K252480 · CapsoVision, Inc. · NEZ · Dec 19, 2025 · Gastroenterology, Urology
Device Facts
Record ID
K252480
Device Name
CE Deliver (DLV)
Applicant
CapsoVision, Inc.
Product Code
NEZ · Gastroenterology, Urology
Decision Date
Dec 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1300
Device Class
Class 2
Indications for Use
The disposable CE Deliver is a 2.5mm single-sheathed device indicated for transendoscopic delivery to the stomach or duodenum of CapsoCam Plus capsule endoscope (CE) devices. This device is intended for patients who are either unable to swallow the capsule or unable to pass the capsule beyond the pylorus in sufficient time to complete the desired diagnostic evaluation.
Device Story
CE Deliver is a single-use, disposable, non-sterile catheter-based accessory for CapsoCam Plus capsule endoscopes. It consists of a flexible catheter, a detachable capsule holder, and a proximal handle. Used in conjunction with a standard upper endoscope (instrument channel ≥2.8mm), the device is inserted through the endoscope channel. A physician loads the capsule onto the distal holder, guides it under direct endoscopic visualization to the stomach or duodenum, and uses a wire-based 'push-off' mechanism to release the capsule. Once released, the device is retracted, allowing the capsule to transit the GI tract naturally. The device facilitates diagnostic imaging for patients with swallowing or motility difficulties.
Clinical Evidence
Bench testing only. CapsoVision conducted comprehensive non-clinical performance testing, including packaging validation and design verification and validation (V&V) at baseline and after accelerated aging, confirming performance, durability, and safety for a 12-month shelf life.
Technological Characteristics
Single-use, non-sterile, disposable catheter-based device. Comprised of a flexible catheter, three-ring handle, and detachable capsule holder. Operates via a wire-based 'push-off' mechanism for capsule deployment. Compatible with upper endoscopes having an instrument channel diameter ≥2.8mm. No electronic components or software.
Indications for Use
Indicated for transendoscopic delivery of CapsoCam Plus capsule endoscopes to the stomach or duodenum in patients unable to swallow the capsule or unable to pass it beyond the pylorus in a timely manner.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
December 19, 2025
CapsoVision, Inc
Zane Liu
Director, Regulatory Affairs
18805 Cox Avenue
Suite 250
Saratoga, California 95070
Re: K252480
Trade/Device Name: CE Deliver (DLV)
Regulation Number: 21 CFR 876.1300
Regulation Name: Ingestible Telemetric Gastrointestinal Capsule Imaging System
Regulatory Class: Class II
Product Code: NEZ
Dated: November 21, 2025
Received: November 21, 2025
Dear Zane Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252480 - Zane Liu
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252480 - Zane Liu
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shanil P. Haugen -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252480 | ? |
| Please provide the device trade name(s). | | ? |
| CE Deliver (DLV) | | |
| Please provide your Indications for Use below. | | ? |
| The disposable CE Deliver is a 2.5mm single-sheathed device indicated for transendoscopic delivery to the stomach or duodenum of CapsoCam Plus capsule endoscope (CE) devices. This device is intended for patients who are either unable to swallow the capsule or unable to pass the capsule beyond the pylorus in sufficient time to complete the desired diagnostic evaluation. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252480 Page 1 of 2
| 510(k) #: K252480 | 510(k) Summary | Prepared on: 2025-11-28 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | CapsoVision, Inc | |
| Applicant Address | 18805 Cox Avenue Suite 250 Saratoga CA 95070 United States | |
| Applicant Contact Telephone | 408-866-6358 | |
| Applicant Contact | Ms. Azimun Jamal | |
| Applicant Contact Email | azimun.jamal@capsovision.com | |
| Correspondent Name | CapsoVision, Inc | |
| Correspondent Address | 18805 Cox Avenue Suite 250 Saratoga CA 95070 United States | |
| Correspondent Contact Telephone | 408-416-4142 | |
| Correspondent Contact | Mr. Zane Liu | |
| Correspondent Contact Email | zane.liu@capsovision.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | CE Deliver (DLV) | |
| Common Name | Ingestible telemetric gastrointestinal capsule imaging system | |
| Classification Name | System, Imaging, Gastrointestinal, Wireless, Capsule | |
| Regulation Number | 876.1300 | |
| Product Code(s) | NEZ | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K163495 | AdvanCE Capsule Endoscope Delivery Device | NEZ |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The CapsoVision CE Deliver device is a single use, disposable, non-sterile catheter-based medical device designed to facilitate the transendoscopic delivery of the CapsoCam Plus Capsule Endoscope to the stomach or duodenum. CE Deliver is an accessory to the CapsoCam Plus, and is intended for use in patients who are unable to swallow a capsule endoscope or cannot pass it beyond the pylorus in a timely manner. The CE Deliver is comprised of a thin, flexible catheter with a Capsule Holder at its distal end and a Handle at its proximal end. The Capsule Holder is detachable from the Handle and Catheter by means of a threaded connector. CE Deliver is designed to be used in conjunction with a standard upper endoscope, through which its flexible catheter is inserted into the endoscope's instrument channel, upon which the Capsule Holder may be threaded onto the distal end. The device enables a user to securely load a capsule endoscope onto the distal Capsule Holder, guide it under direct endoscopic visualization to the desired location within the upper gastrointestinal tract, and then precisely release the capsule. After capsule release, the CE Deliver is retracted along with the endoscope, allowing the | | |
{5}
K252480 Page 2 of 2
capsule to proceed naturally through the GI tract.
The CE Deliver is compatible with upper endoscopes with an instrument channel diameter of 2.8mm or larger.
**Intended Use/Indications for Use**
21 CFR 807.92(a)(5)
The disposable CE Deliver is a 2.5mm single-sheathed device indicated for transendoscopic delivery to the stomach or duodenum of CapsoCam Plus capsule endoscope (CE) devices. This device is intended for patients who are either unable to swallow the capsule or unable to pass the capsule beyond the pylorus in sufficient time to complete the desired diagnostic evaluation.
**Indications for Use Comparison**
21 CFR 807.92(a)(5)
The proposed CE Deliver has the same intended use as the predicate device. Both devices are transendoscopic delivery devices intended for the delivery of capsule endoscopes to the stomach or duodenum in patients who are either unable to swallow the capsule or unable to pass the capsule beyond the pylorus in sufficient time to complete the desired diagnostic evaluation.
The Indications for Use of the proposed device and the predicate device are highly similar, delineating the same range of functionalities and target patient population. For each device, the capsule endoscopy compatibility considerations are included within the Indications for Use with a comparable level of detail.
**Technological Comparison**
21 CFR 807.92(a)(6)
The proposed CE Deliver has highly similar technological characteristics to the predicate device, AdvanCE. Both devices are single-use, non-sterile, disposable medical devices designed for transendoscopic delivery of capsule endoscopes. They both consist of a flexible catheter, a three-ring handle, and a detachable capsule holder, and are designed for use with a standard upper endoscope (i.e., within the instrument channel) to facilitate accurate positioning and deployment of the capsule within the upper GI tract. Both devices utilize the same technology - i.e., deployment wire-based "push off" mechanism - to facilitate the deployment of capsules from the capsule holder into the GI tract.
Overall, the fundamental scientific technology and operating principles of the two devices are the same, and the minor differences consist primarily of the design of the capsule holder (i.e., the holder of each device optimizes for its range of compatible capsule endoscopes) and other differences in the implementation of the same basic design. These differences do not raise different questions of safety or effectiveness, as they do not alter the fundamental mechanism of action or introduce new risks.
**Non-Clinical and/or Clinical Tests Summary & Conclusions**
21 CFR 807.92(b)
CapsoVision has conducted comprehensive non-clinical performance tests on the CE Deliver device, including packaging validation and design verification & validation (V&V) at baseline and after accelerated aging. The tests confirm that the system's design outputs meet requirements, supporting its performance, durability, and safety for its labeled 12-month shelf life.
Overall, the results of the performance tests and assessments of the proposed device support that the CE Deliver is at least as safe and as effective as the predicate device for the same intended use, and confirm that any differences do not raise different questions of safety or effectiveness. The results of these tests therefore support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.