Yomi S Robotic System

K252376 · Neocis, Inc. · PLV · Oct 28, 2025 · Dental

Device Facts

Record IDK252376
Device NameYomi S Robotic System
ApplicantNeocis, Inc.
Product CodePLV · Dental
Decision DateOct 28, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4120
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Yomi S Robotic System (Yomi S) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi S is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi S Robotic System. The output of YomiPlan is to be used with the Yomi S Robotic System.

Device Story

Yomi S is a robotic navigational system for dental implant surgery and alveoplasty. Input: patient CBCT scan. Process: YomiPlan software creates 3D anatomical model for pre-operative planning; intra-operatively, system uses physical linkages (Patient Splint, Tracker End Effector) to track patient movement. Robotic arm with 7 degrees of freedom holds standard dental drill; provides haptic, visual, and audible feedback to surgeon. Arm constrains drill to planned trajectory; automatically adjusts to patient movement to maintain accuracy. Used in clinical settings by dental surgeons. Surgeon maintains manual control of drill; system prevents deviation from plan. Benefits: improved accuracy of drill placement and surgical guidance.

Clinical Evidence

Bench testing only. Performance verified through human factors software validation, biocompatibility testing (ISO 10993-1:2018), cleaning/sterilization validation (ANSI/AAMI ST98:2022), EMC testing (IEC 60601-1-2), and software/cybersecurity verification and validation. Accuracy is <1 mm RMS.

Technological Characteristics

Robotic arm with 7 degrees of freedom; haptic feedback; patient tracking via physical linkage (Patient Splint, Tracker End Effector). Materials: IXEF HC-1022 NT000, Ti-6AL-4V-ELI, Al 7075-T6, Avaspire AV-651CF30, Silicon Nitride. Connectivity: Wi-Fi/Ethernet. Software: Windows 11 24H2. Sterilization: Moist heat for reusable components. Electrical: Class I, Type B (drill), Type BF (tracker).

Indications for Use

Indicated for partially edentulous and fully edentulous adult patients who qualify for dental implants, including planning and performing guided bone reduction (alveoplasty) of the mandible and/or maxilla.

Regulatory Classification

Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION October 28, 2025 Neocis, Inc. Jorge Fernandes Vice President of Quality and Regulatory 545 NW 26th Street Suite 700 Miami, Florida 33127 Re: K252376 Trade/Device Name: Yomi S Robotic System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument And Accessories Regulatory Class: Class II Product Code: PLV, QRY Dated: July 30, 2025 Received: September 30, 2025 Dear Jorge Fernandes: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252376 - Jorge Fernandes Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252376 - Jorge Fernandes Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S for Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Yomi S Robotic System Page 11 of 85 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252376 | ? | | Please provide the device trade name(s). | | ? | | Yomi S Robotic System | | | | Please provide your Indications for Use below. | | ? | | The Yomi S Robotic System (Yomi S) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi S is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi S Robotic System. The output of YomiPlan is to be used with the Yomi S Robotic System. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | {4} yomi BY NEOCIS # 510(k) Summary K252376 ## I. Submitter Neocis, Inc. 545 NW 26th Street Unit 700 Miami, FL 33127 Tel: 1-855-9NEOCIS Contact Person: Jorge Fernandes, VP, QA-RA Date Prepared: October 24, 2025 ## Device Trade Name: Yomi S Robotic System Common Name: Dental Stereotaxic Instrument Classification Name: Bone cutting instrument and accessories (21 CFR 872.4120) Classification: Class II Product Code: PLV, QRY ## Predicate Devices Primary Predicate: Yomi Robotic System (K231018) ## Indications for Use The Yomi S Robotic System (Yomi S) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi S is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi S Robotic System. The output of YomiPlan is to be used with the Yomi S Robotic System. ## Device Description The Neocis Yomi S is a modified next iteration of the Yomi Robotic System, designed to provide guidance for a dental surgeon during dental implant surgery. Yomi S is a dental stereotaxic medical device (Product Codes PLV, QRY) regulated under 21 CFR 872.4120. The device includes a YomiLink that is placed on the patient prior to the CT scan, and a fiducial array with fiducial markers that is placed on the YomiLink prior to the CT scan so the virtual plan can be related to the physical space of the system. The Guidance Arm secures a standard dental drill, allowing the surgeon to grip the drill as normal. The Guidance Arm does not move unless the surgeon applies a manual force to the drill. The Guidance Arm will constrain the surgeon to drill according to the prescribed surgical plan, preventing deviation. The surgeon is constantly in control of the drilling. The system has a mechanical feedback system that is connected to the YomiLink on the patient, which relays information to the control software in order to Page 1 {5} yomi. BY NEOCIS track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle and position to accommodate the patient movement, which will maintain the accuracy of the drill placement. The Yomi S Robotic System allows the user to plan the surgery virtually in YomiPlan, cleared for use alone on third-party PCs for preplanning. The operative plan is based on a cone beam computed tomography (CBCT) scan of the patient, which is used to create a 3-D model of the patient anatomy in the planning software. The plan is used for the system to provide physical, visual, and audible feedback to the surgeon during the implant site preparation. The Yomi S robotic arm holds and guides a standard FDA- cleared third party powered bone cutting instrument. The patient tracking portion of Yomi S is comprised of linkages from the patient to Yomi S, which include the Patient Splint (YomiLink Teeth or YomiLink Bone), Tracker End Effector (TEE), and the Patient Tracker (PT). In cases where YomiLink Teeth is utilized, it is attached to the contralateral side of the patient's mouth over stable teeth using on-label dental materials prior to the presurgical CBCT scan. In cases where YomiLink Bone is utilized, it is placed using bone screws prior to the presurgical CBCT scan (appropriate local anesthesia is required), or after the scan when using the subject YomiLink Arch device. The subject of this submission is to: introduce the Yomi S Robotic System, a next-generation modification of the Yomi Robotic System, intended to assist dental surgeons by providing guidance during dental implant procedures. ## Comparison of Technological Characteristics The following Table 1 provides a summary of the subject Yomi S Robotic System features compared to the predicate device, Yomi Robotic System (K231018). Page 2 {6} yomi. BY NEOCIS Table 1: Comparison of technological characteristics to the predicate (with a supporting reference device) | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | Use Specifications | | | | | Indications for Use (IFU) | The Yomi S Robotic System (Yomi S) is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi S is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi S Robotic System. The output of YomiPlan is to be used with the Yomi S Robotic System. | Yomi Robotic System is a computerized robotic navigational system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides robotic navigational guidance of the surgical instruments. The system can also be used for planning and performing guided bone reduction (also known as alveoplasty) of the mandible and/or maxilla. Yomi Robotic System is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. When YomiPlan software is used for preplanning on third party PCs, it is intended to perform the planning (pre-operative) phase of dental implantation surgery. YomiPlan provides pre-operative planning for dental implantation procedures using the Yomi Robotic System. The output of YomiPlan is to be used with the Yomi Robotic System. | Equivalent | | Principles of Operation | Yomi S is a dental stereotaxic medical device (Product Codes PLV, QRY) and a powered surgical device for bone cutting (21 CFR 872.4120). Yomi S is a computerized navigational system intended to | Yomi Robotic System is a dental stereotaxic medical device (Product Codes PLV, QRY) and a powered surgical device for bone cutting (21 CFR 872.4120). Yomi Robotic System is a computerized navigational | Equivalent | {7} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The Yomi S is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. The device includes a Patient Splint that is placed on the patient prior to the CT scan, and a fiducial array with fiducial markers that is placed on the Patient Splint prior to the CT scan so the virtual plan can be related to the physical space of the system. The Guidance Arm secures a standard dental drill, allowing the surgeon to grip the drill as normal. The Guidance Arm does not move unless the surgeon applies a manual force to the drill. The Guidance Arm will constrain the surgeon to drill according to the prescribed surgical plan, preventing deviation. The surgeon is constantly in control of the drilling. The system has a mechanical feedback system that is connected to the Patient Splint on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle and position to accommodate the patient movement, which will | system intended to provide assistance in both the planning (pre-operative) and the surgical (intra-operative) phases of dental implantation surgery. The system provides software to preoperatively plan dental implantation procedures and provides navigational guidance of the surgical instruments. The Yomi Robotic System is intended for use in partially edentulous and fully edentulous adult patients who qualify for dental implants. The device includes a Patient Splint that is placed on the patient prior to the CT scan, and a fiducial array with fiducial markers that is placed on the Patient Splint prior to the CT scan so the virtual plan can be related to the physical space of the system. The Guidance Arm secures a standard dental drill, allowing the surgeon to grip the drill as normal. The Guidance Arm does not move unless the surgeon applies a manual force to the drill. The Guidance Arm will constrain the surgeon to drill according to the prescribed surgical plan, preventing deviation. The surgeon is constantly in control of the drilling. The system has a mechanical feedback system that is connected to the Patient Splint on the patient, which relays information to the control software in order to track patient movement. If patient movement occurs during the surgical procedure, the system will respond by altering the prescribed surgical cutting angle and position to accommodate the | | Page 4 {8} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | maintain the accuracy of the drill placement. | patient movement, which will maintain the accuracy of the drill placement. | | | Use Environment | Clinical Setting, Doctor's Office | Clinical Setting, Doctor's Office | Equivalent | | Technology / Performance Characteristics | | | | | Pre-Operative Planning | The dentist places the virtual implant images in the desired location and orientation on the Yomi S Application software, which displays the patient's CT information in 3 dimensions. The precise location, angulations and type of virtual implant are determined based on the intermaxillary relations, bone structure and width, and adjacent critical anatomical structures. | The dentist places the virtual implant images in the desired location and orientation on the Yomi Robotic System Application software, which displays the patient's CT information in 3 dimensions. The precise location, angulations and type of virtual implant are determined based on the intermaxillary relations, bone structure and width, and adjacent critical anatomical structures. | Equivalent | | Intra-Operative Setup | The Yomi S dental handpiece is autoclave sterilized and the Patient Tracker End-Effector is high-level disinfected or sterilized. The end-effectors are attached to their respective arms and the system is draped on both arms with single-use, non-sterile drapes. A single-use barrier film protects the column controls. The User Interface monitor, monitor handles, and cart handles are Intermediate-level disinfected. All other components of the system are subject to general cleaning using an intermediate-level disinfectant wipe. | The Yomi Robotic System dental handpiece is autoclave sterilized and the Patient Tracker End-Effector is high-level disinfected or sterilized. The end-effectors are attached to their respective arms and the system is fully draped with single-use, non-sterile drapes. | Equivalent. All high touch points have been assessed and the applicable level of disinfection determined and validated The testing has shown that the device is as safe and effective as the predicate. | | Intra-Operative Process | The registration is completed during pre-operative steps, identification of the fiducials (Extra Oral Fiducial Array (EOFA)) and Intra Oral Fiducial Array (IOFA)) in the software. The Patient Tracker is attached to the | The registration is completed during pre-operative steps, identification of the fiducials (Extra Oral Fiducial Array (EOFA)) and Intra Oral Fiducial Array (IOFA)) in the software. The Patient Tracker is attached to the | Equivalent | Page 5 {9} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | YomiLink to complete the link between the image and the patient. Using the Yomi S System guidance, the surgeon approaches the site of surgery following the audible, visual, and physical cues given by the system. The surgical implantation procedure is performed by the dentist as guided by the system, in line with preoperative planning. During the surgical procedure, the Yomi S guidance will warn and attempt to prevent any deviation between the planned location and the actual performance. The warnings are visual, audible, and physical. | YomiLink to complete the link between the image and the patient. Using the YRS System guidance, the surgeon approaches the site of surgery following the audible, visual, and physical cues given by the system. The surgical implantation procedure is performed by the dentist as guided by the system, in line with preoperative planning. During the surgical procedure, the YRS guidance will warn and attempt to prevent any deviation between the planned location and the actual performance. The warnings are visual, audible, and physical. | | | Robotic Guide Arm | Guided robotic arm with 7 degrees of freedom | Guided robotic arm with 6 degrees of freedom | Equivalent The subject device provides an additional degree of freedom that is intended to provide additional dexterity and address the occurrence of singularities and joint limits. Testing has been performed to show device is as safe and effective as the predicate. | | Movement Direction | Guided Robotic Arm holds a surgical instrument and provides haptic feedback on position with respect to the plan restricting movement | Guided Robotic Arm holds a surgical instrument and provides haptic feedback on position with respect to the plan restricting movement | Equivalent The subject device provides an additional | Page 6 {10} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | outside of volume predefined during planning. 7 degrees of freedom | outside of volume predefined during planning. 6 degrees of freedom | degree of freedom that is intended to provide additional dexterity and address the occurrence of singularities and joint limits. Testing has been performed to show device is as safe and effective as the predicate. | | Patient affixed tracking parts | YomiLink Arch YomiLink Bone Yomi Link Teeth | YomiLink Arch YomiLink Bone Yomi Link Teeth | Equivalent | | Mating Component Design | Kinematic Mount with threaded metal insert and larger screw thread size | Kinematic Mount with tapped hole and smaller screw thread size | Equivalent. The subject device contains a larger thread for usability and a threaded metal insert to reduce the frequency of cross-threading. Testing has been performed to show that the device as is safe and effective as the predicate. | | Patient Tracking Mechanism | Physical linkage to patient via Patient Tracker (PT), Kinematic Mount (KM), and Tracker End Effector (TEE) | Physical linkage to patient via Patient Tracker (PT), Kinematic Mount (KM), and Tracker End Effector (TEE) | Equivalent | Page 7 {11} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | Visualization | CT Scanned Image (DICOM) | CT Scanned Image (DICOM) | Equivalent | | Accuracy | <1 mm RMS | <1 mm RMS | Equivalent | | Fiducials for CT scan | YomiLink Arch contains fiducial beads and is placed directly on patient arch during the CT scan to provide a reference in the image. Intraoral Fiducial Array contains fiducial beads and is attached to patient splint during the CT scan to provide a reference in the image. | YomiLink Arch contains fiducial beads and is placed directly on patient arch during the CT scan to provide a reference in the image. Intraoral Fiducial Array contains fiducial beads and is attached to patient splint during the CT scan to provide a reference in the image. | Equivalent | | Dental Drill Console | Integrated: Bien Air Motor (Model No. MOT MX-I LED 3RD GEN) (K092214) Bien Air Chiropro L System (Model No. CA 20:1 L Micro-Series) (K983183) | Standalone: Aseptico Drill Motor (Model No. AEU7000LNE-70V) (K030163) SE Analysis Anthogyr Mont Blanc handpiece (Aseptico Model No. AHP-85MBFO-CX) (K070084) | Equivalent. The subject device contains an integrated Drill console supplied by Bien Air that has been user validated, bench tested, and IEC-60601-1 tested to ensure the device is as safe and effective as the predicate. The integration of the drill motor allows for the addition of safety mitigations and controls. | | Energy | | | | | Supply Voltage | 120V | 120V | Equivalent | | Phases | 1 | 1 | Equivalent | | Type of Current | AC | AC | Equivalent | | Rated Frequency (HZ) | 50/60 Hz | 50/60 Hz | Equivalent | Page 8 {12} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | Rated Power Input (VA) | 1200 VA | 600 VA | Equivalent. Although the subject device has an increased power input it is within the commercial specification for breaker control. Testing has been performed to show that the device is as safe and effective as the predicate. | | Types and Ratings of external accessible fuses | 10.0 A for 240 V | 5.0 A for 240 V | Equivalent. The subject device has an increased external fuse rating to support an increase in power output. Testing has been performed to show that the device is as safe and effective as the predicate. | | Type of Protection against Electric Shock | Class I Equipment | Class I Equipment | Equivalent | | Degree of Protection against Electric Shock | Type B Drill motor Type BF Patient Tracker | Type BF | Equivalent. The subject device incorporates a Drill Console | Page 9 {13} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | | | supplied by Bien Air, which has been user validated, bench tested, and IEC-60601-1 tested to show device is as safe and effective as the predicate | | Equipment Suitable for use in the presence of Flammable Mixtures? | No | No | Equivalent | | Mode of Operation | Continuous Operation | Continuous Operation | Equivalent | | Reprocessing, Sterilization, and Shelf Life | | | | | Reprocessing Classification | Single-use and Non-Critical reusable | Single-use and Non-Critical reusable | Equivalent | | Reprocessing Method | General Cleaning, Intermediate-level disinfection, High-level disinfection, and Cleaning & Sterilization | General cleaning, High-level disinfection, and Cleaning & Sterilization | Equivalent. The subject device contains additional cleaning steps. Testing has been performed to show that the device as is safe and effective as the predicate. | | Sterilization Method | Provided unsterile, end user moist heat sterilized | Provided unsterile, end user moist heat sterilized | Equivalent | | Shelf life | Currently, there are no components or ancillary components of the system designed from degradable materials or materials prone to expiration | Currently, there are no components or ancillary components of the system designed from degradable materials or materials prone to expiration | Equivalent | | Biocompatibility | | | | | Patient Contact | Surface device limited duration | Surface device limited duration | Equivalent | Page 10 {14} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | Biocompatibility | Mucosal membrane, tissue, bone, dentin contact | Mucosal membrane, tissue, bone, dentin contact | Equivalent | | Materials | YomiLink Bone (YLB): Body: IXEF HC-1022 NT000 (Natural 50% Glass Fiber Polyarylamide) Sleeve: Ti-6AL-4V-ELI Insert: Al 7075-T6 YomiLink Teeth (YLT): Body: IXEF HC-1022 NT000 (Natural 50% Glass Fiber Polyarylamide) Screw: Al 7075-T6 Insert: Al 7075-T6 YomiLink Arch (YLA): Body: Avaspire AV-651CF30 Epoxy: Loctite EA M-31 CL Insert: Al 7075-T6 IntraOral Fiducial Array (IOFA): Main body: Avaspire AV-651 CF30 Fiducial Beads: Silicon Nitride Epoxy: Loctite EA M-31CL epoxy Screw: Al 7075-T6 | YomiLink Bone (YLB): Body: IXEF HC-1022 NT000 (Natural 50% Glass Fiber Polyarylamide) Sleeve: Ti-6AL-4V-ELI YomiLink Teeth (YLT): Body: IXEF HC-1022 NT000 (Natural 50% Glass Fiber Polyarylamide) Screw: Al 7075-T6 YomiLink Arch (YLA): Body: Avaspire AV-651CF30 Epoxy: Loctite EA M-31 CL IntraOral Fiducial Array (IOFA): Main body: Avaspire AV-651 CF30 Fiducial Beads: Silicon Nitride Epoxy: Loctite EA M-31CL epoxy Screw: Al 7075-T6 | Equivalent. The addition of the insert to YLB, YLT, and YLA does not introduce new materials. The Al 7075-T6 material is being used in the predicate device YLT and IOFA products. Testing has been performed to show device is as safe and effective as the predicate. Materials were added to reduce the risk of cross-threading. | | Software/Firmware & Cybersecurity | | | | | Level of Concern | Enhanced | Major | Equivalent | | Yomi Plan Software Version | v3.0 | v2.6.1 | Equivalent. The software has the same functionality as the predicate software with exception of Voice Controls. Software and Human Factors testing has been verified and validated | Page 11 {15} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | | | to ensure the device is as safe and effective as the predicate. | | Operating System | Windows 11 24H2 | Windows 11 24H2 | Equivalent | | PC Requirements | PC with 64-bit Windows 10 OS or newer with a minimum of 4 GB of RAM and a 2 GHz dual core processor. Local memory (hard drive) should be a minimum of 100 GB with 7200 RPM or SSD. Connectivity requirements include ethernet, Wi-Fi, USB, or CD drive. | PC with 64-bit Windows 10 OS or newer with a minimum of 4 GB of RAM and a 2 GHz dual core processor. Local memory (hard drive) should be a minimum of 100 GB with 7200 RPM or SSD. Connectivity requirements include ethernet, Wi-Fi, USB, or CD drive. | Equivalent | | Yomi Plan Functions | • Load CT Scanned Image • Optimize Image • Plan Procedure (place implant) • Save Surgical Plan • Connect to Control software • Provide Feedback to Surgeon regarding physical location of Drill and Drill components • Select Surgical Phase • Set areas for mechanical restriction during surgical operation • Visualize CT Scanned Image with 2D Slices • Generate Panoramic reconstruction along arch • Visualize Panoramic reconstruction with cross sections along panoramic arch • Map Splint coordinate system to structures in CT Scan • Define anatomical planes • Clip CT Scanned Images • Define Arch for generating panoramic reconstruction • Provide the user with a means to define a nerve | • Load CT Scanned Image • Optimize Image • Plan Procedure (place implant) • Save Surgical Plan • Connect to Control software • Provide Feedback to Surgeon regarding physical location of Drill and Drill components • Select Surgical Phase • Set areas for mechanical restriction during surgical operation • Visualize CT Scanned Image with 2D Slices • Generate Panoramic reconstruction along arch • Visualize Panoramic reconstruction with cross sections along panoramic arch Map Splint coordinate system to structures in CT Scan Define anatomical planes Clip CT Scanned Images Define Arch for generating panoramic reconstruction Provide the user with a means to define a nerve | Equivalent. Subject device and predicate have same functionality with exception of voice controls. The subject device has been verified and validated to ensure the device is as safe and effective as the predicate. | Page 12 {16} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | • Allow the user to plan multiple implants • Measure distances and angles in the plan • Voice recognition and controls | • Allow the user to plan multiple implants • Measure distances and angles in the plan | | | CT to STL alignment | Enables different ways to import STL files so they can be aligned to the coordinate system of the user’s choice (relative to the CT or relative to the choice of another STL) | Importing STL files is limited in alignment regarding which coordinate system is used as the reference. | Equivalent | | Dual Arch Workflow | Uses a single case file to capture both upper and lower arches on a single patient. | Uses two case files to plan each of the upper and lower arches for a single patient in separate files. | Equivalent. Consolidating the two arches of the same patient into a single case file makes to streamline the workflow. | | Restorative Planning | Contains basic generic crown planning functionality using standard crown libraries, with additional functionality around morphing crowns and biocopy (mirroring existing crowns). | Contains basic generic crown planning functionality using standard crown libraries. | Equivalent. Enhancements for more crown planning. | | Patient and robot positioning work volume guidance | Patient and robot positioning guidance includes horizontal, vertical, and angulation guidance. | Patient and robot positioning guidance includes horizontal and vertical guidance only. | Equivalent. | | Baseline of force sensor | Baselining can only be performed if the arm is a certain proximity away from the patient workspace. It is automatically performed during startup and can be executed by the user with a button press. | Baseline can be performed at any arm position. It can be executed by the user with a button press. | Equivalent | | YomiLink Bone (YLB) planning. | YLB planning enables a “snap” function that finds the best fit to the segmented bone in the model. | YLB planning requires the user to place it without any automatic adjustments relative to the bone. | Equivalent | | Proximity threshold for distance from implant to anatomy. | Has a minimum value of 1mm as the distance threshold for generating a warning, eliminating the possibility that the user can proceed without warnings or a threshold too small to generate appropriate warnings. | Has a minimum value of 0mm as the distance threshold for generating a warning, which enables the user to effectively remove warnings. | Equivalent | | Bone reduction planning | Bone reduction can account for soft tissue thickness in addition to the bone removal amount | Bone reduction only accounts for the bone removal amount | Equivalent | | Implant info hover | Shows basic implant information on | Shows basic implant information on | Equivalent | Page 13 {17} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | display | hover plus the max depth that was reached | hover | | | VTK Render Library | VTK library version 9 | VTK library version 8 | Equivalent | | Surgeon Preferences | Implant brand, model, and model specifics are available as settings in surgeon preferences | Implant brand and model are available as settings in surgeon preferences | Equivalent | | Case feedback | Makes available restorative feedback in the case feedback section during shutdown | No restorative feedback in the case feedback section during shutdown. | Equivalent | | Implant library | Contains major implant models in the library plus several new additions | Contains major implant models in the library | Equivalent | | Handpiece gestures | User interface allows for mouse / keyboard / gesture interaction to change between Free, Pause, and Guided modes. Gestures (e.g. double-tap, pull, etc.) enable transition between Free to Pause, Pause to Free, Drill to Free, and Free to Guided. All mode changes are accompanied by visual and audio confirmation cues. | User interface requires mouse / keyboard / gesture interaction to change between Free, Pause, and Guided modes. Gestures (e.g. double-tap) enable transition from Free to Pause mode. All mode changes are accompanied by visual and audio confirmation cues. | Equivalent The addition of more gestures enables the user holding the handpiece to have more direct control over the modes. | | Control software response to guide arm joint limits, singularities, and potential wrist / base collisions | Control software provides warnings regarding joint limits, singularities, and potential wrist / base collisions at an improved response rate | Control software provides warnings regarding joint limits, singularities, and potential wrist / base collisions | Equivalent | | Tracker arm joints | Grounding brushes in the joint interfaces. PCB reliability improvement. Generate fewer false positive errors. | Pogo pins in the joint interfaces. Occasional false positive errors from the joints. | Equivalent. Grounding brushes and PCB improvements enable greater reliability. | | YomiLink Teeth (YLT) | compatible with bite blocks | Not easily compatible with bite blocks | Equivalent | | Intraoral Fiducial Array (IOFA) | IOFA has small, medium, and large sizing. | IOFA only has a single size available. | Equivalent. Allows for accommodation of more patients. | | Implant planning on the Osteotomy Page | Only allows depth adjustments | Allows full range of adjustments | Equivalent Restricting freedom of changes to only the most | Page 14 {18} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | | | common reduces the chance of unintended plan adjustments | | YomiLink Bone (YLB) Proximity Warnings | The YLB warning uses the more appropriate threshold value for determining if a warning is appropriate. | The YLB warning uses various thresholds to determine a collision risk for collisions between implant, screw, and nerve. | Equivalent | | | | | | | Wi-Fi | Always Active | Always Active | Equivalent | | Wireless data transmission over LAN | Yes, via integrated hardware, tested according to: • AAMI TIR69: 2017 Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems. • IEEE ANSI C63.27-2021 American National Standard for Evaluation of Wireless Coexistence | Yes, via integrated hardware, tested according to: • AAMI TIR69: 2017 Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems. • IEEE ANSI C63.27-2017 American National Standard for Evaluation of Wireless Coexistence | Equivalent. A new version of the standard IEEE ANSI C63.27:2021 has been introduced. The device has been tested to ensure it is as safe and effective as the predicate. | | Interface | Windows based Graphical User Interface (GUI) and voice commands | Windows based Graphical User Interface (GUI) | Equivalent. Subject device offers the addition of voice commands that have been user validated and bench tested to ensure the device is as safe and effective as the predicate. | | Cybersecurity | AAMI TIR57:2016 ANSI/AAMI SW96:2023 IEC 80001-5-1 Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions September 27, 2023 | Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Issued October 2, 2014 | Equivalent. The subject device was designed, built, and tested to meet the latest | Page 15 {19} yomi. BY NEOCIS | Technological Characteristics | Subject Device: Yomi S Robotic System | Primary Predicate: Yomi Robotic System with YomiLink Arch (K231018) | Comparison | | --- | --- | --- | --- | | | | | standards and ensure it is as safe and effective as the predicate. | | Electrical Safety | | | | | Electrical Safety | Compliance with the following: IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION ANSI/AAMI ES60601-1:2005®2012 and C1:2009/® 2012 and A2:2010/® 2012 (Consolidated Text) Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance. IEC 80601-2-77 Edition 1.0 2019-07 ISO15223-1:2021-07 ISO 14971:2019 | Compliance with the following: IEC 606011:2005 + Corr. 1 (2006) + Corr. 2 (2007) ANSI/AAMI ES60601-1:2005®2012 and C1:2009/® 2012 and A2:2010/® 2012 (Consolidated Text) Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance. ISO15223-1:2012 BS EN ISO 14971:2012 | Equivalent Testing for subject device was performed to latest standard with declaration of conformity to ASCA. | | Electromagnetic Compatibility (EMC) | 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | Compliance with IEC 60601-1-2:2007 | Equivalent testing for subject device was performed to latest standard with declaration of conformity to ASCA. | Page 16 {20} YOMI. BY NEOCIS # Performance Testing Performance testing was conducted to verify and validate that the subject Yomi S Robotic System software application, hardware components, and accessories meet all relevant functional and system requirements. Therefore, ensuring that the system is as safe and as effective and substantially equivalent to the predicate device. ## Usability Human Factors Validation testing The Yomi S Robotic System underwent human factors software validation in accordance with the FDA's "Applying Human Factors and Usability Engineering to Medical Devices" Final Guidance, (February 2016). ## Biocompatibility testing Biocompatibility evaluation for Yomi S components was conducted in accordance with International Standard ISO-10993-1:2018, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a risk management process. The battery of testing included ISO 10993-5:2009, Biological evaluation of medical devices – Part 5: Test for in vitro cytotoxicity, ISO 10993-10:2021, Biological evaluation of medical devices – Part 10: Test for skin sensitization, ISO10993-12:2021 Biological evaluation of medical devices – Part 12 Sample preparation and reference materials, and ISO 10993-23:2021 Biological evaluation of medical devices – Part 23 Test for irritation. ## Cleaning/ Sterilization Verification and Validation Testing Cleaning was conducted in compliance with Guidance for Industry and FDA Staff – Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling. U.S. FDA March 2015, ANSI/AAMI ST98:2022, Cleaning validation of health care products—Requirements for development and validation of a cleaning process for medical devices, and AAMI TIR12:2020 (R2023): Designing, testing, and labeling medical devices intended for processing by health care facilities: A guide for device manufacturers ## EMC Testing EMC testing has been conducted on the Yomi S in accordance with various recognized industry standards by a recognized third-party organization. IEC 60601-1 Edition 3.2 2020 Consolidated Version, ANSI/AAMI ES60601-1:2005*2012 and C1:2009/* 2012 and A2:2010/* 2012 (Consolidated Text) Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance was used for product safety, IEC 60601-1-2:2014/A1:2020 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests, EN60601-1-2:2015/A1:2021 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests, and 80601-2-77 Medical electrical equipment – Part 2-77: Particular requirements for the basic safety and essential performance of robotically assisted surgical equipment ## Software and Cybersecurity Verification and Validation Testing Software verification and validation (V&amp;V) testing was performed in compliance with FDA's "Guidance for the Content of Premarket Submissions for Device Software Functions" (Jun 2023). As a failure or latent flaw in the software could directly lead to minor injury, the software was assigned an "enhanced" level of concern. Software development and testing followed a robust methodology aligning with IEC 62304 Edition 1.1:2015, the aforementioned FDA software guidance, and FDA's "General Principles of Software Validation" (January 2002). The thorough testing and analysis of results provide strong assurance of the device's intended performance. Cybersecurity development and testing was performed in accordance with Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (June 2025) and Postmarket Management of Cybersecurity in Medical Devices (December 2016). Based upon the information provided within this 510(k) Premarket Notification, we conclude that the Yomi S is substantially equivalent to the identified predicate device. The Subject Device has passed testing for appropriate verification, validation, and performed per its intended use. Therefore, it can be considered substantially equivalent to the Predicate Device Page 17 {21} yomi. BY NEOCIS # Conclusion The subject of this submission is to introduce the Yomi S Robotic System, a next-generation modification of the Yomi Robotic System, intended to assist dental surgeons by providing guidance during dental implant procedures. There are no changes to the intended use compared to the predicate device. The performance testing demonstrates substantially equivalent performance of the subject device as compared to the predicate. Page 18
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...