← Product Code [FLE](/productcode/FLE) · K252307

# Getinge GSS67N Series Steam Sterilizer (K252307)

_Maquet GmbH (A Getinge Group Company) · FLE · Jan 22, 2026 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K252307

## Device Facts

- **Applicant:** Maquet GmbH (A Getinge Group Company)
- **Product Code:** [FLE](/productcode/FLE.md)
- **Decision Date:** Jan 22, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6880
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Getinge GSS67N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS67N Series Steam Sterilizer is available in 3 models differentiated by chamber length: GSS67N Model 6710 (39 inch chamber), GSS67N Model 6713 (51 inch chamber) and GSS67N Model 6717 (67 inch chamber).

## Device Story

Steam sterilizer for healthcare facilities; processes wrapped/unwrapped, porous/nonporous heat/moisture stable items (surgical instruments, linens). Uses pressurized saturated steam; gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. Three models based on chamber length (39", 51", 67"). Control system sequences/monitors cycle phases; provides audible/visual alerts for parameter deviations. Operated by healthcare staff. Output: sterilized items. Benefits: ensures sterility of reusable medical devices. Updated software/cybersecurity controls per FDA guidance; increased load capacity for specific PreVac cycles compared to predicate.

## Clinical Evidence

No clinical data required. Bench testing only. Performance validated per ANSI/AAMI ST8:2013/(R)2018; electrical safety per ANSI/UL 61010-1; EMC per IEC 61326-1 and IEC 61010-2-040. Effectiveness demonstrated by complete kill of biological indicators.

## Technological Characteristics

Vessel material: Stainless Steel SA240. Sterilization agent: Saturated steam. Operation: Gravity displacement and dynamic air removal (pre-vacuum). Connectivity: Not specified. Software: Updated for cybersecurity and software function compliance. Sterilization method: Pressurized steam.

## Regulatory Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

## Predicate Devices

- Getinge GSS67N Series Steam Sterilizer ([K172159](/device/K172159.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

January 22, 2026

Maquet GMBH (a Getinge Group company)
% Barb Smith
Principal Regulatory Affairs Specialist
Getinge
45 Barbour Pond Drive
Wayne, New Jersey 07470

Re: K252307
Trade/Device Name: Getinge GSS67N Series Steam Sterilizer
Regulation Number: 21 CFR 880.6880
Regulation Name: Steam Sterilizer
Regulatory Class: Class II
Product Code: FLE
Dated: December 19, 2025
Received: December 19, 2025

Dear Barb Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}

K252307 - Barb Smith
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K252307 - Barb Smith
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

STEPHEN A. ANISKO -S
Digitally signed by
STEPHEN A. ANISKO -S
Date: 2026.01.22
16:37:42 -05'00'

Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection Control Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K252307

Device Name
Getinge GSS67N Series Steam Sterilizer

Indications for Use (Describe)
The Getinge GSS67N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS67N Series Steam Sterilizer is available in 3 models differentiated by chamber length: GSS67N Model 6710 (39 inch chamber), GSS67N Model 6713 (51 inch chamber) and GSS67N Model 6717 (67 inch chamber).

List of available cycles: See tables below

Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)
Page 1 of 3
PSC Publishing Services (301) 443-6740

{4}

Pre-vacuum cycles. (Dynamic air removal cycles)

|  Cycle Name | Factory Settings |   |   | Load Configuration (Note 1) | Maximum Items per Chamber Length  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Exp. Temp. | Exp. Time | Drying Time |   | 6710 1000 mm 39.4 in | 6713 1300 mm 51.2 in | 6717 1700 mm 66.9 in  |
|  PREVAC 1 135°C | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 2 135°C | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 3 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 4 132.2°C | 132.2°C (270.0°F) | 4 min | 30 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 5 132.2°C | 132.2°C (270.0°F) | 4 min | 5 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 6 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 7 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 8 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 9 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 10 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 11 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 12 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 13 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 14 132.2°C 3) | 132.2°C (270.0°F) | 4 min | 30 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 15 132.2°C 4) | 132.2°C (270.0°F) | 4 min | 5 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 16 132.2°C 4) | 132.2°C (270.0°F) | 4 min | 5 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  Bowie Dick Test | 134°C (273.0°F) | 3 min, 30 sec | 0 min | 1 B&D Test Pack in an EMPTY chamber (other than loading accessories) | 1 Test Pack | 1 Test Pack | 1 Test Pack  |
|  IUSS 1 135°C | 135.0°C (275.0°F) | 3 min | 1 min (Note 4) | Double-wrapped instrument trays (up to 25 lbs per tray) | 1 | 1 | 1  |
|   |   |   |   |  Fabric packs | 1 | 1 | 1  |
|  Leak Test (Note 2) | 131.1°C (268.0°F) | N/A | N/A | Empty Chamber (other than loading accessories) | — | — | —  |

1) The cycle is an identical copy of PREVAC 1.
2) The cycle is an identical copy of PREVAC 2.
3) The cycle is an identical copy of PREVAC 4.
4) The cycle is an identical copy of PREVAC 5.

Page 2 of 3

{5}

Gravity displacement cycles

|  Cycle Name | Factory Settings |   |   | Load Configuration (Note 1) | Maximum Items per Chamber Length  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Exp. Temp. | Exp. Time | Drying Time |   | 6710 1000 mm 39.4 in | 6713 1300 mm 51.2 in | 6717 1700 mm 66.9 in  |
|  GRAVITY 1 121.1°C | 121.1°C (250.0°F) | 30 min | 45 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 8 | 12 | 16  |
|   |   |   |   |  Fabric packs | 18 | 24 | 30  |
|  GRAVITY 2 135°C | 135.0°C (275.0°F) | 10 min | 45 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 8 | 12 | 16  |
|   |   |   |   |  Fabric packs | 18 | 24 | 30  |
|  GRAVITY 3 132.2°C | 132.2°C (270.0°F) | 15 min | 45 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 8 | 12 | 16  |
|   |   |   |   |  Fabric packs | 18 | 24 | 30  |
|  IUSS 2 135°C (Note 7) | 135.0°C (270.0°F) | 10 min | 30 sec (Note 4) | Double-wrapped instrument trays (up to 25 lbs per tray) | 1 | 1 | 1  |
|   |   |   |   |  Fabric packs | 1 | 1 | 1  |
|  IUSS 3 132.2°C (Note 7) | 132.2°C (270.0°F) | 4 min | 1 min (Note 4) | Double-wrapped instrument trays (up to 25 lbs per tray) | 1 | 1 | 1  |
|   |   |   |   |  Fabric packs | 1 | 1 | 1  |

Liquid load cycles

|  Cycle Name | Factory Settings |   |   | Load Configuration | Maximum Items per Chamber Length  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Exp. Temp. | Exp. Time | Cooling Time |   | 6710 1000 mm 39.4 in | 6713 1300 mm 51.2 in | 6717 1700 mm 66.9 in  |
|  Vented Bottles 121°C | 121.1°C (250.0°F) | 45 min | 3kPa/min (0.44 psi/min) (Note 3) | Each container 1000 mL (34 fl oz) or smaller (Note 5 and 6) | 3 | 3 | 3  |

NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.

Only the cycles described in the above tables are cleared for use in United States Health Care settings.

TABLE NOTES:

1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAMI Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8). For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to steam sterilization and sterility assurance in health care facilities.
2. Vacuum leak test parameters are not adjustable.
3. Cooldown rate
4. Items may NOT be dry at the end of the following cycles:
- IUSS 1, 2, 3
- PREVAC 2
- PREVAC 5
Drying time may be added if required.
5. Facility must validate the cycle if the load includes containers larger than 1000 mL (34 fl oz).
6. Use vented or open containers only.
7. The recommended minimum exposure time and temperature for unwrapped, nonporous loads (e.g., metal instruments) that are sterilized for immediate use is 3 minutes at 132°C (270°F) or 135°C (275°F).

Page 3 of 3

{6}

GETINGE

# K252307 510(k) SUMMARY

Getinge GSS67N Series Steam Sterilizer

Submitted by: Maquet GmbH
Kehler Strasse 31
Rastatt DE-BW
Germany 76437
+49 7222 932-0
Dr. Silke Avieny

Contact Person: Barb Smith
Principal Regulatory Affairs Specialist
Phone: (585) 370-6101
Email: barb.smith@getinge.com

Date prepared: January 20, 2026

Proprietary Name: GSS67N Series Steam Sterilizer

Common Name: Steam Sterilizer

Device Classification: Steam Sterilizer (FLE)
Class II, as listed per 21 CFR 880.6880

Predicate Device: Getinge GSS67N Series Steam Sterilizer [K172159]

# Description of Device:

The Getinge GSS67N Series Steam Sterilizer is designed for sterilization of heat and moisture stable materials used in healthcare facilities. There are three model designations to identify three different chamber lengths. The model 6710 is 1000 mm (39") long, model 6713 is 1300 mm (51") long and model 6717 is 1700 mm (67") long.

The Getinge GSS67N Series Steam Sterilizer employs both gravity/downward displacement with positive pulse conditioning and pressure/vacuum pulsing for dynamic air removal. All cycle phases are sequenced and monitored by the control system, providing both audible and visual notification of deviation from certain operating parameters.

Page 1 of 6

{7}

Getinge GSS67N Steam Sterilizer 510(k)

# Indications for Use:

The Getinge GSS67N Series Steam Sterilizer is intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam. The GSS67N Series Steam Sterilizer is available in 3 models differentiated by chamber length: GSS67N Model 6710 (39 inch chamber), GSS67N Model 6713 (51 inch chamber) and GSS67N Model 6717 (67 inch chamber).

List of available cycles: See tables below

Pre-vacuum cycles. (Dynamic air removal cycles)

|  Cycle Name | Factory Settings |   |   | Load Configuration (Note 1) | Maximum Items per Chamber Length  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Exp. Temp. | Exp. Time | Drying Time |   | 6710 1000 mm 39.4 in | 6713 1300 mm 51.2 in | 6717 1700 mm 66.9 in  |
|  PREVAC 1 135°C | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 2 135°C | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 3 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 4 132.2°C | 132.2°C (270.0°F) | 4 min | 30 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 5 132.2°C | 132.2°C (270.0°F) | 4 min | 5 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 6 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 7 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 8 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 9 135°C 1) | 135.0°C (275.0°F) | 3 min | 16 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 10 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 11 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 12 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 13 135°C 2) | 135.0°C (275.0°F) | 3 min | 3 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 14 132.2°C 3) | 132.2°C (270.0°F) | 4 min | 30 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 12 | 16 | 24  |
|  PREVAC 15 132.2°C 4) | 132.2°C (270.0°F) | 4 min | 5 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  PREVAC 16 132.2°C 4) | 132.2°C (270.0°F) | 4 min | 5 min (Note 4) | Fabric packs | 18 | 24 | 30  |
|  Bowie Dick Test | 134°C (273.0°F) | 3 min, 30 sec | 0 min | 1 B&D Test Pack in an EMPTY chamber (other than loading accessories) | 1 Test Pack | 1 Test Pack | 1 Test Pack  |

K252307

{8}

Getinge GSS67N Steam Sterilizer 510(k)

|  IUSS 1 135°C | 135.0°C (275.0°F) | 3 min | 1 min (Note 4) | Double-wrapped instrument trays (up to 25 lbs per tray) | 1 | 1 | 1  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |   |   |  Fabric packs | 1 | 1 | 1  |
|  Leak Test (Note 2) | 131.1°C (268.0°F) | N/A | N/A | Empty Chamber (other than loading accessories) | – | – | –  |

1) The cycle is an identical copy of PREVAC 1.
2) The cycle is an identical copy of PREVAC 2.
3) The cycle is an identical copy of PREVAC 4.
4) The cycle is an identical copy of PREVAC 5.

## Gravity displacement cycles

|  Cycle Name | Factory Settings |   |   | Load Configuration (Note 1) | Maximum Items per Chamber Length  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Exp. Temp. | Exp. Time | Drying Time |   | 6710 1000 mm 39.4 in | 6713 1300 mm 51.2 in | 6717 1700 mm 66.9 in  |
|  GRAVITY 1 121.1°C | 121.1°C (250.0°F) | 30 min | 45 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 8 | 12 | 16  |
|   |   |   |   |  Fabric packs | 18 | 24 | 30  |
|  GRAVITY 2 135°C | 135.0°C (275.0°F) | 10 min | 45 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 8 | 12 | 16  |
|   |   |   |   |  Fabric packs | 18 | 24 | 30  |
|  GRAVITY 3 132.2°C | 132.2°C (270.0°F) | 15 min | 45 min | Double-wrapped instrument trays (up to 25 lbs per tray) | 8 | 12 | 16  |
|   |   |   |   |  Fabric packs | 18 | 24 | 30  |
|  IUSS 2 135°C (Note 7) | 135.0°C (270.0°F) | 10 min | 30 sec (Note 4) | Double-wrapped instrument trays (up to 25 lbs per tray) | 1 | 1 | 1  |
|   |   |   |   |  Fabric packs | 1 | 1 | 1  |
|  IUSS 3 132.2°C (Note 7) | 132.2°C (270.0°F) | 4 min | 1 min (Note 4) | Double-wrapped instrument trays (up to 25 lbs per tray) | 1 | 1 | 1  |
|   |   |   |   |  Fabric packs | 1 | 1 | 1  |

## Liquid load cycles

|  Cycle Name | Factory Settings |   |   | Load Configuration | Maximum Items per Chamber Length  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Exp. Temp. | Exp. Time | Cooling Time |   | 6710 1000 mm 39.4 in | 6713 1300 mm 51.2 in | 6717 1700 mm 66.9 in  |
|  Vented Bottles 121°C | 121.1°C (250.0°F) | 45 min | 3kPa/min (0.44 psi/min) (Note 3) | Each container 1000 mL (34 fl oz) or smaller (Note 5 and 6) | 3 | 3 | 3  |

NOTE: Liquid Cycles are not intended for the sterilization of liquids used for direct patient contact.

Only the cycles described in the above tables are cleared for use in United States Health Care settings.

K252307

{9}

Getinge GSS67N Steam Sterilizer 510(k)

# TABLE NOTES:

1. The load configurations listed in these tables are those used during testing validations of the sterilizer. These configurations follow AAMI Standard ST8: Hospital steam sterilizers where applicable (fabric packs are process challenge devices as described in ANSI/AAMI ST8 and were made to be consistent with the packs described in ANSI/AAMI ST8). For guidance on processing loads in the sterilizer, refer to AAMI Standard ST79: Comprehensive guide to steam sterilization and sterility assurance in health care facilities.

2. Vacuum leak test parameters are not adjustable.

3. Cooldown rate

4. Items may NOT be dry at the end of the following cycles:
- IUSS 1, 2, 3
- PREVAC 2
- PREVAC 5

Drying time may be added if required.

5. Facility must validate the cycle if the load includes containers larger than 1000 mL (34 fl oz).

6. Use vented or open containers only.

7. The recommended minimum exposure time and temperature for unwrapped, nonporous loads (e.g., metal instruments) that are sterilized for immediate use is 3 minutes at 132°C (270°F) or 135°C (275°F).

# Comparisons to Predicate Device:

Similarities between the Getinge GSS67N Series Steam Sterilizer and the identified predicate device (Getinge GSS67N Series Steam Sterilizer K172159) are:

- Intended use is the same: Intended for use by health care facilities to sterilize wrapped and unwrapped, porous and nonporous heat and moisture stable items such as surgical instruments and linens by means of pressurized steam.

- Operating Principle is the same: Saturated steam is the sterilizing agent.

- Materials of construction are the same: Vessel material is Stainless Steel SA240. There is no direct patient contact associated with this device.

- Cycle Types: The cycle types offered are the same; Prevacuum, Gravity, Immediate Use and Liquids (not for sterilization of liquids used directly for patient contact).

- Performance Testing: Programmed factory set cycles were tested per industry standards and guidelines and effectiveness of sterilizer function was demonstrated by complete kill of biological indicators. Getinge Sterilizers have been validated to meet the requirements of ANSI/AAMI ST8 Hospital Steam Sterilizers.

The differences between the Getinge GSS67N Series Steam Sterilizer and the predicate device (Getinge GSS67N Series Steam Sterilizer K172159) are:

K252307

{10}

Getinge GSS67N Steam Sterilizer 510(k)

- The Getinge GSS67N series steam sterilizer has been tested to performance standards using an increased load capacity. The validated load capacity for the 132C and 135C PreVac cycles has been increased as noted in the table below:

|  Cycle | Load Configuration | 6710 | 6713 | 6717  |
| --- | --- | --- | --- | --- |
|  PreVac 1, 3, 6-9 135C | Double Wrapped Instrument Trays (up to 25lbs) | Increase from 8 to12 | Increase from 12 to 16 | Increase from 16 to 24  |
|  PreVac 4,14 132.2C | Double Wrapped Instrument Trays (up to 25lbs) | Increase from 8 to12 | Increase from 12 to 16 | Increase from 16 to 24  |

- The software and software documentation used in the GSS67N Series Steam Sterilizer has been updated to comply with FDA guidance: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions and Content of Premarket Submissions for Device Software Functions.

Summary of Performance Testing:

|  Evaluation | Standards Used | Results  |
| --- | --- | --- |
|  Performance | ANSI/AAMI ST8:2013/(R) 2018 Hospital Steam Sterilizers | Passed - Cycles PREVAC 1 and PREVAC 4 were evaluated for performance with increased load size and demonstrated compliance to the standard  |
|  Electrical Safety (ES) | ANSI/UL 61010-1 3rd Ed dated May 12 2012 with revision through July 19 2019 | Passed – results demonstrated compliance to the standard  |
|  EMC and ES | IEC 61010-2-040:2020 | Passed – results demonstrated compliance to the standard  |
|  EMC | IEC 61326-1 Edition 3.0 2020-10 | Passed – provided GAP analysis testing to bridge the gaps in specific tests covered in IEC 60601-1-2 that are not in IEC 61326-1." Results demonstrated compliance to the applicable sections of the standard  |
|  EMC | IEC 61010-2-040:2020 | Passed – the results demonstrated compliance to the standard  |
|  EMC | FCC 47 CFR 15 and ICES-001 | Passed – the results demonstrated compliance to the standard  |
|  Cybersecurity | FDA guidance: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions | Meets the requirements of the guidance  |
|  Software | FDA guidance: Content of Premarket Submissions for Device Software Functions | Meets the requirements of the guidance  |

Clinical Data:

No clinical data is required for this device classification submission.

K252307

{11}

Getinge GSS67N Steam Sterilizer 510(k)

## Conclusion:

Based on the indications for use, technological characteristics, performance testing, and comparison to predicate device, the subject devices Getinge GSS67N Series Steam Sterilizer has been shown to be as safe, as effective, and to perform as well or better than the legally marketed predicate device GSS67N Series Steam Sterilizer (K172159).

K252307
Page 6 of 6

---

**Source:** [https://fda.innolitics.com/device/K252307](https://fda.innolitics.com/device/K252307)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
