K252477 · Hermes Medical Solutions AB · KPS · Sep 9, 2025 · Radiology
Device Facts
Record ID
K252477
Device Name
Hybrid Viewer (00859873006240)
Applicant
Hermes Medical Solutions AB
Product Code
KPS · Radiology
Decision Date
Sep 9, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Hybrid Viewer is a software application for nuclear medicine and radiology. Based on user input, Hybrid Viewer processes, displays and analyzes nuclear medicine and radiology imaging data and presents the results to the user. The results can be stored for future analysis. Hybrid Viewer is equipped with dedicated workflows which have predefined settings and layouts optimized for specific nuclear medicine investigations. The software application can be configured based on user needs. The investigation of physiological or pathological states using measurement and analysis functionality provided by Hybrid Viewer is not intended to replace visual assessment. The information obtained from viewing and/or performing quantitative analysis on the images is used, in conjunction with other patient related data, to inform clinical management.
Device Story
Software application for nuclear medicine/radiology; processes, displays, and analyzes 2D/3D imaging data. Inputs: imaging studies from patient browsers or PACS. Operations: scrolling, zooming, panning, filtering, motion correction, fusion, registration, triangulation, ROI drawing, synchronization, and mathematical operations. Outputs: processed images and quantitative analysis results for clinical management. Used in clinical settings by healthcare professionals. Provides dedicated workflows for neurology, cardiology, gastroenterology, hepatology, pneumology, osteology, and nephrology. Benefits: assists in assessment of physiological/pathological states; informs clinical decision-making alongside other patient data. Does not replace visual assessment.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluation focused on the Centiloid scale implementation. Validation against 2015 Klunk et al. publication standards showed mean SUVR values for YC-0 (1.0068) and AD-100 (2.0730) within 2% of target. Linear correlation analysis for tracers (Florbetaben, Florbetapir, Flutemetamol) demonstrated high correlation coefficients (0.9648–0.9917) and R² values (0.9907–0.9953).
Technological Characteristics
Software-only medical device. Functions: 2D/3D image processing, fusion, registration, and quantitative analysis. Connectivity: loads data from PACS/patient browsers. Algorithm: includes Centiloid scale calculation for neurology. Software verification/validation conducted per FDA guidance for basic documentation level.
Indications for Use
Indicated for patients of any age/gender undergoing molecular imaging and radiology investigations, including assessment of cardiac blood flow (cardiac disease), brain function (Parkinson's disease, dementia), and lung perfusion/ventilation (pulmonary embolism).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 9, 2025
Hermes Medical Solutions AB
Hanne Grinaker
Chief Quality and Regulatory Affairs Officer
Strandbergsgatan 16
Stockholm, 11251
Sweden
Re: K252477
Trade/Device Name: Hybrid Viewer (00859873006240)
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission Computed Tomography System
Regulatory Class: Class II
Product Code: KPS
Dated: August 7, 2025
Received: August 7, 2025
Dear Hanne Grinaker:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252477 - Hanne Grinaker
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252477 - Hanne Grinaker
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D. O'hara - S
Date: 2025.09.09 11:11:26 -04'00'
Michael O'Hara, Ph.D.
Deputy Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252477 | |
| Device Name Hybrid Viewer | |
| Indications for Use (Describe) Hybrid Viewer is a software application for nuclear medicine and radiology. Based on user input, Hybrid Viewer processes, displays and analyzes nuclear medicine and radiology imaging data and presents the results to the user. The results can be stored for future analysis. Hybrid Viewer is equipped with dedicated workflows which have predefined settings and layouts optimized for specific nuclear medicine investigations. The software application can be configured based on user needs. The investigation of physiological or pathological states using measurement and analysis functionality provided by Hybrid Viewer is not intended to replace visual assessment. The information obtained from viewing and/or performing quantitative analysis on the images is used, in conjunction with other patient related data, to inform clinical management. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Hermes Medical Solutions
510(k) Premarket Notification Submission
HERMES MEDICAL SOLUTIONS
# 510(k) SUMMARY
K252477
The following information is provided as required by 21 CFR 807.92.
Date: 06 August 2025
## SUBMITTER
Name and Address: Hermes Medical Solutions AB
Strandbergsgatan 16
11251 Stockholm
Sweden
Contact Person: Hanne Grinaker
Chief Quality and Regulatory Officer
Phone: +46706604650
qa_ra@hermesmedical.com
## DEVICE
Subject Device Name: Hybrid Viewer
Device Classification: Class II
Regulation Number: 21 CFR 892.1200
Product Code: KPS
## PREDICATE DEVICE
Predicate Device Name: Hybrid Viewer
510(k) number: K241364
Regulation Number: 21 CFR 892.1200
Product Code: KPS
Product: Hybrid Viewer
{5}
Hermes Medical Solutions
510(k) Premarket Notification Submission
HERMES MEDICAL SOLUTIONS
# DEVICE DESCRIPTION
Hybrid Viewer is a software application which provides 2D and 3D viewing, processing and analysis for nuclear medicine investigations.
The studies can be loaded from patient browsers (e.g., GOLD) or PACS (Picture Archiving and Communication System).
Hybrid Viewer provides general tools which include scrolling, zooming, panning, filtering, motion correction, fusion, registration, triangulation drawing regions of interest, synchronization of studies and performing mathematical operations. Specific investigation areas for Hybrid Viewer are Neurology (BRASS), Cardiology, Gastroenterology, Hepatology, Pneumology, Osteology and Nephrology.
# INTENDED USE
Hybrid Viewer is a software application for nuclear medicine and radiology. Based on user input, Hybrid Viewer processes, displays and analyzes nuclear medicine and radiology imaging data and presents the result to the user. The result can be stored for future analysis.
Hybrid Viewer is equipped with dedicated workflows which have predefined settings and layouts optimized for specific nuclear medicine investigations.
The software application can be configured based on user needs.
The investigation of physiological or pathological states using measurement and analysis functionality provided by Hybrid Viewer is not intended to replace visual assessment. The information obtained from viewing and/or performing quantitative analysis on the images is used, in conjunction with other patient related data, to inform clinical management.
Product: Hybrid Viewer
{6}
Hermes Medical Solutions
510(k) Premarket Notification Submission
HERMES MEDICAL SOLUTIONS
# INTENDED PATIENT POPULATION AND MEDICAL CONDITIONS
Hybrid Viewer is intended for patients of any age and gender undergoing molecular imaging investigations.
Intended medical indication is any for which molecular imaging and radiology is performed. Examples of indications for which Hybrid Viewer may be used to inform patient management are assessment of cardiac blood flow using Tc99m in patients with cardiac disease, assessment of brain function in patients with Parkinson's' disease or dementia using Tc99m or amyloid tracers, and assessment of the difference between lung perfusion and ventilation using Tc99m tracers in order to provide a definitive diagnosis of pulmonary embolism.
# TECHNOLOGICAL COMPARISON
The subject device, Hybrid Viewer (version 7.2), employs similar fundamental scientific technology as its predicate device (Hybrid Viewer version 7.0).
- Both are software-only medical devices.
- Both have the same intended use.
- Both have the same indications for use.
- Both are based on the same primary device function of image analysis.
- Both have the same use environment
The difference between version 7.2 and version 7.0 of Hybrid Viewer is the addition of the Centiloid scale for Neurology investigations. Other changes in the new version do not affect clinical workflows or risk analysis. They are bug fixes or intended to improve ease of use without changing the available results and algorithms available.
Product: Hybrid Viewer
{7}
Hermes Medical Solutions
510(k) Premarket Notification Submission
HERMES MEDICAL SOLUTIONS
# PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions”. The “basic documentation” level was found to be applicable to the risk level of the software.
## Non-clinical Performance Evaluation of Algorithms
Non-clinical performance testing for the addition of Centiloid shows that the algorithm performs as expected and results were within pre-set acceptance criteria. The implementation of the Centiloid scale follows the directions given in the original Centiloid publication in 2015 by William E. Klunk et al.
Correlation result after spatial normalization to MNI-152 space are as follows (Level1 as described in the original publication):
The mean SUVR YC-0 calculated value of 1.006792265 falls within the 1.0095 +/- 2% range.
The mean SUVR AD-100 calculated value of 2.072978667 falls within the 2.076 +/- 2% range.
Linear correlation of the calculated value compared to the publication are as follows (Level2 as described in the original publication):
- Florbetaben correlation coefficient is 0.9915, intercept is 0.0252 and R²=0.9956
- Florbetapir correlation coefficient is 0.9648, intercept is 0.0658 and R²=0.9907
- Flutemetamol correlation coefficient is 0.9917, intercept is 0.0281 and R²=0.9953
In conclusion, non-clinical testing supports the safety of the subject device.
Product: Hybrid Viewer
{8}
Hermes Medical Solutions
510(k) Premarket Notification Submission
HERMES MEDICAL SOLUTIONS
# SUBSTANTIAL EQUIVALENCE CONCLUSION
There is no change in the overall safety and effectiveness of Hybrid Viewer version 7.2 compared to the predicate version 7.0. The fundamental design and principles of operation remain the same.
Non-clinical testing supports the safety of the device. Verification and validation testing demonstrates that the device performs as intended.
In conclusion, Hermes Medical Solutions considers Hybrid Viewer (version 7.2) to be substantially equivalent to the predicate device Hybrid Viewer (version 7.0).
Product: Hybrid Viewer
Page | 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.