← Product Code [GEX](/productcode/GEX) · K252369

# Boston Pico755 (K252369)

_Boston Aesthetics, Inc. · GEX · Jan 21, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252369

## Device Facts

- **Applicant:** Boston Aesthetics, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jan 21, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Boston Pico755 is indicated for the removal of tattoos and benign pigmented lesions, including but not limited to Ota nevus, Hori spots (Hori nevus), and melasma. The Boston Pico755 with the 2 mm and 6 mm handpieces and the handpieces with Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I-IV.

## Device Story

Boston Pico755 is a solid-state laser system for dermatological procedures; used in professional healthcare facilities by trained physicians. System components include main unit (power, laser, control, cooling), light guide articulated arm, footswitch, and 8 replaceable handpieces. Operator selects handpiece; system automatically detects model and displays corresponding interface. User adjusts parameters (wavelength, fluence, energy, spot size, repetition rate) via interface; laser energy delivered through articulated arm and handpiece upon footswitch activation. Cooling system manages thermal output. Device facilitates tattoo removal, pigmented lesion treatment, and skin rejuvenation (acne scars/wrinkles) via photomechanical effects. Clinical benefit includes non-invasive aesthetic improvement.

## Clinical Evidence

No human clinical data provided. Evidence consists of bench testing (laser performance metrics: wavelength, pulse width, energy, spot size) and animal studies demonstrating safety and efficacy for tattoo removal comparable to the predicate device. Biocompatibility testing (ISO 10993) confirmed non-cytotoxic, non-irritating, and non-sensitizing properties.

## Technological Characteristics

Solid-state laser; 755nm wavelength; 450ps pulse duration. Materials: Aluminum alloy. Energy source: Mains power (110-220VAC). Connectivity: Standalone. Cooling: Integrated system. Handpieces: 8 types (Zoom S, Zoom X, M, M6, M8, M10, 755x, 755s). Standards: IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2, ISO 10993.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- PicoSure Workstation ([K210226](/device/K210226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

January 21, 2026

Boston Aesthetics Inc.
Cao Hongmei
General Manager
1521 Concord Pike
Suite 201
Wilmington, Delaware 19803

Re: K252369
Trade/Device Name: Boston Pico755
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: December 22, 2025
Received: December 22, 2025

Dear Cao Hongmei:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252369 - Cao Hongmei
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.

{2}

K252369 - Cao Hongmei
Page 3

See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA
L. HITHE - S
Digitally signed by
TANISHA L. HITHE - S
Date: 2026.01.21
15:41:28 -05'00"

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K253269  |   |
|  Device Name BSOTON PICO755  |   |
|  Indications for Use (Describe) The Boston Pico755 is indicated for the removal of tattoos and benign pigmented lesions, including but not limited to Ota nevus, Hori spots (Hori nevus), and melasma. The Boston Pico755 with the 2 mm and 6 mm handpieces and the handpieces with Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I-IV.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

510(k) Summary

This 510(k) Summary is submitted in accordance with the requirements of 21CFR Section 807.92.

The assigned 510(k) Number: K252369

## 1. Administrative Information

**Date of Preparation**
2026-01-16

**Submission Correspondent**

Name: Boston Aesthetics INC
Address: 1521 Concord Pike Suite 201 Wilmington DE 19803
Tel: +001 949-792-8168
E-mail: bsnaesthetics@gmail.com
Contact: Ms. Hongmei Cao

## 2. Device Information

|  Device Name: | Boston PICO755  |
| --- | --- |
|  Model: | Boston PICO755  |
|  Manufacturer: | Boston Aesthetics INC  |
|  Regulation Description: | Laser surgical instrument for use in general and plastic surgery and in dermatology.  |
|  Regulation Medical Specialty: | General & Plastic Surgery  |
|  Regulation Number: | 878.4810  |
|  Product Code: | GEX  |
|  Device Class: | 2  |
|  Classification Name | Powered Laser Surgical Instrument  |

## 3. Predicate Device

|  Device name: | PicoSure Workstation  |
| --- | --- |
|  Manufacturer: | Cynosure LLC  |
|  Regulation Description: | Laser surgical instrument for use in general and plastic surgery and in dermatology.  |
|  Regulation Medical Specialty: | General & Plastic Surgery  |
|  Regulation Number: | 878.4810  |
|  Product code: | GEX  |
|  Device Class: | 2  |
|  Classification Name | Powered Laser Surgical Instrument  |

Page 1 of 5

{5}

510(K) Number: K210226

# 4. Intended Use

The Boston Pico755 is indicated for the removal of tattoos and benign pigmented lesions, including but not limited to Ota nevus, Hori spots (Hori nevus), and melasma. The Boston Pico755 with the 2 mm and 6 mm handpieces and the handpieces with Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I-IV.

# 5. Device Description

The Boston Pico755 is a medical solid-state laser equipment which is indicated for the removal of tattoos and benign pigmented lesions, including but not limited to: Ota nevus, Hori spots (Hori nevus) and melasma. It has handpieces for the treatment of acne scars and wrinkles in skin types I-IV. The device is intended to be used in professional healthcare facilities by trained physicians only.

The subject device Boston Pico755 consists of main unit (including power module, laser module, control system and cooling system), a light guide articulated arm, a footswitch, a handpiece and protective Glasses. It is connected to supply mains directly through undetachable power cord and plug. There are 8 types of replaceable handpieces - Zoom S, Zoom X, M, M6, M8, M10, 755x, 755s, and operators can choose the appropriate handpiece according to treatment needs.

When the system automatically detects the model of the installed handpiece, it will display the corresponding treatment interface specific to that handpiece. Users can view or adjust relevant parameters such as wavelength, fluence, energy, spot size, and repetition rate directly from the interface. After turning on the device, set the Wavelength, Fluence, Frequency, Spot Size, etc. to be used then press the Footswitch in the READY state, the laser will be energized and transferred through the Articulated Arm and handpiece. The cooling system controls the heating caused by the laser output.

# 6. Comparison with predicate device

|  Items | Subject Device | Predicate Device (K210226) | Comparison  |
| --- | --- | --- | --- |
|  Name | Boston PICO755 | PicoSure Workstation | /  |
|  Intended use/Indications for use | The Boston Pico755 is indicated for the removal of tattoos and benign pigmented lesions, including but not limited to: Ota nevus, Hori spots (Hori nevus) and melasma. The Boston Pico755 with the 2mm and 6mm handpieces and the handpieces with Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I-IV. | 755nm: The PicoSure Workstation is indicated for tattoo and benign pigmented lesions removal including but not limited to: Nevus of Ota. Hori macules (nevus of Hori), and Melasma. The PicoSure Workstation with the 2mm and 6mm handpieces and the Focus Array are indicated for the treatment of acne scars and wrinkles in Skin Types I-IV. 532nm: The PicoSure 532nm delivery system is indicated for tattoo removal and benign pigmented lesions removal in Skin Types I-III. 1064nm: The PicoSure 1064nm delivery | Similar  |

{6}

|  Items | Subject Device | Predicate Device (K210226) | Comparison  |
| --- | --- | --- | --- |
|   |  | system is indicated for tattoo and benign pigmented lesions removal. |   |
|  Wavelength | 755nm | 755nm | Same  |
|  Pulse Duration | 450ps (±20%) | 500 - 900ps | Comparable  |
|  Output mode | Single/ Repeat pulse | Single/ Repeat pulse | Same  |
|  Repeat pulse frequency | Single pulse, 1 Hz, 2.5 Hz, 5 Hz, 10 Hz | Single Shot, 1 Hz, 2.5 Hz, 5 Hz, 10 Hz | Same  |
|  Max pulse energy | Zoom S: 300mJ, 200 mJ @ 2mm
Zoom X: 300mJ
M; M6; M8; M10: 300mJ
755x: 232mJ
755s: 148mJ | Max 300mJ | Comparable  |
|  Max fluence/pulse energy | ZoomS: 6.37J/cm²
ZoomX: 1.06 J/cm²
M: 2.31 mJ/μbeam
M6: 2.31 mJ/μbeam
M8: 1.29 mJ/μbeam
M10: 0.83 mJ/μbeam
755X: 2.32 mJ/μbeam
755S: 2.31 mJ/μbeam | 6.37J/cm² (max. among all handpieces) | Comparable for all handpieces  |
|  Treatment spot diameter | Zoom S: 2-5 mm
Zoom X: 6-10 mm
M: 6-10mm, stepping 1mm
M6: 6mm
M8: 8mm
M10: 10mm
755x: 10mm × 10mm
755s: 6mm × 6mm | Zoom 2-6mm;
Fixed 5, 6, 8, 10 mm | Comparable  |
|  Patient Contacting Material | Aluminum Alloy | 316 Stainless Steel | Comparable  |
|  Device dimension | 942mm × 380mm × 1025mm
(L×W×H, not including light guide articulated arm) | 109 cm x 56cm x 107cm
Note: Height with arm extended 158cm | Comparable  |
|  Device Weight | About 115kg | 375 lb (171 kg) | Comparable  |
|  Electrical parameters | 110-220VAC ±10%;10A, 50/60 Hz
Input power rate: 2400 VA | 200-240 V~, 4.5 kVA, 50/60 Hz,
Single Phase | Comparable  |

The subject device and the predicate device have the same intended use. Although the subject device and the predicate device have several different technological characteristics as noted in the table above, the differences do not raise different questions of safety and effectiveness.

Page 3 of 5

{7}

Page 4 of 5

# 7. Non-Clinical Performance Testing

## 7.1. Electromagnetic Compatibility and Electrical Safety Test

The subject device has passed safety testing in according to following standards.

1) IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012, IEC 60601-1:2005/AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
2) IEC 60601-2-22: 2019 Medical electrical equipment -Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
3) IEC 60825-1: 2014 Safety of laser products - Part 1: Equipment classification and requirements
4) IEC 60601-1-2:2014 + A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances- Requirements and tests
5) IEC TS 60601-4-2: 2016 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems

## 7.2. Biocompatibility Test

Biocompatibility testing was conducted in accordance with the 2020 FDA guidance “Use of International Standard ISO 10993, Biological Evaluation of Medical Device Part 1: Evaluation and Testing.” The testing includes:

1) Cytotoxicity per ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5: Tests for in vitro Cytotoxicity.
2) Sensitization per ISO 10993-10: 2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
3) Irritation per ISO 10993-23: 2021 Biological evaluation of medical devices - Part 23: Tests for irritation
4) Systemic toxicity per ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

The user- contacting materials were shown to be non-cytotoxic, non-irritating and non-sensitizing.

## 7.3. Performance test

Bench Testing was performed to verify the performance of the device and the predicate for all handpieces. The proposed device met the specification and performance characteristics as identified in design control procedures. The energy related specifications of the subject device were found comparable to the corresponding specifications of the predicate device. The performance tests included but not limited to were the measurement of laser wavelength, pulse width, terminal maximum single pulse energy, repeat pulse frequency, spot diameter, micro-beam diameter, and energy per microbeam.

## 7.4 Software verification and validation

Software documentation of the subject device was provided in accordance with the FDA guidance Document- “Content of Premarket Submissions for Device Software Functions: Guidance for Industry and Food and Drug Administration Staff”, which was issued in 06/14/2023 to support a device’s Basic Documentation Level.

{8}

Page 5 of 5

## 8. Animal testing

Animal studies were provided to show that the device can remove black tattoos as safely and as effectively as the predicate device.

## 9. Clinical data

NA.

## 10. Conclusion

The subject device and the predicate device have the same intended use and comparable technological characteristics. The results of the testing described above demonstrate that the subject device is as safe and effective as the predicate device (K210226) and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/device/K252369](https://fda.innolitics.com/device/K252369)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
