultrasonic surgical unit designed to fragment and emulsify subcutaneous fatty tissues using ultrasound frequency vibrations. The LSSA (NP-200) generates ultrasonic vibrations through the piezoelectric effect of PZT ceramic crystals embedded in the LSSA (NP-200) handpieces. When electrical energy is supplied from the main control unit, these PZT crystals rapidly expand and contract, converting electrical signals into mechanical ultrasonic vibrations. These vibrations are transmitted to the attached probe, which delivers the mechanical action required for tissue fragmentation and emulsification.
The LSSA (NP-200) is equipped with an electronic control system that allows the operator to adjust the vibration amplitude and frequency. It offers two operating modes, Continuous Mode and Pulse Mode. And NP-200 consists of the following components: Main body (control unit) / B Handpiece / F Handpiece / Foot switch / Probes / Spanner
6. Indications for Use:
LSSA is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. LSSA is indicated for use in the following surgical specialists when the fragmentation and aspiration of adipose tissue is desired:
- Plastic and Reconstruction Surgery
- Orthopedic Surgery
7. Brief summary of Nonclinical tests and results
1) Safety test
| Requirement – Test | | Standard | Result |
| --- | --- | --- | --- |
| Medical Electrical Equipment
Part 1: General Requirements for Basic Safety and Essential Performance | | IEC 60601-1 | Pass |
| Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | | IEC 60601-1-2 | Pass |
| No. | Test item | Result | |
| 1 | IEC TS 60601-4-2:2024 | Pass | |
| 2 | Ultrasound output power | Pass | |
{6}
K252495, Page
# 2) Biocompatibility
Biocompatibility of the LSSA probe was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communicating device - Tissue/bone/dentin" with a contact duration of "Limited (<24 hours)".
| Test item | Test method / Test criteria | Test result |
| --- | --- | --- |
| Cytotoxicity test | When it was tested according to ISO 10993-5, tests for in vitro cytotoxicity-Test on extracts method, it should satisfy the requirements. | Pass |
| Intracutaneous reactivity test | When it was tested according to ISO 10993-23, Tests for irritation-Animal intracutaneous (intradermal) reactivity test, it should satisfy the requirements. | Pass |
| Skin sensitization test | When it was tested according to ISO 10993-10, Tests for skin sensitization-Guinea pig maximization test (GPMT), it should satisfy the requirements. | Pass |
| Acute systemic toxicity test | When it was tested according to ISO 10993-11, Tests for systemic toxicity-Acute systemic toxicity, it should satisfy the requirements. | Pass |
| Pyrogen Test | When it was tested according to ISO 10993-11, Tests for systemic toxicity-Information on material-mediated pyrogens, it should satisfy the requirements. | Pass |
Biocompatibility of the LSSA handpiece was evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communicating device - Tissue/bone/dentin" with a contact duration of "Limited (<24 hours)".
| Test item | Test method / Test criteria | Test result |
| --- | --- | --- |
| Cytotoxicity test | When it was tested according to ISO 10993-5, tests for in vitro cytotoxicity-Test on extracts method, it should satisfy the requirements. | Pass |
| Intracutaneous reactivity test | When it was tested according to ISO 10993-23, Tests for irritation-Animal intracutaneous (intradermal) reactivity test, it should satisfy the requirements. | Pass |
| Skin sensitization test | When it was tested according to ISO 10993-10, Tests for skin sensitization-Guinea pig maximization test (GPMT), it should satisfy the requirements. | Pass |
| Acute systemic toxicity test | When it was tested according to ISO 10993-11, Tests for systemic toxicity-Acute systemic toxicity, it should satisfy the requirements. | Pass |
| Pyrogen Test | When it was tested according to ISO 10993-11, Tests for systemic toxicity-Information on material-mediated pyrogens, it should satisfy the requirements. | Pass |
# 8. Substantially Equivalent (SE) Comparison
| | Proposed device | Predicate device | Remark |
| --- | --- | --- | --- |
| 510(k) Number | K252495 | K212969 | N/A |
| Common Name | Suction lipoplasty system | Suction lipoplasty system | Same |
| Trade Name | LSSA | LipoSaver™ | N/A |
| Manufacturer | NEWPONG Co., Ltd | LHbiomed | N/A |
| Product Classification | ClassII | ClassII | Same |
| Product code | QPB | QPB | Same |
| Indication for use | LSSA is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. LSSA is indicated for use in the following surgical specialists when the fragmentation and aspiration of adipose tissue is desired: - Plastic and Reconstruction Surgery - Orthopedic Surgery | LipoSaver is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. LipoSaver is indicated for use in the following surgical specialists when the fragmentation and aspiration of soft tissue is desired: - Plastic and Reconstruction Surgery - Orthopedic Surgery | Same |
| Primary Function | Fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. | Fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. | Same |
{7}
K252495, Page
| Operating Frequency | 37.5 kHz±2%: B handpiece
31 kHz±2%: F handpiece | 35kHz~40kHz | Similar(#1) |
| --- | --- | --- | --- |
| Output acoustic power | 0~40W | 0~40W | Same |
| Output mode | Continuous Mode
Pulsed Mode (Duty cycle: 50%) | Continuous Mode
Pulsed Mode (Duty cycle: 50%) | Same |
| --- | --- | --- | --- |
| Probe | Ø3.7, 324mm
Ø2.9, 319mm
Ø1.8, 121.5mm
Ø1.2, 63.5mm
Ø0.9, 64mm | Ø3.7, 330mm
Ø2.9, 315mm
Ø1.9, 121mm | Similar(#2) |
| Performance Data: Safety and Performance | | | |
| | Proposed device | Predicate device | Remark |
| Biocompatibility | Conformed to ISO 10993-1, ISO10993-5, ISO10993-10, and ISO 10993-11 | Conformed to ISO 10993-1, ISO10993-5, ISO10993-10, and ISO 10993-11 | Same |
| Electrical Safety | Conformed to IEC 60601-1 | Conformed to IEC 60601-1 | Same |
| Electromagnetic Compatibility | Conformed to IEC 60601-1-2 | Conformed to IEC 60601-1-2 | Same |
> Equivalence discussion
Similar #1 – Operating Frequency
The proposed device operates within the same ultrasonic surgical frequency band as the predicate device, as defined by IEC 61847, ensuring functional equivalence in its core technological principles. The dual-handpiece configuration of the LSSA is the result of design optimization to maintain resonant stability specifically for the five types of provided probes. Comprehensive performance testing has confirmed that stable output and operational integrity are maintained across all power settings. This configuration does not change the intended use or technological characteristics of the device and does not raise new questions of safety or effectiveness.
Similar #2 – Probe size
The proposed device offers a wider range of probe diameters (Ø0.9 mm to Ø3.7 mm) compared to the predicate device. This allows surgeons to select the most appropriate probe for various treatment areas. All probes were validated through mechanical strength testing, acoustic output measurements, and performance evaluations. These dimensional differences do not alter the intended use or technological characteristics and do not raise new concerns regarding safety or effectiveness.
9. Conclusion
The device was evaluated through functional and performance testing to compare its operation with that of the legally marketed predicate device.
Based on the performance data and the comparison of technological characteristics, it is concluded that LSSA of NEWPONG Co., Ltd. is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.